1. Introduction
To all residents of the European Union
Important environmental information about this product
This symbol on the device or the package indicates that disposal of the device after its lifecycle could harm the environment. Do not dispose of the unit (or batteries) as unsorted municipal waste; it should be taken to a specialized company for recycling. This device should be returned to your distributor or to a local recycling service. Respect the local environmental rules.
If in doubt, contact your local waste disposal authorities.
Thank you for choosing Velleman! Please read the manual thoroughly before bringing this device into service. If the device was damaged in transit, don't install or use it and contact your dealer.
2. Safety Instructions
Keep this device away from rain, moisture, splashing and dripping liquids.
Keep this device away from children and unauthorized users.
Protect this device from shocks and abuse. Avoid brute force when operating the device.
Keep this device away from dust and extreme temperatures.
3. General Guidelines
Refer to the Velleman® Service and Quality Warranty on the last pages of this manual.
This device is battery-operated, using one 3.6 V lithium battery (½ AA).
Familiarise yourself with the functions of the device before actually using it.
All modifications of the device are forbidden for safety reasons. Damage caused by user modifications to the device is not covered by the warranty.
Only use the device for its intended purpose. Using the device in an unauthorised way will void the warranty.
Damage caused by disregard of certain guidelines in this manual is not covered by the warranty and the dealer will not accept responsibility for any ensuing defects or problems.
Do not use outside the technical specifications (see §10).
4. Features
memory for 32 000 readings (16 000 temperature and 16 000 humidity readings) dew point indication status indication
USB interface for setup and data download user-programmable alarm thresholds for temperature and humidity
V. 03 – 12/06/2013 2 ©Velleman nv
DVM171THD analysis software: software in English only, for Windows® XP, Vista,
Windows 7 (32 and 64 bit) multi-mode to start logging long battery life selectable measuring cycle: 2 s, 5 s, 10 s, 30 s, 1 min, 5 min, 10 min, 30 min,
1 h, 2 h, 3 h, 6 h, 12 h, 24 h.
5. Use
1. protective cover
2. USB connector to PC port
3. start button
4. RH and temperature sensors
5. alarm LED (red/yellow)
6. recording LED (green)
7. mounting bracket
1. First install the data logger software on a personal computer, by inserting the enclosed disc into the CD drive and following the instructions on the screen.
2. Remove the protective cover [1] from the data logger, plug the data logger into a free USB slot on the personal computer and start the data logger software.
3. Set up the data logger software and press the Setup button. Refer to the help function of the data logger software (Help > Help topics…) for more information on the use of the software and setup of the data logger.
4. Remove the data logger from the USB slot and place back the protective cover [1].
5. Place the device in a location where recording of ambient temperature and/or relative humidity is required. Refer to the technical specifications for allowed operating temperature range.
V. 03 – 12/06/2013 3 ©Velleman nv
DVM171THD
6. If the data logger is set up to start manually (delayed start), the green LED
[6] will flash two times every 10 seconds. Press the Start button [3] until the green [6] and yellow LED [5] flash once (together). The data logger starts logging and the green LED [6] will flash every 10 seconds (increase this value through the setup procedure to increase battery life).
7. To download the data into the personal computer, start the data logger software, remove the protective cover [1] and plug the data logger into a free
USB slot.
Refer to the help function of the data logger software (Help > Help topics…) for more information on downloading and handling the data from the data logger.
6. Troubleshooting
Both LEDs are off
Logging is not active or battery is low. Start logging or replace battery.
Green REC LED [6], flashes 1 x every 10 seconds
Logging, no alarm condition.
1x
Green REC LED [6], flashes 2 x every 10 seconds
Delayed start mode.
2x
1x
Red ALM LED [5], flashes 1 x every 10 seconds
Logging, humidity lower limit exceeded.
2x
Red ALM LED [5], flashes 2 x every 10 seconds
Logging, humidity upper limit exceeded.
1x
Red ALM LED [5], flashes 1 x every 60 seconds
Low battery. Logging will stop, replace battery and download data.
1x
Yellow ALM LED [5], flashes 1 x every 10 seconds
Logging, temperature lower limit exceeded.
2x
Yellow ALM LED [5], flashes 2 x every 10 seconds
Logging, temperature higher limit exceeded.
1x
Yellow ALM LED [5], flashes 1 x every 60 seconds
Logger memory full. Logging will stop, download data.
V. 03 – 12/06/2013 4 ©Velleman nv
DVM171THD
Notes:
In case of one alarm, the green REC LED [6] and the red or yellow ALM LED
[5] will blink alternately every 10 seconds. When two alarms occur
(e.g. humidity alarm and temperature alarm), only the ALM LED [5] will blink, showing alternately the first and second alarm every 10 seconds.
However, the unit cannot show both the upper and lower alarm. For example, if both the upper and lower humidity limits have been exceeded during one test period, the unit indicates either the upper or the lower alarm. Download the data to your PC to see the details of the alarms.
When the battery is low, all operations are disabled automatically and memory is preserved.
7. Cleaning and Storage
The humidity and temperature sensors [4] are very sensitive. Avoid dust, smoke and other particles to enter the housing. Refer to the paragraph
“Sensor Reconditioning”.
Wipe the data logger regularly with a moist, lint-free cloth. Do not use alcohol or solvents. Do not touch the USB connector.
Do not submerge the data logger in water or any other liquid.
There are no user-serviceable parts.
Contact your dealer for spare parts if necessary.
8. Battery
Warning: Do not puncture batteries or throw them in fire as they may explode. Do not attempt to recharge non-rechargeable batteries. Dispose of batteries in accordance with local regulations. Keep batteries away from children.
To replace the battery, disconnect the device from the computer.
1. Open the battery compartment by using a pointed object (e.g. small screwdriver). Move the pointed object in the direction of the arrow.
2. Gently pull the data logger from the casing.
3. Remove/insert the battery. Only use 3.6 V lithium batteries. Make sure the polarity is as indicated in the casing. When the battery is inserted, the green and yellow LEDs will flash alternately.
V. 03 – 12/06/2013 5 ©Velleman nv
DVM171THD
4. Slide the data logger back in the housing until it snaps into place.
Note: Make sure that you position the housing with the right side upward: the tab [a] on the data logger should fit into the notch [b] in the housing. If you place the housing upside down, pull firmly to remove it and place it back in the correct position.
5. Configure the data logger using the enclosed software.
Note: Leaving the device in the USB port will reduce the battery life.
9. Sensor Reconditioning
Over time, the internal sensor may become compromised as a result of pollutants, chemical vapours, or other environmental conditions. This can lead to inaccurate readings. To recondition the internal sensors, proceed as follows:
1. Heat the data logger to +80 °C (+176 °F), humidity < 5 % for 36 hours.
2. Keep the data logger at +20 to +30 °C (+70 to +90 °F), humidity > 74 % for
48 hours.
10. Technical Specifications
relative humidity relative humidity accuracy overall range
0 % – 20 %
20 % – 40 %
40 % – 60 %
60 % – 80 %
80 % – 100 %
0 – 100 % ± 5.0 %
± 5.0 %
± 3.5 %
± 3.0 %
± 3.5 %
± 5.0 % temperature temperature accuracy overall range -40 °C to 70 °C (-40 °F to 158 °F)
-40 °C to -10 °C (-40 °F to +14 °F) ± 2 °C (± 3.6 °F)
-10 °C to +40 °C (+14 °F to +104 °F) ± 1 °C (± 1.8 °F) dew point
+40 °C to +70 °C (+104 °F to +158 °F) ± 2 °C (± 3.6 °F) overall range -40 °C to 70 °C (-40 °F to 158 °F) dew point accuracy 25 °C, 40 – 100 % RH ± 2.0 °C (± 4.0 °F) logging rate power supply battery life dimensions weight selectable sampling interval, from 2 seconds up to 24 hours, operating temperature -35 °C to 80 °C (-31 °F to 176 °F)
3.6 V lithium ½ AA (order code LS14250, incl.)
± 1 year
101 x 25 x 23 mm
42 g
Use this device with original accessories only. Velleman nv cannot be held responsible in the event of damage or injury resulting from (incorrect) use of this device. For more info concerning this product and the latest version of this manual, please visit our website www.velleman.eu. The information in this manual is subject to change without prior notice.
© COPYRIGHT NOTICE
The copyright to this manual is owned by Velleman nv. All worldwide
rights reserved. No part of this manual may be copied, reproduced, translated or reduced to any electronic medium or otherwise without the prior written consent of the copyright holder.
V. 03 – 12/06/2013 6 ©Velleman nv