17.5.2 Setting Alarm Pausing and Resetting Properties. Mindray VS9 Operator's Manual
Add to My manuals276 Pages
advertisement
![17.5.2 Setting Alarm Pausing and Resetting Properties. Mindray VS9 Operator's Manual | Manualzz 17.5.2 Setting Alarm Pausing and Resetting Properties. Mindray VS9 Operator's Manual | Manualzz](http://s1.manualzz.com/store/data/067700425_1-0a7a5f7ca753834b1ab9c639289712cc-360x466.png)
Item
Med Alarm Interval
Range
16 s to 30 s
Default
20 s
17.5.1.3
Setting Alarm Volume Escalation
The monitor provides the function of automatically escalating the alarm tone volume.
When this function is enabled, if an alarm is not reset within the designated delay time after the alarm occurs, the volume of the alarm tone increases automatically.
To set the alarm volume escalation, follow this procedure:
1.
Select the Main Menu quick key → from the System column select Maintenance
→ input the required password → select .
2.
Select the Alarm tab. The Audio tab is displayed.
3.
Set Auto Increase Volume .
◆ 2 Steps : if an alarm is not reset within the designated delay time after the alarm occurs, the alarm volume automatically increases by two levels.
◆ 1 Step : if an alarm is not reset within the designated delay time after the alarm occurs, the alarm volume automatically increases by one level.
◆ Off : if an alarm is not reset within the designated delay time after the alarm occurs, the volume of the alarm tone does not change.
4.
Select Increase Volume Delay to set the delay time of alarm volume escalation.
NOTE
•
The alarm volume escalation function is not applied to the latched alarms.
17.5.2
Setting Alarm Pausing and Resetting Properties
17.5.2.1
Defining the Pause Function
You can either pause alarms or pause alarm tones. This depends on the pause setting. To set the pause function, follow this procedure:
1.
Select the Main Menu quick key → from the System column select Maintenance
→ input the required password → select .
2.
Select the Alarm tab → Pause/Reset tab.
3.
Set Pause to Alarm Pause or Audio Pause .
17.5.2.2
Setting the Alarm Pause Time/Alarm Tone Pause Time
The alarm pause time or alarm tone pause time can be set to 1 min , 2 min , 3 min , or
Permanent . The default audio pause time is two minutes.
VS 9 Vital Signs Monitor Operator’s Manual
17 - 9
To set the alarm tone pause time, follow this procedure:
1.
Select the Main Menu quick key → from the System column select Maintenance
→ input the required password → select .
2.
Select the Alarm tab
→
Pause/Reset tab.
3.
Set Pause Time .
17.5.2.3
Setting the Priority of Audio Paused Alarms
To select alarm of what priority can be paused or alarm sound of what priority can be paused, follow this procedure:
1.
Select the Main Menu quick key → from the System column select Maintenance
→ input the required password → select .
2.
Select the Alarm tab → Pause/Reset tab.
3.
Set Pause Priority .
◆ All : pressing the Alarm Pause quick key pauses all alarms.
◆ Med & Low : pressing the Alarm Pause quick key pauses alarms of medium and low priority. The high priority alarms will not be paused.
◆ Disable : the Alarm Pause quick key is disabled.
17.5.2.4
Disabling the Options of Prolonging Alarm Pause Time/Alarm Tone
Pause Time
To disable specific options for the clinician when prolonging alarm pause time or the alarm tone pause time, follow this procedure:
1.
Select the Main Menu quick key → from the System column select Maintenance
→ input the required password → select .
2.
Select the Alarm tab → Pause/Reset tab.
3.
Switch off Pause 5 min , Pause 10 min , or Pause 15 min .
17.5.2.5
Setting Alarm Light Status on Alarm Reset
When the alarm system is reset, the monitor presents the alarm light by default, but you can switch off the alarm light. To change the alarm light setting, follow this procedure:
1.
Select the Main Menu quick key → from the System column select Maintenance
→ input the required password → select .
2.
Select the Alarm tab
→
Pause/Reset tab.
3.
Set Alarm Light .
◆ On When Reset : when the alarm system is reset, the alarm tones of the current alarms are switched off, but the alarm lamp remains flashing.
◆ Off When Reset : when the alarm system is reset, both the alarm tone and alarm lamp of the current alarms are switched off.
17 - 10
VS 9 Vital Signs Monitor Operator’s Manual
advertisement
Related manuals
advertisement
Table of contents
- 153 17 Password-protected User Settings
- 153 17.1 Setting the Device Location
- 153 17.1.1 Setting Monitor Information
- 153 17.1.2 Setting Monitor Location
- 153 17.1.3 Enabling the Auto Obtain Bed Number Function
- 154 17.2 Changing Patient Management Settings
- 154 17.2.1 Selecting Displayed Patient Information
- 154 17.2.2 Selecting Query Criteria for Searching Patients in ADT Server
- 155 17.2.3 Auto Discharging a Patient after Monitor Power Off
- 155 17.2.4 Auto Deleting Patient Data after Discharging a Patient
- 155 17.2.5 Clearing All Patient Data
- 156 17.2.6 Setting Patient Location Options for the Standby Screen
- 156 17.3 Changing the Authorization Settings
- 156 17.3.1 Setting Clinician Login
- 156 17.3.2 Selecting Password for User Maintenance
- 157 17.3.3 Selecting Password for Changing Alarm settings
- 157 17.3.4 Selecting Password for Viewing Discharged Patient Data
- 158 17.4 Changing Settings about Saving and Sending Data
- 158 17.4.1 Setting Items about Saving Data
- 159 17.4.2 Setting Items about Sending Data
- 160 17.5 Changing the Alarm Settings
- 160 17.5.1 Setting Alarm Tone Properties
- 161 17.5.2 Setting Alarm Pausing and Resetting Properties
- 163 17.5.3 Latching Physiological Alarms
- 163 17.5.4 Selecting Alarms to be Sent to the Nurse Call System
- 164 17.5.5 Changing Alarm Priority Settings
- 165 17.5.6 Setting the Alarm Delay Time
- 166 17.6 Changing the Standby Settings
- 166 17.6.1 Setting Prompt Message on Standby Screen
- 166 17.6.2 Setting Auto Standby
- 166 17.6.3 Enabling Black Standby Screen
- 167 17.7 Changing CAA Settings
- 167 17.7.1 Selecting the Default Scoring Tool
- 167 17.7.2 Managing the Scoring Tools
- 167 17.7.3 Setting GCS Threshold for Each Consciousness Level
- 168 17.8 Changing Module Settings
- 168 17.8.1 Selecting NIBP Measurement Algorithm
- 168 Module
- 169 17.8.4 Selecting Temp Module
- 169 17.8.5 Setting Temp Measurement Sites
- 169 17.8.6 Adding a New Manual Parameter
- 170 17.8.7 Managing Pain Score
- 171 17.8.8 Setting Parameter Measurement Timeout
- 171 17.8.9 Enabling Outline Font for Suspected Values
- 171 17.9 Setting Parameter Units
- 208 21.3 Temp Accessories
- 208 21.3.1 SmarTemp Temp Probes Available for Purchase
- 208 21.3.2 SmarTemp Temp Probe Wells Available for Purchase
- 208 21.3.3 SmarTemp Probe Covers Available for Purchase
- 208 21.3.4 Exergen Temp Probe Available for Purchase
- 209 Accessories
- 209 Accessory Available for Purchase
- 209 Accessories Also Compatible