KMPi TFF System Instructions for Use

Add to My manuals
54 Pages

advertisement

KMPi TFF System Instructions for Use | Manualzz

PRESSURE CALIBRATE

7.7.

Pressure Calibrate

Pressure Cal

Feed

Pf: 4.84 Psi

Set: 0 5 .00 Psi

Clear

Cal

Quit

KMPi

1.

Ensure proper connection between pressure transducers and Octopus Cable ports, then tare pressure readings on KMPi

2.

Turn on positive air source from a calibrated pressure source and adjust to dispense 5 psi / 0.35

bar of pressure, then turn positive air source off

3.

Connect Feed pressure transducer to be in-line with positive air source—cap the remaining open end of the pressure transducer

4.

Turn on positive air source

5.

Wait for Feed pressure reading to stabilize

6.

If Feed pressure reading differs from calibrated source measurement, select Pressure Calibrate from Settings Menu

7.

Select source of pressure (in this case, “Feed”), then adjust “Set:” pressure reading to reflect that of calibrated source

8.

Save changes by selecting “Cal”

9.

Repeat steps 3 to 8 with the Retentate and Permeate ports

16

KMPi TFF System • PRODUCT INFORMATION AND OPERATING INSTRUCTIONS

advertisement

Key Features

  • Integrated digital pressure monitor
  • Easy-Load Pump Head
  • Multi-language interface
  • Real-time data collection software
  • Automated shut-off controls
  • Flow rates from 0.01-13.0 LPM

Related manuals

Frequently Answers and Questions

What is the maximum flow rate of the KMPi TFF System?
The KMPi TFF System provides flow rates from 0.01-13.0 LPM.
What types of pump heads are compatible with the KMPi TFF System?
The KMPi TFF System can accommodate the KMPi Easy-Load Pump Head and the High-Performance Pump Head.
What is the purpose of the KFComm software?
KFComm software enables real-time data collection for the KMPi TFF System.
What is the range of processes suitable for the KMPi TFF System?
The KMPi TFF System is ideal for processes ranging from 10-1000L.
What is the maximum number of pump heads that can be attached to the KMPi TFF System?
The KMPi TFF System can mount up to two 13 LPM KMPi Easy-Load Pump Heads or one 17 LPM High-Performance Pump Head.

advertisement

Table of contents