PRESSURE ALARM SETTINGS
KMPi
7.8.
Pressure Alarm Settings
The KMPi has a number of safety pressure alarms and stops that may be set in the Alarm Settings menu.
a.
P f
Hi Stop – When P f value ≥ P f
Hi Stop value, the KMPi pump will stop running
b.
P f
Hi Alarm – When P ning f value ≥ P f
Hi Alarm value, the KMPi pump will alarm but continue run-
c.
P f
Lo Alarm – When P ning f value ≤ P f
Lo Alarm value, the KMPi pump will alarm but continue run-
d.
P p
Lo Alarm – When P ning p value ≤ P p
Lo Alarm value, the KMPi pump will alarm but continue run-
e.
P p
Lo Stop – When P p value ≤ P p
Lo Stop value, the KMPi pump will stop running
f.
Silent Alarm – Set to “On” to silence alarms, set to “Off” to play alarm sound
1.
To enable an alarm, highlight the alarm value and press Enter.
2.
Use the directional keys to set the desired alarm value, then press Enter to save the setting.
3.
To disable an alarm, change the value to 0 and KMPi will read the alarm as “OFF.”
7.9. Tubing Calibration
NOTE:
Condition tubing by running at ½ of maximum flow rate of tubing being calibrated for at least 10-15 minutes before conducting tubing calibration.
1.
Select tubing size in Calibration menu.
2.
Press CAL, the system set calibration volume will appear.
3.
Press START / STOP, the pump will use its stored memory to dispense the specified calibration sample quantity. The pump will stop automatically.
4.
Weigh/measure the sample.
5.
Use directional keys to correct the volume on the flashing display.
6.
Press the Enter key to save the calibration setting. Once calibrated, a lower case "c" appears next to the tubing (e.g. 73c).
7.
To confirm tubing calibration, press CHECK to dispense calibration volume using calibrated tubing and weigh dispensed volume.
NOTE:
If the adjusted calibration is high, "Err" will appear in the display. If this occurs, press the CAL control and repeat the calibration procedure. The microprocessor will retain one special calibration value per tubing size, even when power is turned off. The next calibration will replace the existing value.
17
KMPi TFF System • PRODUCT INFORMATION AND OPERATING INSTRUCTIONS