SYSTEM SETTINGS
KMPi
i.
Pump Mode Status Timeout Display – When Auxiliary Scales are connected and powered on, the user will be able to view the Pump Mode Status Timeout Display. The Timeout Display will show current weight information related to the current Pump Mode if the user highlights the double arrow button and does not touch the pump for seven seconds
(see Section 8.2 for
Timeout Display details).
Spectrum: C
Pf:
Pp:
Pr:
0.0 Psi
0.0 Psi
0.0 Psi
TMP: 0.0 Psi
Feed: 585 ml/min
D
F
Pump Mode Resume Icon
Diafiltration Pump Control Icon
Permeate Pump Control Icon
7.6.3. System Settings
a.
Pump Mode – Depending on user’s Auxiliary Components, user may select one of many Pump
Modes for KMPi to operate in that allow for manual, semi-automated, and automated processes
b.
Pressure Calibrate – Using a calibrated pressure source, the user can calibrate pressure transducer readings
Spectrum: CDDC
F:1.00
Pf:
Pp:
0.0 Psi
0.0 Psi
Pr: 0.0 Psi
2.00
TMP: 0.0 Psi
Conc
Feed: 585 ml/min
c.
Alarm and Stops – Safety pressure alarm and stops
d.
Tubing Calibrate – After running the tubing in the pump for ten minutes, the user may calibrate the tubing to ensure accurate fluid dispensing
e.
Aux Pumps – KMPi’s Aux Pump settings must be configured prior to using the Auxiliary Pumps to ensure the KMPi sends the correct commands to each pump
f.
Hardware Setup – Time, date, unit of measure, and some Pump Mode options may be set from the Hardware Setup menu
g.
Diagnostics – Diagnostics are for the factory use only, but may viewed by the user
14
KMPi TFF System • PRODUCT INFORMATION AND OPERATING INSTRUCTIONS
advertisement
Related manuals
advertisement
Table of contents
- 6 1. Introduction
- 7 2. Safety Precautions
- 9 3. Specifications
- 11 4. System Configuration and Major Components
- 11 4.1. KMPi Pump Drive, Pump Head, & Integrated Pressure Monitor
- 11 4.2. KMPi Mounting Hardware
- 11 4.3. KMPi Octopus Cables
- 12 5. System Parts List
- 13 6. Materials of Construction
- 14 7. KMPi Setup and Operation
- 14 7.1. Basic Setup
- 15 7.2. Manual Mode Setup
- 16 7.3. C Mode Setup
- 16 7.4. C/D, C/D/C, CFC Mode Setup
- 17 7.5. C/D/D/C Mode Setup
- 17 7.6. Home Screen
- 18 7.6.1. Keypad
- 18 7.6.2. Main Screen
- 19 7.6.3. System Settings
- 20 7.6.4. Hardware Setup
- 21 7.7. Pressure Calibrate
- 22 7.8. Pressure Alarm Settings
- 22 7.9. Tubing Calibration
- 23 7.10. Firmware Update
- 24 8. Pump Modes
- 25 8.1. Running the KMPi System
- 25 8.2. Pump Mode Status Timeout Display
- 26 9. KMPi Pump Head Setup and Operational
- 26 9.1. Introduction
- 26 9.2. Specifications
- 27 9.3. Installation and Removal
- 29 9.4. Tubing Loading and Unloading
- 30 9.5. Occlusion Adjustment on Easy-Load Pump Head
- 31 9.6. Multi-Channel Mounting
- 31 9.7. Maintenance and Cleaning
- 32 10. Auxiliary Component Setup and Operation
- 32 10.1. Auxiliary Scales
- 32 10.1.1. Installation
- 32 10.1.2. Controls, Indicators and Connectors
- 33 10.2. Auxiliary Pumps
- 34 10.2.1. Installation
- 34 10.2.2. Controls, Indicators and Connectors
- 35 10.2.3. Auxiliary Pump Settings
- 35 10.2.4. Auxiliary Pump Control Icons
- 36 10.3. Automatic Backpressure Valve
- 37 10.3.1. Installation
- 37 10.3.2. Operation
- 37 10.3.3. Updating Automatic Backpressure Valve
- 39 11. KFComm Software
- 39 11.1. KFComm License
- 40 11.2. Installation
- 40 11.2.1. KFComm Software Suite
- 41 11.2.2. USB Connection
- 41 11.3. Instructions for Use
- 41 11.3.1. Opening and Configuring the KFComm Workbook Template
- 42 11.3.2. Data Collection Tools
- 43 11.3.3. Pump Control
- 44 11.3.4. Worksheets Overview
- 45 11.3.5. Shear Lookup Tool
- 45 12. Basic Concepts of Tangential Flow Filtration
- 46 12.1. Introduction
- 46 12.2. Concentration
- 46 12.3. Diafiltration
- 52 12.4. Sample Applications
- 52 12.4.1. Equations
- 52 12.4.2. Concentration Mode