MAINTENANCE. BHM Medical v4

Below you will find brief information for Patient Lift V4 V4. This manual covers installation, operation, and maintenance of the BHM Medical V4 patient lift, a ceiling-mounted lift with a 200kg capacity. It includes safety instructions, descriptions of parts, operation procedures, emergency lowering features, battery information, maintenance checklists, and troubleshooting tips. Learn how to use the FIDO function for pre-programmed positions and how to program the lift settings. The manual also details sling inspection and care.

PDF
Document
BHM Medical V4 Patient Lift User Manual | Manualzz

Advertisement

Advertisement

/

Advertisement

Key features

  • High torque motor

  • Electronic soft-start and soft-stop control

  • Emergency lowering device

  • Emergency stop device

  • Current limiter for circuit protection

  • Adjustable lift speed

  • Emergency brake

  • Strap withstanding 910 kg load

Frequently asked questions

The batteries need to be charged for a minimum of 8 hours before using the lift for the first time. Connect the charger and leave it connected for at least 8 hours.

Pull the red cord gently once. You will hear a click and the green light will turn off. To reactivate the lift, pull the red cord gently once again.

The lifting capacity is 200 kg (440 lb).

Completely unwind the strap and look for any signs of wear, frays, loose threads or other damage. If there is any evidence of damage, do not use the lift. Contact your local representative to have the lift serviced.

The sling should be inspected daily for tears, frayed straps or loose stitches. The manufacturer recommends checking the sling every 2 months for signs of wear.

The V4 and its accessories must be inspected annually by a certified technician. An annual load test must be performed as required for CE MARK MAINTENANCE.