Chapter 2. Instrument Setup

2.6 Using the Calibration Reminder Alarm (PROBE & CAL - CAL REMINDER)

The DMS Go incorporates a timed alarm feature that causes an icon to appear on user-defined input intervals. To use this alarm:

1.

In the

MAIN menu, scroll to the PROBE & CAL submenu using the joystick (

).

2.

Use the joystick ( ) to select the function titled CAL

REMINDER. The CALIBRATION REMINDER CONFIG window opens. Input the mode, or the parameter that will trigger the reminder (

TIME IN MINUTES, NUMBER OF READINGS, POWER ON, and OFF) and then the value the reminder alarm will trigger. Setting this value automatically resets the alarm to trigger on the specified intervals. This function also allows the alarm to be disabled (

OFF).

3.

Press the function toggle or move the joystick ( ) to change the alarm interval.

4.

Press F2 (

CONFIRM) to acknowledge the triggered alarm and reset it to resume normal operation or F 4 (EXIT) to close the window without changing the alarm setting.

Note: The selection of the VELOCITY function also resets the alarm.

70 DMS Go Operator’s Manual

Chapter 3. Creating Data Recorder Files

Chapter 3. Creating Data Recorder Files

The on-board data recorder option of the DMS Go is a flexible, powerful tool for managing ultrasonic thickness data. It can store, evaluate, display, and report on various types of measurement data. The Data Recorder (

DR) menu functions are shown in Figure 22 below.

DMS Go Operator’s Manual

Figure 22: The DR Menu

65