KMPi SETUP AND OPERATION

KMPi

7. KMPi Setup and Operation

7.1.

Basic Setup

NOTE:

See Section 12.4 for Sample Applications.

1. Mount KMPi on a flat, horizontal surface with no more than two Pump Heads attached

(see Section 9 for

Pump Head Setup details).

2. Connect both Octopus Cables to the back of the KMPi.

3. Connect up to three pressure transducers to the Pressure Transducer Octopus Cable ports as suitable for application.

4. Connect up to two Automatic Backpressure Valves (ABV’s) to the Auxiliary Component Octopus Cable’s

“Valve 1” and/or “Valve 2” ports, then connect ABV power cable(s)

(see Section 10.4 for ABV Setup

details).

5. Connect power cable to the KMPi.

6. Follow guidelines and diagrams below to determine which Auxiliary Components are required to operate specific Pump Modes for manual, semi-automated, and automated processes

(see Section 8 for Pump

Mode Setup details).

i.

If using Auxiliary Pump, configure Auxiliary Pump before starting application

(see Section 10.2

for Pump Setup details).

ii. After connecting Auxiliary Components, power on KMPi first before powering on Auxiliary

Components.

7. Connect TFF flowpath to KMPi.

8. Input Concentration Factor/Diafiltration Volume (CF/DV) set-points into KMPis Pump Mode settings to start application.

Pump Mode

Auxiliary Components Required for Pump Modes Tubing

Feed Scale Permeate Scale Diafiltration Pump

(see Section 10.1 for Scale

Setup details)

(see Section 10.1 for

Scale Setup details)

(see Section 10.2 for Pump

Setup details)

Permeate Pump

(see Section 10.2 for Pump

Setup details)

M Mode

C Mode

X*

C/D Mode

X X X X**

C/D/C Mode

C/D/D/C Mode

X

X

X

X

X

X

X**

X

* The Feed Scale is only required in applications using C Mode if user does not want to input starting

Feed Scale values

** The Permeate Pump is only required in applications that necessitate it, such as MF applications

9

KMPi TFF System • PRODUCT INFORMATION AND OPERATING INSTRUCTIONS

advertisement

Key Features

  • Integrated digital pressure monitor
  • Easy-Load Pump Head
  • Multi-language interface
  • Real-time data collection software
  • Automated shut-off controls
  • Flow rates from 0.01-13.0 LPM

Related manuals

Frequently Answers and Questions

What is the maximum flow rate of the KMPi TFF System?
The KMPi TFF System provides flow rates from 0.01-13.0 LPM.
What types of pump heads are compatible with the KMPi TFF System?
The KMPi TFF System can accommodate the KMPi Easy-Load Pump Head and the High-Performance Pump Head.
What is the purpose of the KFComm software?
KFComm software enables real-time data collection for the KMPi TFF System.
What is the range of processes suitable for the KMPi TFF System?
The KMPi TFF System is ideal for processes ranging from 10-1000L.
What is the maximum number of pump heads that can be attached to the KMPi TFF System?
The KMPi TFF System can mount up to two 13 LPM KMPi Easy-Load Pump Heads or one 17 LPM High-Performance Pump Head.