Remote Control. Tonghui TH2515

Add to My manuals
96 Pages

advertisement

Remote Control. Tonghui TH2515 | Manualzz

TH2515 Operation Manual

6) Remote Control

6.1

RS232C Connection

RS-232 standard, also called as asynchronous serial communication standard, has already been widely used for data communication between computers, computer and external equipment. RS is the English abbreviation of Recommended Standard; 232, the standard number. This standard is issued by EIA in 1969, which rules to send one bit in a data line every time.

As most serial interfaces, the serial interface of TH2515 is also not strictly based on RS-232 standard but only uses the smallest subset of this standard. The signals are listed in the following table.

Signal

Transmitted Data

Received Data

Signal Ground Common

Code

TXD

RXD

GND

Connector Pin Number

2

3

5

The reason is that the use of three lines is much more inexpensive and much simpler than that of five lines or six lines, which is the biggest advantage of using serial interface for communication.

Table 6-1 RS-232 Signal and Pin Connector

The connection of the instrument with PC is shown in figure 6-1.

TXD(2) (3) RXD

PC

( Controller ) RXD(3) (2) TXD TH2515

GND(5) (5) GND

Figure 6-1 Connection of the instrument with PC

Figure 6-1 shows that the serial interface pin definition of this instrument is different from that of

9 pin connector used in the computer. User can purchase the serial interface cable from our company.

RS232 interface characterizes with a baud rate ranging from 9600 to 115200, no parity, 8-bit data bit, 1-bit stop bit.

The instrument commands comply with the SCPI standard. When a command string is

42