Daily Check List
THE FOLLOWING PROCEDURES MUST
BE FOLLOWED BEFORE EACH USE:
• Have batteries been charged? Park the lift on the charger whenever the lift is not in use.
• Inspect the lift for any damage. If the lift casing does not look properly aligned, or there are any cracks or other damage on the lift, or any parts are missing – DO NOT USE
IT. Contact your local representative to have the lift serviced.
• Inspect the visible strap for any signs of wear, frays, loose threads or other damage.
If there is any evidence of damage – DO
NOT USE IT. Contact your local representative to have the lift serviced.
• Inspect the sling for tears, frayed straps or loose stitches. If the sling has any of the above damage – DO NOT USE IT. Contact your local representative to have the sling replaced or repaired.
• Inspect the spreader bar for any signs of the cracking or damage.
• Ensure the ring and cotter pins that attach the spreader bar to the strap are secure.
Inspection and Cleaning
Clean the V4 with a soft dry cloth, or a soft cloth lightly moistened with a mild detergent solution.
Do not use any type of solvent that may damage the finish. Do not use cleaning products that contain phenol or chlorine.
Do not immerse lift in water.
To ensure a better rolling surface for the trolley wheels, clean the inside of the track every 4 months. To do so, insert a damp cloth in the opening and slide it from one end of the track to the other.
Always reinstall the rail end stopper (if it has been removed) after servicing.
OPERATION
Strap Inspection
This equipment is built with a strap that can withstand a load of 910 kg (2000 lb.)
Nonetheless, if this strap is damaged or shows signs of wear, the acceptable load on the strap before rupture can drop rapidly and present a danger for the patient or caregiver.
BHM Medical recommends thoroughly inspecting the strap every 2 months as follows:
1. Completely unwind the strap.
2. Look for any signs of wear:
Loose threads in stitched area.
Noticeable discoloration by having a
Lighter color strap than the double thickness stitched area.
Side wear.
Middle wear.
If there is any sign of wear as indicated here or other visual defects, strap should be changed immediately. By continuing to use the lift without changing the strap, caregiver and patient safety is greatly compromised.
In any case, the manufacturer recommends changing the strap at least every two years. By continuing to use the lift without changing the strap, caregiver and patient safety is greatly compromised.
Take note of inspection results in the logbook at the end of this manual.
Handling and Storage
Avoid violent impact during transportation.
The lift should not remain stored for more than 2 months without recharging the batteries.
If you store or ship the V4 lift, turn the power off
(green light off).
.
The manufacturer recommends charging the batteries at least every two weeks even if the lift is not used. This will prevent premature aging of batteries