17.7 Changing CAA Settings. Mindray VS9 Operator's Manual

Add to My manuals
276 Pages

advertisement

17.7 Changing CAA Settings. Mindray VS9 Operator's Manual | Manualzz

1.

Select the Main Menu quick key → from the System column select Maintenance

→ input the required password → select .

2.

Select the Standby tab.

3.

Switch on Black Screen On Standby .

17.7

Changing CAA Settings

17.7.1

Selecting the Default Scoring Tool

To set the default scoring tool for different patient category, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password

select .

2.

Select the CAA tab. The EWS tab is displayed.

3.

From the Select Default Score area, set Default Adult Score , Default Ped Score , and Default Neo Score .

17.7.2

Managing the Scoring Tools

To manage the scoring tool, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password

select .

2.

Select the CAA tab. The EWS tab is displayed.

3.

Select the Manage Score button.

◆ Delete local scoring tools: from the Local page, delete any unnecessary scoring tools.

◆ Import the desired scoring tools to the monitor: from the USB drive page, select the scoring tools on the USB drive, and then select Import .

NOTE

The monitor provides MEWS and NEWS by default. You cannot delete them.

17.7.3

Setting GCS Threshold for Each Consciousness Level

You can configure the threshold and color of each consciousness level. To do so, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance

→ input the required password → select .

2.

Select the CAA tab

GCS tab.

3.

Set high limit, low limit and color for each level.

VS 9 Vital Signs Monitor Operator’s Manual

17 - 15