Identifying The Power System
Power
Output 1
Power
Output 2
LED
Power
Indicator
Fuel Gauge
Monitor Connect
Top view
Fig 13: Power Output Port Connectors
Power
Output
3
Power
Output 4
Power
Management
Software
Connect
(USB Type B)
November 2006
1110227 Revision 2
11
Chapter 3: 35 Amp-Hour Series MC
2
Power Systems User's Guide
Right-hand view
Cooling Fan
Exhaust
Fig 14: Fan and Power Output Disconnect Switch
Power Output Disconnect Switch:
↑ ON
↓ OFF
Rear view
Fig 15: AC Power Input
120V 60Hz AC
Power Input
1110227 Revision 2
MC
2
Power Systems User's Guide
Preparing For Operation
Parts Listing
•
(1) Power system unit (Fig 16)
•
(1) Cord Set (Fig 17 and 18)
Chapter 3: 35 Amp-Hour Series
Fig 16: MC2 Power System Series 35 Amp-Hour
1 Unpack components carefully, inspecting for damage. WARNING: Power System is
HEAVY!
2 Confirm that the Power Output Disconnect Switch is in the ‘ON’ position. For more information, see “
.” Verify that the green LED power indicators light up beside the outputs.
Note:
The number of LEDs to light up will vary based upon the number of outputs purchased.
3 Charge your MC² Power System using a Stinger Medical supplied cord set (Fig 17 and 18) for 24 hours prior to use. For more information, see “
.”
Fig 17: Curly cord Fig 18: Cord reel
November 2006
1110227 Revision 2
13
Chapter 3: 35 Amp-Hour Series MC
2
Power Systems User's Guide
4
Confirm the ‘Charging’ indicator on the fuel gauge is ON (see “ Identifying Fuel Gauge
”).
5 Your Power System is ready for use when the ‘Charged’ indicator is ON.
1110227 Revision 2
MC
2
Power Systems User's Guide Chapter 3: 35 Amp-Hour Series
User-Selectable Switch Settings
Voltage Settings Orientation
• The diagram below illustrates each switch location and which output it controls.
• The output voltage can be set by user between 12–20V on Power Output 1 and
Power Output 3 and between 3.3 – 12V on Power Output 2 and Power Output 4.
• All outputs have a maximum continuous current of 4 amps.
• User-selectable switches are located on the top and left hand side of the power system enclosure.
• Power outlets are intended for workstation mounted equipment ONLY.
Power Output 1 and 3 Power Settings Power Output 2 and 4 Power Settings
Voltage Settings for Power
Output 1 and Power Output 2
Left hand view
November 2006
1110227 Revision 2
15
Chapter 3: 35 Amp-Hour Series MC
2
Power Systems User's Guide
Voltage Settings for Power Output 3 and Power Output 4 (optional)
Top view
1110227 Revision 2
Letöltés