Chapter 5: Charging the Power System. Stinger Medical MC2 26 Amp-Hour, MC2 35 Amp-Hour, MC2, MC2 POWER SYSTEMS
Advertentie
Advertentie
MC
2
Power Systems User's Guide Chapter 5: Charging the Power System
Chapter 5: Charging the Power System
Note:
Where the integrity of the external PROTECTIVE EARTH CONDUCTOR arrangement is in doubt, EQUIPMENT shall be operated from its INTERAL
ELECTRICAL POWER SOURCE.
Power System Charging
In order to charge the power system, plug it in to a 120V 60Hz AC power source using a
Stinger Medical supplied cord set.
Note:
The MC
2
Power System Series 35 Amp-Hour unit is required to have the Power
Output Disconnect Switch in the ON position. For more information, see the
“ Power Disconnect Output Switch Location
” illustration.
Charge Time
Charge times vary depending on the usage of the power system. Typically, as a power system is used, the capacity of the power system diminishes.
Charge time Capacity of charge
6 hours with no load 100%
Optimal power system life occurs when unit is fully charged every cycle. For more information, see “
Identifying Fuel Gauge Indicators .”
Power System Discharging
The power system begins discharging when it is disconnected from the AC power source and powering a peripheral device(s). After the power system reaches its maximum runtime potential, it will disable power to its outputs. The MC² Power System Advanced
Monitoring System will communicate to the host computer and perform an orderly shutdown of the Microsoft
®
Windows
®
Operating System unless the user plugs in the power system for charging (Fig 19).
Fig 19: Low battery indicator
November 2006
1110227 Revision 2
19
Chapter 5: Charging the Power System MC
2
Power Systems User's Guide
Circuit Protection
The MC² Power System is thermally protected and self-resetting. For more information,
see “ No Output Reset Procedure
.” The cooling fan maintains a constant ambient temperature inside the electrical enclosure.
Deep Discharge Protection
Deep discharge of the power system will significantly reduce its cycle life. After the power to the load is disabled, the internal circuitry goes into suspend mode to prevent the power system from reaching a level of deep discharge. To reduce this risk, the MC²
Power System will shut down and cut off the output voltage.
1110227 Revision 2