Roche cobas p 612 LCP1

Manuals and User Guides for Roche cobas p 612 LCP1. We found 72 manuals for free downloads User Guide, User manual

Roche cobas p 612 LCP1 is a compact, high-performance pre-analytical system designed to automate the primary sample processing steps in clinical laboratories. It offers a wide range of capabilities to streamline sample processing, improve efficiency, and ensure the quality of laboratory results.

Key Features

  • Fully automated primary tube sampling
  • Sample type identification using advanced imaging technology
  • Precise aliquoting and recapping of samples
  • Integration with laboratory information systems for seamless workflow
  • Comprehensive monitoring and reporting capabilities
  • User-friendly interface for ease of operation

Other manuals and documents

User Guide

Languages: English

User manual

Languages: English

User manual

Languages: English

User Guide

Languages: English

User Guide

Languages: English

User Guide

Languages: English

User manual

Languages: English

User manual

Languages: English

User manual

Languages: English

User Guide

Languages: English, Japanese, Chinese

User Guide

Languages: Portuguese

User Guide

Languages: Italian

Frequently Answers and Questions

What types of samples can the cobas p 612 LCP1 process?
The system can process a wide range of sample types, including serum, plasma, urine, and cerebrospinal fluid.
How does the system identify sample types?
The system uses advanced imaging technology to identify sample types based on their shape, size, and color.
Can the system aliquot samples into multiple tubes?
Yes, the system can aliquot samples into up to 10 different tubes.
How does the system ensure the accuracy of sample aliquoting?
The system uses precise pipetting technology to ensure accurate and consistent aliquoting.
Can the system integrate with laboratory information systems?
Yes, the system can integrate with laboratory information systems using standard interfaces.
How does the system monitor its performance?
The system continuously monitors its performance and provides comprehensive reporting capabilities.

Questions & Answers

2024-06-27

S S G

How to resolve the CCM motor error?
Check motor temperature, allow it to cool down, check motor control board and motor, or contact a service technician to check motor control board and the motor settings.
What should I do if the error is related to reaching the defined maximum position?
Use the log file to determine which position should be reached. If it is correct, adapt to the maximum position of the corresponding motor/parameter. Press the Repeat button or contact a service technician if the error persists.
When should I contact a service technician for the CCM motor error?
Contact a service technician if the error persists or occurs repeatedly, as some errors may require specialized knowledge and diagnostics to resolve.
2024-06-27

U U D

What are the specific maintenance tasks that need to be performed weekly?
The User Guide provides instructions for weekly maintenance tasks, including cleaning different components of the cobas p 612 pre-analytical system.
2024-06-27

Z Z D

Why should I avoid touching the pipettor pin?
Touching the pipettor pin can introduce contaminants, affect precision and accuracy, and lead to cross-contamination; potentially compromising the integrity of the testing process and accuracy of results.
2024-06-27

Z Z D

What are the reasons not to touch the pipettor pin by hand?
Touching the pipettor pin by hand can introduce contaminants from the skin, cause damage or misalignment, and result in cross-contamination between different samples or reagents.
2024-06-27

U U D

What information is provided in the document?
The document provides detailed information for operators of the cobas p 612 pre-analytical system, including sections on routine operation, maintenance, safety, troubleshooting, configuration, and background information like the measurement principle.
What is included in the guide?
The guide contains chapters on the system description, operation, maintenance, troubleshooting, glossary, and appendix. It also includes important safety information that users must read before operating the system..
What tasks require proper training from Roche Diagnostics?
Operation tasks, maintenance actions, or troubleshooting tasks require proper training from Roche Diagnostics to avoid potential risks.
What topics are covered in the User Guide?
The User Guide covers topics such as the overview of the system hardware, sample flow and processing, software description, steps before routine operation, routine operation workflow, actions during routine operation, steps after routine operation, non-routine operations, maintenance overview, and weekly maintenance tasks.