Metrohm 900 Touch Control Owner's Manual

Add to My manuals
553 Pages

advertisement

Metrohm 900 Touch Control Owner's Manual | Manualzz

900 Touch Control

Manual

8.900.8015EN / 2021-03-26

Metrohm AG

CH-9100 Herisau

Switzerland

Phone +41 71 353 85 85

Fax +41 71 353 89 01 [email protected]

www.metrohm.com

900 Touch Control

Program version 5.900.0043

Manual

8.900.8015EN / 2021-03-26

Technical Communication

Metrohm AG

CH-9100 Herisau [email protected]

This documentation is protected by copyright. All rights reserved.

This documentation has been prepared with great care. However, errors can never be entirely ruled out. Please send comments regarding possible errors to the address above.

Documentation in additional languages can be found on http://documents.metrohm.com

.

Melody for the BEEP command: excerpt from "En Altfrentsche", with kind permission of the Laseyer Quartett, Appenzell.

■■■■■■■■■■■■■■■■■■■■■■ Table of contents

Table of contents

1 Introduction

1

1.1

Instrument description ......................................................... 1

1.2

Titration and measuring modes .......................................... 1

1.3

About the documentation ................................................... 3

1.3.1

Symbols and conventions ........................................................ 3

2 Safety instructions

5

2.1

General notes on safety ....................................................... 5

2.2

Electrical safety ..................................................................... 5

2.3

Tubing and capillary connections ........................................ 6

2.4

Flammable solvents and chemicals ..................................... 6

2.5

Recycling and disposal ......................................................... 7

3 Overview of the instrument

8

4 Installation 10

4.1

Setting up the instrument .................................................. 10

4.1.1

Packaging .............................................................................. 10

4.1.2

Checks .................................................................................. 10

4.1.3

Location ................................................................................ 10

5 Titrations 11

5.1

Dynamic equivalence point titration (DET) ....................... 11

5.2

Monotonic equivalence point titration (MET) .................. 11

5.3

Endpoint titration (SET) ...................................................... 12

5.4

Water determination according to Karl Fischer (KFT) ..... 12

5.5

Titration with constant maintenance of the measured

value (STAT) ......................................................................... 13

5.6

Manual titration (MAT) ...................................................... 14

5.7

Coulometric water content determination according to

Karl Fischer (KFC) ................................................................ 15

5.8

Bromine index determination (BRC) .................................. 15

6 Operation 16

6.1

Switching the instrument on and off ................................ 16

6.2

Fundamentals of operation ............................................... 17

6.2.1

Touch-sensitive screen ........................................................... 17

900 Touch Control ■■■■■■■■ III

Table of contents ■■■■■■■■■■■■■■■■■■■■■■

6.2.2

Display elements and controls ................................................ 18

6.2.3

Status display ......................................................................... 19

6.2.4

Entering text and numbers ..................................................... 20

7 System settings 22

7.1

General system settings ..................................................... 22

7.1.1

Selecting the dialog language ................................................ 22

7.1.2

Setting the date, time and local time ..................................... 23

7.2

System-specific dialog options .......................................... 24

7.3

User administration ............................................................ 28

7.3.1

Editing the user configuration ................................................ 29

7.3.2

Creating an identification profile ............................................ 33

7.3.3

Defining login options ........................................................... 33

7.3.4

Password options .................................................................. 35

7.3.5

Modification options ............................................................. 37

7.3.6

Reasons ................................................................................. 38

7.3.7

Audit Trail .............................................................................. 39

7.4

Measured value display ...................................................... 40

7.5

Acoustic signals .................................................................. 40

8 Titrants 41

8.1

Adding a new titrant .......................................................... 42

8.2

Editing titrant data ............................................................. 43

8.3

Monitoring the working life .............................................. 45

8.4

Dosing unit .......................................................................... 46

8.4.1

Parameters for preparing (PREP) and emptying (EMPTY) ......... 47

8.4.2

Tubing parameters ................................................................. 49

8.4.3

Shift direction of the valve disk .............................................. 52

8.5

Exchange unit ..................................................................... 52

8.5.1

Parameters for the preparation (PREP) .................................... 54

8.5.2

Tubing parameters ................................................................. 55

8.6

GLP test for exchange unit and dosing unit .................... 56

8.7

Titer determination options and data .............................. 58

8.7.1

Titer validity ........................................................................... 58

8.7.2

Properties of the previous titer determinations ....................... 59

9 Reagents 61

9.1

Editing reagent data .......................................................... 62

9.2

Reagent monitoring ........................................................... 62

10 Sensors 66

10.1 Adding a new sensor ......................................................... 67

10.2 Editing the sensor data ...................................................... 68

IV ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ Table of contents

10.3 Monitoring the working life .............................................. 69

10.4 Calibration data (for pH- and ISE electrodes and con-

ductivity measuring cells only) .......................................... 70

10.4.1

Properties of the previous calibrations .................................... 72

10.5 Limit values for the calibration data ................................. 75

10.6 Monitoring the calibration interval (only for pH and ion-selective electrodes and conductivity measuring

cells) ..................................................................................... 77

11 Device manager 79

11.1 Adding a new device .......................................................... 80

11.2 Configuring the instrument ............................................... 80

11.3 Touch Control ...................................................................... 81

11.3.1

E-mail .................................................................................... 82

11.3.2

PC/LIMS report ...................................................................... 83

11.3.3

Shared memory ..................................................................... 84

11.3.4

TCP/IP settings ....................................................................... 86

11.4 Metrohm control instruments ........................................... 87

11.4.1

Properties – Control instrument ............................................. 88

11.4.2

Properties – Measuring input ................................................. 88

11.4.3

Properties – MSB connector ................................................... 90

11.4.4

Properties – Peripheral devices ............................................... 91

11.5 Sample Processor ............................................................... 92

11.5.1

Properties – Sample Processor ................................................ 93

11.5.2

Properties – Tower ................................................................. 93

11.5.3

Properties – Swing Head ........................................................ 95

11.6 Sample racks ..................................................................... 100

11.6.1

Editing rack data .................................................................. 102

11.6.2

Rack adjustment .................................................................. 108

11.7 Printer ................................................................................ 109

11.7.1

PDF settings ......................................................................... 110

11.7.2

Network printer ................................................................... 111

11.7.3

More options ....................................................................... 113

11.8 Balance .............................................................................. 113

11.9 USB/RS-232 adapter ......................................................... 115

11.10 PC keyboard ...................................................................... 117

11.11 Barcode reader ................................................................. 118

12 File manager 121

12.1 Managing files .................................................................. 121

12.1.1

Copying a file ...................................................................... 123

12.1.2

Renaming a file ................................................................... 123

12.1.3

File properties ...................................................................... 124

900 Touch Control ■■■■■■■■ V

Table of contents ■■■■■■■■■■■■■■■■■■■■■■

12.2 External storage medium ................................................ 125

12.3 Creating backups / Restoring data ................................. 127

12.3.1

Restoring data ..................................................................... 127

13 GLP manager 129

13.1 Automatic system test ..................................................... 130

13.2 Test tools ........................................................................... 130

13.3 GLP tests for measurement and titration ....................... 131

13.3.1

Parameter description .......................................................... 132

13.4 System validation ............................................................. 135

13.4.1

Parameter description .......................................................... 136

13.5 System monitoring ........................................................... 139

13.5.1

Service interval .................................................................... 139

13.5.2

Backup interval .................................................................... 139

14 Common variables 141

14.1 Editing common variables ............................................... 142

14.2 Properties of common variables ..................................... 143

14.3 Monitoring validity ........................................................... 144

14.4 Assigning a result automatically to a common varia-

ble ...................................................................................... 145

15 Templates 147

15.1 Sample data ...................................................................... 147

15.1.1

Sample identification list ...................................................... 148

15.1.2

Sample assignment table ..................................................... 149

15.2 Custom result templates .................................................. 151

15.2.1

Editing result templates ....................................................... 152

15.3 Input lines ......................................................................... 155

15.3.1

Editing the input signal ........................................................ 156

15.4 Output lines ...................................................................... 157

15.4.1

Editing the output signal ...................................................... 159

15.5 Custom calibration buffers .............................................. 160

15.5.1

Defining calibration buffers .................................................. 161

15.6 Report header ................................................................... 162

15.7 Custom electrode type ..................................................... 163

15.7.1

Limit values for the electrode rating ..................................... 164

16 Methods 167

16.1 Creating a new method ................................................... 167

16.2 Saving a method ............................................................... 168

VI ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ Table of contents

16.3 Loading a method ............................................................ 169

16.4 Editing a method .............................................................. 170

16.4.1

Inserting a command ........................................................... 171

16.5 Method options ................................................................ 172

16.5.1

Start options ........................................................................ 173

16.5.2

Stop options ........................................................................ 175

16.5.3

Sample data ........................................................................ 176

16.5.4

Method properties ............................................................... 180

16.5.5

Note .................................................................................... 182

16.5.6

Saving a determination automatically ................................... 182

17 Electronic signatures 184

17.1 Signing methods/determinations electronically ............ 184

17.2 Deleting electronic signatures ......................................... 185

18 Control 186

19 Favorites 190

19.1 Creating favorites ............................................................. 191

19.1.1

Editing favorites ................................................................... 191

20 Sample data 194

20.1 Entering sample data in the main dialog ....................... 194

20.2 Requesting sample data at the start of the determina-

tion .................................................................................... 195

21 Sample table 197

21.1 General .............................................................................. 197

21.2 Editing sample data ......................................................... 200

21.3 Properties .......................................................................... 202

22 Determination run 205

22.1 Carrying out a single determination ............................... 205

22.2 Performing a sample series ............................................. 206

22.3 Canceling determinations manually ............................... 207

23 Live modifications 208

23.1 Editing the sample data of the running determination 208

23.2 Editing the sample table while a determination is run-

ning .................................................................................... 209

23.3 Live display ....................................................................... 210

23.4 Live parameters ................................................................ 212

900 Touch Control ■■■■■■■■ VII

Table of contents ■■■■■■■■■■■■■■■■■■■■■■

24 Results and more determination data 214

24.1 More determination data ................................................ 215

24.1.1

Details ................................................................................. 216

24.2 Messages ........................................................................... 218

24.3 Local common variables .................................................. 219

24.4 Determination properties ................................................ 219

24.5 Loading a determination ................................................. 222

24.5.1

Determination list ................................................................ 223

24.6 Saving a determination .................................................... 227

24.7 Curves ................................................................................ 227

24.8 Recalculation and reevaluation ....................................... 229

25 Statistics 231

25.1 Displaying details for a result .......................................... 233

25.2 Deleting statistical data ................................................... 234

25.3 Adding a determination to a determination series ....... 234

26 Result table 236

26.1 Properties .......................................................................... 237

26.2 Saving the result table ..................................................... 241

26.3 Loading the result table ................................................... 241

27 Printing 242

27.1 General report options ..................................................... 244

27.2 Settings of the individual reports ................................... 245

27.3 List of all printable reports .............................................. 246

28 Manual control 250

28.1 Opening and closing the manual control ....................... 251

28.2 Measuring ......................................................................... 252

28.2.1

Parameter description .......................................................... 253

28.3 Dosing ............................................................................... 256

28.3.1

Continuous dosing .............................................................. 258

28.3.2

Dosing fixed volumes ........................................................... 260

28.3.3

Preparing ............................................................................. 262

28.3.4

Emptying ............................................................................. 263

28.3.5

Filling .................................................................................. 264

28.3.6

Replacing reagent ................................................................ 264

28.4 Stirring ............................................................................... 265

VIII ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ Table of contents

28.5 Conditioning a titration cell ............................................. 266

28.5.1

Parameter description .......................................................... 267

28.6 Remote .............................................................................. 267

28.7 Sample Processor ............................................................. 268

28.7.1

Moving the lift ..................................................................... 270

28.7.2

Moving to a rack position .................................................... 272

28.7.3

External positions ................................................................ 274

29 Parameters 278

29.1 Dynamic equivalence point titrations (DET) ................... 278

29.1.1

Start conditions ................................................................... 278

29.1.2

Titration parameters ............................................................ 281

29.1.3

Stop conditions ................................................................... 285

29.1.4

Potentiometric evaluation .................................................... 286

29.1.5

Control instrument .............................................................. 291

29.1.6

Sensor ................................................................................. 292

29.1.7

Dosing device ...................................................................... 293

29.1.8

Stirrer .................................................................................. 294

29.2 Monotonic equivalence point titrations (MET) .............. 295

29.2.1

Start conditions ................................................................... 295

29.2.2

Titration parameters ............................................................ 298

29.2.3

Stop conditions ................................................................... 301

29.2.4

Potentiometric evaluation .................................................... 302

29.2.5

Control instrument .............................................................. 307

29.2.6

Sensor ................................................................................. 308

29.2.7

Dosing device ...................................................................... 309

29.2.8

Stirrer .................................................................................. 310

29.3 Endpoint titrations (SET) .................................................. 311

29.3.1

Start conditions ................................................................... 311

29.3.2

Control parameters .............................................................. 312

29.3.3

Titration parameters ............................................................ 316

29.3.4

Stop conditions ................................................................... 317

29.3.5

Conditioning ....................................................................... 318

29.3.6

Cell ...................................................................................... 320

29.3.7

Control instrument .............................................................. 320

29.3.8

Sensor ................................................................................. 320

29.3.9

Dosing device ...................................................................... 322

29.3.10 Stirrer .................................................................................. 323

29.4 Volumetric Karl Fischer titrations (KFT) .......................... 323

29.4.1

Start conditions ................................................................... 323

29.4.2

Control parameters .............................................................. 325

29.4.3

Titration parameters ............................................................ 328

29.4.4

Stop conditions ................................................................... 329

29.4.5

Conditioning ....................................................................... 330

29.4.6

Cell ...................................................................................... 332

29.4.7

Control instrument .............................................................. 332

29.4.8

Sensor ................................................................................. 333

900 Touch Control ■■■■■■■■ IX

X

Table of contents

■■■■■■■■

■■■■■■■■■■■■■■■■■■■■■■

29.4.9

Dosing device ...................................................................... 334

29.4.10 Stirrer .................................................................................. 335

29.5 Coulometric Karl Fischer titrations (KFC) ....................... 336

29.5.1

Control parameters .............................................................. 336

29.5.2

Titration parameters ............................................................ 338

29.5.3

Stop conditions ................................................................... 339

29.5.4

Conditioning ....................................................................... 339

29.5.5

Cell ...................................................................................... 342

29.5.6

Control instrument .............................................................. 343

29.5.7

Sensor ................................................................................. 343

29.5.8

Stirrer .................................................................................. 344

29.6 Bromine index determination (BRC) ............................... 345

29.6.1

Control parameters .............................................................. 345

29.6.2

Titration parameters ............................................................ 348

29.6.3

Stop conditions ................................................................... 349

29.6.4

Conditioning ....................................................................... 349

29.6.5

Cell ...................................................................................... 351

29.6.6

Control instrument .............................................................. 352

29.6.7

Sensor ................................................................................. 352

29.6.8

Stirrer .................................................................................. 353

29.7 STAT titrations (STAT) ...................................................... 354

29.7.1

Start conditions ................................................................... 354

29.7.2

Control parameters .............................................................. 356

29.7.3

Titration parameters ............................................................ 358

29.7.4

Stop conditions ................................................................... 360

29.7.5

Monitoring .......................................................................... 361

29.7.6

Control instrument .............................................................. 368

29.7.7

Sensor ................................................................................. 368

29.7.8

Dosing device ...................................................................... 369

29.7.9

Stirrer .................................................................................. 372

29.8 Manual titrations (MAT) ................................................... 373

29.8.1

Dosing parameters .............................................................. 374

29.8.2

Control instrument .............................................................. 375

29.8.3

Sensor ................................................................................. 375

29.8.4

Dosing device ...................................................................... 376

29.8.5

Stirrer .................................................................................. 379

29.9 Measurements (MEAS) ..................................................... 379

29.9.1

Measuring parameters ......................................................... 379

29.9.2

Standard addition ................................................................ 384

29.9.3

Control instrument .............................................................. 388

29.9.4

Sensor ................................................................................. 388

29.9.5

Stirrer .................................................................................. 390

29.10 Calibration of sensors (CAL) ............................................ 391

29.10.1 Calibration parameters (CAL pH) .......................................... 391

29.10.2 Calibration parameters (CAL Conc) ....................................... 394

29.10.3 Calibration parameters (CAL Cond) ...................................... 397

29.10.4 Control instrument .............................................................. 400

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ Table of contents

29.10.5 Sensor ................................................................................. 400

29.10.6 Stirrer .................................................................................. 401

29.11 Electrode test for pH electrodes (ELT) ............................ 402

29.11.1 Electrode test parameters .................................................... 403

29.11.2 Control instrument .............................................................. 404

29.11.3 Sensor ................................................................................. 405

29.11.4 Stirrer .................................................................................. 406

29.11.5 Additional information ......................................................... 406

29.12 Evaluations (EVAL) ............................................................ 408

29.12.1 Fixed endpoint evaluation (EVAL FIX-EP) ............................... 410

29.12.2 pK value and half neutralization potential evaluation (EVAL pK/HNP) .............................................................................. 412

29.12.3 Minimum and maximum evaluation (EVAL MIN/MAX) .......... 413

29.12.4 Break point evaluation (EVAL BREAK) ................................... 414

29.12.5 Rate evaluation (EVAL RATE) ................................................ 417

29.13 Calculations ....................................................................... 418

29.13.1 Calculations (CALC) .............................................................. 418

29.13.2 Calculations (CALC LIVE) ...................................................... 425

29.13.3 Formula editor ..................................................................... 427

29.14 Reports (REPORT) ............................................................. 433

29.14.1 General report options ......................................................... 433

29.14.2 Settings of the individual reports .......................................... 434

29.14.3 List of reports ...................................................................... 435

29.15 Dosing and Liquid Handling ............................................ 438

29.15.1 Preparing an exchange or dosing unit (PREP) ........................ 438

29.15.2 Emptying a dosing unit (EMPTY) .......................................... 439

29.15.3 Dosing a specified volume (ADD) ......................................... 440

29.15.4 Liquid Handling (LQH) .......................................................... 445

29.15.5 Monitored dosing (DOS) ...................................................... 448

29.16 Communication ................................................................ 465

29.16.1 Scanning remote lines (SCAN) .............................................. 466

29.16.2 Setting remote lines (CTRL) .................................................. 467

29.16.3 Scanning the RS-232 interface (SCAN RS) ............................. 469

29.16.4 Defining RS-232 commands (CONTROL RS) .......................... 470

29.17 Automation ....................................................................... 470

29.17.1 Rotating sample rack (MOVE) .............................................. 470

29.17.2 Moving the lift (LIFT) ............................................................ 473

29.17.3 Controlling pumps (PUMP) ................................................... 474

29.17.4 Resetting the rack (RACK) .................................................... 475

29.17.5 Defining the sample variable (SAMPLE) ................................ 475

29.17.6 Creating a subsequence (SUBSEQ) ....................................... 476

29.18 Miscellaneous commands ................................................ 479

29.18.1 Controlling a stirrer (STIR) .................................................... 479

29.18.2 Pausing the method run (WAIT) ........................................... 480

29.18.3 Scan data (REQUEST) ........................................................... 480

29.18.4 Defining an acoustic signal (BEEP) ........................................ 481

900 Touch Control ■■■■■■■■ XI

Table of contents ■■■■■■■■■■■■■■■■■■■■■■

29.18.5 Signing a determination (SIGN) ............................................ 482

29.18.6 Canceling the method run (END) .......................................... 482

30 Operation and maintenance 483

30.1 System initialization ......................................................... 483

31 Troubleshooting 484

31.1 Editing methods ............................................................... 484

31.2 Sample table ..................................................................... 484

31.3 Results/Statistics ............................................................... 484

31.4 Printing .............................................................................. 485

31.5 Manual control ................................................................. 485

31.6 Miscellaneous ................................................................... 486

31.7 SET titration ...................................................................... 486

31.8 Volumetric Karl Fischer Titration .................................... 487

31.9 Coulometric Karl Fischer titration ................................... 489

31.10 STAT titration .................................................................... 491

32 Appendix 492

32.1 Manual Dosing Controller, optional accessory for man-

ual titration ....................................................................... 492

32.2 Dosing unit ........................................................................ 493

32.2.1

Maximum dosing and filling rate .......................................... 493

32.2.2

Default parameters for preparing (PREP) and emptying

(EMPTY) ............................................................................... 493

32.3 Exchange unit ................................................................... 494

32.3.1

Maximum dosing and filling rate .......................................... 494

32.3.2

Default parameters for preparing (PREP) ............................... 494

32.4 Stirring rate ....................................................................... 494

32.5 Balance .............................................................................. 495

32.6 Result variables as parameter setting ............................ 496

32.7 Electrode calibration with Sample Processors ............... 497

32.8 Stored buffer series for pH calibration ........................... 499

32.8.1

Metrohm ............................................................................. 499

32.8.2

NIST (according to DIN standard 19266, 2015) .................... 500

32.8.3

DIN (according to DIN standard 19267, 2012) ..................... 501

32.8.4

Fisher .................................................................................. 502

32.8.5

Fluka ................................................................................... 503

32.8.6

Mettler Toledo ..................................................................... 504

32.8.7

Beckmann ........................................................................... 505

32.8.8

Radiometer Analytical .......................................................... 506

32.8.9

Baker ................................................................................... 507

XII ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ Table of contents

32.8.10 Hamilton DURACAL ............................................................. 508

32.8.11 Precisa ................................................................................. 509

32.8.12 Merck CertiPUR 20 / Titrisol ................................................. 510

32.8.13 Merck CertiPUR 25 .............................................................. 511

32.9 Using AuditTrailViewer .................................................... 513

32.9.1

Installing AuditTrailViewer .................................................... 513

32.9.2

Opening the Audit Trail ........................................................ 514

32.9.3

Contents of the Audit Trail ................................................... 515

32.9.4

Filtering the Audit Trail ......................................................... 516

32.9.5

Exporting the Audit Trail ...................................................... 517

32.10 Diagnosis ........................................................................... 517

32.10.1 LCD test .............................................................................. 518

32.10.2 Temperature monitoring ...................................................... 519

32.10.3 Formatting an external storage medium ............................... 519

32.10.4 Removing an external storage medium ................................ 519

32.10.5 Adjusting the touch screen .................................................. 519

32.10.6 Testing the touch screen ...................................................... 520

32.10.7 Software update (loading program versions and language files) .................................................................................... 521

32.10.8 Simulating titration curves ................................................... 525

32.10.9 Service ................................................................................. 527

32.11 Arithmetic algorithms in the Titrando ............................ 527

33 Technical specifications 529

33.1 Touch screen ..................................................................... 529

33.2 Interfaces .......................................................................... 530

33.3 Power supply .................................................................... 530

33.4 Declarations of Conformity ............................................. 530

33.5 Ambient temperature ....................................................... 531

33.6 Dimensions ........................................................................ 531

33.7 Storage capacity ............................................................... 531

34 Accessories

Index

532

533

900 Touch Control ■■■■■■■■ XIII

Table of figures ■■■■■■■■■■■■■■■■■■■■■■

Figure 1

Figure 2

Figure 3

Figure 4

Figure 5

Figure 6

Figure 7

Figure 8

Figure 9

Figure 10

Figure 11

Figure 12

Figure 13

Figure 14

Figure 15

Figure 16

Figure 17

Figure 18

Figure 19

Figure 20

Figure 21

Figure 22

Figure 23

Figure 24

Figure 25

Figure 26

Figure 27

Figure 28

Figure 29

Table of figures

Front 900 Touch Control ................................................................... 8

Rear 900 Touch Control ..................................................................... 9

Reagent dosing for DET ................................................................... 11

Reagent dosing for MET .................................................................. 11

Reagent dosing for SET ................................................................... 12

Reagent dosing for KFT ................................................................... 13

Reagent dosing for STAT .................................................................. 14

Dosing unit – port assignment ......................................................... 51

Exchange unit – tubing connections ................................................ 56

Directory structure on the external storage medium ....................... 126

Live display "Preparing the dosing unit" ......................................... 263

Live display "Preparing the exchange unit" ..................................... 263

Equivalence point recognition and equivalence point numbering within windows ....................................................................................... 290

Tubbs method for determining the equivalence point .................... 291

Equivalence point recognition and equivalence point numbering within windows ....................................................................................... 306

Action "Exit method" or "Cancel command" .................................. 365

Action "Pause" .............................................................................. 365

Action "Wait" ................................................................................ 365

Evaluation of pK value / half neutralization potential ...................... 412

Evaluation of minimum and maximum ........................................... 413

Evaluation of a break point ............................................................ 414

Dosing criterion "Volume/Dosing rate" .......................................... 449

Dosing criterion "Volume/Dosing time" .......................................... 449

Dosing criterion "Dosing rate/Dosing time" .................................... 450

Action "Exit method" or "Cancel command" .................................. 458

Action "Pause" .............................................................................. 458

Action "Wait" ................................................................................ 458

Manual Dosing Controller 6.2107.120 – Overview ......................... 492

Rotational speed depending on the stirring rate ............................. 495

XIV ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 1 Introduction

1 Introduction

1.1 Instrument description

The 900 Touch Control is an operating unit with a touch-sensitive screen.

You can operate the following control instruments with the 900 Touch

Control:

Titrandos

Sample Processors

846 Dosing Interface

867 pH Module

856 Conductivity Module

With the 900 Touch Control, you can conveniently manage all of your titrants, sensors, methods, etc. For example, you can connect a USB flash drive to the USB connector as an external storage medium. You can use this additional memory not only to store your methods and determinations, but also to create a backup together with all of the data and settings of your system. Thanks to the integrated Ethernet connector, the device can be connected to your network, thus offering you the following options, among others:

Saving data to a PC within the network

Printing reports on a network printer

Sending displayed messages as e-mails

1.2 Titration and measuring modes

The 900 Touch Control supports the following titration and measuring modes. Whether a particular mode is available or not depends on the type of the control instrument connected.

DET

Dynamic equivalence point titration. The reagent is added in variable volume steps.

Measuring modes:

– pH (pH measurement)

– U (potentiometric voltage measurement)

– Ipol (voltametric measurement with selectable polarization current)

– Upol (amperometric measurement with selectable polarization voltage)

900 Touch Control ■■■■■■■■ 1

1.2 Titration and measuring modes ■■■■■■■■■■■■■■■■■■■■■■

MET

Monotonic equivalence point titration. The reagent is added in constant volume steps.

Measuring modes:

– pH (pH measurement)

– U (potentiometric voltage measurement)

– Ipol (voltametric measurement with selectable polarization current)

– Upol (amperometric measurement with selectable polarization voltage)

SET

Endpoint titration at one or two specified endpoints.

Measuring modes:

– pH (pH measurement)

– U (potentiometric voltage measurement)

– Ipol (voltametric measurement with selectable polarization current)

– Upol (amperometric measurement with selectable polarization voltage)

STAT

Titration with constant maintenance of the measured value.

Measuring modes:

– pH (pH measurement)

– U (potentiometric voltage measurement)

MAT

Manual titration with manual dosing and manual ending.

Measuring modes (optional):

– pH (pH measurement)

– U (potentiometric voltage measurement)

KFT

Volumetric water content determination according to Karl Fischer.

Measuring modes:

– Ipol (voltametric measurement with selectable polarization current)

– Upol (amperometric measurement with selectable polarization voltage)

KFC

Coulometric water content determination according to Karl Fischer.

Measuring mode:

– Ipol (voltametric measurement with selectable polarization current)

2 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 1 Introduction

BRC

Coulometric bromine index determination. Determining the amount of double bonds in e.g. mineral oils.

Measuring mode:

– Ipol (voltametric measurement with selectable polarization current)

MEAS

Measuring modes:

– pH (pH measurement)

– U (potentiometric voltage measurement)

– Ipol (voltametric measurement with selectable polarization current)

– Upol (amperometric measurement with selectable polarization voltage)

– T (temperature measurement)

– Conc (concentration measurement with or without standard addition)

– Cond (conductivity measurement)

CAL

Electrode calibration.

Measuring mode:

– pH (calibration of pH electrodes)

– Conc (calibration of ion-selective electrodes)

– Cond (calibration of conductivity measuring cells)

– ELT (Electrode test for pH electrodes)

1.3 About the documentation

CAUTION

Please read through this documentation carefully before putting the instrument into operation. The documentation contains information and warnings which the user must follow in order to ensure safe operation of the instrument.

1.3.1 Symbols and conventions

The following symbols and formatting may appear in this documentation:

Cross-reference to figure legend

The first number refers to the figure number, the second to the instrument part in the figure.

900 Touch Control ■■■■■■■■ 3

1.3 About the documentation

Method

File ▶ New

[Continue]

■■■■■■■■■■■■■■■■■■■■■■

Instruction step

Perform the steps one after the other.

Dialog text , parameter in the software

Menu or menu item

Button or key

WARNING

This symbol draws attention to a possible life-threatening hazard or risk of injury.

WARNING

This symbol draws attention to a possible hazard due to electrical current.

WARNING

This symbol draws attention to a possible hazard due to heat or hot instrument parts.

WARNING

This symbol draws attention to a possible biological hazard.

CAUTION

This symbol draws attention to possible damage to instruments or instrument parts.

NOTE

This symbol highlights additional information and tips.

4 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 2 Safety instructions

2 Safety instructions

2.1 General notes on safety

WARNING

Operate this instrument only according to the information contained in this documentation.

This instrument left the factory in a flawless state in terms of technical safety. To maintain this state and ensure non-hazardous operation of the instrument, the following instructions must be observed carefully.

2.2 Electrical safety

The electrical safety when working with the instrument is ensured as part of the international standard IEC 61010.

WARNING

Only personnel qualified by Metrohm are authorized to carry out service work on electronic components.

WARNING

Never open the housing of the instrument. The instrument could be damaged by this. There is also a risk of serious injury if live components are touched.

There are no parts inside the housing which can be serviced or replaced by the user.

Supply voltage

WARNING

An incorrect supply voltage can damage the instrument.

Only operate this instrument with a supply voltage specified for it (see rear panel of the instrument).

900 Touch Control ■■■■■■■■ 5

2.3 Tubing and capillary connections ■■■■■■■■■■■■■■■■■■■■■■

Protection against electrostatic charges

WARNING

Electronic components are sensitive to electrostatic charges and can be destroyed by discharges.

Do not fail to pull the power cord out of the power socket before you set up or disconnect electrical plug connections at the rear of the instrument.

2.3 Tubing and capillary connections

CAUTION

Leaks in tubing and capillary connections are a safety risk. Tighten all connections well by hand. Avoid applying excessive force to tubing connections. Damaged tubing ends lead to leakage. Appropriate tools can be used to loosen connections.

Check the connections regularly for leakage. If the instrument is used mainly in unattended operation, then weekly inspections are mandatory.

2.4 Flammable solvents and chemicals

WARNING

All relevant safety measures are to be observed when working with flammable solvents and chemicals.

Set up the instrument in a well-ventilated location (e.g. fume cupboard).

Keep all sources of flame far from the workplace.

Clean up spilled liquids and solids immediately.

Follow the safety instructions of the chemical manufacturer.

6 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 2 Safety instructions

2.5 Recycling and disposal

This product is covered by European Directive 2012/19/EU, WEEE – Waste

Electrical and Electronic Equipment.

The correct disposal of your old instrument will help to prevent negative effects on the environment and public health.

More details about the disposal of your old instrument can be obtained from your local authorities, from waste disposal companies or from your local dealer.

900 Touch Control ■■■■■■■■ 7

3 Overview of the instrument

1

■■■■■■■■■■■■■■■■■■■■■■

2 3 4 5 6

Figure 1 Front 900 Touch Control

1 Display

Touch screen.

3 Fixed key [Back]

Opens the next-higher dialog page.

5 Fixed key [Print]

Opens the print dialog.

7 Fixed key [STOP]

Cancels the running determination.

7 8

2 Fixed key [Home]

Opens the main dialog.

4 Fixed key [Help]

Opens the online help for the dialog displayed.

6 Fixed key [Manual]

Opens the manual control.

8 Fixed key [START]

Starts a determination.

8 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

1 2 3

3 Overview of the instrument

4 5

Figure 2 Rear 900 Touch Control

1 Ethernet connector (RJ-45)

For connecting to a network.

3 Power switch

Switches the instrument on/off.

5 Connection cable

For the connection of the Touch Control to a control device (socket "Controller").

2 USB connector (type A)

For connecting printers, USB sticks, USB hubs, etc.

4 Type plate

Contains the serial number.

900 Touch Control ■■■■■■■■ 9

4.1 Setting up the instrument ■■■■■■■■■■■■■■■■■■■■■■

4 Installation

The connection between the Touch Control and the control instrument

(Titrando, etc.) is described in the respective manual, as is the configuration of the titration system with its peripheral devices, e.g. stirrers and dosing devices.

4.1 Setting up the instrument

4.1.1 Packaging

The instrument is supplied in protective packaging together with the separately packed accessories. Keep this packaging, as only this ensures safe transportation of the instrument.

4.1.2 Checks

Immediately after receipt, check whether the shipment has arrived complete and without damage by comparing it with the delivery note.

4.1.3 Location

The instrument has been developed for operation indoors and may not be used in explosive environments.

Place the instrument in a location of the laboratory which is suitable for operation and free of vibrations and which provides protection against corrosive atmosphere and contamination by chemicals.

The instrument should be protected against excessive temperature fluctuations and direct sunlight.

10 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 5 Titrations

5 Titrations

5.1 Dynamic equivalence point titration (DET)

Dynamic equivalence point titration is a titration mode for all standard titrations with an s-shaped curve progression. The reagent is added in variable volume steps. The volume steps vary as a function of the slope of the curve. An attempt is made to reach constant measured value changes with each dosing. The optimal volume for dosing is determined from the measured value changes of the previous dosings. Measured value acceptance is measured value drift-controlled (equilibrium titration) or after a waiting time. Equivalence points are evaluated automatically.

U/mV

V/mL

Figure 3 Reagent dosing for DET

5.2 Monotonic equivalence point titration (MET)

Monotonic equivalence point titration is a robust titration mode for titrations with any curve shape and for slow titrations or slow-response electrodes. The reagent is added in constant volume steps. Measured value acceptance is measured value drift-controlled (equilibrium titration) or after a waiting time. Equivalence points are evaluated automatically.

U/mV

V/mL

Figure 4 Reagent dosing for MET

900 Touch Control ■■■■■■■■ 11

5.3 Endpoint titration (SET) ■■■■■■■■■■■■■■■■■■■■■■

5.3 Endpoint titration (SET)

Endpoint titration is a titration mode for rapid routine determinations to a preset endpoint (e.g. titrations in accordance with special standards) and titrations for which reagent overflow must be avoided. The titration termination at the endpoint takes place either volume drift-controlled or after a waiting time. The volume dosed until the endpoint can be used for further calculations (e.g. the content of the sample).

U/mV

Endpoint

Control range

V/mL t/s

Initial dosing

Continuous dosing

Figure 5 Reagent dosing for SET

V/mL

5.4 Water determination according to Karl Fischer (KFT)

Karl Fischer Titration is a method for volumetric water determination. Conditioning is carried out automatically both before and after the actual titration. The reagent dosing is controlled in such a way that a predefined endpoint is reached as quickly and as accurately as possible. The volume steps and the rate of reagent dosing are regulated by the difference between the current measured value and the predefined endpoint. This means that titration is performed more slowly in the control range and that smaller volumes are added. The titration is stopped at the endpoint either driftcontrolled or after a waiting time. The volume dosed until the endpoint is used to calculate the water content of the sample.

12 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

KFT Ipol

U/mV

Control range

5 Titrations

EP

V/mL t/s

V/mL

Initial dosing

Continuous dosing

KFT Upol

I/µA

Control range

EP

V/mL

Figure 6 Reagent dosing for KFT

5.5 Titration with constant maintenance of the measured value (STAT)

In STAT titrations, reagent is added until a preset measured value (control point) has been reached, which is then maintained at a constant level.

Because the substance that is released as a result of the reaction is immediately titrated off with the titrant, the STAT controller has to be able to maintain a preset control point at a constant level until the preset stop condition has been reached, i.e. without oscillating around the control point.

STAT titrations are used, for example, in enzyme analysis where the enzyme activity can be determined from the resulting dosing rate (in mL/min).

Reagent dosing takes place in two phases during the titration:

Dosing outside the control range

The dosing rate increases continuously during this phase. It starts with the Min. rate and increases to the Max. rate. The dosing continues to be carried out at the Max. rate until the Control range is reached. In this range, STAT mode and SET mode control equally.

900 Touch Control ■■■■■■■■ 13

5.6 Manual titration (MAT) ■■■■■■■■■■■■■■■■■■■■■■

Dosing inside the control range

The STAT controller compensates minimal measured value deviations from the control point quickly by dosing in the control range faster than for the SET controller.

Figure 7 Reagent dosing for STAT

5.6 Manual titration (MAT)

Manual titration is a titration mode that is suitable in particular for titrations that are carried out conventionally manually with a Schellbach buret or a Dosimat.

The reagent is added by manual dosing. The endpoint is detected by a color change or with an electrode connected over a measuring signal. The command always has to be canceled manually with the Cancel command button or with the [FILL] key on the Manual Dosing Controller.

An advantage of the MAT titration mode is that the manual reagent addition as well as the measuring signal can be recorded optionally. These data points (time, volume, measured value) can be evaluated after the titration, e.g. to assess the titration quality, especially the accuracy of reaching the transition point.

In addition, the method can be equipped with further commands, e.g.

calculating a result, creating a report or scanning sample data.

The system is FDA compliant and enables a manual titration according to the FDA guidelines with this mode.

14 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 5 Titrations

5.7 Coulometric water content determination according to Karl Fischer (KFC)

The coulometric Karl Fischer titration is a variation of the classic water content determination method according to Karl Fischer.

The conventional method works with a methanolic solution of iodine, sulfur dioxide and a base as buffer substance. When titrating a sample containing water several reactions take place, which can be summarized by the following equation:

H

2

O + I

2

+ [RNH]SO

3

CH

3

+ 2 RN <=> [RNH]SO

4

CH

3

+ 2 [RNH]I

According to the equation above, the I

2

reacts quantitatively with H

2

O.

This chemical relation is the base for the water content determination.

With the coulometric Karl Fischer titration, the necessary iodine is directly and electrochemically generated in the submitted electrolyte ("electronic buret"). Between the amount of electric charge and the amount of generated iodine, there is a strictly quantitative relationship, which is used for high-precision dosing of the iodine. No titer needs to be determined as the coulometric Karl Fischer method is an absolute determination. It must only be ensured that the reaction generating the iodine runs with a 100% current efficiency. This is the case with all the reagents available at present.

The endpoint indication is effected voltametrically by modulating an alternating current of constant strength to a double Pt electrode. This produces a voltage difference between the Pt wires, which significantly declines when lowest amounts of free iodine are present. This circumstance is used for detecting the endpoint of the titration.

5.8 Bromine index determination (BRC)

The bromine index is a measure for the content of olefins in a sample.

The basis for the bromine index determination is the addition of bromine to the double bond of olefins. A bromine molecule is used for every carbon-carbon double bond.

With the bromine index determination, the necessary bromine is directly and electrochemically generated in the submitted electrolyte ("electronic buret"). Between the amount of electric charge and the amount of generated bromine, there is a strictly quantitative relationship, which is used for high-precision dosing of the bromine. No titer needs to be determined as the bromine index determination is an absolute determination.

The bromine index is usually indicated in mg bromine per 100 g sample.

900 Touch Control ■■■■■■■■ 15

6.1 Switching the instrument on and off ■■■■■■■■■■■■■■■■■■■■■■

6 Operation

6.1 Switching the instrument on and off

Switching on the instrument

CAUTION

Peripheral devices (e.g., printers) must be connected and switched on before you switch on the 900 Touch Control.

NOTICE

English is set as the default dialog language when the instrument is switched on for the first time.

Proceed as follows:

1

Press the power switch on the left-hand side of the back panel of the 900 Touch Control.

The 900 Touch Control and the control instrument are initialized and a system test is carried out. This process takes some time.

NOTICE

If a buret unit is connected, then a request appears to carry out the Prepare function:

16 ■■■■■■■■

All tubings and the cylinder are rinsed with the Prepare function.

The preparing of the buret unit is described in chapter 28.3.3, page 262 .

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Confirm the message with [OK] .

The main dialog is displayed:

Switching off the instrument

CAUTION

6 Operation

The 900 Touch Control must be switched off by pressing the power switch on the rear of the instrument before the electricity supply is interrupted. If this is not done, then there is a danger of data loss.

Because the power supply for the Touch Control is provided through the control instrument (Titrando, etc.) you must never disconnect the control instrument from the mains (e.g., by switching off with a connector strip) before you have switched off the Touch Control.

Proceed as follows:

1 Press the power switch on the left-hand side of the back panel of the

900 Touch Control.

The current data is saved and the system is shut down. This process takes just a short time. At the same time, all other instruments connected to the 900 Touch Control via a USB cable are also being switched off.

6.2 Fundamentals of operation

6.2.1 Touch-sensitive screen

The entire 900 Touch Control user interface is touch-sensitive. Simply touch a few of the buttons on the interface to learn how a touch-sensitive screen reacts. You can always return to the main dialog by touching [ ] .

In order to activate an element on the 900 Touch Control user interface, just touch the screen with your fingertip, finger nail, the eraser of a pencil or a stylus (special tool for operating instruments with touch-sensitive screens).

CAUTION

Never touch the touch screen with a pointed or sharp object such as a ballpoint pen.

900 Touch Control ■■■■■■■■ 17

6.2 Fundamentals of operation ■■■■■■■■■■■■■■■■■■■■■■

In the default setting, the software is configured in such a way that an acoustic signal will be generated every time an active control is touched.

This setting can be deactivated in the system settings (see chapter 7.5, page 40) .

6.2.2 Display elements and controls

18 ■■■■■■■■

The following display elements and controls are available:

Table 1 Fixed keys which are always available

[Home] always opens the main dialog.

[Back] saves the entry and opens the next-higher dialog page.

[Help] opens the online help for the dialog displayed.

[Print] opens the printing dialog.

[Manual] opens the manual control.

[Stop] cancels the running determination.

[Start] starts a determination.

The file name of the loaded method, the time and the system status are displayed in the main dialog in the Title bar .

In the other dialogs, the title bar shows the headings of the next upper level and of the displayed dialog. This is an aid for orientation during navigation through the user dialog.

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 6 Operation

Table 2 Screen elements

Buttons open a new dialog when they are tapped.

Inactive buttons with gray lettering indicate that the respective function is not available at the moment.

Input fields open an input dialog when tapped.

Tapping on the selection symbol opens a selection list.

A check box can also be activated or deactivated by tapping on it.

6.2.3 Status display

The current status of the system is displayed in the upper right-hand corner of the title bar.

Table 3 Status displays

The instrument is in normal status.

The working medium is being conditioned.

Conditioning has been paused.

The working medium is conditioned.

A method has been started.

900 Touch Control ■■■■■■■■ 19

6.2 Fundamentals of operation ■■■■■■■■■■■■■■■■■■■■■■

A method has been paused.

An action has been started in manual control.

6.2.4 Entering text and numbers

In the editing dialog for text or numerical input, enter the individual characters by tapping in the input field. The following functions are available:

Text editor

20 ■■■■■■■■

Table 4 Editing functions

[OK] The modification is applied and the editing dialog is exited.

[Cancel] The editing dialog is exited without applying the modification.

[Delete entry]

[ ⌫ ]

[ ⇦ ]

The content of the input field is deleted completely.

The character in front of the cursor is deleted.

[ ⇨ ]

The cursor within the input field is shifted to the left by one character.

The cursor within the input field is shifted to the right by one character.

[a…z] The lower-case letters are displayed. The label changes to [A…Z] . The upper-case letters are displayed again by tapping.

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

[0…9]

[Special characters]

6 Operation

Numbers and mathematical characters are displayed.

Special characters are displayed. You can use the button [More] to navigate through all available characters.

Number editor

900 Touch Control

Table 5 Editing functions

[OK] The modification is applied and the editing dialog is exited.

[Cancel] The editing dialog is exited without applying the modification.

[Delete entry]

[off]

The content of the input field is deleted completely.

If not only numbers but also special values (e.g., off ) can be entered, then the corresponding buttons will be shown to the right of the numerical keypad.

[R1] For many parameters, a result previously defined in the method can also be entered in place of a number

(see chapter 32.6, page 496) . You can

select the result variable by touching [R1] .

NOTICE

A commercially available USB keyboard can be connected to make it easier for text and numerical input. The key assignment is described in

chapter 11.10, page 117 .

■■■■■■■■ 21

7.1 General system settings ■■■■■■■■■■■■■■■■■■■■■■

7 System settings

Main dialog: System ▶ System settings

This chapter describes the various system settings and configurations.

Selecting the dialog language

(see chapter 7.1.1, page 22)

.

Setting the date and time

(see chapter 7.1.2, page 23) .

Defining system-specific dialog options (see chapter 7.2, page 24)

.

User administration

(see chapter 7.3, page 28) .

Defining settings for the measured value display (see chapter 7.4, page

40)

.

Configuring settings for acoustic signals

(see chapter 7.5, page 40)

.

7.1 General system settings

7.1.1 Selecting the dialog language

The user interface is available in several languages. In addition to the two default dialog languages English and German , additional languages can be selected.

Proceed as follows to select the dialog language:

1 Opening the system settings

In the main dialog, tap on [System] .

Tap on [System settings] .

The System / System settings dialog is displayed.

22 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 7 System settings

2 Selecting the dialog language

Tap on the Dialog language list box and select the desired language.

3 Saving the settings

Tap on the fixed keys [ ] or [ ] .

The main dialog is displayed in the respective dialog language.

7.1.2 Setting the date, time and local time

The Touch Control displays the date and time in accordance with ISO standard 8601.

Proceed as follows to set the date and time:

1 Opening the system settings

In the main dialog, tap on [System] .

Tap on [System settings] .

The System / System settings dialog is displayed.

2 Entering the date

Tap on the input field for the date.

The editor opens.

Enter the current date in the format YYYY-MM-DD and confirm with [OK] .

The arrow keys [ ⇦ ] and [ ⇨ ] are used to move the cursor to the left or to the right by one character.

The input is saved and the editor is closed.

3 Entering the time

Tap on the input field for the time.

The editor opens.

Enter the current time in the format hh:mm:ss (24-hour format) and confirm with [OK] .

The arrow keys [ ⇦ ] and [ ⇨ ] are used to move the cursor to the left or to the right by one character.

The input is saved and the editor is closed.

900 Touch Control ■■■■■■■■ 23

7.2 System-specific dialog options ■■■■■■■■■■■■■■■■■■■■■■

4 Enter the local time

Tap on the list box Local time - UTC and select the difference from the UTC (Coordinated Universal Time).

The selection off means that the time is saved with no difference from the UTC.

5 Saving the settings

Tap on the fixed keys [ ] or [ ] .

The time settings are saved.

7.2 System-specific dialog options

Main dialog: System ▶ System settings ▶ Dialog options

A wide array of different titration systems can be operated with Touch

Control. Accordingly, the Touch Control displays far more functions than are required for any particular system.

Dialog

24 ■■■■■■■■

If you work without a login function, in this dialog you can define whether the system should generally be operated in expert mode or in routine mode. If you work with the login function activated, you must define this setting separately for each user

(see chapter 7.3.1, page 29)

.

Dialog mode in which the user may operate the system.

Selection Expert dialog | Routine dialog

Expert dialog

All functions that are supported by the connected system are available.

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 7 System settings

Routine dialog

The user dialog can be limited for routine operations. Only selected

functions are available (see Configuring the routine dialog, page 26) .

[Command list]

Block unneeded method commands (see "Blocking unneeded commands and fixed keys", page 25) .

[Fixed keys]

Block unneeded fixed keys (see "Blocking unneeded commands and fixed keys", page 25)

.

[Routine dialog]

Configure functions for the routine dialog (see "Configuring the routine dialog", page 26)

.

Selecting the dialog mode

Proceed as follows to change the dialog mode:

1 Select the dialog mode

Open the selection list Dialog and select either Expert dialog or

Routine dialog .

2 Saving the settings

Tap on the fixed keys [ ] or [ ] .

The setting will apply to all dialogs.

NOTICE

If you have selected Routine dialog and if the routine dialog was configured in such a way that the dialog System settings / Dialog options is blocked, then you can switch back over to the expert dialog as follows:

Operation without login function:

In the main dialog, enter User = Metrohm .

Operation with login function:

A user who works with expert dialog must log in.

Blocking unneeded commands and fixed keys

This following configurations apply for both dialog modes: routine dialog and expert dialog.

900 Touch Control ■■■■■■■■ 25

7.2 System-specific dialog options ■■■■■■■■■■■■■■■■■■■■■■

Blocking commands

Proceed as follows to block unneeded commands:

1 Display the command list

Tap on the [Command list] button.

The list of all command groups is displayed.

2 Deactivate command groups

Deactivate those command groups which are not permitted to be used.

3 Saving the settings

Tap on the fixed keys [ ] or [ ] .

All deactivated commands appear grayed out in the method editor and cannot be used for creating methods.

Blocking fixed keys

Proceed as follows to block unneeded fixed keys:

1 Display fixed keys which can be blocked

Tap on the [Fixed keys] button.

All fixed keys which can be blocked are displayed.

2 Deactivate fixed keys

Deactivate those fixed keys which are not permitted to be used.

3 Saving the settings

Tap on the fixed keys [ ] or [ ] .

Deactivated fixed keys cannot be used.

Configuring the routine dialog

A suitable Standard configuration has already been saved for routine operations:

Methods can only be loaded, but not modified or created.

Determinations cannot be recalculated.

You can readjust this standard configuration by disabling additional functions or re-enabling disabled functions.

26 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 7 System settings

NOTICE

The configuration of the routine dialog applies for all routine users. You also have the option of defining user-specific routine settings. To do this, you must work with the login function activated and create an

identification profile for each user (see chapter 7.3.2, page 33)

.

Proceed as follows to modify the configuration for the routine dialog:

1 Opening the dialog

Tap on the [Routine dialog] button.

The list of all buttons in the main dialog, in the manual control, etc. is displayed:

900 Touch Control

2 Deactivate buttons

Deactivate those buttons which are not permitted to be used.

All deactivated buttons will appear grayed-out, i.e. they are inactive.

3 Deactivate other functions

Many additional buttons and parameters can be disabled in the dialogs of [System] , [Load method] , [Control] , [Edit parameters] and [Results] . The corresponding option must be activated in order to enable these buttons.

4 Saving the settings

Tap on the fixed keys [ ] or [ ] .

■■■■■■■■ 27

7.3 User administration ■■■■■■■■■■■■■■■■■■■■■■

All deactivated functions will appear grayed-out, i.e. they are inactive.

7.3 User administration

Main dialog: System ▶ System settings ▶ User admin.

All of the functions of the user administration are described in this chapter. No user administration is mandatory for the simple operation of a titration system. If, however, you wish to be in compliance with FDA Guidance 21 CFR Part 11 , then you must use the functions of the user administration. Additional information regarding the requirements of the FDA

Guidance can be found in the document Compliance Guide Touch Control .

NOTICE

If you work with the login function activated, then the user administration is accessible only for users with administrator rights. This means that you must ensure that at least two users have administrator rights so at least one of them will be available. Keep the access rights for a user with administrator rights in a safe place so that they are accessible in an emergency.

28 ■■■■■■■■

The following data is displayed in the user list for each user:

Name

Dialog mode in which the user may operate the system.

Status

You can use the user list two different ways:

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 7 System settings

Operation with login function:

If you work with the login function activated, i.e. if each user must log into the system before starting work, then only those users entered in the list can log in. The user who is currently logged in is shown in the main dialog.

Operation without login function:

If you work without the login function, the users whose names are entered in the list can be selected in the main dialog or a user name can be entered. This makes it possible to document who has operated the titration system.

[Login options]

Define the settings for the login, Audit Trail, etc.

(see chapter 7.3.3, page

33) .

[Create ID profile]

Create an identification profile for the selected user on a storage medium

(see chapter 7.3.2, page 33)

.

[New]

Adding a new user to the list

(see chapter 7.3.1, page 29)

.

[Delete]

Delete the selected user from the list.

NOTICE

Once users have worked with the login function and password protection, they can no longer be deleted, even if the password protection is disabled again. The status of these users must be set to inactive

(requirement of FDA Guidance 21 CFR Part 11 ).

The last user with administrator rights cannot be deleted.

[Edit]

Editing the data of the selected user

(see chapter 7.3.1, page 29)

.

7.3.1 Editing the user configuration

User list: User ▶ New / Edit

900 Touch Control ■■■■■■■■ 29

7.3 User administration ■■■■■■■■■■■■■■■■■■■■■■

User

30

Full name

Dialog

Status

■■■■■■■■

The designation of the user is used for unambiguous identification, e.g.

the company internal shorthand symbol or the personal number. The user name is printed out in all reports containing determination data and stored in the determination file. Each file always contains the name of the user who created it and the name of the last user to edit it.

Entry

Default value

24 characters maximum empty

Complete name of the user.

Entry

Default value

24 characters maximum empty

Dialog mode in which the user may operate the system.

Selection Expert dialog | Routine dialog

Expert dialog

All functions that are supported by the connected system are available.

Routine dialog

The user dialog can be limited for routine operations. Only selected functions are available

(see Configuring the routine dialog, page 26) .

Status of the user. Users can be deactivated. This function is useful, for instance, if the user is no longer authorized to operate the system or no longer works for the company.

Selection

Default value active | inactive active

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 7 System settings active

The user is authorized to operate the system.

inactive

The user is not authorized to operate the system and can no longer log in.

Admin. rights on | off (Default value: off )

If this option is activated, then the user has administrator rights.

NOTICE

The last user with administrator rights cannot be deleted anymore.

[Favorites]

Saving methods and sample tables as user-specific favorites (see chapter

19, page 190)

.

[Signature method]

Assign the permission to use and sign methods (see ""Edit user / Signature method" dialog", page 31)

.

[Signature determ.]

Assign the permission to use and sign determinations

(see ""Edit user /

Signature determination" dialog", page 32)

.

"Edit user / Signature method" dialog

User: Edit ▶ Signature method

Authorizations for using and signing methods can be defined in this dialog. These settings are only effective if you are working with the login function enabled and password protection.

Use only released methods on | off (Default value: off )

If this option is activated, then the user may only start methods that have been released.

Review methods (signature level 1) on | off (Default value: off )

If this option is activated, then the user may only sign methods which have the status saved . The method is assigned the status reviewed .

900 Touch Control ■■■■■■■■ 31

7.3 User administration ■■■■■■■■■■■■■■■■■■■■■■

Release methods (signature level 2) on | off (Default value: off )

If this option is activated, then the user may only sign methods which have the status reviewed . The method is assigned the status released .

NOTICE

If the two options Review methods and Release methods are selected for a user, then this user can sign different methods at either level 1 or level 2, but not the same method at both level 1 and level 2.

Delete signatures on | off (Default value: off )

If this option is activated, then the user may delete the signatures of a method which has been released. The method is assigned the status saved . The signatures of a method can only be deleted if it has the status released .

"Edit user / Signature determination" dialog

User: Edit ▶ Signature determ.

Authorizations for signing determinations can be defined in this dialog.

These settings are only effective if you are working with the login function enabled and password protection.

Review determinations (signature level 1) on | off (Default value: off )

If this option is activated, then the user may only sign determinations at the first level. The determination is assigned the status reviewed .

Release determinations (signature level 2) on | off (Default value: off )

If this option is activated, then the user may only sign determinations at the second level. The determination is assigned the status released .

Delete signatures on | off (Default value: off )

If this option is activated, then the user may delete the signatures of a determination which has been released. The signatures of a determination can only be deleted if it has the status released .

32 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 7 System settings

7.3.2 Creating an identification profile

User list: User ▶ Create ID profile

If you plan to carry out the login with an identification profile (see chapter

7.3.3, page 33)

, then you must first create an identification profile on a storage medium for each user. A check can then be made at the time of login as to whether or not the user does exist and whether or not he or she is working in the expert dialog or in the routine dialog. After a successful login the routine dialog settings stored on the card are loaded.

NOTICE

In addition to the user name, the current routine dialog settings are also stored in this identification profile. This means you can define user-specific routine dialog settings for each user. However, you must configure them

(see "Configuring the routine dialog", page 26)

, before you create the identification profile.

Before you create the identification profile, check whether the desired routine dialog settings are active.

1 Plug in a storage medium

Plug in the USB storage medium.

2 Select a user

In the user list, select the user for whom the profile is to be created.

3 Create the identification profile

Tap on [Create ID profile] .

The user configuration and the current routine dialog settings are saved.

7.3.3 Defining login options

Main dialog: System ▶ System settings ▶ User admin. ▶ Login options

There are a number of different ways to log onto the system:

Without login

Login via user name

Login via user name and password

Login via identification profile

Login via identification profile and password

900 Touch Control ■■■■■■■■ 33

7.3 User administration ■■■■■■■■■■■■■■■■■■■■■■

NOTICE

If you work with the login function activated, then the user administration is accessible only for users with administrator rights. This means that you must ensure that at least two users have administrator rights so at least one of them will be available. Keep the access rights for a user with administrator rights in a safe place so that they are accessible in an emergency.

NOTICE

If you exit this dialog with [ ] or [ ] , and if you have selected one of the login variants Login via user name or Login via identification profile , then the login dialog will open automatically and you must also log in to the system.

This means that you must make sure that you have first defined all the users and created the identification profiles before you activate the login function.

Login via user name on | off (Default value: off )

If this option is activated, then the user must log in with his or her unambiguous identification.

Login via identification profile on | off (Default value: off )

If this option is activated, then the login will take place via USB storage medium with the identification profile stored on it.

34 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 7 System settings

Password required on | off (Default value: off )

If this option is activated, then the user must enter a password in addition to his or her user name or identification profile.

Logout automatically on | off (Default value: off )

If this option is activated, then the user will be logged out automatically after the specified time.

Input range 1 to 60 min

Login only for the same user on | off (Default value: off )

If this option is activated, then only the same user may log in again after he or she has logged out. Users with administrator rights can, however, log in at any time.

[Audit Trail]

Define the settings for recording an Audit Trail (see chapter 7.3.7, page

39) .

[Reasons]

Predefining a list of reasons which can be selected for the modification/

signature of a method or determination (see chapter 7.3.6, page 38) .

[Modific. options]

Defining the modifications for which a reason is required

(see chapter

7.3.5, page 37)

.

[Password options]

Define the settings for the password, see following chapter.

7.3.4 Password options

Main dialog: System ▶ System settings ▶ User admin. ▶ Login options ▶ Password options

You can make various settings for password entry in the password options.

900 Touch Control ■■■■■■■■ 35

7.3 User administration ■■■■■■■■■■■■■■■■■■■■■■

Minimum password length

Minimum number of characters of the passwords.

Input range

Default value

1 to 10

1

No. of entry attempts

If the user has logged in incorrectly this many times, then it will automatically be deactivated. It can only be reactivated by a user with administrator rights.

Input range

Selection

Default value

2 to 5 off off

Special characters required on | off (Default value: off )

If this option is activated, then the password must contain one of the following special characters: ° § + ¦ @ * # ç % & ¬ ( ) = ' ^ ` ~ ] [ } { - _ : . ; , >

< £ !

Password expires on | off (Default value: off )

If this option is activated, then the user must define a new password after the time specified. A password that has already been used cannot be used again.

Input range

Default value

1 to 999 days

365 days

36 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 7 System settings

Forgotten password

CAUTION

If a user has forgotten his password, a new user name must be defined.

The same user name can only be used again after a re-installation and re-creating the user list.

Proceed as follows:

1 Create a backup

Create a backup (see chapter 12.3, page 127) .

2 Executing a system initialization

Switch off the Touch Control and execute a system initialization (see chapter 30.1, page 483) .

3 Restore backed-up data

Use the function Restore to reload the data from the backup into your system

(see chapter 12.3.1, page 127) .

Deactivate the options User list and System settings / User admin.

.

4 Switch on the Touch Control

Switch the Touch Control back on again after a few seconds.

5 Restore the user list

Create the user list again and redefine the login options.

7.3.5 Modification options

Main dialog: System ▶ System settings ▶ User admin. ▶ Login options ▶ Modific. options

In the dialog Login options / Modification options , you can define for which actions a reason must be entered. These reasons are documented in the Audit Trail

(see chapter 7.3.7, page 39)

together with the modification. The reason for the last modification is shown in the properties of the method or determination.

900 Touch Control ■■■■■■■■ 37

7.3 User administration ■■■■■■■■■■■■■■■■■■■■■■

NOTICE

The reasons are only requested when working with activated login function and password.

Saving modified method on | off (Default value: off )

If this option is activated, then a reason must be given when saving a method modification.

Recalculating determination on | off (Default value: off )

If this option is activated, then a reason must be given when determinations are being recalculated.

7.3.6 Reasons

Main dialog: System ▶ System settings ▶ User admin. ▶ Login options ▶ Reasons

In the dialog Login options / Reasons , you can create a selection list containing reasons from which a selection can be made when signing and modifying methods and determinations. Some reasons have already been provided.

[New]

[Delete]

Add a new reason to the list.

Delete the selected reason from the list.

38 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

[Edit]

Change the designation of the selected reason.

7 System settings

7.3.7 Audit Trail

Main dialog: System ▶ System settings ▶ User admin. ▶ Login options ▶ Audit Trail

An Audit Trail is an automatically generated log of all user activities. An

Audit Trail contains precise logs of user actions (date, time, user, action, etc.). Recording an Audit Trail is important for compliance with FDA Guidance 21 CFR Part 11 when using PC programs. A step-by-step description of how you must proceed to remain in compliance with the FDA Guidance is included in the document Compliance Guide Touch Control . The Audit

Trail is saved to the internal memory.

You can also use the Audit Trail function specifically to record the data which are of interest to you.

You can view, filter and export the Audit Trail on a PC with the supplied software program AuditTrailViewer . Details concerning utilization of the

AuditTrailViewer can be found in the Appendix (

chapter 32.9, page 513

).

If necessary, you can delete all of the entries in the Audit Trail with

[Delete Audit Trail] . Do not fail to create a backup before doing so, however.

Security log on | off (Default value: off )

If this option is activated, then the following events will be recorded: user login/logout, password changes, automatic deactivation of users and messages displayed during the login procedure.

User administration log on | off (Default value: off )

If this option is activated, then all changes with respect to user administration will be recorded (modifying user list/user data, changing login options, etc.).

Method log on | off (Default value: off )

If this option is activated, then all changes with respect to methods and determinations will be recorded (saving, deleting, renaming, copying and loading methods; editing method parameters; starting, stopping, pausing determinations).

900 Touch Control ■■■■■■■■ 39

7.4 Measured value display

Data log

■■■■■■■■■■■■■■■■■■■■■■ on | off (Default value: off )

If this option is activated, then the following events will be recorded: changing sample data; settings in the Control dialog, recalculating determinations, changing statistics data, changing the titer/concentration of a titrant, changing the calibration data of a sensor, changing the value of a common variable.

System log

■ on | off (Default value: off )

If this option is activated, then the program start and the displayed messages of the following types will be recorded:

: General warning messages

: Error messages

7.4 Measured value display

Main dialog: System ▶ System settings ▶ Meas. value display

The number of decimal places can be defined for pH values and voltages.

This setting refers exclusively to the display of the measured values in the live display and in the manual control. The values will, however, always be stored with their full accuracy.

7.5 Acoustic signals

Main dialog: System ▶ System settings ▶ Acoustic signals

You can define acoustic signals in order to direct attention to particular events. You can define signals for the following events:

Wrong manipulation

An acoustic signal will sound each time an invalid action is attempted

(e.g. pressing [ ] again when Help is open).

Display a message

A short beep will sound each time a message appears on the display.

This informs the user that the message must be confirmed.

Button contact

Each time a button is touched on the touch screen, this will be confirmed by an acoustic signal.

External data input

An acoustic signal will provide confirmation each time data is received from external devices (e.g. balance, barcode reader).

40 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 8 Titrants

8 Titrants

Main dialog: System ▶ Titrants

This chapter describes how you can create a list of titrants used in the system. Titrants can be used in intelligent buret units or in nonintelligent buret units. Intelligent buret units have a built-in data chip on which the data for the titrant is stored. This data is automatically read out during attachment and entered in the titrant list.

900 Touch Control

The titrant list can contain a maximum of 30 titrants. The following data is specified for each titrant:

Designation

Cylinder volume

Type

– EU : exchange unit without data chip

– IEU : exchange unit with integrated data chip

– DU : dosing unit without data chip

– IDU : dosing unit with integrated data chip

MSB connector of the dosing device/control instrument (only when exchange unit/dosing unit is attached)

Titrants in exchange units/dosing units with integrated data chips are depicted in green lettering.

The following titrant data is stored in the titrant list:

Name

Each titrant in the system is identified by its unambiguous name.

Concentration

Current titer

Working life

■■■■■■■■ 41

8.1 Adding a new titrant ■■■■■■■■■■■■■■■■■■■■■■

Data on exchange/dosing unit:

– Parameters for the function PREP

– Length and diameter of the tubings

– Port assignment of the dosing unit

– Cylinder volume

– Serial number etc.

– etc.

NOTICE

If data is read out from the data chip, then a check is made whether the titrant list already contains a titrant of the same type with the identical serial number. If this is the case, then the older data set will always be overwritten by the new data set, no matter whether the data set in the titrant list or the data set on the data chip is the most recent one.

[New]

[Delete]

[Edit]

Adding a new titrant to the list (see chapter 8.1, page 42) .

Delete the selected titrant from the list.

Editing the data of the selected titrant

(see chapter 8.2, page 43)

.

8.1 Adding a new titrant

Before you can use a titrant, you must add it to the titrant list. To do this, use the button [New] .

Exchange unit/dosing unit with data chip:

The exchange unit or dosing unit must be attached. All dosing devices on which non-configured exchange or dosing units have been detected are included in a selection list. Tapping on the button [Edit] opens the properties dialog, see following chapter.

Exchange unit/dosing unit without data chip:

The properties dialog is opened after the dosing unit type has been selected, see following chapter.

42 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

8.2 Editing titrant data

Titrant list: Titrants ▶ New / Edit

8 Titrants

Titrant

Concentration

Comment

Titer

All of the data for the selected titrant is displayed in the dialog Titrants /

Edit .

The designation of the titrant is used for unambiguous identification.

Entry

Selection

24 characters maximum

Selection of frequently used titrants

Concentration of the titrant.

Input range

Default value

–999999999 to 9999999999

1.000

Unit of the concentration.

Entry

Selection

Default value

10 characters maximum

µmol/mL | mmol/L | mol/L | g/L | mg/L | mg/mL |

µg/L | ppm | % | mEq/L mol/L

Entry 24 characters maximum

Titer of the titrant.

Input range

Default value

–999999999 to 9999999999

1.000

900 Touch Control ■■■■■■■■ 43

8.2 Editing titrant data ■■■■■■■■■■■■■■■■■■■■■■

Unit of the titer.

Entry

Default value

Selection

NOTICE

10 characters maximum empty

µmol/mL | mmol/L | mol/L | g/L | mg/L | mg/mL |

µg/L | ppm | % | mEq/L

If you modify the titer or the concentration of the titrant in a loaded determination at a later date and would then like to recalculate the determination with the corrected value, then you must modify the value in the determination data under View data / Titrant data

(see "Calibration and titrant data", page 218)

.

Date titer det.

[Working life]

[Dosing unit]

[Exchange unit]

[Titer options]

Date and time of the last titer determination. For new titrants, the time the preparation was made is specified until after the first time a titer determination has been carried out.

Defining the working life of the titrant

(see chapter 8.3, page 45) .

This button is only displayed for Type = IDU or DU .

Defining the properties of the dosing unit used

(see chapter 8.4, page

46)

.

This button is only displayed for Type = IEU or EU .

Defining the properties of the exchange unit used (see chapter 8.5, page

52)

.

Displaying the properties for titer determination (see chapter 8.7, page

58)

.

44 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 8 Titrants

8.3 Monitoring the working life

Titrant: Edit ▶ Working life

In the dialog Edit titrant / Working life , you can define the time interval after which the titrant must be replaced. This is particularly important if your titrant has a limited working life. If you do not wish to monitor the working life, then you can enter only the date of manufacture for documentation purposes.

Preparation date

Date on which the reagent was manufactured or the bottle was opened.

For new titrants, the time the preparation was made will be specified.

Format: YYYY:MM:DD

Monitoring on | off (Default value: off )

If this parameter is activated, then the working life will be monitored.

Working life

If you define a time interval for the working life, then the Expiry date will be tracked automatically.

Input range

Default value

1 to 999 days

999 days

Expiry date

If you define an expiry date, then the Working life will be tracked automatically.

Format: YYYY:MM:DD

Action

Selection of the action which is carried out when the time interval has expired.

Selection

Default value

Display message | Document message | Cancel determination

Display message

For all three options it is documented in the determination data (see dialog More determination data / Messages ), that the time interval has been expired.

900 Touch Control ■■■■■■■■ 45

8.4 Dosing unit ■■■■■■■■■■■■■■■■■■■■■■

Display message

A message is displayed. You can select whether you want to continue with the determination or cancel the run.

Document message

In the determination data it will be documented, that the time interval has been expired.

Cancel determination

The determination is stopped.

8.4 Dosing unit

Titrant: Edit ▶ Dosing unit

Name

Order number

Serial number

You can edit data for the dosing unit of the titrant in this dialog.

Designation of the exchange or dosing unit.

Entry

Default value

24 characters maximum empty

Order number of the exchange or dosing unit. It will be read out automatically on units with integrated data chips.

Entry

Default value

24 characters maximum empty

Serial number of the exchange or dosing unit. It will be read out automatically on units with integrated data chips.

Entry 8 digits maximum

46 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 8 Titrants

Cylinder volume

Cylinder volume of the dosing unit. It will be read out automatically on dosing units with integrated data chips.

Selection

Default value

2 |

20

5 | 10 | 20 | 50

Serial no. cyl.

Serial number of the dosing cylinder. It will be read out automatically on exchange or dosing units with integrated data chips. The number can be changed manually at any time, e.g. when a cylinder is replaced.

Entry 8 digits maximum

[Valve disk]

Specifying the shift direction of the valve disk

(see chapter 8.4.3, page

52) .

[GLP test]

Defining the time interval for the GLP test (see chapter 8.6, page 56) .

[PREP param.]

Entering the parameters for the preparation (see chapter 8.4.1, page

47) .

[Tubing param.]

Entering the parameters for the connected tubing (see chapter 8.4.2, page 49)

.

8.4.1 Parameters for preparing (PREP) and emptying (EMPTY)

Titrant: Edit ▶ Dosing unit ▶ PREP param.

In the dialog Dosing unit / PREP parameters , you can adjust the parameters for the execution of the Prepare (command PREP) and

Empty (command EMPTY) functions. The Prepare function is used to rinse the cylinder and tubing of the dosing unit and fill it air bubble-free.

You should carry out this function before the first determination or once a day. The EMPTY function empties the cylinder and the tubings of the dosing unit.

Dosing port PREP/EMPTY

Dosing port through which the cylinder contents are ejected.

Selection

Default value

Dosing port 1 | Dosing port 2 | Fill port | Special port

Dosing port 1

900 Touch Control ■■■■■■■■ 47

8.4 Dosing unit ■■■■■■■■■■■■■■■■■■■■■■

Dosing rate Dos. port 1

Rate used for the aspiration and ejection of the reagent via dosing port 1.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum dosing rate depends on the cylinder volume

(see chapter 32.2, page 493)

.

Dosing rate Dos. port 2

Rate used for the aspiration and ejection of the reagent via dosing port 2.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum dosing rate depends on the cylinder volume

(see chapter 32.2, page 493)

.

Dosing rate Fill port

Rate used for the aspiration and ejection of the reagent via fill port.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum dosing rate depends on the cylinder volume

(see chapter 32.2, page 493)

.

Dosing rate Spec.port

Rate used for the aspiration and ejection of the reagent via the special port.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum dosing rate depends on the cylinder volume

(see chapter 32.2, page 493)

.

48 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 8 Titrants

8.4.2 Tubing parameters

Titrant: Edit ▶ Dosing unit ▶ Tubing param.

You can enter the length and diameter of the connected tubings in the dialog Dosing unit / Tubing parameters . The values which have already been entered correspond to the dimensions of the supplied standard tubings. In addition, the port assignment can be modified.

NOTICE

These parameters are important for the correct execution of the Preparing (PREP command) and Emptying (EMPTY command) functions, because the volumes of the tubing connections are taken into account.

Dosing port 1

Port

Port to be used as dosing port 1 for the PREP and EMPTY

(see figure 8, page 51)

functions.

Selection

Default value

Port 1

Port 1

| Port 2 | Port 3 | Port 4

Length

Length of the tubing.

Input range

Default value

0.0 to 999.9 cm

40.0 cm

The setting 0.0

means that this tubing will neither be rinsed nor emptied.

Diameter

Diameter of the tubing.

Input range

Default value

0.0 to 9.9 mm

2.0 mm

Dosing port 2

Port

Port to be used as dosing port 2 for the PREP and EMPTY

functions (see figure 8, page 51) .

Selection

Default value

Port 1

Port 3

| Port 2 | Port 3 | Port 4

900 Touch Control ■■■■■■■■ 49

8.4 Dosing unit

Length

■■■■■■■■■■■■■■■■■■■■■■

Diameter

Length of the tubing.

Input range

Default value

0.0 to 999.9 cm

0.0 cm

The setting 0.0

means that this tubing will neither be rinsed nor emptied.

Diameter of the tubing.

Input range

Default value

0.0 to 9.9 mm

2.0 mm

Fill port

Port

Length

Diameter

Port to be used as fill port for the PREP and EMPTY

functions (see figure

8, page 51)

.

Selection

Default value

Port 1

Port 2

| Port 2 | Port 3 | Port 4

Length of the tubing.

Input range

Default value

0.0 to 999.9 cm

25.0 cm

The setting 0.0

means that this tubing will neither be rinsed nor emptied.

Diameter of the tubing.

Input range

Default value

0.0 to 9.9 mm

2.0 mm

Special port

Port

Length

Port to be used as special port for the PREP and EMPTY functions

(see figure 8, page 51) .

Selection

Default value

Port 1

Port 4

| Port 2 | Port 3 | Port 4

Length of the tubing.

50 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Diameter

Input range

Default value

8 Titrants

0.0 to 999.9 cm

0.0 cm

The setting 0.0

means that this tubing will neither be rinsed nor emptied.

Diameter of the tubing.

Input range

Default value

0.0 to 9.9 mm

2.0 mm

4

5

1

2

3

Figure 8 Dosing unit – port assignment

1 VENT

This port is set up for the deaeration of the reagent bottle. An adsorber tube (filled with desiccant) is usually mounted here.

3 Port 2

The default definition of this port is the filling port. A riser tube is usually mounted to it.

5 Port 3

The default definition of this port is dosing port 2.

2 Port 1

The default definition of this port is dosing port 1.

4 Port 4

Air is suctioned through this port during the

Emptying function.

900 Touch Control ■■■■■■■■ 51

8.5 Exchange unit ■■■■■■■■■■■■■■■■■■■■■■

8.4.3 Shift direction of the valve disk

Titrant: Edit ▶ Dosing unit ▶ Valve disk

In this dialog, you can specify the shift direction of the valve disk.

Direction

Shift direction of the valve disc.

Selection

Default value descending | ascending | automatic | not over automatic descending

The ports are moved to in descending order.

ascending

The ports are moved to in ascending order.

automatic

The ports are moved to by the shortest path.

not over

Define a protected port.

Not over port

This parameter can only be edited with Direction = not over .

Define a protected port if the valve disc is not to be rotated over this port

(useful with pipetting functions). The protected port can, however, be moved to directly.

Selection

Default value

1 | 2 | 3 | 4

4

8.5 Exchange unit

Titrant: Edit ▶ Exchange unit

52 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

You can edit data for the exchange unit of the titrant in this dialog.

8 Titrants

Name

Designation of the exchange or dosing unit.

Entry

Default value

24 characters maximum empty

Order number

Order number of the exchange or dosing unit. It will be read out automatically on units with integrated data chips.

Entry

Default value

24 characters maximum empty

Serial number

Serial number of the exchange or dosing unit. It will be read out automatically on units with integrated data chips.

Entry 8 digits maximum

Cylinder volume

Cylinder volume of the exchange unit. It will be read out automatically on exchange units with integrated data chips.

Selection

Default value

1 |

20

5 | 10 | 20 | 50

Serial no. cyl.

Serial number of the dosing cylinder. It will be read out automatically on exchange or dosing units with integrated data chips. The number can be changed manually at any time, e.g. when a cylinder is replaced.

Entry 8 digits maximum

[GLP test]

Defining the time interval for the GLP test (see chapter 8.6, page 56) .

[PREP param.]

Entering the parameters for the preparation (see chapter 8.5.1, page

54) .

[Tubing param.]

Entering the parameters for the connected tubing (see chapter 8.5.2, page 55)

.

900 Touch Control ■■■■■■■■ 53

8.5 Exchange unit ■■■■■■■■■■■■■■■■■■■■■■

8.5.1 Parameters for the preparation (PREP)

Titrant: Edit ▶ Exchange unit ▶ PREP param.

In the dialog Exchange unit / PREP parameters , you can adjust the parameters for the execution of the Prepare function (command PREP).

This function is used to rinse the cylinder and tubings of the exchange unit and fill it air bubble-free. You should carry out this function before the first determination or once a day.

Volume

Volume of titrant dosed during a rinsing cycle.

Input range

Selection

Default value

0.00000 to 99999.9 mL

Cylinder volume

Cylinder volume

Cylinder volume

The entire cylinder volume is being dosed.

Cycles

Number of rinsing cycles. We recommend carrying out at least two rinsing cycles in order to remove all air bubbles.

Selection

Default value

1

2

| 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9

Dosing rate

The rate at which dosing takes place.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum dosing rate depends on the cylinder volume

(see chapter 32, page 492) .

Filling rate

Rate at which the dosing cylinder is filled.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum filling rate depends on the cylinder volume (see chapter

32, page 492) .

54 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 8 Titrants

8.5.2 Tubing parameters

Titrant: Edit ▶ Exchange unit ▶ Tubing param.

You can enter the length and diameter of the connected tubings in the dialog Exchange unit / Tubing parameters . The values which have already been entered correspond to the dimensions of the supplied standard tubings.

Dosing tip

Tubing to the dosing tip

(92 )

.

Length

Length of the tubing.

Input range

Default value

0.0 to 999.9 cm

40.0 cm

Diameter

Diameter of the tubing.

Input range

Default value

0.0 to 9.9 mm

2.0 mm

Dosing cylinder

Tubing to the dosing cylinder

(93 ) .

Length

Length of the tubing.

Input range

Default value

0.0 to 999.9 cm

13.0 cm

Diameter

Diameter of the tubing.

Input range

Default value

0.0 to 9.9 mm

2.0 mm

Reagent bottle

Tubing to the reagent bottle (91 )

.

Length

Length of the tubing.

Input range

Default value

0.0 to 999.9 cm

25.0 cm

900 Touch Control ■■■■■■■■ 55

8.6 GLP test for exchange unit and dosing unit

Diameter

Diameter of the tubing.

Input range

Default value

0.0 to 9.9 mm

2.0 mm

1

2

3

■■■■■■■■■■■■■■■■■■■■■■

Figure 9 Exchange unit – tubing connections

1 Tubing to the reagent bottle

3 Tubing to the dosing cylinder

2 Tubing to the dosing tip

8.6 GLP test for exchange unit and dosing unit

Titrant: Edit ▶ Dosing unit / Exchange unit ▶ GLP test

In the dialog Exchange unit / GLP test or Dosing unit / GLP test , respectively, you can define the time interval after which a GLP test must be carried out again for the exchange unit or dosing unit.

GLP test date

Date on which the last GLP test was carried out.

Format: YYYY:MM:DD

56 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 8 Titrants

Monitoring on | off (Default value: off )

If this parameter is activated, then the time interval after which a GLP test has to be carried out again will be monitored.

GLP test interval

If you define a time interval for the GLP test, then the date in Next GLP test will be tracked automatically.

Input range

Default value

1 to 999 days

999 days

Next GLP test

If you define a date for the next GLP test, then the GLP test interval will be tracked automatically.

Format: YYYY:MM:DD

Action

Selection of the action which is carried out when the time interval has expired.

Selection

Default value

Display message | Document message | Cancel determination

Display message

For all three options it is documented in the determination data (see dialog More determination data / Messages ), that the time interval has been expired.

Display message

A message is displayed. You can select whether you want to continue with the determination or cancel the run.

Document message

In the determination data it will be documented, that the time interval has been expired.

Cancel determination

The determination is stopped.

900 Touch Control ■■■■■■■■ 57

8.7 Titer determination options and data ■■■■■■■■■■■■■■■■■■■■■■

8.7 Titer determination options and data

Titrant: Edit ▶ Titer options

Detailed information concerning the titer determination is displayed in the

Edit titrant / Titer options dialog:

Titer method

Method by which the titer was determined. If the titer was entered manually, then manual will be displayed.

User

User who carried out the titer determination.

Statistical data

The following information is also displayed for automatically assigned titers if the mean value of the results has been saved as the titer

(see

"Save as titer", page 153)

:

– n (titer det.)

Number of titer determinations.

– s abs

Absolute standard deviation

– s rel

Relative standard deviation

[Validity]

Defining the time interval for the titer validity (see chapter 8.7.1, page

58)

.

[History]

Displaying information about the last ten titer determinations (see chapter

8.7.2, page 59)

.

8.7.1 Titer validity

Titrant: Edit ▶ Titer options ▶ Validity

In the dialog Titer options / Validity , you can define the time interval after which the titer must be determined again.

Date titer det.

Date and time of the last titer determination. For new titrants, the time the preparation was made is specified until after the first time a titer determination has been carried out.

Monitoring on | off (Default value: off )

If this parameter is activated, then the titer validity will be monitored.

58 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 8 Titrants

Titer validity

Next titer determ.

If you define a date for the next titer determination, then the time interval for the Titer validity will be tracked automatically.

Format: YYYY:MM:DD

Action

If you define a time interval for the validity of the titer, then the date in

Next titer determ.

will be tracked automatically.

Input range

Default value

1 to 999 days

999 days

Selection of the action which is carried out when the time interval has expired.

Selection

Default value

Display message | Document message | Cancel determination

Display message

For all three options it is documented in the determination data (see dialog More determination data / Messages ), that the time interval has been expired.

Display message

A message is displayed. You can select whether you want to continue with the determination or cancel the run.

Document message

In the determination data it will be documented, that the time interval has been expired.

Cancel determination

The determination is stopped.

8.7.2 Properties of the previous titer determinations

Dialog "Titer options / History"

Titrant: Edit ▶ Titer options ▶ History

The date, time and titer of the last ten titer determinations are displayed in tabular form in the dialog Titer options / History . Titers that were determined automatically will be displayed in green; manually entered titer values will be displayed in black with the designation (m) . You can delete these entries, e.g. if you have opened a new bottle.

900 Touch Control ■■■■■■■■ 59

8.7 Titer determination options and data ■■■■■■■■■■■■■■■■■■■■■■

NOTICE

If multiple determinations have been carried out in order to determine the titer, then only one entry will be made in the history.

[Delete History]

[Graph]

Delete the entire history.

Open the diagram of the titer values, see following chapter.

Dialog "History / Graph"

Titrant: Edit ▶ Titer options ▶ History ▶ Graph

[Limits]

In this diagram, titer values are plotted against the date of the titer determination. You can define warning limits (blue dashed lines) and intervention limits (red dashed lines). These limits will not, however, be monitored.

Define warning and intervention limits.

60 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 9 Reagents

9 Reagents

Main dialog: System ▶ Reagents

This chapter describes how you can create a list of the reagents used in the system. Depending on the usage a distinction is made between two types of reagents:

Reagent for volumetric determinations

Reagents for coulometric determinations

[New]

[Delete]

[Edit]

The list of reagents gives the designation and type of each reagent.

Add a new reagent to the list, see following chapter.

Delete the selected reagent from the list.

Edit the data of the selected reagent, see following chapter.

900 Touch Control ■■■■■■■■ 61

9.1 Editing reagent data

9.1 Editing reagent data

Main dialog: System ▶ Reagents ▶ New / Edit

■■■■■■■■■■■■■■■■■■■■■■

Reagent

The designation of the reagent is used for unambiguous identification.

Entry 24 characters maximum

Comment

Entry 24 characters maximum

[Reagent monitoring]

Set the parameters for the reagent monitoring, see following chapter.

9.2 Reagent monitoring

The conditions for the monitoring of the reagent are defined in the dialog

Edit reagent / Reagent monitoring .

62 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 9 Reagents

Number of determ.

The number of determinations to be carried out with a certain amount of reagent depends on the type of sample and its amount.

Input range

Selection

Default value

1 to 999 off off

Working life

If one of the following values is reached, then the reagent must be replaced. The values are checked in the following cases: at the start of the determination.

at the end of the determination.

Working life of the reagent.

Input range

Selection

Default value

1 to 999 days off off

Volume

This parameter is only visible for volumetric reagents.

Volume of titrant dosed.

Input range

Selection

Default value

1.0 to 999.9 mL off off

Reagent capacity

This parameter is only visible with coulometric reagents.

Water capacity of the reagent.

Input range 1 to 9999 mg

900 Touch Control ■■■■■■■■ 63

9.2 Reagent monitoring ■■■■■■■■■■■■■■■■■■■■■■

Selection

Default value off off

Drift

This parameter is only visible with coulometric reagents.

During conditioning the measured drift has to be in the following range for 2 min: ' specified drift value + 50 µg/min '.

Input range

Selection

Default value

1 to 999 µg/min off off

[Status]

Display the status overview of the current values of the reagent monitoring.

[Reagent replacement]

Edit the parameters for the reagent exchange.

"Reagent monitoring / Status" dialog

The current reagent monitoring values are displayed in this dialog.

[Reset]

Reset the values to zero.

Dialog "Reagent monitoring / Reagent replacement"

The parameters for the reagent replacement are defined in this dialog.

Reagent replacement

The reagent can either be exchanged manually or automatically.

Selection

Default value manual manual

| auto manual

If a monitored parameter has reached the limit set, a message is being displayed. Then the reagent has to be exchanged manually.

auto

If a monitored parameter has reached the limit set, the method defined below is started automatically.

Memory

This parameter can only be edited with Reagent replacement = auto .

Memory location the method is loaded from. All memory locations are selectable, even if they are currently not accessible.

64 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 9 Reagents

Selection Internal memory | External memory 1 | External memory 2 | Shared memory

Internal memory Default value

Shared memory

Shared directory in the network.

Method

This parameter can only be edited with Reagent replacement = auto .

Method used for emptying the titration cell.

NOTICE

Make sure that the memory is accessible.

Entry

Selection

32 characters maximum

Selection of stored methods

900 Touch Control ■■■■■■■■ 65

■■■■■■■■■■■■■■■■■■■■■■

10 Sensors

Main dialog: System ▶ Sensors

This chapter describes how you can create a list of the sensors used in the system.

66 ■■■■■■■■

Five standard sensors are defined in the sensor list: pH electrode , Metal electrode , Fluoride electrode , Temperature sensor and Conductivity sensor . These sensors cannot be deleted or renamed. A maximum of

25 additional sensors can be added to these sensors.

For each sensor, the following data is displayed in the sensor list:

Designation

Type

– pH : pH electrode

– Metal : Metal electrode

– ISE : Ion-selective electrode

– Temp.

: Temperature sensor

– Cond.

: Conductivity measuring cell

– Other : Other sensor type

Measuring input/control instrument (only for intelligent sensors if they are connected)

Intelligent sensors are also indicated by IS and are depicted in green lettering.

The following sensor data is stored in the list of sensors:

Name

Each sensor in the system is identified by its unambiguous name.

Calibration data (for pH electrodes, ISE electrodes and conductivity measuring cells only)

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 10 Sensors

Calibration interval (for pH electrodes, ISE electrodes and conductivity measuring cells only)

Working life etc.

NOTICE

If data is read out from the data chip of an intelligent sensor, then a check is made whether the sensor list already contains a sensor of the same serial number. If this is the case, then the older data set will always be overwritten by the new data set, no matter whether the data set in the sensor list or the data set on the data chip is the most recent one.

[New]

[Delete]

[Edit]

Adding a new sensor to the list

(see chapter 10.1, page 67) .

Delete the selected sensor from the list.

Editing the data of the selected sensor

(see chapter 10.2, page 68) .

10.1 Adding a new sensor

Before you can use a sensor, you must add it to the sensor list. To do this, use the button [New] .

Conventional sensors:

The properties dialog is opened after the sensor type has been selected, see following chapter. The following sensor types can be selected:

– pH electrode

– Metal electrode (Pt electrode, Ag Titrode, Ag/AgCl reference electrode, etc.)

– Ion-selective electrode

– Other sensor, e.g. Spectrosense

– Temperature sensor

– Conductivity measuring cell

Intelligent sensors (also known as iTrodes):

If the 854 iConnect with iTrode is connected to a control instrument, then the sensor is automatically entered in the sensor list and can be configured, see following chapter.

900 Touch Control ■■■■■■■■ 67

10.2 Editing the sensor data

10.2 Editing the sensor data

Sensor list: Sensor ▶ New / Edit

■■■■■■■■■■■■■■■■■■■■■■

Sensor

Ion

Order number

Serial number

Comment

All of the data for the selected sensor is displayed in the dialog Sensors /

Edit .

The designation of the sensor is used for unambiguous identification.

Entry 24 characters maximum

This parameter is only visible for electrodes of the type ISE .

Display of the ion and its charge.

Order number of the sensor. With intelligent sensors it is read out automatically.

Entry

Default value

24 characters maximum empty

Serial number of the sensor. With intelligent sensors it is read out automatically.

Entry 8 digits maximum

Entry 24 characters maximum

68 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 10 Sensors

[Working life]

Defining the working life of the sensor

(see chapter 10.3, page 69) .

[Limit values]

[Calibration interval]

This button is displayed only for pH and ISE electrodes and conductivity measuring cells.

Define the time interval for the next calibration

(see chapter 10.6, page

77) .

[Calibration data]

This button is displayed only for pH and ISE electrodes and conductivity measuring cells.

Define the limit values for monitoring the slope, the electrode zero point

or the cell constant (see chapter 10.5, page 75)

.

This button is displayed only for pH and ISE electrodes and conductivity measuring cells.

Display the properties for titer determination (see chapter 10.4, page

70) .

10.3 Monitoring the working life

Sensor: Edit ▶ Working life

In the dialog Edit sensor / Working life , you can define the time interval after which the sensor must be replaced. If you do not wish to monitor the working life, then you can enter only the date of manufacture for documentation purposes.

Start-up

Date on which the sensor was used for the first time.

Format: YYYY:MM:DD

Monitoring on | off (Default value: off )

If this parameter is activated, then the working life will be monitored.

Working life

If you define a time interval for the working life, then the Expiry date will be tracked automatically.

Input range

Default value

1 to 999 days

999 days

900 Touch Control ■■■■■■■■ 69

10.4 Calibration data (for pH- and ISE electrodes and conductivity measuring cells only) ■■■■■■■■■■■■■■■■■■■■■■

Expiry date

If you define an expiry date, then the Working life will be tracked automatically.

Format: YYYY:MM:DD

Action

Selection of the action which is carried out when the time interval has expired.

Selection

Default value

Display message | Document message | Cancel determination

Display message

For all three options it is documented in the determination data (see dialog More determination data / Messages ), that the time interval has been expired.

Display message

A message is displayed. You can select whether you want to continue with the determination or cancel the run.

Document message

In the determination data it will be documented, that the time interval has been expired.

Cancel determination

The determination is stopped.

10.4 Calibration data (for pH- and ISE electrodes and conductivity measuring cells only)

Sensor: Edit ▶ Calibration data

70 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 10 Sensors

Detailed information concerning calibration is displayed in this dialog Edit sensor / Calibration data .

Slope

Slope of the electrode.

pH electrodes:

Input range

Default value

–999.9 to 999.9 %

100.0 %

ISE electrodes:

Input range –999.9 to 999.9 mV

The default value depends on the charge of the ion.

pH(0)

This parameter is only visible with pH electrodes.

pH value of the electrode at 0 mV. Apart from the slope, pH(0) is the second characteristic of the calibration curve.

Input range

Default value

–20.000 to 20.000

7.000

E(0)

This parameter is only visible with ISE electrodes.

Electrode zero point. Apart from the slope, E(0) is the second characteristic of the calibration curve.

Input range

Default value

–2000.0 to 2000.0 mV

0.0 mV c(blank)

This parameter is only visible with ISE electrodes.

Blank value concentration.

Input range

Default value

–999999999 to 9999999999

0.00

Variance

This parameter is only visible with ISE electrodes.

The variance is only calculated, if the calibration has been carried out with at least three standards.

Cell constant

This parameter is only visible for conductivity measuring cells.

Cell constant.

900 Touch Control ■■■■■■■■ 71

10.4 Calibration data (for pH- and ISE electrodes and conductivity measuring cells only) ■■■■■■■■■■■■■■■■■■■■■■

Input range

Default value

0.001 to 500.000 /cm

1.000 /cm

The following data cannot be edited:

Electrode test (only with intelligent sensors)

Result of the electrode test.

Calibration temp

Temperature at which the calibration was carried out.

If the temperature was measured manually during the calibration, then

(manual) will also be displayed. If the temperature was measured with a connected temperature sensor, then the sensor type ( (Pt1000) or

(NTC) ) will be displayed.

Calibration date

Date and time of the last calibration. For new sensors, the time the preparation was made is specified until after the first time a calibration has been carried out.

Cal. method

Method with which the sensor was calibrated. If the calibration data was entered manually, then manual will be displayed.

User

User who carried out the calibration.

Measuring input (only with intelligent sensors)

The type and the serial number of the measuring input with which the calibration was carried out.

[Initial data]

This button is only displayed for intelligent sensors.

Display the initial calibration data determined at the time of the Metrohm quality control.

[GLP test]

This button is displayed only for pH electrodes.

Define the time interval for the GLP test

(see "Dialog "Calibration data /

GLP test"", page 73) .

[History]

Display information about the last ten calibrations

(see "Dialog "Calibration data / History"", page 72)

.

10.4.1 Properties of the previous calibrations

Dialog "Calibration data / History"

Sensor: Edit ▶ Calibration data ▶ History

The date, time and calibration data of the last ten calibrations are displayed in tabular form in the dialog Calibration data / History . Calibrations that were carried out automatically will be displayed in green; man-

72 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 10 Sensors ually entered calibration data will be displayed in black with the designation (m) .

[Delete History]

Delete the entire history.

[Graph slope]

Open the diagram of the electrode slopes, see following chapter.

[Graph pH(0)]

Open the diagram of the electrode zero points, see following chapter.

[Graph E(0)]

Open the diagram of the electrode zero points, see following chapter.

[Graph cell const.]

Open the diagram of the cell constants, see following chapter.

Dialog "History / Graph"

Sensor: Edit ▶ Calibration data ▶ History ▶ Graph Slope / pH(0) /

E(0) / Cell const.

[Limits]

900 Touch Control

In this diagram, either slope, pH(0), E(0) or the cell constant is plotted against the date of the calibration. You can define warning limits (blue dashed lines) and intervention limits (red dashed lines). These limits will not, however, be monitored.

Define warning and intervention limits.

Dialog "Calibration data / GLP test"

Sensor: Edit ▶ Calibration data ▶ GLP test

In the dialog Calibration data / GLP test , you can define the time interval after which a GLP test must be carried out again for the sensor.

■■■■■■■■ 73

10.4 Calibration data (for pH- and ISE electrodes and conductivity measuring cells only) ■■■■■■■■■■■■■■■■■■■■■■

GLP test date

Monitoring on | off (Default value: off )

If this parameter is activated, then the time interval after which a GLP test has to be carried out again will be monitored.

GLP test interval

If you define a time interval for the GLP test, then the date in Next GLP test will be tracked automatically.

Input range

Default value

1 to 999 days

999 days

Next GLP test

Date on which the last GLP test was carried out. After you have carried out an electrode test (ELT command), the date of the electrode test is automatically entered into this field. However, you also can enter the date manually.

Format: YYYY:MM:DD

If you define a date for the next GLP test, then the GLP test interval will be tracked automatically.

Format: YYYY:MM:DD

Action

Selection of the action which is carried out when the time interval has expired.

Selection

Default value

Display message | Document message | Cancel determination

Display message

For all three options it is documented in the determination data (see dialog More determination data / Messages ), that the time interval has been expired.

Display message

A message is displayed. You can select whether you want to continue with the determination or cancel the run.

Document message

In the determination data it will be documented, that the time interval has been expired.

Cancel determination

The determination is stopped.

74 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

10.5 Limit values for the calibration data

Sensor: Edit ▶ Limit values

10 Sensors

Monitoring slope

You can define the following limit values in the dialog Edit sensor /

Limit values :

Slope (pH and ion-selective electrodes)

Electrode zero point (pH and ion-selective electrodes)

Cell constant (conductivity measuring cells)

These values are monitored during the calibration. If these limits are infringed, then a message will be displayed and you can decide whether to accept the calibration data or not.

on | off (Default value: off )

If this parameter is activated, then the slope will be monitored.

Lower limit pH electrodes:

Input range

Default value

ISE electrodes:

Input range

–999.9 to 999.9 %

96.0 %

–999.9 to 999.9 mV

The default value depends on the charge of the ion:

–25.0 mV (–2), –55.0 mV (–1), 55.0 mV (+1), 25.0

mV (+2).

900 Touch Control ■■■■■■■■ 75

10.5 Limit values for the calibration data ■■■■■■■■■■■■■■■■■■■■■■

Upper limit pH electrodes:

Input range

Default value

ISE electrodes:

Input range

–999.9 to 999.9 %

101.0 %

–999.9 to 999.9 mV

The default value depends on the charge of the ion:

–31.0 mV (–2), –61.0 mV (–1), 61.0 mV (+1), 31.0

mV (+2).

Monitoring pH(0) on | off (Default value: off )

This parameter is only available with pH electrodes.

If this parameter is activated, then the electrode zero point pH(0) will be monitored.

Lower limit

Input range

Default value

–20.000 to 20.000

6.750

Upper limit

Input range

Default value

–20.000 to 20.000

7.250

Monitoring E(0) on | off (Default value: off )

This parameter is only available with ISE electrodes.

If this parameter is activated, then the electrode zero point E(0) will be monitored.

Lower limit

Input range

Default value

–2000.0 to 2000.0 mV

–2000.0 mV

Upper limit

Input range

Default value

–2000.0 to 2000.0 mV

2000.0 mV

Monitoring cell constant on | off (Default value: off )

This parameter is only available for conductivity measuring cells.

76 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 10 Sensors

If this parameter is activated, then the cell constant will be monitored.

Lower limit

Input range

Default value

0.001 to 500.000 /cm

0.850 /cm

Upper limit

Input range

Default value

0.001 to 500.000 /cm

1.150 /cm

10.6 Monitoring the calibration interval (only for pH and ion-selective electrodes and conductivity measuring cells)

Sensor: Edit ▶ Calibration interval

In the dialog Edit sensor / Calibration interval , you can define the time interval after which the sensor must be recalibrated.

Calibration date

Date of the last calibration.

Monitoring on | off (Default value: off )

If this parameter is activated, then the validity of the calibration will be monitored.

Calibration interval

If you define a time interval for the validity of the calibration, then the date in Next calibration will be tracked automatically.

Input range

Default value

1 to 999 days

7 days

Next calibration

If you define a date for the next calibration, then the Calibration interval will be tracked automatically.

Format: YYYY:MM:DD

Action

Selection of the action which is carried out when the time interval has expired.

900 Touch Control ■■■■■■■■ 77

10.6 Monitoring the calibration interval (only for pH and ion-selective electrodes and conductivity measuring cells)

Selection

Default value

■■■■■■■■■■■■■■■■■■■■■■

Display message | Document message | Cancel determination

Display message

For all three options it is documented in the determination data (see dialog More determination data / Messages ), that the time interval has been expired.

Display message

A message is displayed. You can select whether you want to continue with the determination or cancel the run.

Document message

In the determination data it will be documented, that the time interval has been expired.

Cancel determination

The determination is stopped.

78 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager

11 Device manager

Main dialog: System ▶ Device manager

This chapter describes how you can configure the Touch Control, the control instruments (Titrando, Sample Processor etc.) and the connected peripheral devices. A detailed description of the hardware installation can be found in the installation manual for the control instrument.

[New]

[Delete]

[Edit]

900 Touch Control

The instrument list gives the name and type of every instrument. Connected control instruments (Titrando, Dosing Interface, Sample Processor, etc.) with the peripheral devices (dosing devices, stirrers, etc.) connected to the MSB connector are entered in the list automatically. A USB/RS-232 adapter will also be automatically recognized and entered in the instrument list with default settings. A printer is entered in the instrument list in the default settings. You must enter a PC keyboard, a barcode reader or a balance in the instrument list yourself.

Add a new device to the list (see chapter 11.1, page 80)

.

Delete the selected device from the list.

NOTICE

Devices which are recognized automatically cannot be deleted from the list while they are still connected.

Configure the selected device

(see chapter 11.2, page 80) .

■■■■■■■■ 79

11.1 Adding a new device ■■■■■■■■■■■■■■■■■■■■■■

11.1 Adding a new device

Of the following device types, you can enter one device each in the device list, even if it is not yet connected:

Titrando (three instruments maximum)

Sample Processor

Balance

Barcode reader

USB/RS-232 adapter

PC keyboard

Proceed as follows:

1 Displaying the device selection

Tap on [New] .

2 Selecting the instrument

Tap on the button for the desired device.

The new device is entered in the list.

11.2 Configuring the instrument

Instrument list: Instrument ▶ Edit

The data stored for an instrument depends on the type of instrument. You can define an instrument name and a comment for each instrument. The instrument name of the Touch Control is printed out in the standard report header.

The description of the individual instruments can be found in the following specific chapters:

Touch Control (see chapter 11.3, page 81)

Metrohm control instruments

(see chapter 11.4, page 87)

Sample Processor (see chapter 11.5, page 92)

Printer (see chapter 11.7, page 109)

Balance

(see chapter 11.8, page 113)

PC keyboard

(see chapter 11.10, page 117)

USB/RS-232 adapter (see chapter 11.9, page 115)

Barcode reader (see chapter 11.11, page 118)

80 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

11.3 Touch Control

Instrument list: 900 Touch Control ▶ Edit

11 Device manager

Device name

This designation is used for identification purposes when selecting control devices (command, manual control).

Entry 24 characters maximum

Comment

Entry 24 characters maximum

Switch off display

If this time interval has expired without the Touch Control having been operated, then the display will be switched off. Touching the display will switch it back on again at any time.

Input range

Selection

Default value

1 to 999 min off off

Program version

Program version of the software.

Serial number

Shows the serial number of the device.

Control Remote Box

Shows to which control device and MSB connector the Remote Box is connected.

900 Touch Control ■■■■■■■■ 81

11.3 Touch Control ■■■■■■■■■■■■■■■■■■■■■■

The "Control Remote Box" is the interface via which the system can be started and stopped externally. If multiple Remote Boxes are connected, then the one that is recognized first when the program starts will be used as the "Control Remote Box."

Selection Name of the control device / Number of the

MSB connector | Not available

11.3.1 E-mail

900 Touch Control: Edit ▶ E-mail

The system allows you to send displayed messages as e-mails. The Touch

Control must be connected to a network for this to function. The following types of messages can be sent:

: general warning messages

: error messages

Configuring e-mail dispatch

Proceed as follows so that messages can be sent as e-mails:

1 Activating the option

In the instrument properties of the 900 Touch Control, tap on the

[E-mail] button.

In the Edit device / E-mail dialog, activate the Send the following messages as e-mail: option.

2 Configuring e-mail addresses

Tap on the [E-mail settings] button.

The E-mail / Settings dialog is displayed.

Enter the addresses of the mail server, the sender and the desired recipient.

Parameter description

Send the following messages as e-mail: on | off (Default value: off )

If this parameter is activated, then messages with the following symbols will be sent as e-mails:

: General warning messages

: Error messages

82 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager

Send only during running determination on | off (Default value: on )

If this parameter is activated, then messages will be sent as e-mails only if a determination is running. Deactivate this parameter if messages are also to be sent in normal status.

Mail server

Address of the mail server for outgoing mail, e.g. mail.metrohm.ch. You can find the address of the mail server either in your e-mail program settings or obtain it from your IT department.

Entry

Default value

50 characters maximum empty

Sender

E-mail address of the sender. This address must be formatted as an e-mail address, but need not necessarily correspond to an existing e-mail account, e.g. [email protected].

Entry

Default value

50 characters maximum empty

Recipient

The messages will be sent to this e-mail address.

Entry

Default value

50 characters maximum empty

11.3.2 PC/LIMS report

900 Touch Control: Edit ▶ PC/LIMS report

You can generate a machine-readable report with all of the important data concerning a determination, which is referred to as a PC/LIMS report.

This report can be saved as a TXT file (according to ISO/IEC 8859-1) or as a

UTF-8 file or sent to a terminal program or a LIMS via an RS-232 interface:

■ manually with the [ ]

fixed key (see chapter 27, page 242) .

automatically at the end of a determination

(see chapter 16.5.6, page

182) .

The file name of the report file is constructed as follows: PC_LIMS_Report-

ID1-YYYYMMDD-hhmmss.txt

. A detailed description of the contents of the PC/LIMS report can be found in the PC/LIMS Report Guide .

Memory

Memory location where the PC/LIMS report is stored as a file. The report will be saved in the directory pc_lims_report . This directory will be created

900 Touch Control ■■■■■■■■ 83

11.3 Touch Control ■■■■■■■■■■■■■■■■■■■■■■ the first time a PC/LIMS report is generated. All three memory locations are listed as possible selections, even if they cannot be accessed at the moment.

Selection

Default value off | External memory 1 | External memory 2 |

Shared memory off off

The report will not be saved as a file.

Shared memory

The report will be saved in a shared directory on the network. The shared directory is selected in the Edit device / Shared memory dia-

log (see chapter 11.3.3, page 84)

.

RS-232

The RS-232 interface via which the PC/LIMS report is sent. The interface parameters are adjusted in the Edit device / Port parameters dialog

(see chapter 11.9, page 115)

.

Selection

Default value off | COM 1 | COM 2 off off

The report will not be sent via an RS-232 interface.

COM 2

This interface is inactive.

Coding

Format in which the PC/LIMS report is coded and stored.

Selection

Default value

ISO 8859-1

ISO 8859-1

| UTF-8

ISO 8859-1

This format is recommended for all languages that use the extended

ASCII code (e.g. German, English, Spanish, etc.).

UTF-8

This format is required for all languages that do not use the extended

ASCII code (e.g. Russian, Chinese, Korean, etc.).

11.3.3 Shared memory

900 Touch Control: Edit ▶ Shared memory

If you have your Touch Control connected to your network, then you can specify in this dialog a shared memory location on a PC within your network for the purpose of saving data (methods, determinations, etc.).

84 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager

CAUTION

If the computer on which you share a memory location does not have the same subnet as the Touch Control, then a WINS server must be present. The computer must be entered in this server. In order for the communication to work, the SMB protocol version 1 (SMBv1) must be activated. Since this function is not standard anymore in Windows 10, it has to be activated manually.

Computer

Host name of the computer on which a memory location is to be shared.

If you are working with Windows, then you will find the host name of the computer as follows: In the Windows command prompt window, enter the command ipconfig -all . The host name is listed together with other parameters of the computer.

Do not under any circumstances enter an IP address in this input field.

Entry

Default value max. 32 characters empty

Share name

Share name of the shared memory location (file directory) on the abovespecified computer. Please note that the Share name of a file directory often does not match the name of the file directory. The share name can be found on the Release tab in the Properties dialog of the shared file directory.

Entry

Default value max. 32 characters empty

Domain

Network domain in which the above-specified computer is located. If you are working with Windows, then you will find the domain name as follows: In the Windows command window , enter the command ipconfig

-all . The domain is listed together with other parameters of the computer.

Leave the field empty if the computer is not located within a domain.

Entry

Default value max. 32 characters empty

User

User name of the user authorized to access the shared memory location.

Entry

Default value max. 32 characters empty

900 Touch Control ■■■■■■■■ 85

11.3 Touch Control ■■■■■■■■■■■■■■■■■■■■■■

Password

Password of the user configured on the computer.

Entry

Default value max. 32 characters empty

[Connect]

Establish the network connection. If the connection has been set up correctly, then all of the input fields will become inactive and the label switches to [Disconnect] . The network connection can be disconnected with this.

11.3.4 TCP/IP settings

900 Touch Control: Edit ▶ TCP/IP settings

If you have connected your Touch Control to your network, then you have to define the network-relevant settings in this dialog. The Touch Control requires an IP address so that it is identifiable on the network. The Touch

Control can acquire the IP address either dynamically from a DHCP server or you can enter the address directly.

Get IP address automatically (DHCP) on | off (Default value: on )

If this parameter is activated, then the Touch Control will obtain its IP address directly from a DHCP server. In this case, the remaining parameters can no longer be edited.

IP address

IP address for the Touch Control. IP addresses are 32-bit numbers and are written as sequences of four decimals, each separated by a period, e.g.

"10.157.212.8".

Entry

Default value x.x.x.x

192.168.10.11

"x" is a decimal between 0 and 255.

Subnet mask

The net mask or subnet mask, together with the IP address, indicates to which network the device to be connected belongs. Subnet masks are 32bit numbers and are written as sequences of four decimals, each separated by a period.

Entry

Default value x.x.x.x

255.255.255.0

86 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Default gateway

11 Device manager

IP address for the standard gateway. A gateway sets up connections to several networks. It is located in the same subnet as the device to be configured.

Entry

Default value x.x.x.x

192.168.10.10

MAC address

Unambiguous address of the network adaptor for identification within a network. This parameter cannot be edited.

NetBIOS name

Unambiguous designation of the device within the network. This parameter cannot be edited.

11.4 Metrohm control instruments

The following Metrohm devices can be used as control instruments:

Titrandos

Sample Processors

846 Dosing Interface

867 pH Module

856 Conductivity Module

The following specifications apply for all of the control instruments mentioned above.

Additional device configurations for the Sample Processor are described in

a separate chapter (see chapter 11.5, page 92)

.

900 Touch Control

The dialog shows the control instrument with its measuring inputs, the

MSB connectors and connected peripheral devices (dosing device, stirrer,

■■■■■■■■ 87

11.4 Metrohm control instruments ■■■■■■■■■■■■■■■■■■■■■■

Remote Box). If the control instrument is not connected, then the properties dialog of the control instrument will be displayed directly

(see chapter

11.4.1, page 88)

.

[Properties]

Open the properties dialog of the highlighted entry.

The following properties can be displayed and edited to a certain extent:

Properties of the control instrument

(see chapter 11.4.1, page 88)

Properties of the measuring inputs

(see chapter 11.4.2, page 88)

This dialog is available only for devices which have a built-in measuring input.

Properties of the MSB connectors

(see chapter 11.4.3, page 90)

Properties of the peripheral devices on the MSB connector (see chapter

11.4.4, page 91)

11.4.1 Properties – Control instrument

Instrument list: Control instrument ▶ Edit ▶ Properties

Device name

This designation is used for identification purposes when selecting control devices (command, manual control).

Entry 24 characters maximum

Comment

Entry 24 characters maximum

Program version

Program version of the instrument software.

Serial number

Shows the serial number of the device.

11.4.2 Properties – Measuring input

Instrument list: Control instrument ▶ Edit ▶ Measuring input

1/2 ▶ Properties

88 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager

In this dialog, you define the type of temperature sensor you have connected to the selected measuring input.

ADC type

Type of analog-digital converter.

Serial number

Serial number of the measuring interface.

Thick film type

Type of the thick film.

Temperature sensor

Type

The instrument supports the use of two different temperature measurement techniques:

NTC (Negative Temperature Coefficient)

Pt1000 (Platinum resistance)

Select here the type that has been connected to the instrument. If an NTC sensor is used, then it is also necessary to enter two sensor characteristics.

These characteristics are listed in the specifications of the sensor.

Selection

Default value

Pt1000 | NTC

Pt1000

R (25 °C)

This parameter can only be edited with Type = NTC .

Nominal resistance of the NTC sensor at 25 °C.

900 Touch Control ■■■■■■■■ 89

11.4 Metrohm control instruments

Input range

Default value

■■■■■■■■■■■■■■■■■■■■■■

1000 to 99999 ohm

30000 ohm

The default value applies to Metrohm sensors with an NTC sensor.

B value

This parameter can only be edited with Type = NTC .

Material constant of the NTC sensor. B values of NTC sensors are frequently based on different reference temperatures (usually 25 °C and

50…100 °C). When entering the B value the influence of the second reference temperature is negligible in comparison with the measuring accuracy of an NTC sensor.

Input range

Default value

1000 to 9999 K

4100 K

The default value applies to Metrohm sensors with an NTC sensor. If no B value is given for your sensor then you can retain the default value.

11.4.3 Properties – MSB connector

Instrument list: Control instrument ▶ Edit ▶ MSB connector 1…

4 ▶ Properties

90 ■■■■■■■■

In this dialog, you can define when the request to carry out the Prepare function for connected dosing devices is to be displayed. This setting applies for all dosing devices of the selected MSB connector. The manual control contains a description of how to prepare the exchange unit and/or the dosing unit

(see chapter 28.3.3, page 262) .

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager

Switch on

Attach an exchange or dosing unit on | off (Default value: on )

If this parameter is activated, then you will be requested to prepare the dosing device when the exchange/dosing unit is attached.

Time interval on | off (Default value: on )

If this parameter is activated, then you will be requested to prepare the dosing device when the Touch Control is switched on.

on | off (Default value: off )

Activate this parameter if you wish to receive a regular request to prepare the dosing unit.

Input range

Default value

0.1 to 999.9 h

12.0 h

11.4.4 Properties – Peripheral devices

Instrument list: Control instrument ▶ Edit ▶ Peripheral device ▶ Properties

The following data is displayed for the connected dosing devices and stirrers:

Type

The display includes the device number, e.g. "800" (Dosino) or "804"

(Ti Stand)

Serial number

In the case of the 700 Dosino and the 685 Dosimat, the serial number cannot be read out.

900 Touch Control ■■■■■■■■ 91

11.5 Sample Processor ■■■■■■■■■■■■■■■■■■■■■■

11.5 Sample Processor

NOTICE

The following settings apply to all Sample Processors.

[Racks]

[Running time]

[Properties]

The dialog shows the Sample Processor with its tower (or towers), the

MSB connectors and connected peripheral devices (dosing device, stirrer,

Remote Box). If the control instrument is not connected, then the properties dialog of the control instrument will be displayed directly

(see chapter

11.5.1, page 93)

.

The settings of the MSB connectors and the connected peripheral devices

are described in Chapter "Properties – MSB connector", page 90

.

Configure sample racks (see chapter 11.6, page 100) .

Configure the running time meter.

The running time meter adds together the time while the Sample Processor is "busy" i.e. when an action is being carried out. A message with the prompt to service the device is displayed after expiry of the time limit set here. We recommend to carry out a service after 1,000 operating hours. A reset of the running time meter can only be carried out by a

Metrohm service engineer.

Open the properties dialog of the highlighted entry.

92 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager

11.5.1 Properties – Sample Processor

Instrument list: Sample Processor ▶ Edit ▶ Properties

Device name

This designation is used for identification purposes when selecting control devices (command, manual control).

Entry 24 characters maximum

Comment

Entry 24 characters maximum

Program version

Program version of the instrument software.

Serial number

Shows the serial number of the device.

Rack name

Name of the currently attached sample rack.

[Adjustment data]

Display the internal adjustment data (EEPROM data) of the Sample Processor.

CAUTION

Do not modify and save these settings under any circumstances. This data is required by the service engineer in case of positioning problems.

11.5.2 Properties – Tower

Instrument list: Sample Processor ▶ Edit ▶ Tower 1/2 ▶ Properties

900 Touch Control ■■■■■■■■ 93

11.5 Sample Processor ■■■■■■■■■■■■■■■■■■■■■■

NOTICE

In the case of a Sample Processor with 2 towers, Tower 1 is the tower on the right and Tower 2 is the tower on the left, when viewed from the front.

Maximum stroke path

Lowest permissible lift position for automatic and manual operation.

This is a safety setting. A correct specification of this value can prevent the glass of an electrode from breaking, because a titration head can not be moved below the position specified.

Input range

Default value

1 to 235 mm

235 mm

Min. beaker radius

Smallest permissible radius of the sample vessels for automatic and manual operation.

This is a safety setting. It is intended to prevent a completely loaded, wide titration head from being moved into a narrow sample vessel.

Input range

Selection

Default value off

No check takes place.

1.0 to 100.0 mm (Increment: 0.1)

If the lift is to be moved to the work position, a check is made to see whether the beaker radius

defined for the attached rack (see "Beaker radius samples", page 102) is less than this value.

off off

94 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager

Rotation increment

Amount by which the sample rack can be rotated forward or backward relative to its current position. This parameter is required in the command

MOVE for the settings Destination = rotate + and rotate .

Input range

Default value

0 to 353.99 °

5.00 °

Axial distance

Distance between the axis of rotation of the sample rack and the swing axis of the robotic arm.

for 814 USB Sample Processor:

Input range

Default value

100.0 to 300.0 mm

166.0 mm for 815 Robotic USB Sample Processor XL, 855 Robotic Titrosampler:

Input range 100.0 to 300.0 mm

Default value 196.0 mm

11.5.3 Properties – Swing Head

Instrument list: Sample Processor ▶ Edit ▶ Swing Head ▶ Properties

[Ext. pos. 1-4]

[Robotic arm]

The dialog Edit device / Properties shows a list of all external positions with the assigned swing angle and the specific work position for each.

Edit the settings that apply for all external positions

(see "Properties –

External positions 1-4", page 96)

.

Edit the robotic arm settings (see "Properties – Robotic arm", page 97) .

900 Touch Control ■■■■■■■■ 95

11.5 Sample Processor

[Adjustment data]

[Edit]

■■■■■■■■■■■■■■■■■■■■■■

Display the overview of the EEPROM data of the Swing Head. This dialog cannot be edited.

Edit specific settings of the selected external position (see "Properties –

External position", page 98)

.

Properties – External positions 1-4

Swing Head: Properties ▶ Ext. pos. 1-4

Shift position

Rinse position

Properties which apply to all four external positions can be defined in the dialog Properties / Edit external positions 1-4 .

Lift position at which the robotic arm rotates to the external positions.

Input range

Default value

0 to 'Maximum stroke path' mm

0 mm

A lift position of 0 mm corresponds to the "home position", i.e. the lift is located at the upper stop position. The maximum stroke path is defined in the properties of the tower

(see "Maximum stroke path", page 94)

.

Lift position used for rinsing.

96 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Swing increment

Input range

Default value

11 Device manager

0 to 'Maximum stroke path' mm

0 mm

A lift position of 0 mm corresponds to the "home position", i.e. the lift is located at the upper stop position. The maximum stroke path is defined in the properties of the tower

(see "Maximum stroke path", page 94) .

Amount by which the robotic arm can be swung relative to its current position. This parameter is used in the command MOVE for the settings

Destination = swing + and swing .

Input range

Default value

0.00 to 180.00 °

10.00 °

Properties – Robotic arm

Swing Head: Properties ▶ Robotic arm

Properties of the robotic arm can be defined in the dialog Properties /

Robotic arm .

Swing offset

The swing offset is a physical angle offset of a specific robotic arm model.

The required values can be found in the leaflet for the robotic arm.

Input range

Default value

–270.00 to 270.00 °

0.00 °

Maximum swing angle

Useable swing range of the robotic arm. Each robotic arm model displays a different value on the basis of its construction. The range can also be reduced if necessary. You will find the required values in the leaflet for the robotic arm.

900 Touch Control ■■■■■■■■ 97

11.5 Sample Processor

Swing radius

Rotation offset

Swing direction

■■■■■■■■■■■■■■■■■■■■■■

Input range

Default value

0.00 to 330.00 °

60.00 °

Maximum swing radius of the robotic arm. The swing radius is dependent on the length of the robotic arm and is, together with the axial distance

(see "Axial distance," page 95) , the most important variable for precise

movement to a rack position. You will find the required values in the leaflet for the robotic arm.

Input range

Default value

30.00 to 300.00 mm

110.00 mm

The rotation offset is the offset from the center of the tower to the center of the robotic arm. This value does not usually need to be changed. If a

Swing Head is to be mounted on the tower with a lateral offset, then this value can be determined by the service technician when the rack is adjusted.

Input range

Default value

–270.00 to 270.00 °

0.00 °

The swing direction of the robotic arm depends on its type. For a 2-tower model, the robotic arm must be defined as right-swinging on Tower 1 and as left-swinging on Tower 2.

Selection

Default value -

+ | -

+

Left-swinging.

-

Right-swinging.

Properties – External position

Swing Head: Properties ▶ Edit

98 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager

Angle

Work position

Properties which apply only to the selected external position can be defined in the dialog Properties / Edit external position .

Swing angle for the selected external position.

Input range

Default value

(Offset) to (Offset + max. swing range) °

60.00 °

The offset is made up of a design-dependent angle

(approx. 8…9°) together with the robotic arm offset from the robotic arm properties. The maximum swing range is also defined under the robotic arm

properties (see "Properties – Robotic arm", page

97) .

Work position for the selected external position.

Input range 0 to 'maximum stroke path' mm

A lift position of 0 mm corresponds to the "home position", i.e. the lift is located at the upper stop position. The maximum stroke path is defined in the properties of the tower

(see "Maximum stroke path", page 94) .

900 Touch Control ■■■■■■■■ 99

11.6 Sample racks ■■■■■■■■■■■■■■■■■■■■■■

11.6 Sample racks

Instrument list: Sample Processor ▶ Edit ▶ Racks

[Load]

[Copy]

[Delete]

[Edit]

100 ■■■■■■■■

In the dialog Edit device / Sample racks , you will find the list of configured racks. The attached rack is displayed in green. New racks can be loaded or created and existing ones can be edited or deleted in this dialog window.

The following data is displayed in the list:

Rack

Name of the sample rack. Metrohm standard racks are designated by their order numbers.

Positions

Number of positions on the rack.

Code

The rack code corresponds to the arrangement of magnets on the base of the rack and is read in by the Sample Processor in order to recognize the rack.

Load a new sample rack (see "Loading the sample rack", page 101)

.

Create a new sample rack by copying an existing rack (see "Creating a new sample rack", page 102) .

Delete the selected sample rack from the list.

Edit the data of the selected sample rack (see chapter 11.6.1, page 102)

.

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager

Loading the sample rack

When you receive the file for a new sample rack from your Metrohm representative, you can easily import this file into your existing system. Proceed as follows:

1 Copying the file to an external storage medium

Copy the file to a subdirectory of "Files".

If this structure is not maintained, the new rack will not be found because the software directly accesses subdirectories of "Files".

Plug in the external storage medium at the Touch Control.

2 Display the list of saved sample racks

In the Edit device / Sample racks dialog, tap on the [Load] button.

The selection of file groups on the external storage medium is displayed. If only one group is available, then the list of the saved sample rack files will be displayed directly.

Select the group with the desired sample rack.

Tap on [Show files] .

The list with the saved sample rack files is opened.

3 Load the sample rack file

Select the desired file.

Tap on [Load] .

The new sample rack is now loaded and appears in the list of available racks.

900 Touch Control ■■■■■■■■ 101

11.6 Sample racks ■■■■■■■■■■■■■■■■■■■■■■

Creating a new sample rack

Own sample racks can be simply and conveniently created by copying an existing sample rack. Proceed as follows:

1 Copy the existing rack

In the dialog Edit device / Sample racks , select a sample rack which is to be used as a template.

Tap on [Copy] .

The Sample rack / Copy dialog is displayed.

2 Enter the rack name and rack code

In the field New rack name , enter a name for the new rack.

In the field New rack code , 110000 is suggested. As a rule, this rack code is used for special racks. Apply this code or enter a new rack code and confirm with [OK] .

The new sample rack appears in the list of available racks.

11.6.1 Editing rack data

Sample rack list: Rack ▶ Edit

You can edit the data of the selected rack in the dialog Sample rack /

Edit rack data .

Beaker radius samples

Actual radius of the sample vessels at the general sample positions of the rack.

This beaker radius may not be less than the minimum beaker radius defined in the tower properties

(see "Min. beaker radius," page 94) . If the

102 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager lift is to be moved to the work position, then these two values will be compared with one another.

Input range

Selection

Default value

1.0 to 100.0 mm off off off

No check takes place.

Beaker sensor

Each time a sample position is to be moved to with the MOVE command, the beaker sensor checks whether a vessel is present. In the MOVE command, you define the action that takes place if the beaker sensor does not detect a vessel at the position being moved to.

Selection

Default value

Tower | Robotic arm | off off

Robotic arm

A Swing Head with beaker sensor must be mounted. In addition, a suitable work position must be defined for the lift, so that the robotic arm touches the sample vessel. The work position is moved to for the purpose of beaker detection.

off

No check takes place.

1. Calibration pos.

Position of the first calibration solution for automatic calibration with a

Sample Processor (see "Definition of the calibration positions", page 498) .

Input range

Selection

Default value

1 to "highest rack position"

It is imperative that the remaining buffers/standards be placed on the rack positions directly following.

Special beaker 1…n | off off

Special beaker 1…n

If a special beaker is selected as the first calibration position, then the number of buffers/standards used for calibration will determine the number of special beakers to be defined. Any rack position can be defined as a special beaker. It is, however, preferable to set them at high rack positions in order to be able to begin sample series at rack position 1. The special beakers are moved to in ascending order.

Rack offset

The rack offset is a production-related tolerance value between the upper and lower sections of the rack. This value is determined when the rack is adjusted and is displayed here. It can be changed if necessary.

900 Touch Control ■■■■■■■■ 103

11.6 Sample racks ■■■■■■■■■■■■■■■■■■■■■■

Input range

Default value

–10.00 to 10.00 °

0.00 °

[Adjust rack]

Adjust the rack (see chapter 11.6.2, page 108) .

[Lift pos. tower 1]/[Lift pos. tower 2]

Define rack specific lift positions

(see "Lift positions Tower 1/2", page

104) .

[Special beakers]

Define rack positions as special beakers

(see "Editing special beakers", page 105)

.

Lift positions Tower 1/2

Sample rack list: Rack ▶ Edit ▶ Lift pos. tower 1 / Lift pos. tower 2

Work position

You can define rack-specific lift positions in the dialog Edit rack data /

Lift positions tower 1/2 . These then apply for all rack positions except those that have been defined as special beakers.

These positions can be moved to directly under manual control and with the LIFT command. Only lift positions within the maximum stroke path can be entered. This is defined in the device properties of the tower.

NOTICE

These lift positions can also be assigned directly in manual control after

moving to the desired lift height (see chapter 28.7.1, page 270)

.

At this lift position the electrodes, stirrer and buret tips are optimally positioned for work.

104 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Rinse position

Shift position

Special position

Input range

11 Device manager

0 to 'maximum stroke path' mm

A lift position of 0 mm corresponds to the "home position", i.e. the lift is located at the upper stop position. The maximum stroke path is defined in the properties of the tower

(see "Maximum stroke path", page 94)

.

Lift position used for rinsing.

Input range

Default value

0 to 'Maximum stroke path' mm

0 mm

A lift position of 0 mm corresponds to the "home position", i.e. the lift is located at the upper stop position. The maximum stroke path is defined in the properties of the tower

(see "Maximum stroke path", page 94) .

The lift is raised to this position with each rotational movement of the rack whenever it is located at a lower lift position.

Input range 0 to 'maximum stroke path' mm

A lift position of 0 mm corresponds to the "home position", i.e. the lift is located at the upper stop position. The maximum stroke path is defined in the properties of the tower

(see "Maximum stroke path", page 94) .

For sample positions, you can define an additional lift position for special applications.

Input range 0 to 'maximum stroke path' mm

A lift position of 0 mm corresponds to the "home position", i.e. the lift is located at the upper stop position. The maximum stroke path is defined in the properties of the tower

(see "Maximum stroke path", page 94) .

Editing special beakers

Sample rack list: Rack ▶ Edit ▶ Special beakers

900 Touch Control ■■■■■■■■ 105

11.6 Sample racks ■■■■■■■■■■■■■■■■■■■■■■

[Edit]

A maximum of 16 rack positions can be defined as special beakers for each sample rack. The dialog Edit rack data / Special beakers shows a list of all special beakers with their assigned rack positions.

Edit the data of the selected special beaker, see the following.

Rack position

You can edit the data of the selected special beaker in the dialog Special beaker / Edit .

Number of rack position for selected special beaker. Any rack position can be defined as a special beaker. It is, however, preferable to set them at high rack positions in order to be able to begin sample series at rack position 1. Rack positions defined as special beakers can no longer be used as sample positions.

Input range 0 to 'maximum number of rack positions'

106 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager

Work pos. tower 1/2

Work position for the selected special beaker. One specific work position can be defined for tower 1 and tower 2.

Input range 0 to 'maximum stroke path' mm

A lift position of 0 mm corresponds to the "home position", i.e. the lift is located at the upper stop position. The maximum stroke path is defined in the properties of the tower

(see "Maximum stroke path", page 94)

.

Beaker radius

Actual radius of the selected special beaker.

This beaker radius may not be less than the minimum beaker radius defined in the tower properties

(see "Min. beaker radius," page 94)

. If the lift is to be moved to the work position, then these two values will be compared with one another.

Input range

Selection

Default value off

No check takes place.

1.0 to 100.0 mm off off

Beaker sensor

Each time this special beaker is to be moved to with the MOVE command, the beaker sensor checks whether a vessel is present. In the MOVE command, you define the action that takes place if the beaker sensor does not detect a vessel at the position being moved to.

Selection

Default value

Tower | Robotic arm | off off

Robotic arm

A Swing Head with beaker sensor must be mounted. In addition, a suitable work position must be defined for the lift, so that the robotic arm touches the sample vessel. The work position is moved to for the purpose of beaker detection.

off

No check takes place.

900 Touch Control ■■■■■■■■ 107

11.6 Sample racks ■■■■■■■■■■■■■■■■■■■■■■

11.6.2 Rack adjustment

Instrument list: Sample Processor ▶ Edit ▶ Racks ▶ Edit ▶ Adjust rack

If necessary, each sample rack can be finely adjusted, i.e. the rack offset is determined in the direction of rotation. The adjustment of a rack is usually not necessary. However, if the exact positioning of a robotic arm is necessary for a high degree of accuracy (e.g. for very small sample beakers), then a fine adjustment can be carried out.

Proceed as follows:

1 Open the dialog for rack adjustment

In the device properties of the Sample Processor, open the list of the configured sample racks.

Select the attached rack and tap on [Edit] .

The Sample rack / Edit rack data dialog is displayed.

Tap on the [Adjust rack] button.

The Edit rack data / Adjust rack dialog is displayed:

108 ■■■■■■■■

2 Carry out the adjustment

Tap on the [Start] button (NOT on the [ ] fixed key).

The rack is being initialized. The rack then moves to position 1 and the lift is lowered to the work position.

If necessary, the lift position can be corrected with the arrow keys

[ ⇧ ] and [ ⇩ ] .

Now use the arrow keys [ ⇦ ] and [ ⇨ ] to rotate the rack so that the robotic arm or the center of the titration head is positioned exactly above the center of rack position 1.

Finish the adjustment with [OK] .

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

The lift is moved to the uppermost stop.

11.7 Printer

Instrument list: Printer ▶ Edit

11 Device manager

Device name

Comment

Printer

900 Touch Control

The list of devices always includes a printer, even if no corresponding device is connected. In this dialog, you also define when you would like to have a report generated as a PDF file.

Here you can enter a device name of your choice.

Entry 24 characters maximum

Entry 24 characters maximum

Selection of the printer type.

Selection PCL Inkjet | PCL Laser | HP | Epson | Epson new | Canon | CUSTOM NEOS | off

PCL Inkjet for HP DeskJet printers.

PCL Laser for HP LaserJet printers.

CUSTOM NEOS

Due to the narrow paper size, only reports marked with # in the "List of all printable reports"

(see chapter 27.3, page 246) can be printed.

off

The reports will not be printed out on paper.

■■■■■■■■ 109

11.7 Printer

Connector

■■■■■■■■■■■■■■■■■■■■■■

Selection of the connection type for the printer.

Selection

Default value

USB

USB

| Ethernet

USB

Local printer at a USB interface.

Ethernet

Network printer.

NOTICE

The Touch Control prints the reports with a fixed resolution of 300 dpi. If you are using a printer with a resolution of 360 dpi (or a multiple thereof, e.g. an Epson), then the text will be printed out somewhat smaller than with printers with a resolution of 300 dpi (or a multiple thereof, e.g. a Canon or HP).

11.7.1 PDF settings

Printer: Edit ▶ PDF settings

The settings for saving a report as a PDF file are defined in this dialog.

Memory

Memory location where the PDF file will be saved. The report will be saved in the directory PDF_Report . This directory is created when a PDF file is generated. The file name is appended with the date and time when the determination was started and the date and time when the PDF was generated (e.g.: Identification 1 -20130320-141044-20130320-141220).

The first date and time set indicates the start of the determination, the second date and time set indicates the generation of the PDF file. This prevents previously created PDF files from being overwritten each time a new file is created, e.g. when a result is recalculated and the report is generated again.

All three memory locations are listed as possible selections, even if they cannot be accessed at the moment.

Selection

Default value off | External memory 1 | External memory 2 |

Shared memory off off

The report will not be saved as a PDF file.

110 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager

Shared memory

The report will be saved in a shared directory on the network. The shared directory is selected in the Edit device / Shared memory dialog

(see chapter 11.3.3, page 84) .

Copy or extract content allowed on | off (Default value: on )

If this option is activated, then content can be copied or deleted from the

PDF file.

Printing allowed on | off (Default value: on )

If this option is activated, then the PDF file can be printed.

Change the document allowed on | off (Default value: off )

If this option is activated, then the PDF file can be edited.

Add or change comments allowed on | off (Default value: on )

If this option is activated, then comments can be added to the PDF file.

11.7.2 Network printer

Printer: Edit ▶ Network printer

If you have connected your Touch Control to your network, then you can specify a network printer for your reports in this dialog.

NOTICE

If you have shared a memory location on a PC within your network and

if you have configured the Touch Control accordingly (see chapter

11.3.3, page 84)

, then the settings for the parameters Domain , User and Password will be applied and can then no longer be modified in this dialog.

900 Touch Control ■■■■■■■■ 111

11.7 Printer

Print server

Share name

Domain

User

■■■■■■■■■■■■■■■■■■■■■■

CAUTION

If the computer on which the network printer is configured does not have the same subnet as the Touch Control, then a WINS server must be present. The computer must be entered in this server. In order for the communication to work, the SMB protocol version 1 (SMBv1) must be activated. Since this function is not standard anymore in Windows

10, it has to be activated manually.

Host name of the print server or the computer on which the network printer is configured. If you are working with Windows, then you will find the host name as follows: In the Windows command prompt window, enter the command ipconfig -all . The host name is listed together with other parameters of the print server or the computer.

Do not under any circumstances enter an IP address in this input field.

Entry

Default value max. 32 characters empty

Share name of the network printer on the above-specified computer.

Please note that the Share name of a network printer often does not match the name of the printer. The share name can be found on the

Release tab in the Properties dialog of the released network printer.

Entry

Default value max. 32 characters empty

Network domain in which the above-specified computer is located. If you are working with Windows, then you will find the domain name as follows: In the Windows command window , enter the command ipconfig

-all . The domain is listed together with other parameters of the computer.

Leave the field empty if the computer is not located within a domain.

Entry

Default value max. 32 characters empty

User name of the user authorized to access the shared network printer.

Entry

Default value max. 32 characters empty

112 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Password

Password of the user configured on the computer.

Entry

Default value max. 32 characters empty

11.7.3 More options

Paper size

Selection of the paper size.

NOTICE

11 Device manager

The paper size selected will be ignored when the report is printed on a

CUSTOM NEOS printer.

Selection

Default value

A4 (210 mm x 297 mm) | Letter (216 mm x

279 mm)

A4 (210 mm x 297 mm)

Color on | off (Default value: off )

If this option is activated, then the report will be printed out in color.

Spooler on | off (Default value: on )

If this option is activated, then you can continue to work while the printer is printing. If this option is deactivated, then the Touch Control will be blocked while the printout is being produced.

11.8 Balance

List of devices: Balance ▶ Edit

900 Touch Control ■■■■■■■■ 113

11.8 Balance ■■■■■■■■■■■■■■■■■■■■■■

Device name

Comment

Balance type

RS-232

You define the type of balance and its connector in the dialog Device manager / Edit .

Here you can enter a device name of your choice.

Entry 24 characters maximum

Entry 24 characters maximum

If you have connected a balance, then you have to specify the balance type here.

Selection

Default value

AND | Mettler | Mettler AT | Mettler AX |

Ohaus | Precisa | Sartorius | Shimadzu

Sartorius

RS-232 interface to which the balance is connected. The interface parameters are adjusted in the Edit device / Port parameters dialog

(see chapter 11.9, page 115)

.

Selection

Default value

COM 1 | COM 2

COM 1

COM 2

The interface is disabled.

The following table indicates the balance type that needs to be selected for the balance model:

114 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager

Balance

AND

Mettler AB, AE, AG, AM,

AJ, PE, PM, PJ, PR, XP, XS

Mettler AT

Mettler AX, MX, UMX, PG,

AB-S, PB-S

Ohaus Voyager, Explorer,

Analytical Plus

Precisa

Sartorius

Shimadzu BX, BW

Balance type

AND

Mettler

Mettler AT

Mettler AX

Ohaus

Precisa

Sartorius

Shimadzu

11.9 USB/RS-232 adapter

List of devices: USB/RS-232 adapter ▶ Edit ▶ COM interface ▶ Edit

As a rule, balances are equipped with a serial RS-232 interface. To connect a balance, you will require the 6.2148.050 cable. You can adjust the interface parameters in the dialog Edit device / Port parameters . These settings must match the settings on the connected device.

NOTICE

If you make changes to parameter settings in this dialog, then you must switch the Touch Control off and back on in order for the changes to take effect.

900 Touch Control ■■■■■■■■ 115

11.9 USB/RS-232 adapter

Baud rate

Data bits

Parity

Stop bits

Handshake

■■■■■■■■■■■■■■■■■■■■■■

Transfer rate in characters per second.

Selection

Default value

1200

57600 | 115200

9600

| 2400 | 4800 | 9600 | 19200 | 38400 |

Number of data bits.

Selection

Default value

7 | 8

8

Type of parity testing.

Selection

Default value even | odd | none none

Number of stop bits.

Selection

Default value

1 | 2

1

Type of data transfer protocol.

NOTICE

In case of communication problems, try the software handshake ( Software (XON/XOFF) ).

Selection none | Software (XON/XOFF) | Hardware (DTR/

CTS)

Hardware (DTR/CTS) Default value

Software (XON/XOFF)

Use the software handshake when you send a PC/LIMS report via an

RS-232 interface.

116 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager

11.10 PC keyboard

Instrument list: PC keyboard ▶ Edit

A commercially available USB keyboard can be connected to make text and numerical input easier. It will be recognized automatically and entered in the list of devices with default settings.

Device name

Here you can enter a device name of your choice.

Entry 24 characters maximum

Comment

Keyboard layout

Entry 24 characters maximum

Define the country-specific keyboard layout here.

Selection

Default value

English US | German DE | French FR | Spanish

ES | German CH

English US

To enter texts and numbers with the PC keyboard, the appropriate text and number input dialog must be opened on the Touch Control. Only the following keys on the PC keyboard have a function:

Table 6 Key assignment

Function in the editing dialog PC keyboard key

[Cancel] [Esc]

900 Touch Control ■■■■■■■■ 117

11.11 Barcode reader ■■■■■■■■■■■■■■■■■■■■■■

[

Function in the editing dialog PC keyboard key

Enter respective character

[OK]

⌫ ]

Line break (for multi-line text input)

Numbers, letters and special characters + Shift

[Enter] / [ ↲ ]

[ ← ] (backspace)

[Enter] / [ ↲ ]

The cursor is moved up or down by one character.

The cursor is moved to the left or to the right by one character.

[

[ ↑ ] [ ↓ ]

← ] [ → ]

[F1]

[F2]

[F3]

[F4]

[F5]

[F6]

[F7]

NOTICE

The lettering of the USB keyboard may differ from above lettering, depending on the country-specific keyboard used.

11.11 Barcode reader

List of devices: Barcode reader ▶ Edit

118 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 11 Device manager

A barcode reader can be connected to read in sample data or other texts.

A connected barcode reader will be recognized automatically and entered in the list of devices with default settings.

You will hear an acoustic signal as confirmation that a character string has been transmitted by the barcode reader and accepted.

NOTICE

The data will be read only if the system is in normal status, i.e. when no determination is running.

Device name

Here you can enter a device name of your choice.

Entry 24 characters maximum

Comment

Entry 24 characters maximum

Barcode input target

Selection of the input field for the character string read in by the barcode reader.

Selection Active input field | Method | Identification 1

Identification 2 | Sample size

|

Active input field

The character string is entered in the input field of the opened text- or number-input dialog.

Method

The character string is entered in the input field Method .

900 Touch Control ■■■■■■■■ 119

11.11 Barcode reader

Keyboard layout

■■■■■■■■■■■■■■■■■■■■■■

Identification 1

The character string is entered in the input field Identification 1 .

Identification 2

The character string is entered in the input field Identification 2 .

Sample size

The character string is entered in the input field Sample size . Character strings containing characters other than numbers and decimal separators will be ignored.

Specify the country-specific keyboard layout for the emulation of the PC keyboard. This setting must match the setting on the barcode reader (see documentation for the barcode reader).

Selection

Default value

English US | German DE | French FR | Spanish

ES | German CH

English US

120 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 12 File manager

12 File manager

Main dialog: System ▶ File manager

The saved methods, determinations, sample tables, etc. are managed in the file manager. You can also create a backup of your system (all data and settings). Similarly, an existing backup can be reloaded.

The file memory is organized as follows:

Internal memory

The following files can be stored in the internal memory:

– Methods

External memory

You can use a USB flash drive as an auxiliary storage medium, for instance. The following files can be stored on an external memory:

– Backup

– Methods

– Determinations

– Sample tables

– Result tables

– Reports as *.pdf

– PC/LIMS report

12.1 Managing files

File manager: Internal memory / External memory 1 / External memory 2

The saved files can be organized into groups. These groups are comparable to file directories on your PC, although unlike your PC, only one level is possible.

NOTICE

The file names must be unique for each memory location, i.e. you cannot save two files with the same name, even if they are in different groups.

900 Touch Control ■■■■■■■■ 121

12.1 Managing files ■■■■■■■■■■■■■■■■■■■■■■

NOTICE

If you use an external storage medium with the FAT or FAT32 file system, then you can save a maximum of 999 files per group. If you find it necessary to store more than 999 files in a single group, then you must reformat the storage medium with the file system ExFAT

(see chapter

32.10.3, page 519)

.

[Show all]

[Show files]

Display the list of all the files in the selected memory location.

Display the list of all the files in the selected group.

[Rename]

[Properties]

Rename the selected file

(see chapter 12.1.2, page 123) .

Display the file properties (see chapter 12.1.3, page 124)

.

122 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

[Copy]

12 File manager

Copy the selected file to a different memory location

(see chapter 12.1.1, page 123)

.

[Delete]

Delete the selected file.

[Load]

Load the selected file.

12.1.1 Copying a file

Proceed as follows to copy a file:

1 Selecting the file

Select the desired file.

Tap on [Copy] .

The selection of memory locations is displayed.

2 Copy the file

Tap on the desired memory location.

Only memory locations currently being accessible are active.

The file is copied. The file group is retained, i.e. the group will be created again in the new memory location if it does not already exist there.

12.1.2 Renaming a file

Proceed as follows to rename a file:

1 Selecting the file

Select the desired file.

Tap on [Rename] .

The Show files / Rename dialog is displayed.

2 Change the file name

Tap on the input field File name .

The text editor opens.

Enter a new file name (max. 32 characters) and confirm with

[OK] .

Tap on [OK] .

The file is saved under the new name.

900 Touch Control ■■■■■■■■ 123

12.1 Managing files ■■■■■■■■■■■■■■■■■■■■■■

12.1.3 File properties

File manager: Internal memory / External memory 1 / External memory 2 ▶ Show all / Show files ▶ Properties

Detailed information concerning the file is displayed in the Show files /

Properties dialog. They can be edited to a certain extent.

File name

Name of the file.

File group

Entry or selection of the group in which the file will be saved. If you enter a new name, then the file will be moved into the new group.

Write protection

If this option is activated, then the file cannot be saved, deleted, moved or renamed. This is only an internal write protection feature and is independent of the write protection of the operating system on your computer.

Created by

User who saved the file initially.

Created on

Date and time at which the file was saved for the first time.

Last saved by

User who saved the file most recently.

Last saved on

Date and time at which the file has been saved most recently. Additionally, the version of the file is displayed. The version number will be increased by one each time the file is saved.

Size

File size.

Program version

Program version of the software with which the file was last saved.

124 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

12.2 External storage medium

12 File manager

[Backup]

[Restore]

[Memory info]

Creating a backup of all data and settings on this storage medium

(see chapter 12.3, page 127) .

Load the backup. This function is active only if a backup is available

(see chapter 12.3, page 127) .

Display information on the storage medium, see following chapter.

Storage medium – Information

Detailed information about the storage medium is displayed in the dialog

External memory 1/2 / Memory info :

Name

Designation of the storage medium.

Write protection

If this option is activated, then the file cannot be saved, deleted, moved or renamed. This is only an internal write protection feature and is independent of the write protection of the operating system on your computer.

Storage capacity

Total capacity of the storage medium.

Used memory

Storage capacity used on the storage medium.

Free memory

Free storage capacity on the storage medium.

900 Touch Control ■■■■■■■■ 125

12.2 External storage medium ■■■■■■■■■■■■■■■■■■■■■■

Directory structure

The directory structure appears as follows on the external storage medium:

Figure 10 Directory structure on the external storage medium

Backup All of the files of the backup are stored in this directory. The directory will be created the first time a backup is created.

Files pc_lims_report

This directory contains all of the groups as well as the files contained in them (methods, determinations, etc.).

PC/LIMS reports are stored in this directory as

TXT files. The directory will be created the first time a PC/LIMS report is printed out.

PDF_Report

Profiles

SwUpdates

PDF reports are stored in this directory. The directory will be created the first time a report is saved as a PDF file.

The identification profile is stored in this directory. The directory will be created the first time an identification profile is generated.

Files for software updates must be saved in this directory.

126 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 12 File manager

12.3 Creating backups / Restoring data

File manager: External memory 1 / External memory 2

You can use the [Backup] function to easily create a backup containing all the data and settings of your system. You should make a backup at regular intervals in order to avoid data loss.

NOTICE

Only one backup can be created on a storage medium.

If a backup has already been stored on the medium, then it will be overwritten when this function is carried out again.

12.3.1 Restoring data

You can use the function [Restore] to restore either a complete backup or only certain data.

CAUTION

All of the methods in the internal memory will be deleted.

Backups are not backward compatible, i.e. backups of the latest version cannot be restored using earlier versions.

Proceed as follows to restore backed-up data:

1 Select the storage medium

Connect the storage medium on which the backup is located.

Select the storage medium under System ▶ File manager .

2 Select data

Tap on [Backup] .

The selection of the data that can be restored separately is displayed (see the following).

Deselect the data sets which are not to be restored.

Tap on [Load data] .

The selected data blocks are restored.

3 Complete the restoration

Switch the Touch Control off and on again.

900 Touch Control ■■■■■■■■ 127

12.3 Creating backups / Restoring data ■■■■■■■■■■■■■■■■■■■■■■

Restorable data blocks

The following data blocks can be loaded individually:

Methods

All of the methods stored in the internal memory.

Sample table

Current sample table.

Result table

Current result table.

Current determination data

All the data for the current determination (including the method with which the determination was carried out).

Control

Settings in the control dialog.

User list

Settings for each user under System settings / User administration .

System settings / User admin.

All of the system settings, including device-specific dialog configuration and dialog options for the command list and fixed keys, device-specific settings for the user administration (login options, password options and Audit Trail).

Titrants

All titrants with all their data.

Sensors

All sensors with all their data.

Device data

All of the devices configured in the device manager, with all of the data.

GLP data

All data from the GLP manager.

Deactivate if the backup is to be loaded onto a different system.

Common variables

All common variables.

Templates

All templates for sample data, result calculations, calibration buffers, etc.

Routine dialog settings

Current routine dialog settings (see Dialog options / Routine dialog ).

Sample racks

The sample racks present in the properties of the USB sample processor.

Own subsequences

All subsequences created and stored by the user.

128 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 13 GLP manager

13 GLP manager

Main dialog: System ▶ GLP manager

In the GLP manager you can document data for various GLP tests. The results of the automatic system tests carried out after switching on are also documented.

The following tests can be documented:

GLP test "Measurement" (see chapter 13.3, page 131)

GLP test "Titration" (see chapter 13.3, page 131)

System validation (see chapter 13.4, page 135)

In addition, you can also:

Create a list with your test tools

(see chapter 13.2, page 130)

.

Define a service interval for having routine maintenance carried out by

a Metrohm service technician (see chapter 13.5, page 139) .

Define an interval for the regular performance of backups

(see chapter

13.5, page 139) .

Further information about quality management and validation is also given in the documentation series Quality Management with Metrohm and

Application Bulletin AB 252 (Validation of Metrohm titrators (potentiometric) in accordance with GLP/ISO 9001).

[Test tools]

The table shows the last time that each test was carried out and when the next test is to be carried out. A test is entered in the list the first time that it is documented.

Configuring the test tools for GLP tests

(see chapter 13.2, page 130) .

900 Touch Control ■■■■■■■■ 129

13.1 Automatic system test

[GLP tests HW/SW]

■■■■■■■■■■■■■■■■■■■■■■

Document the GLP tests "Measurement" and "Titration"

(see chapter

13.3, page 131) .

[System validation]

Document the system validation

(see chapter 13.4, page 135)

.

[Monitoring]

Defining the time intervals for system maintenance and backups (see chapter 13.5, page 139)

.

[View test data]

Display the data of the selected test.

13.1 Automatic system test

The system test is carried out automatically when the Touch Control is switched on. The result of each individual test is shown in green if no error has occurred. If a result is shown in red, then an error occurred when the respective test was carried out. Switch the Touch Control off and back on again. If the error still occurs pleased notify Metrohm Service.

Print report at system start on | off (Default value: off )

If this parameter is activated, then the result of the system test will be printed out automatically when the Touch Control is switched on.

13.2 Test tools

Main dialog: System ▶ GLP manager ▶ Test tools

You can create a list of test tools to be used in the tests in the GLP manager.

The following test tools have already been defined:

767 Calibrated Reference : Device for checking measuring inputs and electrode cable.

822 Titr.Curve Simulator : Curve simulator for checking the hardware and software.

773 pH/mV Simulator : Device for checking measuring inputs and electrode cable.

868 UR Generator : Device for checking measuring inputs and electrode cable.

[New]

Add a new test tool to the list.

130 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

[Delete]

Delete the selected test tool from the list.

[Edit]

Change the designation of the selected test tool.

13 GLP manager

13.3 GLP tests for measurement and titration

Main dialog: System ▶ GLP manager ▶ GLP tests HW/SW ▶ GLP test "Measurement" / GLP test "Titration"

You can document the GLP tests for measurements and titrations in the dialog GLP manager / GLP tests Hardware/Software and its subdialogs. The following procedure describes the GLP test "Measurement," but also applies for the GLP test "Titration".

Proceed as follows:

1 Open the properties dialog

In the System / GLP manager dialog, tap on the [GLP tests

HW/SW] button and then on the [GLP test "Measurement"] button.

900 Touch Control

2 Edit data

Define the test method, the test results, etc. (see "Dialogs "GLP tests HW/SW / GLP test "Measurement"" and "GLP tests HW/SW /

GLP test "Titration""", page 132)

.

3 Define the hardware used

Tap on [Hardware] .

■■■■■■■■ 131

13.3 GLP tests for measurement and titration ■■■■■■■■■■■■■■■■■■■■■■

Select the hardware used for the test (see "Dialogs "GLP test

"Measurement" / Hardware" and "GLP test "Titration" / Hardware"", page 134)

.

Tap on the fixed key [ ] .

The dialog GLP tests HW/SW / GLP test "Measurement" appears again.

4 Define the test interval

Tap on [GLP test interval] .

Enter the time interval or the date for the next GLP test (see "Dialogs "GLP test "Measurement" / Test interval" and "GLP test

"Titration" / Test interval"", page 133)

.

Tap on the fixed key [ ] .

13.3.1 Parameter description

Dialogs "GLP tests HW/SW / GLP test "Measurement"" and

"GLP tests HW/SW / GLP test "Titration""

Method

Method the GLP test has been carried out with.

Entry

Selection

32 characters maximum

Selection of methods stored in the internal memory

User

User who carried out the GLP test.

Entry

Selection

24 characters maximum

Selection of configured users

Test date

Date on which the GLP was carried out.

Format: YYYY:MM:DD

Test result

Result of the test.

Selection

Default value

Test OK | Test not OK

Test OK

Comment

Entry 24 characters maximum

132 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 13 GLP manager

[GLP test interval]

Defining the time interval for the GLP test (see "Dialogs "GLP test "Measurement" / Test interval" and "GLP test "Titration" / Test interval"", page

133) .

[Hardware]

Documenting the hardware with which the GLP test was carried out

(see

"Dialogs "GLP test "Measurement" / Hardware" and "GLP test "Titration" / Hardware"", page 134)

.

Dialogs "GLP test "Measurement" / Test interval" and "GLP test "Titration" / Test interval"

Test date

Monitoring on | off (Default value: off )

If this parameter is activated, then the time interval after which a GLP test has to be carried out again will be monitored.

GLP test interval

If you define a time interval for the GLP test, then the date in Next GLP test will be tracked automatically.

Input range

Default value

1 to 999 days

999 days

Next GLP test

Date on which the GLP was carried out.

Format: YYYY:MM:DD

If you define a date for the next GLP test, then the GLP test interval will be tracked automatically.

Format: YYYY:MM:DD

Action

Selection of the action which is carried out when the time interval has expired.

Selection

Default value

Display message | Document message | Cancel determination

Display message

For all three options it is documented in the determination data (see dialog More determination data / Messages ), that the time interval has been expired.

900 Touch Control ■■■■■■■■ 133

13.3 GLP tests for measurement and titration ■■■■■■■■■■■■■■■■■■■■■■

Display message

A message is displayed. You can select whether you want to continue with the determination or cancel the run.

Document message

In the determination data it will be documented, that the time interval has been expired.

Cancel determination

The determination is stopped.

Dialogs "GLP test "Measurement" / Hardware" and "GLP test

"Titration" / Hardware"

Test tool

Selection of the test tool. Test tools are defined at GLP manager ▶ Test tools .

Selection Selection of configured test tools

Control device

Selection of the control device with which the GLP test has been carried out.

Selection Selection of configured control devices

Measuring input

Selection of the measuring input used for the GLP test. The selection is not dependent on whether the control device has one or two measuring interfaces.

Selection

Default value

1 | 2

1

Dosing device

Selection of the MSB connector to which the dosing device is connected.

Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4

134 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 13 GLP manager

13.4 System validation

Main dialog: System ▶ GLP manager ▶ System validation

In the dialog GLP manager / System validation and its subdialogs, you can document the system validation results and define the time interval after which validation must be carried out again.

Proceed as follows:

1 Open the properties dialog

In the System / GLP manager dialog, tap on the [System validation] button.

900 Touch Control

2 Edit data

Define the method, the result, etc. (see ""GLP manager / System validation" dialog", page 136) .

3 Define the validation interval

Tap on [Validation interval] .

Enter the time interval or the date for the next system validation

(see "Dialog "System validation / Validation interval"", page 137)

.

Tap on the fixed key [ ] .

The dialog GLP manager / System validation appears again.

4 Enter statistical data

Tap on [Test statistics] .

Enter the statistical data for the last system validation (see ""System validation / Test statistics" dialog", page 138) .

Tap on the fixed key [ ] .

■■■■■■■■ 135

13.4 System validation ■■■■■■■■■■■■■■■■■■■■■■

The dialog GLP manager / System validation appears again.

5 Insert a note

Tap on [Note (SOP)] .

Enter a brief description, e.g. a summary of the SOP (standard operating procedure) according to which the system validation

was carried out (see ""System validation / Note (SOP)" dialog", page 137) .

Tap on the fixed key [ ] .

13.4.1 Parameter description

"GLP manager / System validation" dialog

Method

Method with which the system validation has been carried out.

Entry

Selection

32 characters maximum

Selection of methods stored in the internal memory

User

User who carried out the system validation.

Entry

Selection

24 characters maximum

Selection of configured users

Test date

Date on which the system validation was carried out.

Format: YYYY:MM:DD

Test result

Result of the test.

Selection

Default value

Test OK | Test not OK

Test OK

Comment

Entry 24 characters maximum

[Note (SOP)]

Enter a brief description, e.g. a summary of the SOP (standard operating procedure) according to which the system validation has been carried out.

[Validation interval]

Defining the time interval for the system validation (see "Dialog "System validation / Validation interval"", page 137)

.

136 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

[Test statistics]

13 GLP manager

Documenting the statistical data of the system validation

(see ""System validation / Test statistics" dialog", page 138) .

"System validation / Note (SOP)" dialog

In this dialog, you can enter a brief text, e.g. a summary of the SOP (standard operating procedure) according to which the system validation was carried out.

Dialog "System validation / Validation interval"

Last validation

Date on which the last system validation was carried out.

Format: YYYY:MM:DD

Monitoring on | off (Default value: off )

If this parameter is activated, then the time interval after which a system validation has to be carried out again will be monitored.

Validation interval

If you define a time interval for the system validation, then the date in

Next validation will be tracked automatically.

Input range

Default value

1 to 999 days

999 days

Next validation

If you define a time interval for the next system validation, then the Validation interval will be tracked automatically.

Format: YYYY:MM:DD

Action

Selection of the action which is carried out when the time interval has expired.

Selection

Default value

Display message | Document message | Cancel determination

Display message

For all three options it is documented in the determination data (see dialog More determination data / Messages ), that the time interval has been expired.

900 Touch Control ■■■■■■■■ 137

13.4 System validation

Number (n)

Mean value s abs s rel d rel a sys

■■■■■■■■■■■■■■■■■■■■■■

Display message

A message is displayed. You can select whether you want to continue with the determination or cancel the run.

Document message

In the determination data it will be documented, that the time interval has been expired.

Cancel determination

The determination is stopped.

"System validation / Test statistics" dialog

In the dialog System validation / Test statistics , you can document the statistical data for the last system validation.

Number of determinations carried out.

Input range

Default value

1 to 9999999999 empty

Mean value out of single results.

Input range

Default value

–999999999.00000 to 999999999.00000 empty

Absolute standard deviation of the results.

Input range

Default value

0.00000 to 999999999.00000 empty

Relative standard deviation of the results.

Input range

Default value

0.00000 to 100.00000 % empty %

Systematic deviation of the results.

Input range

Default value

0.00000 to 100.00000 % empty %

Systematic error.

Input range

Default value

0.00000 to 9999999999.00000 empty

138 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 13 GLP manager

13.5 System monitoring

13.5.1 Service interval

Main dialog: System ▶ GLP manager ▶ Monitoring ▶ Service interval

In the dialog GLP manager / Service interval , you can define the time interval for the maintenance of the system by the Metrohm Service department. The service interval is checked each time the system is started.

Last service

Date on which the last servicing was carried out.

Format: YYYY:MM:DD

Monitoring on | off (Default value: off )

If this parameter is activated, then the time interval after which system maintenance must be carried out again will be monitored.

Service interval

If you define a time interval for the system maintenance, then the date in

Next service will be tracked automatically.

Input range

Default value

1 to 999 days

999 days

Next service

If you specify a date for the next system maintenance, then the Service interval will be tracked automatically.

Format: YYYY:MM:DD

13.5.2 Backup interval

Main dialog: System ▶ GLP manager ▶ Monitoring ▶ Backup interval

In the dialog GLP manager / Backup interval , you can define the time interval for backups. The backup interval is checked each time the system is started.

Last backup

Date on which the last backup was created.

Format: YYYY:MM:DD

900 Touch Control ■■■■■■■■ 139

13.5 System monitoring ■■■■■■■■■■■■■■■■■■■■■■

Monitoring on | off (Default value: off )

If this parameter is activated, then the time interval after which a backup has to be created again will be monitored.

Backup interval

If you define a time interval for the creation of backups, then the date in

Next backup will be tracked automatically.

Input range

Default value

1 to 999 days

999 days

Next backup

If you specify a date for the next backup, then the Backup interval will be tracked automatically.

Format: YYYY:MM:DD

140 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 14 Common variables

14 Common variables

Main dialog: System ▶ Common variables

You can save 25 method-independent variables , or common variables.

These variables can be used in future calculations (as variables CV01…

CV25 ). Common variables are useful, e.g. for the following applications:

Determination of a blank value which will be taken into account during the content determination of the sample.

Determination of the content of a standard solution, which will be taken into account during the content determination of the sample.

[Delete]

[Edit]

The designation and the value (including the unit) are displayed for every common variable in the list.

Delete the selected common variable from the list.

Edit the data of the selected common variable, see following chapter.

900 Touch Control ■■■■■■■■ 141

14.1 Editing common variables ■■■■■■■■■■■■■■■■■■■■■■

14.1 Editing common variables

List of common variables: Common Variable ▶ Edit

The common variables can be modified as follows:

Edit manually, see the following.

Automatic assignment from the determination run. A calculation result must be configured accordingly for this purpose

(see chapter 14.4, page 145) .

Name

Value

Designation of the common variable.

Entry

Default value

24 characters maximum empty

Value of the common variable.

Input range

Default value

–999999999 to 9999999999 empty

Unit of the common variable.

Entry

Default value

Selection

Default value

10 characters maximum empty

% | mol/L | mmol/L | g/L | mg/L | mg/mL | mg/100 g | ppm | g | mg | µg | mL | µL | mg/ piece | °C | µg/min | mL/min | µL/min

%

142 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Next assignment

14 Common variables

This parameter is displayed only if a validity has been defined for the common variable.

Date on which the validity of the common variable expires.

Format: YYYY:MM:DD

[Validity]

Defining the time interval for the validity of the common variable (see chapter 14.3, page 144) .

[Properties]

Display the properties of the common variable, see following chapter.

14.2 Properties of common variables

Common variable: Edit ▶ Properties

Detailed information concerning the common variable is displayed in the

Edit common variables / Properties dialog:

Status

Status of the common variable. If the time interval for the validity has expired, then invalid will be displayed.

Method

Method with which the value has been assigned to the common variable. If the value was entered manually, then manual will be displayed.

Method status (only for automatic assignment from the determination run)

Determination status (only for automatic assignment from the determination run)

Last assignment

Date and time of the last assignment.

User

User who assigned the value to the common variable.

900 Touch Control ■■■■■■■■ 143

14.3 Monitoring validity ■■■■■■■■■■■■■■■■■■■■■■

14.3 Monitoring validity

Common variable: Edit ▶ Validity

In the dialog Edit common variables / Validity , you can define the time interval after which a new value must be assigned to the common variable.

Last assignment

Date on which the common variable was last assigned a value.

Format: YYYY:MM:DD

Monitoring on | off (Default value: off )

If this parameter is activated, then the time interval after which the common variable must be assigned a new value will be monitored.

Validity

Next assignment

If you specify a date for the next assignment, then the time interval for the

Validity will be tracked automatically.

Format: YYYY:MM:DD

Action

If you define a time interval for the validity of the common variable, then the date in Next assignment will be tracked automatically.

Input range

Default value

1 to 999 days

999 days

Selection of the action which is carried out when the time interval has expired.

Selection

Default value

Display message | Document message | Cancel determination

Display message

For all three options it is documented in the determination data (see dialog More determination data / Messages ), that the time interval has been expired.

Display message

A message is displayed. You can select whether you want to continue with the determination or cancel the run.

144 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 14 Common variables

Document message

In the determination data it will be documented, that the time interval has been expired.

Cancel determination

The determination is stopped.

14.4 Assigning a result automatically to a common variable

NOTICE

This instruction is based on the assumption that the method contains a calculation command with a calculation.

Proceed as follows to assign a result to a common variable:

1 Opening the editing dialog of the result

In the command list, select the command CALC .

Tap on the [Edit command] button.

Select the result whose value is to be assigned to a common variable and tap on [Edit] .

The editing dialog of the result is displayed.

2 Define result options

Tap on the [Result options] button.

900 Touch Control

Activate the parameter Save as common variable .

Define Variable = CV01…CV25 .

■■■■■■■■ 145

14.4 Assigning a result automatically to a common variable ■■■■■■■■■■■■■■■■■■■■■■

3 Saving the settings

Tap on the fixed keys [ ] or [ ] .

In the future, the calculated result will be assigned to the selected common variable (result name, value and unit).

146 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 15 Templates

15 Templates

Main dialog: System ▶ Templates

You have the option of defining system-specific templates . You can use these templates when editing the respective data.

You can create the following templates:

Sample data

(see chapter 15.1, page 147)

Create sample identifications and sample assignments.

Custom result templates

(see chapter 15.2, page 151)

Define formulas for result calculations.

Input lines

(see chapter 15.3, page 155)

Define input signals on the remote interface.

Output lines

(see chapter 15.4, page 157)

Defining output signals on the remote interface.

Custom calib. buffers

(see chapter 15.5, page 160)

Define a buffer series for the calibration of pH electrodes.

Report header

(see chapter 15.6, page 162)

Create a system-specific report header that contains information concerning the laboratory, for example.

Custom electrode type

(see chapter 15.7, page 163)

Define customized limit values for the electrode test of pH electrodes.

15.1 Sample data

Main dialog: System ▶ Templates ▶ Sample data

900 Touch Control

You can create the following templates in the dialog Templates / Sample data :

■■■■■■■■ 147

15.1 Sample data ■■■■■■■■■■■■■■■■■■■■■■

Sample identifications (see chapter 15.1.1, page 148)

List with sample identifications. If you must enter the sample identification before starting a determination, then you can select the entries contained in this list.

Sample assignments (see chapter 15.1.2, page 149)

You can assign a particular method to a sample identification.

Use sample identification list on | off (Default value: off )

Activate this parameter so that the sample identifications defined in the list will be displayed as selections.

Use sample assignment table on | off (Default value: off )

Activating/deactivating the use of the sample assignment table.

NOTICE

If this parameter is activated, the following will no longer be possible:

Carrying out determinations without their sample identification being defined in the sample assignment table.

Defining a method in the sample table. Previously defined methods will be ignored.

Assignment ident.

Identification which is used as assignment identification to load the correct method.

Selection

Default value

Identification 1 | Identification 2

Identification 1

Request assignment identification on | off (Default value: off )

If this parameter is activated, then the sample identification will be requested automatically at the start of a determination.

15.1.1 Sample identification list

Main dialog: System ▶ Templates ▶ Sample data ▶ Sample identification list

148 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 15 Templates

You can create a system-specific list with sample identifications in the dialog Sample data / Sample identification list . This list makes it easier for you to enter frequently used sample identifications, i.e. this list is available to you for entering them (main dialog, command REQUEST , etc.). It may be a good idea to define the unchanging part of the identification as a template and to add the variable part during sample data input.

[New]

Add a new sample identification to the list.

[Delete]

Delete the selected sample identification from the list.

[Edit]

Change the designation of the selected sample identification.

15.1.2 Sample assignment table

Main dialog: System ▶ Templates ▶ Sample data ▶ Sample assignment table

900 Touch Control ■■■■■■■■ 149

15.1 Sample data

[New]

[Delete]

[Edit]

Identification

Memory

Method

■■■■■■■■■■■■■■■■■■■■■■

In the sample assignment table, a particular method is assigned to a sample identification. In this way, you ensure that your samples will be processed with the correct method; mix-ups are not longer possible. When you start a determination, you need only enter the sample identification; the method is loaded automatically.

Add a new sample assignment to the list, see following chapter.

Delete the selected sample assignment.

Edit the selected sample assignment, see following chapter.

Editing the sample assignment

Identification of the sample.

NOTICE

You can place an * as a wildcard at the beginning or end of the character string. Doing so allows you to prefix or suffix a sequential number, for example, to the identification, which will be ignored when methods are being assigned.

Entry

Selection

24 characters maximum

Selection of defined identifications in the sample identification list

Memory location the method is loaded from. All memory locations are selectable, even if they are currently not accessible.

Selection

Default value

Internal memory | External memory 1 | External memory 2 | Shared memory

Internal memory

Shared memory

Shared directory in the network.

Method that is loaded when a sample with the above-specified identification is processed. You can also enter a method that does not yet exist in the selected memory. When a determination is started there is a check whether the method is available.

150 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Entry

Selection

15 Templates

32 characters maximum

Selection of methods stored in the selected memory

15.2 Custom result templates

Main dialog: System ▶ Templates ▶ Custom result templates

[New]

[Delete]

[Edit]

900 Touch Control

In the dialog Templates / Custom result templates , you can define the result calculations which can be loaded in the calculation command

CALC . With the exception of the following points, creating a result tem-

plate is identical to editing a calculation in the command CALC (see chapter 29.13.1.2, page 421)

:

No result variable can be assigned to the template.

The definition of limit values is not possible.

NOTICE

Up to nine wildcards can be inserted in the calculation formula, the

Variables F1…F9 . Use these variables, e.g. for the molar mass of your analyte. If you have loaded a result template with these wildcards in the calculation command, then you will automatically be prompted to enter the numerical values.

Add a new result template to the list, see following chapter.

Delete the selected result template from the list.

Edit the data of the selected result template, see following chapter.

■■■■■■■■ 151

15.2 Custom result templates ■■■■■■■■■■■■■■■■■■■■■■

15.2.1 Editing result templates

List of result templates: Result template ▶ New / Edit

Result name

The result name is the text which will be shown in the results display and in the report.

Entry

Default value max. 24 characters

R

Calculation formula

Shows the calculation formula. A special editor is opened for the definition

(see chapter 29.13.3, page 427) .

Entry

Default value max. 100 characters empty

Decimal places

Number of decimal places used to display the result.

Input range

Default value

0 to 5

2

Result unit

The result unit is displayed and saved together with the result.

Entry

Selection

Default value max. 10 characters

% mg/100 g | ppm | g | mg | µg | mL | µL | mg/ piece | °C | µg/min | mL/min | µL/min

%

| mol/L | mmol/L | g/L | mg/L | mg/mL |

[Note]

Entering a note on the calculation.

[Note for wizard]

Enter a note on the calculation. This note is displayed when loading the result template.

[Result options]

Defining additional settings for the calculation.

"Edit result template / Result options" dialog

In the dialog Edit result template / Result options , settings for how to process the calculated result are defined.

152 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 15 Templates

Variable for mean value

If the statistic calculations have been activated (see method options), the mean value of the single results will be saved as variable SMN1 to SMN9.

As default value, always the first free variable is being displayed.

Selection SMN1 … SMN9 | off off

For the result not statistic calculations are carried out.

Save as titer

The result can be saved as titer for the selected titrant.

NOTICE

The titer is assigned to the titrant defined in the last titration command before the calculation. Please note that the CALC command containing the titer assignment is inserted after the titration command with which the titer is determined.

Selection

Default value

Single value | Mean value | off off

Single value

The result of the current determination is saved as titer.

Mean value

If statistic calculations are carried out for the result, the current mean value of the determination series is saved.

off

The result is not saved as titer.

Save as common variable on | off (Default value: off )

The calculated result can be saved as a method-independent variable, a so-called common variable. The result is then also available in other methods for calculations. In all cases, only the single value will be saved, even if the statistics function is enabled.

Variable

Selection of the common variable to which the result is being assigned.

Selection CV01 … CV25

900 Touch Control ■■■■■■■■ 153

15.2 Custom result templates ■■■■■■■■■■■■■■■■■■■■■■

Display result

Save result in result table on | off (Default value: off )

The calculated result can be saved in the result table. This may be advisable if e.g. the results of all determinations carried out on a particular day are to be displayed clearly. A maximum of nine results from a determination can be saved in the result table.

Precision on | off (Default value: on )

If you deactivate this parameter, the result is neither displayed in the result dialog nor printed in the result report. This can be advisable for intermediate results.

Setting, with which accuracy the result is used in additional calculations.

Selection

Default value

Round

Round

| Truncate | Full precision

Round

The result is rounded to the defined number of decimal places (commercial rounding, in accordance with the US Pharmacopeia USP). If the digit at the first dropped decimal place is 1, 2, 3 or 4 , then it will be rounded off; if this digit is 5, 6, 7, 8 or 9 , then it will be rounded up.

Negative digits will be rounded in accordance with their amount, i.e.

away from zero.

Truncate

The result is cut to the number of decimal places defined.

Full precision

The result is used with full accuracy (floating point number either in

"single precision" (32 bit) or in "double precision" (64 bit), according to the standard IEEE 754).

154 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

15.3 Input lines

Main dialog: System ▶ Templates ▶ Input lines

15 Templates

Table 7 List of predefined input signals

Signal name Input signal

Cond OK

End1

******1*

****1***

End2

EndMeter

Ready*

Ready1

Ready2

In the dialog Templates / Input lines , you can define the input signals at the remote interface as a template. You can select these templates in the command SCAN . The list can contain a maximum of 20 templates.

*1******

***11***

**1****1

*******1

**1*****

Function queries "Cond OK" condition of device.

waits for the EOD pulse from Device 1 (Titrino,

Titrando).

waits for the EOD pulse from Device 2.

waits for the EOD pulse from 780/781 pH/Ion meter (stirrer 1 will be switched on during the waiting period).

queries the "Ready" condition of Devices 1 and 2

(Titrino, Titrando).

The status of devices working in parallel can be scanned with this signal. In this case the "Ready" line of both devices must be statically set (i.e. constantly) at the end of a determination. Devices which only transmit a brief impulse cannot be controlled in parallel.

queries the "Ready" condition of Device 1.

queries the "Ready" condition of Device 2.

900 Touch Control ■■■■■■■■ 155

15.3 Input lines ■■■■■■■■■■■■■■■■■■■■■■

Signal name

Sample ready

Input signal

***1****

Function waits for a stepping pulse, of, for instance, a connected Sample Processor, as soon as it is ready.

[New]

Add a new input signal to the list, see following chapter.

[Delete]

Delete the selected input signal from the list.

[Edit]

Edit the data of the selected input signal, see following chapter.

15.3.1 Editing the input signal

List of input signals: Input signal ▶ New / Edit

Signal name

Designation of the template.

Entry 24 characters maximum

Input signal

Entering the desired bit pattern:

Entering the bit pattern:

0 = line inactive

1 = line active

* = retain line status

The input lines are always numbered from right to left, i.e. with the signal

*******1 line 0 is expected to be active.

NOTICE

We recommend masking lines that are of no interest or for which no defined condition can be predicted with an asterisk ( * ).

Entry

Default value

Bit pattern of exactly 8 characters

********

156 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

15.4 Output lines

Main dialog: System ▶ Templates ▶ Output lines

15 Templates

In the dialog Templates / Output lines , you can define the output signals at the remote interface as a template. You can select these templates in the command CTRL . The list can contain a maximum of 20 templates.

Table 8 List of predefined output signals

Signal name Output signal

Continue *********p****

Init

Meter Cal C

Meter Cal pH

Meter enter

Meter Mode C

Meter Mode I

Meter Mode pH

00000000000000

*********1001*

*********0101*

*********1111*

*********1000*

*********0100*

*********0001*

Function sends a stepping pulse to the connected Sample

Processor.

initializes the remote interface.

switches the 781 pH/Ion meter to concentration calibration.

switches the 780/781 pH/Ion meter to pH calibration and starts the calibration.

simulates the [Enter] key of the 780/781 pH/Ion meter; mandatory for pH calibration in order to start the measurement of the second buffer.

switches the 781 pH/Ion meter to concentration measurement.

switches the 780/781 pH/Ion meter to voltametric measurement with polarization current and starts the measurement.

switches the 780/781 pH/Ion meter to pH measurement and starts the measurement.

900 Touch Control ■■■■■■■■ 157

15.4 Output lines ■■■■■■■■■■■■■■■■■■■■■■

Signal name

Meter Mode T

Meter Mode U

Start device*

Start device1

Start device2

Start Dos*

Output signal

*********0010*

*********0011*

********p****p

*************p

********p*****

*****p*p******

Function switches the 780/781 pH/Ion meter to temperature measurement and starts the measurement.

switches the 780/781 pH/Ion meter to voltage measurement and starts the measurement.

starts Devices 1 and 2 (e.g. Titrino, Titrando, etc.)

*) .

The signal is transmitted as a short pulse of 200 ms.

starts Device 1 (e.g. Titrino, Titrando, etc.) *) .

The signal is transmitted as a short pulse of 200 ms.

starts Device 2 (e.g. Titrino, Titrando, etc.) *) .

The signal is transmitted as a short pulse of 200 ms.

starts Dosimat at Devices 1 and 2 (Titrino via "activate").

The signal is transmitted as a short pulse of 200 ms.

Start Dos1 *******p****** starts Dosimat at Device 1 (Titrino via "activate").

The signal is transmitted as a short pulse of 200 ms.

Start Dos2 *****p******** starts Dosimat at Device 2 (Titrino via "activate").

The signal is transmitted as a short pulse of 200 ms.

*) In the case of the 780/781 pH/Ion meter, a result report is triggered.

[New]

Add a new output signal to the list, see following chapter.

[Delete]

Delete the selected output signal from the list.

[Edit]

Edit the data of the selected output signal, see following chapter.

158 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 15 Templates

15.4.1 Editing the output signal

List of input signals: Output signal ▶ New / Edit

Signal name

Designation of the template.

Entry 24 characters maximum

Output signal

Entering the desired bit pattern:

Entering the bit pattern:

0 = line inactive

1 = line active

* = retain line status p = set pulse

The output lines are always numbered from right to left, i.e. with the signal *************1 line 0 is set.

NOTICE

We recommend masking lines that are of no interest or for which no defined condition can be predicted with an asterisk ( * ).

Entry

Default value

Bit pattern of exactly 14 characters

**************

Pulse length

Duration of the sent pulses.

Input range

Default value

100 to 1000 ms

200 ms

900 Touch Control ■■■■■■■■ 159

15.5 Custom calibration buffers ■■■■■■■■■■■■■■■■■■■■■■

15.5 Custom calibration buffers

Main dialog: System ▶ Templates ▶ Custom calib. buffers

Buffer 1 pH

Buffer 2...5 pH

[Temp. table]

In the Templates / Custom calibration buffers dialog, you can define a buffer series of your own for pH calibrations with a maximum of five calibration buffers. Automatic buffer recognition is active for this buffer.

pH value of the first calibration buffer.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control instrument used.

Input range

Selection

Default value

–20.000 to 20.000 off off

See Buffer 1 pH .

Display the temperature table for the selected buffer, see following chapter.

160 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 15 Templates

15.5.1 Defining calibration buffers

The Custom calibration buffers / Temperature table dialog lists the pH values of the buffer in the temperature range 0 - 95 °C.

[Edit pH value]

Enter the pH value for the selected temperature.

"Temperature table / Edit pH value" dialog

In this dialog, the pH values of the buffer are defined at various temperatures. Enter the pH values for the temperature range in which you will carry out your pH calibration and pH measurement. If you do not know the pH values at individual temperatures, they will be calculated automatically by means of linear interpolation.

Temperature in °C

Input range pH value

0.0 to 95.0 °C (Increment: 5.0)

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control device used.

Input range

Selection

Default value

–20.000 to 20.000 off off

900 Touch Control ■■■■■■■■ 161

15.6 Report header ■■■■■■■■■■■■■■■■■■■■■■

15.6 Report header

Main dialog: System ▶ Templates ▶ Report header

In the dialog Templates / Report header , you can create a system-specific report header that contains information concerning the laboratory, for example. The report header can contain a maximum of four lines of 46 characters each and will always be printed out in front of the standard report header. The standard report header contains the instrument type, serial number and program version of the software, instrument name and printing date.

Print logo on | off (Default value: off )

If this parameter is activated, then the logo defined below will be printed out in the right margin.

Logo

Selection

Default value

Metrohm logo | Custom logo

Metrohm logo

Custom logo

Select this setting if your logo is to be printed out in the report header.

How to create a logo is described below.

Applying the custom logo

You can create a graphics file of your own and use it as a logo in the report header in place of the Metrohm logo.

NOTICE

Observe the size of the graphics object. The maximum size of your graphics file is 64 kB. We recommend a maximum of 500 × 200 pixels.

Proceed as follows to create your own logo:

1 Creating the logo

Create a graphics file with the desired logo in an image processing program.

Save the graph as a JPG file under the file name

"CustomImage.jpg".

162 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 15 Templates

2 Copying the file to an external storage medium

Copy the file "CustomImage.jpg" to the directory "900".

If this structure is not maintained, the file will not be found because the software directly accesses the "900" directory.

Plug in the external storage medium at the Touch Control.

3 Loading the graphics file

In the Templates / Report header dialog, tap on the [Load] button.

The file is copied to the internal memory.

Switch the instrument off and back on again.

4 Applying the custom logo

In the Templates / Report header dialog, activate the Print logo option.

Define Logo = Custom logo .

Your logo will now be printed in the report header of all reports in the future.

15.7 Custom electrode type

Main dialog: System ▶ Templates ▶ Custom electrode type

For the electrode test of pH electrodes, you can define your own values instead of using the limit values that are saved. This can be helpful for special applications, e.g. when other quality requirements are demanded of the pH electrode or when working with a reference electrode that pushes the offset potential Uoff beyond the required –15 to +15 mV (standard electrode).

Limit values can be defined for the following electrode ratings:

Excellent electrode

Good electrode

Usable electrode

Lower limit Uoff

Lower limit value for the offset voltage, i.e. the voltage at pH = 7.0. The value applies for all ratings.

Input range

Default value

–999 to 999 mV

–15 mV

900 Touch Control ■■■■■■■■ 163

15.7 Custom electrode type ■■■■■■■■■■■■■■■■■■■■■■

Upper limit Uoff

Upper limit value for the offset voltage, i.e. the voltage at pH = 7.0. The value applies for all ratings.

Input range

Default value

–999 to 999 mV

15 mV

15.7.1 Limit values for the electrode rating

You can define limit values for the three electrode ratings Excellent electrode , Good electrode and Usable electrode . These values represent maximum values that may not be exceeded (exception: Min. slope = minimum value).

Streaming potential

The streaming potential is the difference between the measured voltages in stirred and unstirred solution.

for very good electrodes:

Input range –999.9 to 999.9 mV

Default value 2.5 mV for good electrodes:

Input range

Default value

–999.9 to 999.9 mV

3.0 mV for usable electrodes:

Input range

Default value

–999.9 to 999.9 mV

4.0 mV

Drift

The drift in stirred solution is determined for each buffer. The sum of these values is compared with this value here.

for very good electrodes:

Input range 0.1 to 9.9 mV/min

Default value 2.0 mV/min for good electrodes:

Input range

Default value

0.1 to 9.9 mV/min

2.5 mV/min for usable electrodes:

Input range

Default value

0.1 to 9.9 mV/min

3.0 mV/min

164 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Min. slope

Minimum slope of the pH electrode.

for very good electrodes:

Input range

Default value

0.1 to 999.9 %

96.5 % for good electrodes:

Input range

Default value

0.1 to 999.9 %

96.0 % for usable electrodes:

Input range

Default value

0.1 to 999.9 %

95.0 %

Max. slope

Maximum slope of the pH electrode.

for very good electrodes:

Input range

Default value

0.1 to 999.9 %

101.0 % for good electrodes:

Input range

Default value

0.1 to 999.9 %

102.0 % for usable electrodes:

Input range

Default value

0.1 to 999.9 %

103.0 %

Response time

15 Templates

The voltage measured after three minutes in stirred solution serves as a comparison value for determining the response time. The response time is the time after which the measured voltage is within ±1 mV of this comparison value.

for very good electrodes:

Input range

Default value

0 to 999 s

45 s for good electrodes:

Input range

Default value

0 to 999 s

50 s

900 Touch Control ■■■■■■■■ 165

15.7 Custom electrode type for usable electrodes:

Input range

Default value

0 to 999 s

60 s

■■■■■■■■■■■■■■■■■■■■■■

166 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

16 Methods

16.1 Creating a new method

Proceed as follows to create a new method:

1 Opening the method table

In the main dialog, tap on [Load method] .

Tap on [New method] .

The method table with the stored templates opens:

16 Methods

2 Loading the method

Tap on [Empty method] .

or

Select the desired template and tap on [Load template] .

The method is now loaded. New method is displayed in the main dialog in the title line.

If a new method has been created, then the individual parameters can be modified with [Edit parameters] .

900 Touch Control ■■■■■■■■ 167

16.2 Saving a method ■■■■■■■■■■■■■■■■■■■■■■

16.2 Saving a method

If you modify method parameters, then you can save these as your own method.

To save a method, proceed as follows:

1 Open the command sequence

In the main dialog, tap on [Edit parameters] .

The command sequence of the method is displayed:

2 Modifying/applying the method name

Tap on [Save method] .

A method name will be suggested for new methods. If the method has already been saved once, then the method name will be displayed.

168 ■■■■■■■■

Applying the name:

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 16 Methods

Tap on [Save] .

The method will be saved and the command sequence is displayed.

Entering a new name:

Tap on the input field File name .

The text editor opens.

Enter a new file name (max. 32 characters) and confirm with

[OK] .

Tap on [Save] .

The method will be saved and the command sequence is displayed.

16.3 Loading a method

Proceed as follows to load a method:

1 Opening the method table

In the main dialog, tap on [Load method] .

The selection of memory locations is displayed. This selection will be skipped if only the internal memory is accessible.

Select the memory location where the desired method is saved.

The selection of file groups is displayed (see chapter 12.1, page

121) . If only one group is available, then the method table will be

displayed directly.

Select the group with the desired method.

Tap on [Show files] .

The method table with the stored methods opens:

900 Touch Control

2 Selecting the method

Select the desired method.

■■■■■■■■ 169

16.4 Editing a method ■■■■■■■■■■■■■■■■■■■■■■

3 Loading the method

Tap on [Load] .

The method is now loaded. The method name is displayed in the main dialog in the title line.

NOTICE

The data of the current determination will be deleted when you load a method.

16.4 Editing a method

Main dialog: Edit parameters

[Save method]

[Method options]

In the dialog Parameters / Sequence , you can define and edit the method commands which are executed successively during a determination.

The following data is displayed in the command list for each command:

Line number

Name

Comment

You can adjust the comment in accordance with your requirements.

Save the current method (see chapter 16.2, page 168)

.

Define various settings which refer to the entire method, e.g. statistics, automatic saving of determinations, settings for sample data, etc.

(see chapter 16.5, page 172)

.

170 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

[Insert command]

16 Methods

Insert a new method command. It is inserted before the selected command.

NOTICE

Not all commands are available for subsequences. Commands which cannot be inserted into subsequences are disabled.

[Delete command]

Delete the selected method command.

[Edit command]

Edit the selected method command.

16.4.1 Inserting a command

A method is comprised of individual commands. When you start a determination, the commands are executed one after the other.

The following constraints apply:

Maximum of 99 commands

Maximum of nine commands for titrations, measurements and calibrations

Inserting a command

To insert a command in a method, proceed as follows:

1 Open the command overview

In the main dialog, tap on [Edit parameters] .

Tap on the [Insert command] button.

900 Touch Control ■■■■■■■■ 171

16.5 Method options ■■■■■■■■■■■■■■■■■■■■■■

This dialog contains all of the commands which can be inserted into a method run, organized into thematic groups.

2 Select a command group

Tap on the desired command group.

In the case of [Report] , the command is inserted directly into the command list. For all other command groups, the available commands will be displayed (titration / measuring commands, etc.).

3 Inserting a command

Tap on the desired command.

The dialog is closed and the command is displayed in the command list.

16.5 Method options

Main dialog: Edit parameters ▶ Method options

172 ■■■■■■■■

Method options are settings which apply to the method as a whole and not to an individual command.

The following settings are available:

Activate/deactivate statistics calculations

Save determination automatically

Sign a method electronically

Check the method when the determination starts

Define a fixed sample size etc.

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 16 Methods

Statistics on | off (Default value: off )

If this parameter is activated, then statistics calculations will be carried out for all of the defined results. The prerequisite here, however, is that a vari-

able for the mean value is defined in the result options (see "Dialog "Edit calculation / Result options"", page 423) .

Number of determinations

The number of determinations that are carried out for the statistics calculations.

If an additional determination has to be added to the determination series, because one determination has been incorrect, for example, then

this can be accomplished in the statistical overview (see chapter 25, page

231) .

Input range

Default value

2 to 20

3

[Save automat.]

Define the settings for automatically saving the determination (see chapter 16.5.6, page 182) .

[Properties]

Display the properties of the current method and sign electronically (see chapter 16.5.4, page 180)

.

[Note]

Enter a note on the method (see chapter 16.5.5, page 182)

.

[Start/Stop options]

Edit functions which are carried out immediately after the method is

started or stopped, see chapter Start options, page173 and

chapter Stop options, page 175

.

[Sample data]

Specify method-specific settings for the sample data, e.g. designation of

the sample identifications, limits for the sample size (see chapter 16.5.3, page 176)

.

16.5.1 Start options

Main dialog: Edit parameters ▶ Method options ▶ Start/Stop options ▶ Start options

In the dialog Method options / Start options , you can configure the functions which are carried out when a determination is started.

900 Touch Control ■■■■■■■■ 173

16.5 Method options ■■■■■■■■■■■■■■■■■■■■■■

Method check at start on | off (Default value: on )

If this option is activated, then the following tests are carried out at the method start.

Are the required control and peripheral devices connected?

Are the titrant and the sensor on hand?

Have the monitoring intervals expired?

etc.

NOTICE

We recommend that the method check only be deactivated if the titration must begin immediately after the start of the determination (e.g.

with very rapid reactions such as enzyme-catalyzed reactions). If the method check is deactivated, it could happen that the determination run will be interrupted, e.g. because devices, titrants or sensors are not on hand.

The following settings can be implemented for a connected Sample Processor:

Rack reset

■ on | off (Default value: off )

If this option is activated, then the sample rack is initialized at the method start. The following actions are carried out:

The lift is moved upwards.

The sample rack is rotated to the starting position.

The rack code of the rack attached is being read out and the corresponding rack data is transferred to the Sample Processor.

The sample variable is being reset to the value 1.

Check rack

Definition of the sample rack which must be attached at the time of the method start. This ensures that the method will be carried out only with this sample rack.

NOTICE

The sample rack can however only be checked if the Rack reset option is activated.

174 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 16 Methods

Selection

Default value no | Selection of configured sample racks no

Increase sample variable automatically on | off (Default value: on )

The sample variable describes the current position of the sample on the rack of the Sample Processor. If this option is activated, then the sample variable is automatically increased by 1 at the end of each determination.

This option must be deactivated if the sample variable is to be altered specifically with a SAMPLE command.

Request rack position at start on | off (Default value: off )

If this option is activated, then the rack position of the first sample to be processed is requested.

16.5.2 Stop options

Main dialog: Edit parameters ▶ Method options ▶ Start/Stop options ▶ Stop options

In the dialog Method options / Stop options , you can define the actions which are carried out when a method is canceled.

The method can be canceled as follows:

Manual stop with the fixed key [ ]

Stop because of an error

Stop by remote signal via the Control Remote Box

Switch off pumps on | off (Default value: on )

If this option is activated, then all of the connected pumps will be switched off.

Switch off stirrers on | off (Default value: on )

If this option is activated, then all of the connected stirrers will be switched off.

900 Touch Control ■■■■■■■■ 175

16.5 Method options ■■■■■■■■■■■■■■■■■■■■■■

NOTICE

Do not fail to deactivate this option for the currently loaded method if you are conditioning a titration cell at the same time in manual control.

Otherwise, the stirrer of the titration cell will also be switched off if the currently loaded method is canceled. This would result in an overtitration of the working medium.

Set remote lines

Selection of the signal out of the templates or entering the required bit pattern. Templates are defined under System ▶ Templates ▶ Output lines .

NOTICE

A line set active is not being reset automatically, not even at the end of the determination.

Entering the bit pattern:

0 = line inactive

1 = line active

* = retain line status p = set pulse

The output lines are always numbered from right to left, i.e. with the signal *************1 line 0 is set. With a pulse, the length is set to 200 ms.

If you wish to set pulses with other lengths, you have to define a corresponding template.

Entry

Default value

Selection

Bit pattern containing exactly 14 characters or a max. of 24 characters for the name of the template

**************

Selection of the templates defined

16.5.3 Sample data

Main dialog: Edit parameters ▶ Method options ▶ Sample data

You can define the following method-specific settings in the dialog

Method options / Sample data :

Modify the designation for the sample identifications

Define a fixed sample size

176 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 16 Methods

Identification 1, Identification 2 on | off (Default value: on )

If this option is activated, then the input field for the sample identification in the main dialog will be displayed.

Designation

Define limits for the sample size

If you define limits, these will be monitored at the start and at the end of the determination.

Below you will find an example demonstrating the effects on the main

dialog of the settings in this dialog (see "Modifying sample data", page

178) .

A designation of your own choosing can be defined for the input field for each method.

Entry 16 characters maximum

Fixed sample size on | off (Default value: off )

Activate this parameter if the same sample amount is always to be used for all determinations. If this is case, then you can define the sample size here. It will be displayed in the main dialog, but it can no longer be changed there.

Sample size

This parameter can only be edited when Fixed sample size is activated.

Input range

Default value

–999999999 to 9999999999

1.0

Unit of sample size.

Selection

Default value g g

| mg | µg | mL | µL | pieces

900 Touch Control ■■■■■■■■ 177

16.5 Method options ■■■■■■■■■■■■■■■■■■■■■■

NOTICE

You can also use the parameter Fixed sample size if you would like to establish a default for the sample size.

Example: For a particular determination, you require 10 mL of sample.

Now and again, however, you have too little sample available and you must perform the determination with 5 mL, for instance. In this case, proceed as follows:

1.

Activate the parameter Fixed sample size .

2.

Define the value and unit for the sample size, in the above example, for instance, [10 mL] .

3.

Deactivate the parameter Fixed sample size again.

4.

Save method.

⇨ Each time you load this method, 10 mL is entered as the sample size, but in the example mentioned you can enter the lesser quantity of 5 mL.

[Sample size limits]

This button is accessible only if Fixed sample size is deactivated.

Define the limit values for the sample size (see "Defining limit values for the sample size", page 179) .

Modifying sample data

This instruction is for the purpose of clarifying the settings which are possible in this dialog. The following settings should be made:

Change the designation for the input field of the first sample identification.

Hide the input field for the second sample identification.

Define a fixed sample size.

Proceed as follows:

1 Change the designation for the input field

Tap on the input field Identification 1 .

The text editor opens.

Change the designation in Batch and confirm with [OK] .

Deactivate the parameter Identification 2 .

2 Define a fixed sample size

Activate the parameter Fixed sample size .

178 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Enter the value 10 mL as Sample size .

3 Displaying the main dialog

Tap on the fixed key [ ] .

The data is saved and the adjusted main dialog appears.

16 Methods

Only Identification 1 is shown with the title "Batch." The fixed sample size "10 mL" is shown, but cannot be edited.

Defining limit values for the sample size

In the dialog Sample data / Sample size limits , you can define the limit values for the sample size. These values are not monitored during sample data input, but rather:

■ when the determination is started.

when the determination is finished.

upon automatic sample data query via the command REQUEST .

when the determination is being recalculated.

An entry is made in the message list automatically when these limits are

infringed (see chapter 24.2, page 218) .

Monitoring sample size limits on | off (Default value: off )

If this parameter is activated, then the sample size will be monitored.

Lower limit

Input range

Default value

–999999999 to 9999999999

0

900 Touch Control ■■■■■■■■ 179

16.5 Method options ■■■■■■■■■■■■■■■■■■■■■■

Upper limit

Input range

Default value

–999999999 to 9999999999

9999999999

Display message on | off (Default value: off )

If this parameter is activated, a message is displayed in case the limits are infringed. You can select whether you want to continue with the determination or cancel the run.

If this parameter is deactivated, then the message will be entered only in the message list of the determination.

Timeout

The message is displayed during this time. Afterwards, the determination is continued automatically.

Input range

Default value

Selection

0 to 999999 s

30 s off off

The determination will not be continued until after the message has been confirmed.

16.5.4 Method properties

Main dialog: Edit parameters ▶ Method options ▶ Properties

Detailed information concerning the method is displayed in the Method options / Properties dialog and the methods can be signed electronically here:

Method status

Current status of the method.

– new

The method has been newly created and not yet saved.

– saved

The current method has been saved.

– modified

The current method has been modified.

– reviewed

The current method has been signed at level 1.

– released

The current method has been signed at level 2.

Created by

User who created the method.

180 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 16 Methods

Created on

Date and time at which the method was created.

Last saved by

User who saved the method most recently.

Last saved on

Date and time at which the method was most recently saved. The version of the method is displayed as well. The version number will be increased by one each time the file is saved.

The following data is displayed only if the option Saving modified method under Login options / Modification options is activated:

Modified by

User who modified and saved the method most recently.

Modified on

Date and time at which the method was most recently modified and saved.

Reason

Reason for the modification of the method.

Comment

Comment for the modification of the method.

The following information is only shown if the method has been signed at level 1:

Reviewed by

User who signed the method at level 1.

Reviewed on

Date and time at which the method has been signed at level 1.

Reason

Reason for the signing of the method.

Comment

Comment for the signing of the method.

The following information is only shown if the method has been signed at level 2:

Released by

User who signed the method at level 2.

Released on

Date and time at which the method has been signed at level 2.

Reason

Reason for the signing of the method.

Comment

Comment for the signing of the method.

[Delete signature]

Delete all of the signatures for the method or the determination. This but-

ton is inactive if level 2 has not yet been signed (see chapter 17.2, page

185) .

900 Touch Control ■■■■■■■■ 181

16.5 Method options

[Sign]

■■■■■■■■■■■■■■■■■■■■■■

Sign the method or determination. This button is active only when working with activated login function and password

(see chapter 17.1, page

184) .

16.5.5 Note

Main dialog: Edit parameters ▶ Method options ▶ Note

In this dialog, you can enter a short text, e.g. important information about carrying out the determination.

[Display options]

Definition when the note is being displayed.

Automatically after loading the method on | off (Default value: off )

If this parameter is activated, the note is displayed when loading the method. It can otherwise only be read in this dialog.

16.5.6 Saving a determination automatically

Main dialog: Edit parameters ▶ Method options ▶ Save automat.

In the dialog Method options / Save automatically , you can specify whether the determination is to be saved automatically and/or whether a

PC/LIMS report is to be created.

Save determination automatically on | off (Default value: off )

If this parameter is activated, then the determination will be saved automatically.

Memory

Memory location where the determination will be saved.

Selection

Default value

External memory 1

Shared memory

External memory 1

| External memory 2 |

Shared memory

The determination will be saved in a shared directory on the network.

The shared directory is selected in the Edit device / Shared memory

dialog (see chapter 11.3.3, page 84)

.

Group

Directory where the determination will be saved.

Entry 32 characters maximum

182 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 16 Methods

Selection

Default value

Selection of available directories | empty empty

File name

File name of the determination. The file name is always extended to include the date and the time of day (YYYYMMDD-hhmmss) in order to ensure that the file names are unique for all determinations.

Entry

Selection

Default value max. 16 characters

Identification 1 | Identification 2 | Method

Identification 1

Identification 1

The first 16 characters of the text which was entered in the main dialog for Identification 1 + YYYYMMDD-hhmmss.

Identification 2

The first 16 characters of the text which was entered in the main dialog for Identification 2 + YYYYMMDD-hhmmss.

Method

The first 16 characters of the text, of the method name +

YYYYMMDD-hhmmss.

Write protection on | off (Default value: on )

If this parameter is activated, then the file cannot be saved, deleted, or renamed. This is only an internal write protection feature and is independent of the write protection of the operating system on your computer.

This write protection feature protects the saved determination data against accidental modifications or modifications by unauthorized persons.

Create PC/LIMS report on | off (Default value: off )

If this parameter is activated, then a machine-readable report with all of the important data for the determination is created, a so-called PC/LIMS report.

The settings are implemented in the device manager

(see chapter 11.3.2, page 83)

.

When you save the PC/LIMS report as a file, the file name will be generated automatically as follows: PC_LIMS_Reportfirst 16 characters of the value of the parameter File name-Date-Time.txt

. If the value of the parameter File name contains one or more of the following characters, then these will be automatically converted in the file name into the character "_": / \ : * ? " < > |.

900 Touch Control ■■■■■■■■ 183

17.1 Signing methods/determinations electronically ■■■■■■■■■■■■■■■■■■■■■■

17 Electronic signatures

17.1 Signing methods/determinations electronically

Methods: Main dialog ▶ Edit parameters ▶ Method options ▶ Properties ▶ Sign

Determinations: Results dialog ▶ More data ▶ Properties ▶ Sign

You can sign methods or determinations in the dialog Electronic signature . The following information is saved along with each electronic signature:

User (complete name)

Date and time

Reason

Comment

Methods can only be signed if they have the status saved . The status of a method can be found in the dialog Method options / Properties .

Signature

Selection of the level at which the determination should be signed.

Selection Review (signature level 1) level 2)

| Release (signature

A maximum of three signatures from up to three different users can given at each level. However, it is not possible for a user to sign at both level 1 and level 2.

Review (signature level 1)

If the method has never been signed before, then only level 1 can be selected.

Release (signature level 2)

If the method has been signed three times at level 1 or already once at level 2 then only level 2 can be selected.

User

Identification of the user who signs.

Password

Password of the user.

184 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 17 Electronic signatures

Reason

Selection of the reasons for the signature. The list with reasons is compiled in the user administration

(see chapter 7.3.6, page 38)

.

Selection Selection of configured reasons

Comment

Entry 24 characters maximum

[Change password]

Change the current password.

[Sign]

Sign the method or determination with the data entered above.

17.2 Deleting electronic signatures

If a released method must be revised, e.g. in order to adapt to altered conditions, then the signatures must be deleted. The method can otherwise not be modified. In the case of released determinations, the signatures can also be deleted. Signatures, however, cannot be deleted until after the method/determination has been signed at Level 2.

Signatures are deleted automatically in the following cases:

A method signed at Level 1 is being altered.

A determination signed at Level 1 is being recalculated.

User

Identification of the user who deletes the signatures.

Password

Password of the user.

Reason

Selection of the reasons for the deletion of the signature. The list with reasons is compiled in the user administration

(see chapter 7.3.6, page 38) .

Selection Selection of configured reasons

Comment

Entry 24 characters maximum

[Delete signature]

Delete all the signatures for the method or the determination. The method is assigned the status saved .

900 Touch Control ■■■■■■■■ 185

18 Control

Main dialog: Control

■■■■■■■■■■■■■■■■■■■■■■

Statistics

Sample table

In the dialog Control , the settings for the execution of a single determination or of one sample series are defined.

If you work with the login function activated, you can log off the system in this dialog. The login dialog will then be shown immediately.

on | off (Default value: off )

With this option, you can enable or disable the statistics calculations for individual determinations

(see "Statistics", page 188) .

NOTICE

The parameter Statistics in the method options will not be changed as a result. This setting here will be adjusted automatically to match the one in the method options.

on | off (Default value: off )

If this option is activated, the sample data for a sample series can be

entered in a table (see chapter 21, page 197) .

186 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Autostart

18 Control on | off (Default value: off )

If this option is activated, a new determination is started automatically at the end of a determination. This continues until the number specified has been reached (siehe Number of autostarts ).

NOTICE

If you deactivate this option while a determination is running, then that determination will be ended, although the next one will however also not be started. This can be used to cancel a sample series.

Number of autostarts

Number of automatic starts.

Input range

Selection

Default value

1 to 9999

Table

Table

Table

The number of automatic starts corresponds to the number of samples in the sample table.

Sample number

The sample number is primarily used for documentation. Each time a determination is started, it is increased by one. The sample number can, however, also be entered manually. Each time the Touch Control is switched on, the value is reset to zero.

Input range

Default value

0 to 9,999

0

[Logout]

User logout (visible only when work is done with activated login function).

[Favorites]

Save methods/sample tables as a favorite

(see chapter 19, page 190)

.

Depending on whether you are working with or without login function, user-specific or common favorites can be created. On the main dialog, a button is created for each favorite. This way, a method or a sample table can be started with one single mouse click.

[Delete table]

Delete the sample table completely.

[Reset autostart]

Reset the value of the autostart counter to zero.

900 Touch Control ■■■■■■■■ 187

■■■■■■■■■■■■■■■■■■■■■■

[Delete statistics]

Delete all statistics data.

The statistics data should be deleted manually if a sample series has been canceled and a new sample series is to be started.

The statistics data is deleted automatically in following cases:

■ when all of the determinations of the determination series have been carried out and a new determination has been started afterwards.

when a new method is loaded.

when a determination is loaded (the method with which the determination was carried out is loaded simultaneously with the determination).

Statistics

In the dialog Control , you can deactivate the statistics calculations for individual determinations. This option is required primarily in order to insert the processing of an " urgent sample " during the processing of a sample series when the urgent sample is not to be included in the statistics calculations.

Interrupting a determination series for an "urgent sample"

If the same method is required for the "urgent sample" as is being used for the sample series, then you need only deactivate the option Statistics and then reactivate it after the determination is completed. Proceed as follows if you need to process the "urgent sample" with a different method:

1 Deactivate statistics calculation

Deactivate the option Statistics .

2 Saving the determination

Save the current determination of the sample series (see chapter

24.6, page 227)

.

3 Carrying out the determination

Load the method for the "urgent sample".

Carry out the determination.

4 Load the last determination of the sample series

Reload the previously saved last determination.

The determination, the method used for it and the current statistical data are loaded. The option Statistics is activated again and the

188 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 18 Control value of the statistics counter corresponds to that in effect before the interruption.

5 Continuing the sample series

Carry out the remaining determinations in the sample series.

900 Touch Control ■■■■■■■■ 189

■■■■■■■■■■■■■■■■■■■■■■

19 Favorites

Main dialog: Control ▶ Favorites

Main dialog: System ▶ System settings ▶ User admin. ▶ Edit ▶ Favorites

This chapter describes how you can create favorites. Favorites are comparable to the favorites/bookmarks in your Internet browser. In the main dialog, a button is created for each favorite. By pressing a single key, you can trigger an action without having to navigate through different dialogs.

190 ■■■■■■■■

The following objects can be saved as favorites:

Methods

Sample tables

User-specific or common favorites can be created, depending on whether you are working with or without the login function.

Operation with login function:

If you are working with the login function activated, user-specific favorites can be created. These can only be used by a certain user.

User-specific favorites are created as follows:

– in the user administration by a user with administrator rights

( System ▶ System settings ▶ User admin. ▶ Edit ).

– in the dialog Control for the user who is logged in.

Operation without login function:

If you are working without the login function, common favorites can be created. These favorites are available for all users. Common favorites are created in the Control dialog.

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

19.1 Creating favorites

19 Favorites

The list of favorites gives the designation and position of the button on the main dialog of each favorite configured.

[New]

Add a new favorite to the list, see following chapter.

[Delete]

Delete the selected favorite from the list.

[Edit]

Edit the properties of the selected favorite, see following chapter.

19.1.1 Editing favorites

900 Touch Control ■■■■■■■■ 191

19.1 Creating favorites ■■■■■■■■■■■■■■■■■■■■■■

Position

On the main dialog, a button is created for each favorite. These buttons are ordered in three rows at fixed positions. The position 1 can be found top left.

Input range

Default value

1 to 14

1

Name

The designation of the favorite is used for unambiguous identification.

Entry 24 characters maximum

Type

Definition as to whether the favorite represents a single method or a complete sample table.

Selection

Default value

Method

Method

| Sample table

Memory

Memory location the method or the sample table is loaded from. Only the currently accessible memory locations are selectable.

Selection

Default value

Internal memory | External memory 1 | External memory 2 | Shared memory

Internal memory

Shared memory

The favorite will be saved in a shared directory on the network. The shared directory is selected in the Edit device / Shared memory dia-

log (see chapter 11.3.3, page 84)

.

Method / Sample table

Method or sample table of the favorite.

Selection Selection of stored methods/sample tables

[More options]

Parameterize the autostart function, see following chapter.

More options

The autostart function is parameterized under [More options] .

192 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 19 Favorites

Autostart on | off (Default value: off )

If this option is activated, a new determination is started automatically at the end of a determination. This continues until the number specified has been reached (siehe Number of autostarts ).

Number of autostarts

This option can only be edited when Autostart is activated.

Number of automatic starts.

Input range

Default value

Selection

1 to 9999

1

Table

Table

The number of automatic starts corresponds to the number of samples in the sample table.

900 Touch Control ■■■■■■■■ 193

20.1 Entering sample data in the main dialog ■■■■■■■■■■■■■■■■■■■■■■

20 Sample data

You can enter the sample data (identification, sample size, etc.) in a variety of ways:

Directly in the main dialog.

Using the sample table. This is particularly useful with sample series.

The sample table is a table in which the sample data for up to 999

samples can be entered (see chapter 21, page 197) .

Automatic request immediately after the start of the determination (see chapter 20.2, page 195) .

You can also send the sample size and the unit from a connected balance in any case. With some balances, the sample identification and method

can be also sent (see chapter 32.5, page 495)

.

20.1 Entering sample data in the main dialog

For a sample, you can enter the sample data directly in the main dialog,

even while the determination is running (see chapter 23, page 208) .

Identification 1

Sample identification. The sample identification can be used in calculations as the variable CI1

. If you have defined a sample identification list (see chapter 15.1.1, page 148)

then the entries can be selected here.

Entry

Default value

24 characters maximum empty

194 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Identification 2

20 Sample data

Sample identification. The sample identification can be used in calculations as the variable CI2 . If you have defined a sample identification list

(see chapter 15.1.1, page 148)

then the entries can be selected here.

Entry

Default value

24 characters maximum empty

Sample size

Sample size. The value of the sample size can be used in calculations as the variable C00 .

Input range

Default value

–99999999 to 9999999999

1.0

Unit of sample size.

Entry

Selection

Default value

6 characters maximum g g

| mg | µg | mL | µL | pieces

Unit defined in method:

The unit defined in the method is ignored. The unit entered in the sample table is used.

Balance connected and configured:

Value from the connected balance.

Fixed sample size defined in method:

The value will be entered in the corresponding line when the method is loaded. The existing entry will be overwritten.

20.2 Requesting sample data at the start of the determination

The sample data can be automatically requested immediately after the start of the determination in order to ensure that the sample data entry is not forgotten. This automatic inquiry is requisite for reweighing.

900 Touch Control ■■■■■■■■ 195

20.2 Requesting sample data at the start of the determination ■■■■■■■■■■■■■■■■■■■■■■

To accomplish this, you must insert and configure the command

REQUEST

in the method (see chapter 29.18.3, page 480) .

If the Hold sequence parameter is activated, then the run will be paused and must be continued with [Continue] after the sample data has been entered. If the Hold sequence parameter is deactivated, then the determination will continue in the background until the measured data is required for further processing. This dialog will be displayed until the entering of the sample data is confirmed with [Continue] , even if the titration/measurement is already completed. This ensures that the sample data is available for calculations.

If a fixed sample size has been defined in the method (see chapter 16.5.3, page 176)

, then it will be displayed at the time of the entry but it will not be editable.

196 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 21 Sample table

21 Sample table

21.1 General

Main dialog: Sample table

The sample table is a table in which the sample data for up to 999 samples can be entered. If you are processing larger-sized sample series, it makes sense to enter the sample data in this table.

Activating the sample table

Proceed as follows to enable the sample table:

1 Opening the control dialog

In the main dialog, tap on [Control] .

The control dialog is displayed:

2 Activating the sample table

Activate the Sample table option.

Tap on the fixed key [ ] .

The main dialog is displayed:

900 Touch Control ■■■■■■■■ 197

21.1 General ■■■■■■■■■■■■■■■■■■■■■■

The display will show the number of determinations already carried out and the total number of sample lines containing data. The sample table is still empty in this example.

The sample table contains numbered lines; each line represents one determination. In the standard settings, the first sample identification and the sample size of each sample are displayed. In the properties, you define the columns to be displayed

(see chapter 21.3, page 202)

.

198

[Load/ Save]

[Properties]

[Insert line]

[Delete]

■■■■■■■■

Load a saved sample table

(see "Loading a sample table", page 200) or

save the current sample table

(see "Saving a sample table", page 199) .

Edit the properties of the sample table

(see chapter 21.3, page 202)

.

Insert a new line above the line selected.

Delete the selected line from the sample table.

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

[Edit]

Edit the data of the selected line (see chapter 21.2, page 200)

.

21 Sample table

Saving a sample table

NOTICE

Sample tables can only be saved on external storage media or on a shared file system.

Proceed as follows to save a sample table:

1 Opening the dialog

In the Sample table dialog, tap on the [Load/ Save] button.

The Sample table / Load/Save dialog is displayed.

If no external storage medium is plugged in or if no file system is shared, then [Save] is disabled.

2 Modify/apply a name

Tap on [Save] .

A name is suggested for new sample tables. If the sample table has already been saved, the previous name will be displayed.

Applying the name:

Tap on [Save] .

The sample table is saved.

Entering a new name:

Tap on the input field File name .

The text editor opens.

Enter a new file name (max. 32 characters) and confirm with

[OK] .

Tap on [Save] .

The sample table is saved.

900 Touch Control ■■■■■■■■ 199

21.2 Editing sample data ■■■■■■■■■■■■■■■■■■■■■■

Loading a sample table

Proceed as follows to load a sample table:

1 Display the list of saved sample tables

In the Sample table dialog, tap on the [Load/ Save] button.

The Sample table / Load/Save dialog is displayed.

If no external storage medium is plugged in or if no file system is shared, then [Load] is disabled.

Tap on [Load] .

The selection of memory locations is displayed. This selection will be skipped if only one memory can be accessed.

Select the memory location where the desired sample table is stored.

The selection of file groups is displayed

(see chapter 12.1, page

121)

. If only one group is available, then the list of the saved sample tables will be displayed directly.

Select the group with the desired sample table.

Tap on [Show files] .

The list with the saved sample table is opened.

2 Select the sample table

Select the desired sample table.

3 Loading a sample table

Tap on [Load] .

The sample table is now loaded.

21.2 Editing sample data

200 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 21 Sample table

At the very top you will see the line number of the selected line. In this example, the sample data of the first line is displayed. You can scroll between individual data sets with the keys [–] and [+] .

Method

Method used for processing the sample.

If you use the sample assignment table

(see chapter 15.1.2, page 149)

, then the method can no longer be defined here.

Selection

Default value

Selection of stored methods | empty empty empty

The currently loaded method is used.

Identification 1

Sample identification. The sample identification can be used in calculations as the variable CI1 . If you have defined a sample identification list

(see chapter 15.1.1, page 148)

then the entries can be selected here.

Entry

Default value

24 characters maximum empty

Identification 2

Sample identification. The sample identification can be used in calculations as the variable CI2 . If you have defined a sample identification list

(see chapter 15.1.1, page 148)

then the entries can be selected here.

Entry

Default value

24 characters maximum empty

Sample size

Sample size. The value of the sample size can be used in calculations as the variable C00 .

Input range

Default value

–99999999 to 9999999999

1.0

Unit of sample size.

Entry

Selection

Default value

6 characters maximum g g

| mg | µg | mL | µL | pieces

Unit defined in method:

The unit defined in the method is ignored. The unit entered in the sample table is used.

900 Touch Control ■■■■■■■■ 201

21.3 Properties ■■■■■■■■■■■■■■■■■■■■■■

Balance connected and configured:

Value from the connected balance.

Fixed sample size defined in method:

The value will be entered in the corresponding line when the method is loaded. The existing entry will be overwritten.

NOTICE

The sample size limits defined in the method are not monitored when the sample data is being entered in the sample table.

21.3 Properties

Method memory

Display in table

Memory location the method is loaded from. All memory locations are selectable, even if they are currently not accessible.

Selection

Default value

Internal memory | External memory 1 | External memory 2 | Shared memory

Internal memory

Shared memory

Shared directory in the network.

Selection of the columns to be displayed in the sample table.

Selection

Default value

Method | Identification 1

Identification 1

| Identification 2

202 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 21 Sample table

Method

Only the assigned method is displayed for each sample.

Identification 1

The first sample identification and the sample size are displayed for each sample.

Identification 2

The second sample identification and the sample size are displayed for each sample.

Number of lines

Maximum number of lines which can be present in the sample table.

Input range

Default value

2 to 999

99

Lock editing on | off (Default value: off )

If this option is activated, then the sample data can no longer be altered.

Display options

Configure the sample data to be displayed in the dialog Sample table /

Edit

(see ""Properties / Display options" dialog", page 203) .

Editing options

Specify the settings for the processing of the sample data

(see ""Properties / Editing options" dialog", page 203) .

Identifications

Define the designations for the input fields of Identification 1 and Identification 2

(see "Dialog "Properties / Identifications"", page 204)

.

"Properties / Display options" dialog

Sample table: Properties ▶ Display options

In this dialog, you can configure which of the following sample data can be edited in the sample table:

Method

Identification 1

Identification 2

Sample size

"Properties / Editing options" dialog

Sample table: Properties ▶ Editing options

In this dialog, you can define the settings so that the input of the sample data becomes simpler and faster.

900 Touch Control ■■■■■■■■ 203

21.3 Properties ■■■■■■■■■■■■■■■■■■■■■■

End line with

Selection as to with which input the editing dialog for the next sample is automatically displayed.

Selection

Default value

Sample size | Identification 1 | Identification

2 | manual

Sample size

■ manual

The editing dialog for the next sample can be displayed with the [New sample] button in the editing dialog.

If, for example, you have to enter similar data for each sample, then you can automatically copy the data which is identical for each sample into the next line:

Method

Identification 1

Identification 2

Sample size

Dialog "Properties / Identifications"

Sample table: Properties ▶ Identifications

In the dialog Properties / Identifications , you can alter the designations for the sample identifications in accordance with specific methods.

204 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 22 Determination run

22 Determination run

This chapter provides a description of how to carry out a determination, i.e. how to process a single sample or a sample series. The current sample data is used for calculations.

22.1 Carrying out a single determination

When you carry out a determination, you can enter the sample data in

various ways (see chapter 20, page 194)

. The following instructions describe how to enter the sample data in the main dialog.

To carry out a single determination, proceed as follows:

1 Loading the method

See Chapter "Loading a method", Page 169

.

2 Preparing the sample

NOTICE

Calculate the amount of the sample in such a way that a titrant consumption of 10…90% of the cylinder volume will result.

Weigh in or measure the sample in a sample vessel.

Add solvent.

Place the sample vessel on the stirrer.

Immerse the electrode and buret tip in the solution.

3 Entering the sample size

Tap on the input field Sample size in the dialog.

The number editor opens.

Enter the sample size and confirm with [OK] .

4 Starting a determination

Tap on the fixed key [ ] .

900 Touch Control ■■■■■■■■ 205

22.2 Performing a sample series ■■■■■■■■■■■■■■■■■■■■■■

Sequence of the determination

After a determination is started, a check is made as to

– whether all of the devices required are connected and ready for operation.

– whether all titrants and sensors are present.

– whether the monitoring intervals for titrants, sensors and common variables are still valid.

– if a sample assignment table is activated (see chapter 15.1.2, page 149)

: whether the entered sample identification is defined in the sample assignment table. If this is the case, then the assigned method will be loaded automatically and the actual determination will be started.

NOTICE

Manual titrations (MAT command):

Start of titration : Because of the technical implementation involved, a few microliters are already dosed at the start of a manual titration. The possibility of deviation from the actual endpoint volume is excluded, however, because the already dosed volume is taken into account.

End of titration : A manual titration must always be ended with the

[Cancel command] button (NOT with the [ ] fixed key) or with the

[FILL] key on the Manual Dosing Controller. This is the only way to ensure that subsequent method commands will be carried out.

22.2 Performing a sample series

You can use the following functions with sample series:

Statistics for multiple determinations

The statistics calculations are enabled in the method options (see chapter 16.5, page 172)

.

Sample table

You can enter the sample data for an entire sample series in the sample

table (see chapter 21, page 197) .

206 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 22 Determination run

Sequence of the sample series

After a determination is started, a check is made as to

– whether all of the devices required are connected and ready for operation.

– whether all titrants and sensors are present.

– whether the monitoring intervals for titrants, sensors and common variables are still valid.

– if a sample assignment table is activated

(see chapter 15.1.2, page 149) :

whether the entered sample identification appears in the sample assignment table. If this is the case, then the assigned method will be loaded automatically and the actual determination will be started. If a different method has been entered in the sample table, then it will be ignored.

At the end of the determination,

– the line in the sample table is deleted.

– the next determination is started (if you are working with the autostart function enabled

(see chapter 18, page 186) ).

22.3 Canceling determinations manually

You can cancel a determination at any time with the fixed key [ ] . The command being carried out at that moment is canceled and only calculations and reports are executed or created.

If you are working with the autostart function activated

(see chapter 18, page 186)

, the entire series will be canceled.

NOTICE

If you do not wish to cancel the entire determination, but only the current command, use the function [Cancel command] in the live display.

The subsequent method commands are carried out normally.

900 Touch Control ■■■■■■■■ 207

23.1 Editing the sample data of the running determination ■■■■■■■■■■■■■■■■■■■■■■

23 Live modifications

23.1 Editing the sample data of the running determination

The sample data can be entered or modified in the main dialog while a determination is running. The sample data entered at the end of the determination is always used in calculations.

Proceed as follows to edit the sample data:

1 Display the main dialog

Tap on the fixed key [ ] .

The main dialog is displayed. The determination continues to run in the background.

2 Edit the sample data

Edit the sample data and apply with [OK] .

3 Display the live dialog

Tap on [Live display] .

The live dialog is displayed once again.

NOTICE

Make sure that the input dialogs are closed before the determination is ended. Otherwise the determination will have to be recalculated.

208 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 23 Live modifications

23.2 Editing the sample table while a determination is running

You can insert new lines or delete existing ones or edit sample data while a determination is running.

NOTICE

To ensure that no problems arise during the run and that the current data is always available for the calculation, we recommend that the input dialogs always be closed.

Editing the sample table

Proceed as follows to edit the sample table:

1 Display the main dialog

Tap on the fixed key [ ] .

The main dialog is displayed. The determination continues to run in the background.

2 Select the sample data

Tap on Sample table .

The sample table is displayed. The first line always contains the sample data of the running determination.

Select the desired line.

Tap on the button [Edit] .

3 Edit the sample data

Edit the sample data and apply with [OK] .

NOTICE

In addition to the sample data, the method can also be modified, except in cases where the determination is running.

4 Display the live dialog

Tap on the fixed key [ ] .

The main dialog is displayed.

Tap on [Live display] .

900 Touch Control ■■■■■■■■ 209

23.3 Live display

The live dialog is displayed once again.

23.3 Live display

Main dialog: Live display

■■■■■■■■■■■■■■■■■■■■■■

210 ■■■■■■■■

The current method and the system status are displayed in the title line.

The currently running command is displayed directly underneath the title line.

The following functions are available to you while a determination is running:

[Hold]

Pause the determination. The label changes to [Continue] . This can be used to continue the run.

[Cancel command]

Cancel the current method command. The next command is started.

[View] (only for titrations/measurements)

To modify the settings for the curve and measured value display, see below.

[Sample data] (visible only if a titration or a measurement is not currently running)

Switch to the main dialog in order to edit the sample data.

[Edit parameters]

Edit the method options and live parameters (see chapter 23.4, page

212)

.

[Results]

Display the results dialog (see chapter 24, page 214)

.

Dialog "View"

Main dialog: Live display ▶ View

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 23 Live modifications

View

[Meas. value options]

Type of curve display.

Selection Curve | Curve & meas. values | Measured values | Curve & sample data

Curve & meas. values Default value

Curve

Only the curve is displayed.

Curve & meas. values

The measured values are displayed to the right next to the curve.

Measured values

Only the measured values are displayed.

Curve & sample data

The sample data is displayed to the right next to the curve.

Define settings for the display of the measured values.

A maximum of three different values can be displayed at the same time.

The selection depends on the titration/measuring mode. The settings are saved separately for each mode.

[Curve options]

Define the settings for the curve display (see "Curve options", page 228)

.

The settings are saved separately for each titration/measuring mode and apply not only for the live display but also for the curve display in the results dialog.

900 Touch Control ■■■■■■■■ 211

23.4 Live parameters ■■■■■■■■■■■■■■■■■■■■■■

23.4 Live parameters

Live display: Edit parameters

Certain method parameters can be edited while a determination is being carried out. The modified parameters are taken into account at once. If you modify, for instance, the start conditions after the start volume has been dosed, then these modifications will not be taken into account until the next determination.

NOTICE

If the live modifications are also to be used in the future, then the method must be saved after the determination has been completed.

The live modifications are discarded as soon as you load a new method.

If you are working with the sample table and a different method is used for the next sample, then the live modifications will be discarded.

Proceed as follows to modify method parameters during an ongoing determination:

1 Select the method command

Tap on Edit parameters .

The command sequence is displayed.

212 ■■■■■■■■

[Edit command] is enabled only for those commands for which the live parameters can be edited.

All of the functions except for the method properties can be edited with [Method options] .

Select the desired method command.

Tap on [Edit command] .

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

2 Edit the method parameters

Change the desired parameters accordingly.

3 Display the live dialog

Tap on the fixed key [ ] .

The main dialog is displayed.

Tap on [Live display] .

The live dialog is displayed once again.

23 Live modifications

900 Touch Control ■■■■■■■■ 213

■■■■■■■■■■■■■■■■■■■■■■

24 Results and more determination data

Main dialog: Results

Detailed specifications concerning the current determination are displayed in the Results dialog:

[Result table]

[Statistics]

Current method

Method with which the determination was carried out.

Determination

Name of the determination. Displayed only if the determination has been saved.

Determination time

Date and time at which the determination was carried out.

Results

Calculated results. If monitoring of the result limits is activated, then the result will be highlighted in color:

– Green lettering: The result is within the limits

– Red lettering: The result is outside the limits

Commands

List of the data-generating commands (titrations, measurements, etc.).

The most important data is displayed for each command.

Open result table

(see chapter 26, page 236)

. The result table is a table in which the results for several determinations can be saved.

Display the statistical overview of a determination series (see chapter 25, page 231)

.

214 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

[Recalculate]

24 Results and more determination data

Recalculate the determination. The procedure will be executed immedi-

ately. Recalculation cannot be undone (see chapter 24.8, page 229) .

[Sign]

Sign the current determination.

If the method contains a command SIGN , then this button will be displayed instead of [Recalculate] . The determination can be signed directly on level 1 (with a second command also on level 2).

[Curve]

Display the curve of the current determination

(see chapter 24.7, page

227) .

[More data]

Display detailed specifications for the determination (determination properties, measuring point list, endpoint details, etc.), save determination

(see chapter 24.1, page 215) .

24.1 More determination data

Results dialog: More data

[Load/ Save]

[Properties]

900 Touch Control

All of the commands for which you can display other data (titrations, measurements, evaluations, etc.) are listed in the Results / More determination data dialog.

Load saved determinations (see chapter 24.5, page 222) and save current

determination (see chapter 24.6, page 227)

.

Display the properties of the current determination

(see chapter 24.4, page 219)

.

■■■■■■■■ 215

24.1 More determination data

[Local com. var.]

■■■■■■■■■■■■■■■■■■■■■■

(available only with loaded determinations)

Display the list of the common variables defined at the time of the determination

(see chapter 24.3, page 219) .

[Messages]

Display list with all of the messages which occurred during the determination

(see chapter 24.2, page 218) .

[View data]

Display all of the details for the selected command.

24.1.1 Details

Results dialog: More data ▶ View data

All of the variables generated by the determination are listed in the More determination data / View data dialog. These can be used in calculations.

[Calibration data]

(available only with loaded determinations)

Display calibration data of the sensor used which was valid at the time of the determination

(see "Calibration and titrant data", page 218) .

[Titrant data]

[Endpoint details]

(available only with loaded determinations)

Display titrant data which was valid at the time of the determination (see

"Calibration and titrant data", page 218)

.

Display details for each endpoint found (see "Endpoint details", page

217) .

216 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 24 Results and more determination data

[Measuring point list]

Display measuring point list of titrations and measurements (see "Measuring point list", page 217)

.

Measuring point list

Results dialog: More data ▶ View data ▶ Measuring point list

Results dialog: Curve ▶ Display curve ▶ Measuring point list

The individual measured values are listed in the View data / Measuring point list dialog. You can configure which columns are displayed ( [View] button). These settings are saved separately for each mode (DET, MET, etc.).

Column 1 / Column 2 / Column 3

Quantity displayed in the columns 1, 2 and 3. Both the selection of the quantities and the default setting depend on the mode.

Selection Volume | Measured value | Water | Bromine |

ERC | Delta meas.value

| Time | Temperature |

Drift | Volume drift | Signal drift | Index | Monitoring

Endpoint details

Results dialog: More data ▶ View data ▶ Endpoint details

Results dialog: Curve ▶ Display curve ▶ Endpoint details

Detailed information concerning each equivalence point or endpoint determined is displayed in the View data / Endpoint details dialog.

The volume, the measured value, the equivalence point criterion ERC (only

DET, MET), the time and the temperature are specified for each equivalence point or endpoint determined.

900 Touch Control ■■■■■■■■ 217

24.2 Messages ■■■■■■■■■■■■■■■■■■■■■■

The equivalence point criterion determined (ERC = Equivalence point Recognition Criterion) is a measure for the size of the jump in the titration curve. The ERC determined is compared with the EP criterion defined in the method for the evaluation of the titration curve

see "Evaluation and equivalence point criterion with DET", page 290

and "Evaluation and equivalence point criterion with MET", page 307)

. The ERC is thus an important quantity when you need to adjust the parameters for the evaluation of the equivalence points.

Calibration and titrant data

Results dialog: More data ▶ View data ▶ Calibration data

Results dialog: More data ▶ View data ▶ Titrant data

The calibration data or titrant data which was current at the time the determination was performed is displayed in the dialogs View data / Calibration data and View data / Titrant data .

NOTICE

Recalculation:

The concentration and titer of the titrant can be modified retroactively.

The values displayed here are used for the recalculation of a loaded determination. The values saved in the dialog System / Titrants are not changed as a result.

24.2 Messages

Results dialog: More data ▶ Messages

If messages occurred during a determination, they will be displayed in the

More determination data / Messages dialog. The following data for the message is indicated under [Details] :

Message number

Unique identification number.

Time

Moment that the message occurred.

Message

Message text.

218 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 24 Results and more determination data

24.3 Local common variables

Results dialog: More data ▶ Local com. var.

The list of common variables as it existed at the time of the determination is displayed in the More determination data / Local common variables dialog.

NOTICE

Recalculation:

The common variables can be modified retroactively. The values displayed here are used for the recalculation of a loaded determination.

The list of common variables in the System / Common variables dialog is not changed as a result.

24.4 Determination properties

Results dialog: More data ▶ Properties

Detailed information concerning the determination is displayed in the

More determination data / Properties dialog:

User

User who carried out the determination.

Sample number

Number which was entered in the dialog Control at the start of the determination.

Start counter

This value, just like the sample number, is increased by one each time a determination starts. In contrast to the sample number, this value cannot be modified by the user.

Duration determ.

Duration of the entire determination.

900 Touch Control ■■■■■■■■ 219

24.4 Determination properties ■■■■■■■■■■■■■■■■■■■■■■

Method status

Current status of the method.

– new

The method has been newly created and not yet saved.

– saved

The current method has been saved.

– modified

The current method has been modified but not saved yet (e.g

live modification).

– reviewed

The current method has been signed at level 1.

– released

The current method has been signed at level 2.

Determination

Name of the determination. Displayed only with saved determinations.

End of determin.

Way in which the determination was ended.

– Regular without errors

The determination was automatically ended at the end of the method.

– Manual stop

The determination has been canceled with the fixed key [ ] .

– Error

The determination has been canceled due to an error.

Status of deter.

– original

The determination has not been recalculated.

– recalculated

The determination has been recalculated.

– loaded

The determination has been loaded.

– saved

The determination has been saved. The version number indicates how many times the file has been saved.

– reviewed

The determination has been signed at level 1.

– released

The determination has been signed at level 2.

The following data is displayed only if the option Recalculating determination under Login options / Modification options is activated:

Recalculated by

User who recalculated the determination.

Recalculated on

Date and time at which the determination was recalculated.

220 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 24 Results and more determination data

Reason

Reason for the recalculation of the determination.

Comment

Comment for the recalculation of the determination.

The following information is displayed only if the determination has been signed at level 1:

Reviewed by

User who signed the determination at level 1.

Reviewed on

Date and time at which the determination was signed at level 1.

Reason

Reason for the signing of the determination.

Comment

Comment for the signing of the determination.

The following information is displayed only if the determination has been signed at level 2:

Released by

User who signed the determination at level 2.

Released on

Date and time at which the determination was signed at level 2.

Reason

Reason for the signing of the determination.

Comment

Comment for the signing of the determination.

Used devices

Hardware that was used for the determination.

– Touch Control with program version of the software and the serial number.

– Control instrument with the program version of the device software and the serial number.

– Measuring inputs with ADC type und serial number.

– MSB connector with peripheral devices connected to it (dosing device, stirrer).

[Delete signature]

Delete all of the signatures for the method or the determination. This but-

ton is inactive if level 2 has not yet been signed (see chapter 17.2, page

185) .

[Sign]

Sign the method or determination. This button is active only when work-

ing with activated login function and password (see chapter 17.1, page

184) .

900 Touch Control ■■■■■■■■ 221

24.5 Loading a determination ■■■■■■■■■■■■■■■■■■■■■■

24.5 Loading a determination

Proceed as follows to load a determination:

1 Opening the determination list

In the results dialog, tap on [More data] .

Tap on [Load/ Save] .

Tap on [Load] .

The selection of memory locations is displayed. This selection will be skipped if only one external storage medium is available.

Select the memory location where the desired determination is saved.

The selection of file groups is displayed

(see chapter 12.1, page

121)

. If only one group is available, then the determination list will be displayed directly.

Select the group with the desired determination.

Tap on [Show files] .

The determination list with the stored determination opens:

222 ■■■■■■■■

2 Selecting a determination

Select the desired determination.

3 Loading the determination

Tap on [Load] .

The determination is now loaded. The properties are displayed in the results dialog.

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 24 Results and more determination data

NOTICE

When you load a determination, the method with which the determination was carried out will be loaded automatically.

For this reason, save any modifications which may have been made to the method currently loaded before you load a determination.

24.5.1 Determination list

Results dialog: More data ▶ Load/ Save ▶ Load ▶ Show files ▶ View

[Previous 100]

[Next 100]

[Properties]

In the Show files / View dialog you can depict the individual determinations in greater detail. The sorting criterion and the number of saved determinations are displayed at the very top. The list contains numbered lines; each line represents one determination. In the default settings, the first sample identification and the first calculated result of each determination are displayed. In the properties, you define which data is displayed in the columns and the criteria according to which the determinations are

sorted (see "Properties", page 224)

. In addition, you can define search filters to aid you in finding the desired determinations

(see "Filter criteria", page 225)

.

Display the previous 100 determinations.

Display the next 100 determinations.

Configure the display of the columns and define filter criteria.

900 Touch Control ■■■■■■■■ 223

24.5 Loading a determination

[Delete]

Delete the selected file.

[Load]

Load the selected file.

Properties

Determination list: Properties

■■■■■■■■■■■■■■■■■■■■■■

Column 1

Column 2

Sort by

In the View / Properties dialog, you can define which data is displayed in the columns and the criteria according to which the determinations are to be sorted.

Selection of what is to be displayed in the first column.

Selection

Default value

Identification 1 | Identification 2 | Determ.

time | Result name | Method | Sample size |

User

Identification 1

Selection of what is to be displayed in the second column.

Selection Result 1…9

The numbering of the results corresponds to the order in which the results were calculated in the determination run.

Selection according to which criterion the list is to be sorted.

224 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 24 Results and more determination data

Selection

Default value

Identification 1 | Identification 2 | Determ.

time | Result | Method | Sample size | User

Determ. time

Identification 1

Sort alphabetically according to the first sample identification.

Identification 2

Sort alphabetically according to the second sample identification.

Determ. time

Sort chronologically according to the determination time.

Result

Sort according to the numerical value of the result.

Method

Sort alphabetically according to the method name.

Sample size

Sort according to the numerical value of the sample size.

User

Sort alphabetically according to the user who carried out the determination.

Sort sequence

Selection

Default value ascending | descending descending

[Filter]

Specify the filter criteria for the list.

Filter criteria

Determination list: Properties ▶ Filter

900 Touch Control

You can set filters in the Properties / Filter dialog. A filter is a rule defining what is to be displayed in the determination list. If a filter has been set,

■■■■■■■■ 225

24.5 Loading a determination ■■■■■■■■■■■■■■■■■■■■■■ this is indicated with a corresponding symbol in the Show files / View dialog.

Filter

Define the criterion according to which the list is to be filtered. The only lines which will be displayed are those which fulfil the filter criterion.

Selection

Default value no filter | Identification 1 | Identification 2 |

Date | Result | Method | Sample size | User no filter no filter

The list itself will not be filtered.

Identification 1

In the input field is equal to , select or enter the first sample identification according to which filtering is to be carried out.

Identification 2

In the input field is equal to , select or enter the second sample identification according to which filtering is to be carried out.

Date

In the input fields Date from … to , specify the period of time according to which filtering is to be carried out.

Result

In the input fields Value from … to , specify the value range according to which filtering is to be carried out.

Method

In the field is equal to , select or enter the method name according to which filtering is to be carried out.

Sample size

In the input fields Value from … to , specify the value range according to which filtering is to be carried out.

User

In the field is equal to , select or enter the user according to whom filtering is to be carried out.

NOTICE

The filter can be applied to empty fields as well, for example, if no sample identification has been entered. In this case, select only the desired filter criterion and leave the fields below empty.

226 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 24 Results and more determination data

24.6 Saving a determination

Results dialog: More data ▶ Load/ Save ▶ Save

Determinations can be saved as follows:

Manually in this dialog.

The suggested default setting for file names is the first 16 characters of

Identification 1, followed by the determination time in the format

YYYYMMDD-hhmmss. Extending the file name with the determination time has the advantage that the file names of all determinations are unique.

Automatically at the end of the determination (see chapter 16.5.6, page 182)

.

The following data is stored for each determination:

Calculated results

Measuring point lists, equivalence points, curves and variables which were generated during the determination

Determination properties

Statistics (if carried out)

Method with which the determination was carried out

Common variables (values and properties of all common variables at the time of the determination)

Titrant data of the titrant used

Calibration data of the sensor used

24.7 Curves

Results dialog: Curve

900 Touch Control

The curve of the current determination is displayed in the Results / Curve dialog. You can configure the curve display.

■■■■■■■■ 227

24.7 Curves ■■■■■■■■■■■■■■■■■■■■■■

[Endpoint details]

[Measuring point list]

Display measuring point list of titrations and measurements

(see "Measuring point list", page 217) .

[Curve options]

Display details for each endpoint found (see "Endpoint details", page

217) .

Define the settings for the curve display

(see "Curve options", page 228)

.

The settings are saved separately for each titration/measuring mode and apply not only for the live display but also for the curve display in the results dialog.

Curve options

Results dialog: Curve ▶ Curve options x axis y1 axis

You can make the settings for the curve display in the Curve / Options dialog. These settings are saved separately for each mode (DET, MET, etc.) and also apply for the live display. Both the selection of the quantities and the default setting depend on the mode.

Quantity which is shown on the x axis of the graph.

Selection Volume | Measured value meas.value

| Time | Temperature | Volume drift | Signal drift

| ERC | Delta

Quantity which is shown on the y1 axis of the graph.

228 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 24 Results and more determination data

Selection Volume | Measured value | ERC | Delta meas.value

| Time | Temperature | Volume drift | Signal drift y2 axis

Quantity which is shown on the y2 axis of the graph.

Selection

Default value none | Volume | Measured value meas.value

| Time | Temperature | Volume drift | Signal drift none

| ERC | Delta

Color

Selection of a color for the curves.

Grid on | off (Default value: on )

If this parameter is activated, grid lines are shown.

Display measuring points on | off (Default value: off )

If this parameter is activated, the curve is shown with the single measured values.

24.8 Recalculation and reevaluation

Results dialog: Recalculate

You can use the [Recalculate] function to recalculate and reevaluate determinations. This is necessary in the following cases:

If you have modified the evaluation parameters in a way that changes the recognized endpoints or equivalence points.

If you have modified the calculations.

If you have modified the variables used in calculations, e.g. sample size, titer or common variables.

NOTICE

Neither the recalculation nor the reevaluation can be undone.

We therefore recommend that the original determination be saved

beforehand (see chapter 24.6, page 227)

. The best way of doing this is to use the Save determination automatically function

(see chapter

16.5.6, page 182)

.

900 Touch Control ■■■■■■■■ 229

24.8 Recalculation and reevaluation ■■■■■■■■■■■■■■■■■■■■■■

The following method commands are executed once again with the

[Recalculate] function:

DET and MET titrations:

The potentiometric evaluation will be carried out again.

EVAL commands:

All evaluations are carried out again. You can also insert new evaluations retroactively.

CALC command:

All calculations are carried out again. You can also define new calculations retroactively.

If you subsequently modify the titrant in a method, then the variables

TITER and CONC of the "new" titrant will be used. If a result is assigned to the variable TITER in a calculation, then the titer of the "new" titrant will also be overwritten.

NOTICE

If you modify the calibration data of the sensor used in the System /

Sensors dialog, these modifications will not be taken into account.

Changing the sensor in the method also has no influence on the measured data.

If a determination has been recalculated, this will be documented in the

determination properties (see chapter 24.4, page 219) .

Recalculation/reevaluation of loaded determinations

With loaded determinations, it is assumed that they were carried out a long time ago and that the titrant data and common variables of the system no longer correspond to those of the determination. This is why the most important titrant data of the titrants used and the list of common variables are stored with each determination. This data will be used if you recalculate a loaded determination. If you wish to recalculate the result with a corrected common variable or with a corrected titer, you must therefore modify the respective values in the More determination data / Local common variables or More determination data / View data dialog.

If a result is assigned to the variable TITER or to a common variable , then the titer or common variable which is currently valid in the system will not be overwritten until after a query has been confirmed.

If you modify the titrant retroactively in the method, this will not be taken into account.

The recalculated results of loaded determinations are not saved in the result table.

230 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 25 Statistics

25 Statistics

Main dialog: Results ▶ Statistics

In the dialog Results , you can use [Statistics] to display the statistical overview of a determination overview. You can statistically evaluate a maximum of nine results calculated in a determination. A statistical series can contain a maximum of 20 determinations.

Activating the statistics function

Proceed as follows to activate the statistics function:

1 Editing the method options

In the main dialog, tap on [Edit parameters] .

Tap on [Method options] .

900 Touch Control

Activate the Statistics option.

Under Number of determinations , enter the desired number of determinations.

The results for which statistical calculations are performed are defined in the calculation command CALC

(see "Variable for mean value", page 153)

.

■■■■■■■■ 231

■■■■■■■■■■■■■■■■■■■■■■

[Add determ.]

[Details]

In this dialog, the method with which the determination was carried out is displayed in the first line. The second line shows how many determinations have already been carried out and the total number of determinations to be carried out. The name of the result and the mean value are shown for each result.

Add a further determination to the determination series (see chapter 25.3, page 234)

.

Display the statistics results for the selected result and the single results from all determinations, see next chapter.

NOTICE

If you recalculate a determination, the statistics for this determination are also recalculated.

232 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

25.1 Displaying details for a result

Results dialog: Statistics ▶ Details

25 Statistics

[Sample data]

[Determ. on/off]

[Result on/off]

In the dialog Statistics / Details , the following details are displayed:

Result name

Mean value

On the right-hand side of the dialog the number of single results from which the mean value was calculated is displayed. In this example, it is

3.

Absolute standard deviation ( s abs )

Relative standard deviation ( s rel )

Result and sample size of each single determination

Results that could not be calculated are indicated with invalid and ignored.

Display the sample data of the selected determination.

Remove all of the results of the selected determination from the statistics.

All of the entries of this line are marked with an asterisk (*), the statistics will be recalculated automatically.

Remove the selected result from the statistics. The result is marked with an asterisk (*), the statistics will be recalculated automatically.

900 Touch Control ■■■■■■■■ 233

25.2 Deleting statistical data ■■■■■■■■■■■■■■■■■■■■■■

NOTICE

If the mean value has been assigned to a common variable or to the variable TITER (see command CALC ), then the determination must be recalculated manually

(see chapter 24.8, page 229) .

25.2 Deleting statistical data

In the dialog Control , you can delete all of the statistics data manually with the function [Delete statistics] . The statistics data is deleted automatically in the following cases:

■ when all of the determinations of the determination series have been carried out and a new determination has been started afterwards.

when a new method is loaded (even if it is the same method)

If the sample table or the sample assignment table is used, the statistics data is only deleted when another method is loaded.

25.3 Adding a determination to a determination series

You can use the function [Add determ.] to add an additional sample to a determination series, e.g. because a determination was faulty and had to be removed from the statistics.

234 ■■■■■■■■

The number of determinations by which the set statistics counter has been increased is shown above the list in brackets. This modification is also displayed in the main dialog and in the statistics report.

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 25 Statistics

NOTICE

The number of samples defined in the method options does not, however, change as a result.

900 Touch Control ■■■■■■■■ 235

26 Result table

Main dialog: Results ▶ Result table

■■■■■■■■■■■■■■■■■■■■■■

[Load/ Save]

[Properties]

[Delete all]

The result table is suitable for displaying in detail the results of all of the determinations performed on a single day, for instance. You can save the results of up to 200 determinations. A maximum of 9 results from a single determination can be saved in the result table. You define which results from this table are to be saved using the CALC command. Because the size of the display is not sufficient to display all of the data at the same time, it is practical to print out the result table.

The sorting criterion and the number of determinations from which results are saved in the result table are displayed at the top. The list contains numbered lines; each line represents one determination. In the default settings, the first sample identification and the first calculated result of each determination are displayed. In the properties, you define which data is displayed in the columns and the criteria according to which the determi-

nations are sorted (see chapter 26.1, page 237) . In addition, you can

define search filters to aid you in finding the desired determinations (see

"Filter criteria", page 239) .

Load a saved result table

(see chapter 26.3, page 241)

or save the current

result table (see chapter 26.2, page 241)

.

Configure the display of the columns and define filter criteria.

Delete the result table completely.

236 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

[Delete]

Delete the selected line from the result table.

[Details]

26 Result table

Display details concerning the selected determination.

"Result table / Details" dialog

The following determination data is displayed in this dialog:

User

User who carried out the determination.

Method

Method with which the determination was carried out.

Determ. time

Date and time at which the determination was carried out.

Sample data (Identification 1 and 2 and sample size).

All of the results of this determination saved in the result table.

26.1 Properties

Result table: Properties

Column 1

900 Touch Control

In the Result table / Properties dialog, you can select the data to be displayed with each determination as well as the sorting criterion for the list.

Selection of what is to be displayed in the first column.

Selection

Default value

Identification 1 | Identification 2 | Determ.

time | Result name | Method | Sample size |

User

Identification 1

■■■■■■■■ 237

26.1 Properties ■■■■■■■■■■■■■■■■■■■■■■

Column 2

Selection of what is to be displayed in the second column.

Selection Result 1…9

The numbering of the results corresponds to the order in which the results were calculated in the determination run.

Sort by

Selection according to which criterion the list is to be sorted.

Selection

Default value

Identification 1 time | Result | Method | Sample size | User

Determ. time

| Identification 2 | Determ.

Identification 1

Sort alphabetically according to the first sample identification.

Identification 2

Sort alphabetically according to the second sample identification.

Determ. time

Sort chronologically according to the determination time.

Result

Sort according to the numerical value of the result.

Method

Sort alphabetically according to the method name.

Sample size

Sort according to the numerical value of the sample size.

User

Sort alphabetically according to the user who carried out the determination.

Sort sequence

Selection

Default value ascending | descending descending

Max. number of lines

If the result table contains the quantity of lines defined here, then no further results can be saved. If this is the case, save the result table and create a new one.

Input range

Default value

10 to 200

100

[Filter]

Specify the filter criteria for the list.

238 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

[Delete automat.]

26 Result table

Define the settings for the automatic deletion of the result table

(see

"Deleting the result table automatically", page 240)

.

Filter criteria

Result table: Properties ▶ Filter

Filter

900 Touch Control

You can set filters in the Properties / Filter dialog. A filter is a rule defining what is to be displayed in the result table. If a filter has been set, this is indicated with a corresponding symbol in the Results / Result table dialog.

Define the criterion according to which the list is to be filtered. The only lines which will be displayed are those which fulfil the filter criterion.

Selection

Default value no filter | Identification 1 | Identification 2 |

Date | Result | Method | Sample size | User no filter no filter

The list itself will not be filtered.

Identification 1

In the input field is equal to , select or enter the first sample identification according to which filtering is to be carried out.

Identification 2

In the input field is equal to , select or enter the second sample identification according to which filtering is to be carried out.

Date

In the input fields Date from … to , specify the period of time according to which filtering is to be carried out.

■■■■■■■■ 239

26.1 Properties ■■■■■■■■■■■■■■■■■■■■■■

Result

In the input fields Value from … to , specify the value range according to which filtering is to be carried out.

Method

In the field is equal to , select or enter the method name according to which filtering is to be carried out.

Sample size

In the input fields Value from … to , specify the value range according to which filtering is to be carried out.

User

In the field is equal to , select or enter the user according to whom filtering is to be carried out.

NOTICE

The filter can be applied to empty fields as well, for example, if no sample identification has been entered. In this case, select only the desired filter criterion and leave the fields below empty.

Deleting the result table automatically

Result table: Properties ▶ Delete automat.

In the Properties / Delete automatically dialog, you define when the contents of the result table is to be automatically deleted.

Delete oldest line only on | off (Default value: off )

If this option is activated at the same time that the result table is full, the oldest entry will be deleted as soon as the results of a new determination are saved.

Time when result table is deleted:

Switch on on | off (Default value: off )

If this option is activated, then the result table will be deleted each time you switch on the Touch Control.

Start a new sample series on | off (Default value: off )

If this option is activated, then the result table will be deleted when you start a new sample series (with activated autostart function!).

240 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 26 Result table

Save result table on | off (Default value: on )

If this option is activated, the result table will be deleted as soon as you have saved it.

26.2 Saving the result table

NOTICE

When you save the result table, the complete determination will not be saved with each entry. If you wish to save the determination data for each determination, use the Save determination automatically function

(see chapter 16.5.6, page 182)

.

26.3 Loading the result table

Proceed as follows to load a result table:

1 Opening the current result table

In the results dialog, tap on [Result table] .

Tap on [Load/ Save] .

Tap on [Load] .

The selection of memory locations is displayed. This selection will be skipped if only one external storage medium is available.

Select the memory location where the desired result table is saved.

The selection of file groups is displayed (see chapter 12.1, page

121) . If only one group is available, then the list of the saved

result tables will be displayed directly.

Select the group with the desired result table.

Tap on [Show files] .

The list with the saved result tables opens.

2 Selecting the result table

Select the desired result table.

3 Loading the result table

Tap on [Load] .

The result table is now loaded and will be displayed immediately.

900 Touch Control ■■■■■■■■ 241

■■■■■■■■■■■■■■■■■■■■■■

27 Printing

Reports can be printed out as follows:

Manually with the [ ] fixed key in this dialog.

The fixed key is disabled while a determination is running.

Automatically in the determination run with the command REPORT

(see chapter 29.14, page 433)

.

The Titrando manual describes how to connect a printer to a Titrando sys-

tem. The configuration of the printer is described in Chapter 11.7, page

109 .

[Report seq. as def. in method]

Print out all of the reports which are defined in the loaded method.

[Print screenshot]

Activate the "Print Screen" function. The current screen contents will be printed out the next time the fixed key [ ] is tapped on.

[Report options]

This screenshot shows the printing dialog associated with the main dialog.

The sample data is entered in the main dialog, which is why you can print out the sample data from there in a context-sensitive manner with [Current sample data] .

The following buttons are also available in the other print dialogs:

Defining the general settings for the report output.

[Clear spooler]

Delete all outstanding printing orders.

242 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

[More reports]

Open the list of available reports.

27 Printing

More reports

Fixed key [ ] : More reports

The dialog Print / More reports displays a list of all of the available reports.

[System reports]

[Method reports]

NOTICE

If you are using a CUSTOM Neos printer, you can print only the reports marked with a # . The designation on the reports is only visible once the CUSTOM NEOS printer is selected (see screenshot above).

Opening the list of reports for the system settings.

NOTICE

If you are using a CUSTOM NEOS printer, you cannot print system reports.

Opening the list of reports for the current method.

900 Touch Control ■■■■■■■■ 243

27.1 General report options ■■■■■■■■■■■■■■■■■■■■■■

NOTICE

If you are using a CUSTOM NEOS printer, you cannot print method reports.

[Send]

[Edit]

[Print]

Create a PC/LIMS report.

Edit the settings of the selected report.

Print the selected report.

27.1 General report options

Fixed key [ ] : Report options

The settings for the report printout can be adjusted in the dialog Print /

Report options to indicate, for example, whether a report header, a signature line or a frame is to be printed.

Report header

Defining the output of the report header. The report header contains general information such as device type, serial number, program version and the printing date.

NOTICE

You can also define your own report header, which is printed in addition to this report header ( System ▶ Templates ▶ Report header ).

Selection

Default value off | once | on each page on each page off

The report header will not be printed.

once

The report header will only be printed on the first page.

on each page

The report header will be printed on every page.

244 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Signature line

27 Printing

Output of a special line for date and signature. This line will be printed at the very bottom of every page.

Selection

Default value off off

| once | on each page off

No signature line will be printed.

once

The signature line will only be printed on the last page.

on each page

The signature line will be printed on every page.

Frame on | off (Default value: on )

If this parameter is activated, a frame is printed as lateral marking.

27.2 Settings of the individual reports

Fixed key [ ] : More reports ▶ Report ▶ Edit

Command REPORT: Report ▶ Edit

Settings can be edited for the following reports (for details, see online help):

Result report

Curve

Definition of the curve size, display of the individual measuring points and grid lines, etc.

Measuring point list

Definition of the method command to which the measuring point list is to be printed.

Calculations

Definition of the accuracy with which the variables used as well as the results calculated are printed.

Statistics short

Definition of the time at which the report is to be printed (with each determination or only at the end of a sample series).

Statistics overview

Definition of the time at which the report is to be printed (with each determination or only at the end of a sample series).

Sample table

900 Touch Control ■■■■■■■■ 245

27.3 List of all printable reports ■■■■■■■■■■■■■■■■■■■■■■

Result table

Definition of the time at which the report is to be printed (with each determination, at the end of a sample series or at the end of the sample table).

Used devices

Variables

Monitoring

27.3 List of all printable reports

The following reports can be printed out with the fixed key [ ] :

NOTICE

Report

Result report

Curve

Measuring point list

Calculations

Used devices

Variables

If you have selected a CUSTOM NEOS printer, you can print only the reports that are marked with a # on the display.

Contents

# Report with determination properties, sample data, calculated results, etc. If there are multiple determinations, the statistics will also be printed out.

# Curve report. The settings for the curve print-out are adopted from the curve options of the live display.

# Measuring point list report.

# Details concerning the calculations carried out (parameter settings in accordance with the CALC command and the calculated results).

The devices used for the determination, as displayed in the

More determination data / Properties dialog.

All determination variables, as displayed in the results dialog.

Details for the monitored quantities (STAT, DOS only).

Monitoring

Statistics

Statistics short # Summary of the statistics calculations. The number of determinations, the mean value, and the absolute and relative standard deviations are printed out for each result.

246 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 27 Printing

Report

Statistics overview

Contents

# Detailed statistical overview. The sample data and all individual results are printed out for each determination. The number of determinations, the mean value, and the absolute and relative standard deviations are printed out for each result.

PC/LIMS

PC/LIMS report Machine-readable report with all of the important data for a determination. This report can be saved as a TXT or

UTF-8 file or sent to a terminal program or a LIMS via an

RS-232 interface. The definition is carried out in the device manager

(see chapter 11.3.2, page 83) .

Sample data

Current sample data

Sample table

Sample data of the current determination, context-sensitive only from the main dialog.

List of all determinations in the sample table with the respective sample data, as entered in the sample table.

Result table

Result table List of all determinations in the result table with results and with the determination data, as saved in the result table.

More reports ▶ Method reports

Method sequence Method properties and the list of all method commands.

Parameters full

Titration & measuring param.

Modified parameters

Method properties and options, all method commands with all parameters.

All parameters which no longer have their default settings will be printed in bold . All parameters which have been modified in comparison to the stored version of the method are indicated by an *.

Method properties; titration, measurement and calibration commands with all parameters.

All parameters which no longer have their default settings will be printed in bold . All parameters which have been modified in comparison to the stored version of the method are indicated by an *.

Method properties, all method parameters which have been modified in comparison to the stored version of the method.

900 Touch Control ■■■■■■■■ 247

27.3 List of all printable reports ■■■■■■■■■■■■■■■■■■■■■■

Report Contents

Non default parameters

More reports ▶ System reports

System settings

Method properties, all method parameters which no longer have any default settings.

System settings Settings for acoustic signals and accuracy of the measured value display.

Dialog options

User list

Settings for routine dialog and expert dialog.

List with all of the users configured in the system, together with their data, context-sensitive only from the user list.

Login options Login options, context-sensitive only from the respective dialog.

Titrants

Titrant list

All titrant data short

All titrant data full

Titrant data short

Titrant data full

List of all the titrants configured in the system.

The most important titrant data of all titrants (name, concentration, titer, last titer determination).

All of the titrant data of all of the titrants (working life, titer options, data concerning the exchange unit/dosing unit).

The most important titrant data of the selected titrant, context-sensitive only from the editing dialog.

All titrant data of the selected titrant, context-sensitive only from the editing dialog.

Sensors

Sensor list

All sensor data short

All sensor data full

Sensor data short

Sensor data full

List of all the sensors configured in the system.

The most important sensor data for all sensors (name, calibration data).

All of the sensor data for all of the sensors (working life, complete calibration data, calibration interval).

The most important sensor data of the selected sensor, context-sensitive only from the editing dialog.

All of the sensor data of the selected sensor, context-sensitive only from the editing dialog.

Device manager

Device list

All device properties

List of all devices configured in the system.

Properties of all the devices configured in the system.

248 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 27 Printing

Report

Device properties

GLP manager

GLP data

Templates

Templates sample data

Result template list

All result templates details

All data stored in the GLP manager.

Common variables

Common variable list

All common variable properties

List of all the common variables defined in the system, together with their most important data (name, value, status).

Properties of all common variables (name, value, validity, status).

Common variable properties Properties of the selected common variable (name, value, validity, status), context-sensitive only from the editing dialog.

Local Common variable list List with the common variables available at the time of the determination, together with their most important data

(name, value, status), context-sensitive only from the dialog "More determination data / Local common variables".

Local common var. properties

Properties of the selected common variable (name, value, validity, status), context-sensitive only from the dialog

"More determination data / Local common variables".

Sample identification list and sample assignment table.

List with all of the user-generated result templates.

Details of the all of the user-generated result templates

(calculation formula, result options, note).

Result template details

Input/Output lines

Custom calibration buffers

Contents

Properties of the selected device, context-sensitive only from the editing dialog.

Details of the selected result template, context-sensitive only from the editing dialog.

List with all of the defined input and output lines at the remote interface (name, bit pattern).

Temperature tables for all defined custom calibration buffers.

Rack tables

Sample rack list List with all of the sample racks configured in the system, together with designation, number of positions and rack code.

900 Touch Control ■■■■■■■■ 249

■■■■■■■■■■■■■■■■■■■■■■

28 Manual control

250

[Measure]

[Dosing]

■■■■■■■■

The following functions are available in the manual control:

Carrying out manual measurements (see chapter 28.2, page 252) .

Measuring modes: pH (potentiometric pH measurement)

U (potentiometric voltage measurement)

Ipol (voltametric measurement with selectable polarization current)

Upol (amperometric measurement with selectable polarization voltage)

T (temperature measurement)

Conc (concentration measurement)

Cond (conductivity measurement)

NOTICE

Only active if a Dosino or dosing unit is connected.

Manual dosing (see chapter 28.3, page 256)

.

The following dosing functions are available:

Preparing the exchange unit or dosing unit

Emptying the dosing unit

Filling the dosing cylinder of the exchange unit/dosing unit

Dosing a specified volume

Dosing continuously

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

[Conditioning]

28 Manual control

Conditioning the volumetric or coulometric titration cell manually (see chapter 28.5, page 266) .

[Stir]

Controlling the stirrer manually (see chapter 28.4, page 265) .

[Remote]

Scanning the input lines on the remote interface and setting the output lines

(see chapter 28.6, page 267) .

[Sample Processor]

Controlling the Sample Processor manually (see chapter 28.7, page 268)

.

28.1 Opening and closing the manual control

Opening the manual control

1 Opening the dialog window

Tap on the fixed key [ ] .

Manual control opens:

900 Touch Control

2 Select the function

Tap on the button for the desired function.

If the function can be carried out by several devices, a dialog to select the device will appear first; otherwise the dialog for the selected function will appear immediately.

■■■■■■■■ 251

28.2 Measuring ■■■■■■■■■■■■■■■■■■■■■■

Closing the manual control

1

Tap on the fixed key [ ] .

NOTICE

Manual control can also be exited when a manual sequence has been started but has not yet ended. The fact that a sequence has been started in manual control can be recognized by the corresponding sym-

bol in the title line (see chapter 6.2.3, page 19)

.

28.2 Measuring

Manual control ▶ Measure

252 ■■■■■■■■

Manual measurements can be carried out with the function [Measure] .

Proceed as follows:

1 Selecting a sensor

Select the desired sensor from the sensor list.

All of the sensors available in the sensor list are displayed. Sensors are defined under System ▶ Sensors .

Select the measuring input at which the sensor is connected.

2 Select a measuring mode

Select the desired measuring mode.

Only those measuring modes are displayed which are advisable for the selected sensor.

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 28 Manual control

3 Enter a measuring temperature

Enter the measuring temperature if no temperature sensor is connected. If a temperature sensor is connected, then the temperature will be measured automatically.

This temperature is used for automatic temperature compensation with pH measurements.

4 Starting the measurement

Tap on [Start] .

The measured value display is opened:

The ongoing measurement is canceled with [Stop] . The label switches once again to [Start] .

28.2.1 Parameter description

"Manual control / Measure" dialog

Sensor

Selection of the sensor from the sensor list. Sensors are defined under

System ▶ Sensors .

Selection Selection of configured sensors

Measuring input

Selection of the measuring input the sensor is connected to. The selection depends on whether the control device has one or two measuring interfaces.

Selection

Default value

1 | 2

1

900 Touch Control ■■■■■■■■ 253

28.2 Measuring

Measuring mode

■■■■■■■■■■■■■■■■■■■■■■

Selection of the measuring mode. Only those measuring modes are displayed which are advisable for the selected sensor.

Selection pH | U | Ipol | Upol | Conc | Cond | T pH potentiometric pH measurement

U potentiometric voltage measurement

Ipol voltametric measurement with selectable polarization current

Upol amperometric measurement with selectable polarization voltage

Conc concentration measurement as direct measurement

Cond conductivity measurement

T temperature measurement

Temperature

This parameter is not available for conductivity measuring cells.

Temperature entered manually. If a temperature sensor is connected, then the temperature will be measured continuously. This value is used for temperature correction in pH and concentration measurements.

Input range

Default value

–20.0 to 150.0 °C

25.0 °C

Measuring temp.

[Temp. compens.]

[Info sensor]

This parameter is available only with conductivity measurements.

Temperature entered manually. If a temperature sensor is connected then the temperature will be measured continuously.

Input range

Default value

–20.0 to 150.0 °C

25.0 °C

This button is available only with conductivity measurements.

Edit the parameters for converting the measured conductivity to a reference temperature.

Display information on the sensor.

254 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 28 Manual control

"Measure / Temperature compensation" dialog

The parameters for converting the measured conductivity to a reference temperature are defined under [Temp. compens.] .

Reference temp.

Temp. compensation

Type of temperature compensation.

Selection constant | ISO / DIN constant

The temperature compensation is carried out with the value entered for the temperature coefficient.

ISO / DIN

For ground water, spring water and surface water, the temperature coefficients according to standard ISO 7888:1985 (German version:

DIN EN 27888:1993) are stored in the system.

Temp. coefficient

The measured conductivity is converted to this temperature.

Input range

Default value

Selection

–20.0 to 150.0 °C

25.0 °C off off

The temperature compensation is disabled.

This parameter can only be edited with Temp. compensation = constant .

Temperature coefficient belonging to the reference temperature entered above. With this temperature coefficient, the conductivity is converted to the reference temperature.

Input range

Default value

0.00 to 9.99 %/°C

0.00 %/°C

900 Touch Control ■■■■■■■■ 255

28.3 Dosing ■■■■■■■■■■■■■■■■■■■■■■

28.3 Dosing

Manual control ▶ Dosing

The following dosing functions are available:

Preparing the exchange or dosing unit

(see chapter 28.3.3, page 262)

Emptying the dosing unit

(see chapter 28.3.4, page 263)

Filling the dosing cylinder of the exchange unit/dosing unit

(see chapter 28.3.5, page 264)

Dosing a specified volume

(see chapter 28.3.2, page 260)

Dosing continuously

(see chapter 28.3.1, page 258)

Selecting the dosing device

If several dosing devices are connected to a control instrument, then the dialog for selecting a dosing device will be opened.

[Dosing device X]

Selection of the dosing device. The MSB connector and the type of the connected dosing device are displayed.

Various information is shown next to the button:

Titrant

For dosing devices of the type 8XX with an intelligent exchange unit or dosing unit the titrant name is read off the data chip.

Cylinder volume

For dosing devices of the type 6XX or 7XX only the cylinder volume is displayed.

---

For exchange or dosing units which are not or not correctly attached.

Manual busy

When a dosing function is being carried out.

256 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

[All dosing devices]

28 Manual control

Carrying out the functions [Prepare] , [Empty] and [Fill] at the same time with several dosing devices of the control device.

Selecting the dosing function

If only one dosing device is connected, then the properties of this dosing device will be shown directly. This information varies according to the type of dosing device:

Type 8XX with integrated data chip:

If an exchange or dosing unit with integrated data chip is attached, the data stored on the data chip is displayed.

Type 8XX without integrated data chip:

If an exchange unit without integrated data chip is attached, then the titrant can be selected from the titrant list. We recommend always selecting the titrant. This ensures that the parameters defined for the titrant for the preparing are used. Titrants are defined under System ▶ Titrants .

900 Touch Control ■■■■■■■■ 257

28.3 Dosing

[Empty]

■■■■■■■■■■■■■■■■■■■■■■

Empty the cylinder and the tubings of the dosing unit

(see chapter 28.3.4, page 263)

.

[Prepare]

Rinse the cylinder and the tubings of the exchange / dosing unit (see chapter 28.3.3, page 262)

.

[Fill]

Fill the cylinder of the exchange / dosing unit (see chapter 28.3.5, page

264) .

[Dosing fixed vol.]

Dose a specified volume

(see chapter 28.3.2, page 260)

.

[Dosing]

Dose continuously (see chapter 28.3.1, page 258)

.

28.3.1 Continuous dosing

Manual control: Dosing ▶ Dosing

With the function [Dosing] , dosing is continuous for as long as you hold the button [Dose] pressed down.

258 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Proceed as follows:

1 Configure the dosing function

NOTICE

28 Manual control

The dosing and filling rates should be decreased for viscous liquids.

The maximum dosing and filling rates depend on the cylinder

volume (see chapter 32, page 492) .

Enter the dosing rate.

Enter the filling rate.

2 Start dosing

Tap on [Dose] .

The volume dosed is displayed. When the volume of one cylinder has been dosed, the dosing cylinder will be refilled automatically.

3 Fill the cylinder

Tap on [Fill] .

The dosing cylinder is filled. The displayed volume value will be reset to 0.0000 mL.

NOTICE

Dosing rate

If you exit the dialog with the fixed key [ ] , the dosing cylinder is filled automatically.

Parameter description

Rate at which it is dosed.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum | dynamic maximum maximum

The maximum dosing rate depends on the cylinder volume (see chapter 32, page 492) .

900 Touch Control ■■■■■■■■ 259

28.3 Dosing ■■■■■■■■■■■■■■■■■■■■■■ dynamic

This setting can only be selected when the dosing cylinder is filled. The dosing is being carried out faster and faster until the maximum dosing rate is reached (starting with 1 mL/min and then doubling the dosing rate every 1.5 s).

Filling rate

Rate at which the dosing cylinder is filled.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum filling rate depends on the cylinder volume (see chapter

32, page 492) .

28.3.2 Dosing fixed volumes

Manual control: Dosing ▶ Dosing fixed vol.

260 ■■■■■■■■

You can dose a particular volume with the [Dosing fixed vol.] function.

Proceed as follows:

1 Configure the dosing function

NOTICE

The dosing and filling rates should be decreased for viscous liquids.

The maximum dosing and filling rates depend on the cylinder

volume (see chapter 32, page 492)

.

Enter the desired volume.

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 28 Manual control

Enter the dosing rate.

Enter the filling rate.

2 Start dosing

Tap on [Start] .

The volume display is shown. When the volume of one cylinder has been dosed, the dosing cylinder will be refilled automatically.

Volume

Dosing rate

900 Touch Control

Continuous dosing is paused with [Hold] . The label changes to

[Continue] . It can be used to resume the run.

3 Fill the cylinder

With the default settings

(see "Fill automatic.", page 262) the dosing

cylinder is filled automatically.

Otherwise:

Tap on [Fill] .

The dosing cylinder is filled. The displayed volume value will be reset to 0.0000 mL.

Parameter description

Volume which is dosed.

Input range

Default value

0.00000 to 99999.9 mL

0.10000 mL

The rate at which dosing takes place.

Input range 0.01 to 166.00 mL/min

■■■■■■■■ 261

28.3 Dosing ■■■■■■■■■■■■■■■■■■■■■■

Selection

Default value maximum maximum maximum

The maximum dosing rate depends on the cylinder volume

(see chapter 32, page 492) .

Filling rate

Rate at which the dosing cylinder is filled.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum filling rate depends on the cylinder volume (see chapter

32, page 492) .

Fill automatic.

on | off (Default value: on )

If this parameter is activated, then the dosing cylinder is being automatically filled at the end of the dosing.

28.3.3 Preparing

Manual control: Dosing ▶ Prepare

The Prepare function is used to rinse and fill the cylinder and tubings of the exchange or dosing unit air bubble-free. You should carry out this function before the first determination or once a day.

If the titrant is selected (see "Selecting the dosing function", page 257) ,

the parameters defined for the titrant for preparing/emptying and the tubing parameters will be used (dosing unit only). If the titrant is not selected,

default parameters will be used (see Chapter 32.2.2, page 493 and

Chapter 32.3.2, page 494) .

The following two figures show the live display of an exchange unit and a dosing unit:

262 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 28 Manual control

Figure 11 Live display "Preparing the dosing unit"

Figure 12 Live display "Preparing the exchange unit"

28.3.4 Emptying

Manual control: Dosing ▶ Empty

NOTICE

The function Empty is possible only with dosing units.

The cylinders and the tubings of the dosing unit are emptied with this function.

If the titrant is selected

(see "Selecting the dosing function", page 257) ,

the parameters defined for the titrant for preparing/emptying and the tubing parameters will be used. If the titrant is not selected, default parame-

ters will be used (see Chapter 32.3.2, page 494 and

Chapter 32.2.2, page

493

).

900 Touch Control ■■■■■■■■ 263

28.3 Dosing ■■■■■■■■■■■■■■■■■■■■■■

28.3.5 Filling

Manual control: Dosing ▶ Fill

You can use the function [Fill] to fill the dosing cylinder of the exchange unit/dosing unit manually. The filling rate cannot be configured; the maximum filling rate will be used.

28.3.6 Replacing reagent

NOTICE

You can easily change the reagent in a dosing unit without no contact with the chemicals using the functions Prepare and Empty .

Proceed as follows:

1 Execute the function [Empty] .

2 Attach the dosing unit to the bottle with a suitable solvent.

3 Execute the function [Prepare] .

The last traces of the reagent in the cylinder and tubings are removed with the solvent.

4 Execute the function [Empty] again.

5 Attach the dosing unit to the bottle with the new reagent.

6 Execute the function [Prepare] .

The dosing unit can now be used.

264 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

28.4 Stirring

Manual control ▶ Stir

28 Manual control

Stirring rate

The MSB connector and type of stirrer is indicated for each connected stirrer.

You can control a connected stirrer manually with the function [Stir] .

Proceed as follows:

1 Setting the stirring rate

Tap on the [–] or [+] button.

Each time one of the buttons is tapped, the stirring rate is reduced or increased by one level.

2 Switching on the stirrer

Tap on [Start] .

The stirrer is started.

3 Switching off the stirrer

Tap on [Stop] .

The stirrer is stopped.

Parameter description

Setting the stirring rate. It can be set in steps of –15 to +15. The default setting 8 corresponds to 1000 rpm. The formula for calculating the rotational speed is specified in

chapter 32.4, page 494

.

900 Touch Control ■■■■■■■■ 265

28.5 Conditioning a titration cell ■■■■■■■■■■■■■■■■■■■■■■

The algebraic sign of the stirring rate changes the direction in which the stirring is done. When the stirrer is viewed from above, this means:

"+": counterclockwise rotation

"–": clockwise rotation

Input range

Default value

–15 to 15

8

NOTICE

If a stirrer is used in the determination run, it can still be manually controlled at the same time. For example, you can use the manual control to reduce the stirring rate of a stirrer which has been started in the determination run.

28.5 Conditioning a titration cell

Manual control ▶ Conditioning

In manual control, you have the option of permanently conditioning one titration cell for Karl Fischer titrations while carrying out a Karl Fischer titration in a second titration cell.

Proceed as follows:

1 Create a method

Create a Karl Fischer method.

Save the method.

2 Load the method

Select the memory location where the desired method is saved.

Tap on the input field File name .

The method table with the saved methods is opened.

Select the desired method.

3 Start the conditioning

Tap on [Start] .

The working medium is titrated to the endpoint. The status is kept stable.

266 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 28 Manual control

28.5.1 Parameter description

Memory

Memory location the method is loaded from. All memory locations are selectable, even if they are currently not accessible.

Selection

Default value

Internal memory | External memory 1 | External memory 2 | Shared memory

Internal memory

Shared memory

Shared directory in the network.

File name

Selection of the method.

Selection Selection of stored methods

28.6 Remote

Manual control ▶ Remote

Inputs

Outputs

900 Touch Control

With the function [Remote] , you can define manual output signals to the remote interface of a connected Remote Box. It is not possible to define input signals; the current status of the input lines is, however, displayed.

Current status of the input lines.

Selection of the signal out of the templates or entering the required bit pattern. Templates are defined under System ▶ Templates ▶ Output lines .

■■■■■■■■ 267

28.7 Sample Processor ■■■■■■■■■■■■■■■■■■■■■■

Entering the bit pattern:

0 = line inactive

1 = line active

* = retain line status p = set pulse

The output lines are always numbered from right to left, i.e. with the signal 00000000000001 line 0 is set. With a pulse, the length is set to 200 ms. If you wish to set pulses with other lengths, you have to define a corresponding template.

Entry

Default value

Selection

Bit pattern containing exactly 14 characters or a max. of 24 characters for the name of the template

00000000000001

Selection of the templates defined

[Set]

Set a defined output signal.

28.7 Sample Processor

Manual control ▶ Sample Processor

268 ■■■■■■■■

With the function [Sample Processor] , you can control a connected

Sample Processor manually. The following functions are available:

Rotating a sample rack

Moving the lift

Defining specific lift positions (work position, rinse position, etc.)

Switching pumps on and off

Initializing the rack

Defining external positions outside of the sample rack (only with Swing

Head)

900 Touch Control

[

[

■■■■■■■■■■■■■■■■■■■■■■ 28 Manual control

Tower for the manual control.

Selection Tower 1 | Tower 2

Tower 1

View from the front, the right tower.

Tower 2

View from the front, the left tower. This selection is only available for

Sample Processors with two towers.

Lift position

Current lift position.

Rack position

Current rack position. If the rack is not set to a valid position, e.g. after a rack reset, - is displayed to indicate this.

NOTICE

Note that the rack has to be set to a valid position for all lift movements! Otherwise a corresponding error message will be displayed.

Move to a valid rack position with [Previous] or [Next] . Alternatively, you can carry out a MOVE command.

[Home position]

[Work position]

⇩ ]

]

[Previous]

[Next]

Move the lift to the home position (corresponds to lift position 0 mm).

Move the lift to the work position of the current rack position or the external position.

The work position can directly be defined here

(see "Assigning lift positions", page 271)

or in the device manager

(see "Lift positions Tower 1/2", page 104)

.

Continuously move the lift upwards as long as the button is pressed down. The lift rate used is defined in the dialog Sample Processor / Lift .

Continuously move the lift downwards as long as the button is pressed down. The lift rate used is defined in the dialog Sample Processor / Lift .

Move to the previous rack position.

Move to the next rack position.

900 Touch Control ■■■■■■■■ 269

28.7 Sample Processor

[Reset rack]

■■■■■■■■■■■■■■■■■■■■■■

Initialize the rack. This function is identical with the command RACK . The following actions are carried out:

The lift is moved upwards.

The sample rack is rotated to the starting position.

The rack code of the rack attached is being read out and the corresponding rack data is transferred to the Sample Processor.

The sample variable is being reset to the value 1.

[Pump]

Switch on and off the pumps mounted or connected to the tower.

[Lift]

Move the lift to any or a specific lift position and assigning lift positions

(see chapter 28.7.1, page 270)

.

[Rack position]

Move the rack to any sample position or to defined positions which are

defined as special beakers (see chapter 28.7.2, page 272)

.

[External position]

This button is only enabled when a Swing Head is mounted to the Sample

Processor.

Swing the robotic arm to an external position outside the sample rack and

assign the swing angle and the lift position (see chapter 28.7.3, page

274) .

28.7.1 Moving the lift

Manual control: Sample Processor ▶ Lift

270 ■■■■■■■■

In the dialog Sample Processor / Lift , you can move the lift to any lift position. You can also define frequently used positions (work position, rinse position, etc.) as specific lift positions and then to move to them with

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 28 Manual control

ease (see "Assigning lift positions", page 271)

. This definition is possible in this dialog for the attached rack, and in the device manager for all of the racks in the list.

Current lift pos.

Current lift position.

Move to lift pos.

Selecting a predefined lift position or enter any lift position.

Input range

Selection

Default value

0 to 'maximum stroke path' mm

A lift position of 0 mm corresponds to the "home position", i.e. the lift is located at the upper stop position. The maximum stroke path is defined in the properties of the tower

(see "Maximum stroke path", page 94)

. If a higher value is entered, an error message will be displayed.

Work position | Shift position | Rinse position |

Special position

Work position

Lift rate

Rate at which the lift is moved in the manual control.

Input range

Default value

5 to 25 mm/s

25 mm/s

[Assign lift pos.]

Assign the current lift position to a preset lift position.

Assigning lift positions

Specific lift positions are stored separately for every sample rack used as well as for tower 1 and tower 2. The following specific lift positions are available:

General rack positions

For general rack positions you can define one work position, one shift position and one rinse position as well as one special position.

Special beaker positions

A specific work position can be defined for each special beaker. The definitions of the general rack positions are used for the shift position, rinse position and special position of the affected tower.

900 Touch Control ■■■■■■■■ 271

28.7 Sample Processor ■■■■■■■■■■■■■■■■■■■■■■

External positions (only with robotic arm and mounted Swing Head,

see Chapter 28.7.3, page 274 )

For each of the four possible external positions, a specific work position can be defined. Shift position and rinse position can only be defined for all four external positions commonly. A special position is not possible.

Proceed as follows to assign the current lift position to a specific lift position:

1 Move to lift position

Enter the desired lift position in mm and tap on [Start] .

The lift moves to the desired lift position.

2 Assign lift position

Tap on [Assign lift pos.] .

The Lift / Assign lift position dialog is displayed:

Select the desired specific lift position and tap on [Assign] .

The current lift position is assigned to the specific lift position.

28.7.2 Moving to a rack position

Manual control: Sample Processor ▶ Rack position

272 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 28 Manual control

Current rack pos.

Rack position

Shift rate

Shift direction

In the dialog Sample Processor / Rack position , you can move to any position on the attached rack.

Current rack position. If the rack is not set to a valid position, e.g. after a rack reset, - is displayed to indicate this.

Desired rack position.

Input range

Default value

Selection

1 to Number of positions on the rack attached.

1

Special beaker 1…16

Special beaker 1…16

The selection depends on how many rack positions are defined as spe-

cial beakers (see "Editing special beakers", page 105) .

Rate at which the sample rack is moved.

Input range

Default value

3 to 20 °/s

20 °/s

Direction in which the rack is moved.

Selection

Default value auto | + | auto auto

A shift direction, with which the shorter way has to be passed, is automatically selected.

900 Touch Control ■■■■■■■■ 273

28.7 Sample Processor ■■■■■■■■■■■■■■■■■■■■■■

+

Counterclockwise rotation.

-

Clockwise rotation.

Swing rate

Rate at which the robotic arm is swung in the manual control.

Input range

Default value

10 to 55 °/s

55 °/s

28.7.3 External positions

Manual control: Sample Processor ▶ External position

274

Lift position

Angle

■■■■■■■■

NOTICE

This dialog is accessible only if you have a Swing Head with robotic arm attached to the Sample Processor.

In the dialog Sample Processor / External position , you can swivel the robotic arm to an external position outside of the sample rack. As with the rack positions, here you can also define up to four frequently used external positions as specific positions. You can assign specific lift positions to these positions, analogous to the rack positions.

External positions are required if you have, for example, a measuring vessel or a titration cell mounted in addition to the rack.

Current lift position.

Current angle position of the robotic arm.

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

[ ⇧ ]

28 Manual control

Only enabled, when the robotic arm has been moved to a specific external position with [Start] .

Continuously move the lift upwards as long as the button is pressed down. The lift rate used is defined in the dialog Sample Processor / Lift .

[ ⇩ ]

Only enabled, when the robotic arm has been moved to a specific external position with [Start] .

Continuously move the lift downwards as long as the button is pressed down. The lift rate used is defined in the dialog Sample Processor / Lift .

[ ⇦ ]

Swing the robotic arm in clockwise direction as long as the button is pressed down.

[ ⇨ ]

Swing the robotic arm in counterclockwise direction as long as the button is pressed down.

Swing rate

Rate at which the robotic arm is swung in the manual control.

Input range

Default value

10 to 55 °/s

55 °/s

Move to ext. pos.

Selecting a predefined position or entering any swing angle.

Input range

Selection

(Offset) to (Offset + max. swing range) °

The offset is made up of a design-dependent angle

(approx. 8...9°) together with the robotic arm offset from the robotic arm properties. The maximum swing range is also defined under the robotic arm

properties (see "Properties – Robotic arm", page

97) .

External position 1…4

[Assign Ext. pos.]

Assign the current angle position of the robotic arm to an external position as swing angle.

[Assign lift pos.]

Assign the current lift position to a preset lift position.

900 Touch Control ■■■■■■■■ 275

28.7 Sample Processor ■■■■■■■■■■■■■■■■■■■■■■

Assigning swing angles and lift positions

For each of the four possible external positions, a specific work position can be defined. Shift position and rinse position can only be defined for all four external positions commonly. A special position is not possible.

Proceed as follows to assign a swing angle to an external position and to assign the current lift position to a specific lift position:

1 Move to the external position

Swing the robotic arm to the desired position with the arrow keys

[ ⇦ ] or [ ⇨ ] .

2 Assign the external position

Tap on [Assign Ext. pos.] .

The External position / Assign swing angle dialog is displayed:

276 ■■■■■■■■

Select the desired external position and tap on [Assign] .

The current angle position of the robotic arm is assigned to the external position.

3 Move to the external position

Tap on [ ] .

The next higher-level dialog is displayed.

Select Move to ext. pos.

= External position X (X = 1…4) and tap on [Start] .

The robotic arm swings to the desired position.

4 Move to lift position

Move the lift to the desired height with the arrow keys [ ⇧ ] or [ ⇩ ] .

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

5 Assign lift position

Tap on [Assign lift pos.] .

The Lift / Assign lift position dialog is displayed:

28 Manual control

Select the desired specific lift position.

Tap on [Assign] .

The current lift position is assigned to the specific lift position.

900 Touch Control ■■■■■■■■ 277

29.1 Dynamic equivalence point titrations (DET) ■■■■■■■■■■■■■■■■■■■■■■

29 Parameters

NOTICE

For most of the numerical parameters, the result of a calculation can also be utilized in place of a number. A description of this can be found in the appendix

(see chapter 32.6, page 496) .

29.1 Dynamic equivalence point titrations (DET)

29.1.1 Start conditions

The parameters that are carried out before the start of titration are defined under [Start conditions] . The start conditions are processed in the following sequence:

1.

Start volume

2.

Start measured value

3.

Start slope

4.

Pause

Start volume

Volume that is dosed prior to the start of the titration.

Input range

Default value

0.00000 to 9999.99 mL

0.00000 mL

Dosing rate

Rate at which the start volume is dosed.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum dosing rate depends on the cylinder volume

(see chapter 32, page 492) .

Pause

Waiting time, e.g. for the electrode to settle down after the start or a reaction time after the dosing of a start volume. The time entered here is only running when all start conditions have been fulfilled.

Input range

Default value

0 to 999999 s

0 s

278 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Start meas. value

29 Parameters

More start conditions

Before starting the titration, it will be dosed until this measured value is reached. If the start measured value is reached by the dosing of a start volume, then the titration starts directly.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control instrument used.

Measuring mode pH:

Input range

Selection

Default value

–20.000 to 20.000 off off

Measuring mode U, Ipol:

Input range –2000.0 to 2000.0 mV

Selection

Default value off off

Measuring mode Upol:

Input range

Selection

Default value

–200.0 to 200.0 µA off off

Start slope

Before starting the titration, it will be dosed until this slope (measured value per volume) is reached. If the start slope is reached by the dosing of a start volume, then the titration starts directly.

Measuring mode pH:

Input range

Selection

Default value

00.000 to 9,999 pH/mL off off

Measuring mode U, Ipol:

Input range 0 to 999 mV/mL

Selection

Default value off off

Measuring mode Upol:

Input range

Selection

Default value

0 to 99 µA/mL off off

900 Touch Control ■■■■■■■■ 279

29.1 Dynamic equivalence point titrations (DET)

Dosing rate

■■■■■■■■■■■■■■■■■■■■■■

Rate to be dosed with until the start measured value or the start slope is reached.

Input range

Default value

Selection

0.01 to 166.00 mL/min

5.00 mL/min maximum maximum

The maximum dosing rate depends on the cylinder volume

(see chapter 32, page 492) .

Initial measured value

The [Initial meas. value] is measured before carrying out the start conditions . This value is used for the titration direction determination , which is necessary for the start measured value and stop measured value parameters.

Signal drift

Maximum permissible drift for the measured value acceptance, i.e. maximum change of the measured value per minute. For this value, a suitable maximum waiting time is automatically calculated except you already have modified this waiting time.

Measuring mode pH, U and Ipol:

Input range 0.1 to 999.0 mV/min

Selection

Default value off off off

Measured value acceptance will take place after the maximum waiting time has elapsed.

Measuring mode Upol:

Input range 0.01 to 99.90 µA/min

Selection

Default value off off off

Measured value acceptance will take place after the maximum waiting time has elapsed.

Min. waiting time

The measured value is not accepted until the minimum waiting time has elapsed, even if the signal drift has already been reached. The minimum waiting time is only important for drift-controlled measurements.

280 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Input range

Default value

0 to 999999 s

0 s

Max. waiting time

If the signal drift has been switched off or has not yet been reached, then the measured value will be accepted when the maximum waiting time has elapsed. As default value, a suitable waiting time for the signal drift is automatically calculated.

Input range

Default value

0 to 999999 s

1 s

29.1.2 Titration parameters

Under [Titration parameters] , the parameters influencing the run of the entire titration are defined.

Titration rate

Three predefined sets of parameters can be selected for the titration rate.

Selection

Default value slow | optimal optimal

| fast | user slow

For titrations in which the finest details are also to be visible. This could however also lead to an increase in noise, which could result in unwanted equivalence points.

optimal

For all standard titrations. The parameters have been optimized for the most frequent applications.

fast

For fast and less critical titrations.

user

The individual titration parameters can be modified.

NOTICE

Select optimal as titration rate when you are developing a new titration method. This parameter is suitable for virtually all titrations and only needs adaptation in special cases.

The settings of the individual titration rates are listed in

table 9, page 284 .

Temperature

Temperature entered manually. The temperature is being continuously measured when a temperature sensor is connected and when Temp.

900 Touch Control ■■■■■■■■ 281

29.1 Dynamic equivalence point titrations (DET) ■■■■■■■■■■■■■■■■■■■■■■ measurement = automatic or continuous is defined (see sensor dialog of the command). This value is used for temperature correction in pH measurements.

Input range

Default value

–20.0 to 150.0 °C

25.0 °C

User-defined parameters

These parameters are only accessible when Titration rate = user .

Meas. point density

A small value means small volume increments, i.e. a high measuring point density. The curve then shows all the finest details which also include noise; this could cause unwanted equivalence points to be found. A larger value, i.e. a smaller measuring point density, permits quicker titrations. If you are using a dosing device with a small cylinder volume then a smaller measuring point density value may be beneficial. However, you should also set a smaller signal drift and a higher EP criterion at the same time.

Input range

Default value

0 to 9

4

Min. increment

This smallest permitted volume increment is added at the start of the titration and with steep curves in the region of the equivalence point. Very small values should only be used if a low titrant consumption is expected; otherwise unwanted equivalence points could be evaluated.

Input range

Default value

0.05 to 999.90 µL

10.00 µL

Max. increment

A maximum volume increment should be selected in the following cases:

■ when titration consumption is very low up until the equivalence point is reached.

when a start volume is dosed up until shortly before the equivalence point is reached.

when the change of direction in the jumping range is very abrupt, because otherwise it could easily happen that an excessively large volume will be dosed in the region of the equivalence point.

The value should not be less than 1/100 cylinder volume.

Input range

Selection

Default value

0.1 to 9999.9 µL off off

282 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

NOTICE

It is not advisable to select similar volumes for the minimum and the maximum increment. Monotonic equivalence point titration (MET) is appropriate for these applications.

Dosing rate

Signal drift

Rate at which the volume increments are dosed.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum dosing rate depends on the cylinder volume (see chapter 32, page 492) .

Maximum permissible drift for the measured value acceptance, i.e. maximum change of the measured value per minute. This type of titration is often referred to as equilibrium titration. For this value, a suitable maximum waiting time is automatically calculated except you already have modified this waiting time.

NOTICE

A constant measured value is often only reached after a certain time, as mixing and the reaction itself require a certain time. The response time of an electrode can also increase with time, i.e. reaching a constant measured value takes longer and longer. Drift-controlled measured value acceptance is particularly advisable in such cases, as the measured values are only accepted when equilibrium has almost been reached.

Measuring mode pH, U and Ipol:

Input range

Default value

Selection

0.1 to 999.0 mV/min

50.0 mV/min off off

Measured value acceptance will take place after the maximum waiting time has elapsed. This can be useful when the titration reaction proceeds slowly or the electrode is slow to respond.

900 Touch Control ■■■■■■■■ 283

29.1 Dynamic equivalence point titrations (DET) ■■■■■■■■■■■■■■■■■■■■■■

Measuring mode Upol:

Input range 0.01 to 99.90 µA/min

Default value

Selection

50.00 µA/min off off

Measured value acceptance will take place after the maximum waiting time has elapsed. This can be useful when the titration reaction proceeds slowly or the electrode is slow to respond.

Min. waiting time

The measured value is not accepted until the minimum waiting time has elapsed, even if the signal drift has already been reached. The minimum waiting time is only important for drift-controlled measurements.

Input range

Default value

0 to 999999 s

0 s

Max. waiting time

If the signal drift has been switched off or has not yet been reached, then the measured value will be accepted when the maximum waiting time has elapsed. As default value, a suitable waiting time for the signal drift is automatically calculated.

Input range

Default value

0 to 999999 s

26 s

Table 9 Default values of the predefined titration rates for DET

Titration rate slow optimal

Meas. point density

Min. increment

Max. increment

Dosing rate

Signal drift

– pH, U and Ipol

– Upol

Min. waiting time

Max. waiting time

2

10.00 µL off maximum

20.0 mV/min

20.0 µA/min

0 s

38 s

4

10.00 µL off maximum

50.0 mV/min

50.0 µA/min

0 s

26 s fast

6

30.00 µL off maximum

80.0 mV/min

80.0 µA/min

0 s

21 s

284 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

29.1.3 Stop conditions

The conditions for canceling the titration are defined under [Stop conditions] .

Stop volume

The titration is stopped when the specified volume has been dosed since the start of the titration. This volume should be adjusted to the size of the titration vessel in order to prevent the contents from running over.

Input range

Default value

Selection

0.00000 to 9999.99 mL

100,000 mL off

Stop meas. value

The titration is canceled when the specified measured value has been reached since the start of the titration.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control instrument used.

Measuring mode pH:

Input range

Selection

Default value

–20.000 to 20.000 off off

Measuring mode U, Ipol:

Input range

Selection

Default value

–2000.0 to 2000.0 mV off off

Measuring mode Upol:

Input range –200.0 to 200.0 µA

Selection

Default value off off

Stop EP

The titration is stopped when the specified number of equivalence points has been found.

Input range

Default value

Selection

1 to 9

9 off

900 Touch Control ■■■■■■■■ 285

29.1 Dynamic equivalence point titrations (DET) ■■■■■■■■■■■■■■■■■■■■■■

Volume after EP

This volume will be added when the number of equivalence points defined under Stop EP has been found. The curve shape after the equivalence point can also be seen this way.

Input range

Selection

Default value

0.01000 to 9999.99 mL off off

Stop time

The titration is stopped when the specified time has elapsed since the start of the titration.

Input range

Selection

Default value

0 to 999999 s off off

Filling rate

Rate at which the dosing cylinder is filled.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum filling rate depends on the cylinder volume (see chapter

32, page 492) .

29.1.4 Potentiometric evaluation

The parameters for the evaluation of the titration curve are defined under

[Potentiometr. evaluation] .

EP criterion

The Equivalence point Recognition Criterion (ERC) found is compared with this value. Equivalence points whose ERC is less than the value defined here will be ignored.

Input range

Default value

0 to 200

5

EP recognition

This parameter allows you to filter out only the equivalence points that are being sought.

Selection

Default value all | greatest | last | ascending | descending | off all

286 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters all

All equivalence points will be recognized.

greatest

Only the equivalence point with the greatest ERC value, i.e. the steepest jump, will be recognized.

last

Only the last equivalence point will be recognized.

ascending

Only equivalence points with a positive slope of the titration curve will be recognized.

descending

Only equivalence points with a negative slope of the titration curve will be recognized.

off

No evaluation takes place.

Set windows

This parameter allows to recognize equivalence points only in a particular range (so-called window) of the curve. Equivalence points outside this window will not be recognized. A maximum of 9 windows can be defined. Setting windows is advisable when interference and unnecessary equivalence points are to be suppressed.

Selection

Default value

Measured value | Volume | off off

Measured value

The windows are defined on the measured value axis.

Volume

The windows are defined on the volume axis.

off

The entire titration curve is being evaluated.

Setting windows

The list of defined windows is opened with [Set window] .

When the list is opened for the first time, a window over the entire measured value range or volume range is already defined. The windows must not overlap, they only may lie next to one another. For each window, own criteria can be defined for the equivalence point recognition.

900 Touch Control ■■■■■■■■ 287

29.1 Dynamic equivalence point titrations (DET) ■■■■■■■■■■■■■■■■■■■■■■

NOTICE

When you define a result variable instead of a numerical value as limit

value (see chapter 32.6, page 496) , then just one window can be

defined.

[New]

[Delete]

[Edit]

Define a new window. This is only possible when not the entire range is covered yet.

Delete the selected window.

Edit the settings of the selected window.

Lower limit

Measured value or volume for the lower limit.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control instrument used.

Measured value (measuring mode pH):

Input range

Default value

–20.000 to 20.000

–20.000

Measured value (measuring mode U, Ipol):

Input range –2000.0 to 2000.0 mV

Default value –2000.0 mV

Measured value (measuring mode Upol):

Input range

Default value

–200.00 to 200.00 µA

–200.00 µA

Volume:

Input range

Default value

0.00000 to 9999.99 mL

0.00000 mL

Upper limit

Measured value or volume for the upper limit.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges

288 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters can be found in the "Technical specifications" chapter for the control instrument used.

Measured value (measuring mode pH):

Input range –20.000 to 20.000

Default value 20.000

Measured value (measuring mode U, Ipol):

Input range

Default value

–2000.0 to 2000.0 mV

2000.0 mV

Measured value (measuring mode Upol):

Input range –200.00 to 200.00 µA

Default value 200.00 µA

Volume:

Input range

Default value

0.00000 to 9999.99 mL

9999.99 mL

EP criterion

The Equivalence point Recognition Criterion (ERC) found is compared with this value. Equivalence points whose ERC is less than the value defined here will be ignored.

Input range

Default value

0 to 200

5

EP recognition

This parameter allows you to filter out only the equivalence points that are being sought.

Selection

Default value first first

| greatest | last | ascending | descending first

Only the first equivalence point will be recognized.

greatest

Only the equivalence point with the greatest ERC value, i.e. the steepest jump, will be recognized.

last

Only the last equivalence point will be recognized.

ascending

Only equivalence points with a positive slope of the titration curve will be recognized.

descending

Only equivalence points with a negative slope of the titration curve will be recognized.

900 Touch Control ■■■■■■■■ 289

29.1 Dynamic equivalence point titrations (DET) ■■■■■■■■■■■■■■■■■■■■■■

Only one equivalence point will be recognized in a window. The numbering of the equivalence points (EP) is defined by the numbering of the windows (e.g. EP2 in window 2), so that even if EPs are missing, the calculations will still be carried out with the correctly assigned EP volumes.

U/mV U/mV

EP window 2

EP window 1 EP1

EP2

EP window 1 upper limit 1

+

EP1 lower limit 1

V/mL V/mL

Figure 13 Equivalence point recognition and equivalence point numbering within windows

In the first example in the above figure, two equivalence points are recognized in two different windows (EP1 in window 1 and EP2 in window 2).

In the second example, two equivalence points are found in one window, but only the first one is recognized. In order to ensure that the user recognizes that more than one equivalence point was found in the set window,

EP1 is marked with a "+" in the result view. In addition, a corresponding message is entered in the message list.

Evaluation and equivalence point criterion with DET

The equivalence points (EP) are localized in a way similar to the Tubbs method [1][2]. The volume value of the equivalence point (V

E

) is shifted from the inflection point (see arrow) towards the smaller circle of curvature for real asymmetric titration curves.

[1] C. F. Tubbs, Anal. Chem.

1954 , 26 , 1670–1671.

[2] E. Bartholomé, E. Biekert, H. Hellmann, H. Ley, M. Weigert, E. Weise,

Ullmanns Encyklopädie der technischen Chemie , Vol. 5, Verlag Chemie,

Weinheim, 1980, p. 659.

290 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ pH

29 Parameters

V

E

V/mL

Figure 14 Tubbs method for determining the equivalence point

The figure shows that the evaluation still requires measured values from the measuring point list even after the equivalence point.

For the recognition of the EPs found, the set EP criterion is compared to the ERC (Equivalence point Recognition Criterion) found. The ERC is the first derivative of the titration curve combined with a mathematical function that is more sensitive for flat jumps than for steeper ones. EPs whose

ERC is smaller than the defined EP criterion will not be recognized. The

ERC is displayed in the results dialog for each discovered and recognized

EP. If you adjust the EP criterion later in order to recognize more or fewer

EPs, then you can reevaluate the determination ( [Recalculate] function in the results dialog).

29.1.5 Control instrument

The control instrument the determination is carried out with is selected under [Control device] . Control instruments are defined under System ▶ Device manager .

NOTICE

This button is only displayed when several control instruments have been configured.

Control device

Selection of the control instrument from the list of devices. Only those devices are displayed which are able to carry out the command. The control instruments do not have to be connected. This allows to transfer methods more easily from one system to another.

Selection Selection of configured control instruments

900 Touch Control ■■■■■■■■ 291

29.1 Dynamic equivalence point titrations (DET) ■■■■■■■■■■■■■■■■■■■■■■

29.1.6 Sensor

The parameters for the sensor are edited under [Sensor] .

Measuring input

Selection of the measuring input the sensor is connected to. The selection is not dependent on whether the control device has one or two measuring interfaces.

Selection

Default value

1 | 2

1

Sensor

Selection of the sensor from the sensor list. The selection depends on the measuring mode. Sensors are defined under System ▶ Sensors . You can also enter a sensor name which is not contained in the sensor list. When a determination is started there is a check whether the sensor is contained in the sensor list.

Selection Selection of configured sensors

I(pol)

The polarization current is the current that is applied to a polarizable electrode during voltametric measurement. This parameter is available only with I(pol) determinations.

The values that are actually adjustable may vary from the values specified here, depending on the hardware. The binding values can be found in the

Technical specifications chapter for the control instrument used.

Input range

Default value

Selection

–125.0 to 125.0 µA (Increment: 2.5)

5.0 µA

–1.0 µA | 1.0 µA

U(pol)

The polarization voltage is the voltage applied to the polarizable electrode during an amperometric measurement. This parameter is available only with U(pol) determinations.

The values that are actually adjustable may vary from the values specified here, depending on the hardware. The binding values can be found in the

Technical specifications chapter for the control instrument used.

Input range

Default value

–1,250 to 1,250 mV (Increment: 25)

400 mV

292 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Electrode check on | off (Default value: off )

For the following electrodes, an electrode check can be carried out:

■ pH electrodes

Metal electrodes

Ion-selective electrodes

A check is made that the electrode is properly connected and that no short-circuit is present. The electrode check is carried out when this command is started. Note that this electrode check has nothing to do with the electrode test (command ELT ).

Temp. measurement

Type of temperature measurement.

Selection

Default value continuous automatic

| automatic | off continuous

A temperature sensor must be connected. The temperature is measured continuously.

automatic

If a temperature sensor is connected then the temperature will be measured continuously. Otherwise, the temperature entered manually will be used (see dialog of the titration and measuring parameters).

off

The temperature will not be measured. The temperature entered manually is used (see dialog of the titration and measuring parameters).

29.1.7 Dosing device

The parameters for the dosing device are edited under [Dosing device] .

Dosing device

Selection of the MSB connector to which the dosing device is connected.

Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4

Titrant

Selection of the titrant from the titrant list. We recommend always selecting the titrant. This ensures that the correct data (titer, concentration, etc.) is always used for the calculation. Titrants are defined under System ▶ Titrants . You can also enter a name which is not contained in the titrant list. When a determination is started there is a check whether the solution is contained in the list.

900 Touch Control ■■■■■■■■ 293

29.1 Dynamic equivalence point titrations (DET) ■■■■■■■■■■■■■■■■■■■■■■

For buret units with integrated data chip, a check is made in the method run to verify whether the correct titrant has been attached and whether the type of dosing drive matches. For buret units without an integrated data chip, the cylinder volume and the type of dosing drive are checked.

For the selected titrant, the validity of the titer, the working life of the titrant and the time interval for the GLP test of the buret unit are checked at the start of the determination.

Selection

Default value

Selection of configured titrants | not defined not defined not defined

No check takes place.

29.1.8 Stirrer

The parameters for the stirrer are edited under [Stirrer] .

Stirrer

Selection of the MSB connector the stirrer is connected to. Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4 | off off

No stirrer will be used.

Stirring rate

Setting the stirring rate. It can be set in steps of –15 to +15. The default setting 8 corresponds to 1000 rpm. The formula for calculating the rota-

tional speed is specified in chapter 32.4, page 494 . The optimum stirring

rate can be tested in the manual control.

The algebraic sign of the stirring rate changes the direction in which the stirring is done. When the stirrer is viewed from above, this means:

"+": counterclockwise rotation

"–": clockwise rotation

Input range

Default value

–15 to 15

8

Switch off automatically on | off (Default value: on )

If this parameter is activated, the stirrer will be switched off automatically at the end of the titration, measurement, etc.

294 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

29.2 Monotonic equivalence point titrations (MET)

29.2.1 Start conditions

The parameters that are carried out before the start of titration are defined under [Start conditions] . The start conditions are processed in the following sequence:

1.

Start volume

2.

Start measured value

3.

Start slope

4.

Pause

Start volume

Volume that is dosed prior to the start of the titration.

Input range

Default value

0.00000 to 9999.99 mL

0.00000 mL

Dosing rate

Rate at which the start volume is dosed.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum dosing rate depends on the cylinder volume (see chapter 32, page 492) .

Pause

Waiting time, e.g. for the electrode to settle down after the start or a reaction time after the dosing of a start volume. The time entered here is only running when all start conditions have been fulfilled.

Input range

Default value

0 to 999999 s

0 s

More start conditions

Start meas. value

Before starting the titration, it will be dosed until this measured value is reached. If the start measured value is reached by the dosing of a start volume, then the titration starts directly.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control instrument used.

900 Touch Control ■■■■■■■■ 295

29.2 Monotonic equivalence point titrations (MET)

Measuring mode pH:

Input range

Selection

Default value

–20.000 to 20.000 off off

Measuring mode U, Ipol:

Input range

Selection

Default value

–2000.0 to 2000.0 mV off off

Measuring mode Upol:

Input range –200.0 to 200.0 µA

Selection

Default value off off

■■■■■■■■■■■■■■■■■■■■■■

Start slope

Before starting the titration, it will be dosed until this slope (measured value per volume) is reached. If the start slope is reached by the dosing of a start volume, then the titration starts directly.

Measuring mode pH:

Input range

Selection

Default value

00.000 to 9,999 pH/mL off off

Measuring mode U, Ipol:

Input range

Selection

Default value

0 to 999 mV/mL off off

Measuring mode Upol:

Input range 0 to 99 µA/mL

Selection

Default value off off

Dosing rate

Rate to be dosed with until the start measured value or the start slope is reached.

Input range

Default value

Selection

0.01 to 166.00 mL/min

5.00 mL/min maximum maximum

The maximum dosing rate depends on the cylinder volume

(see chapter 32, page 492) .

296 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Initial measured value

The [Initial meas. value] is measured before carrying out the start conditions . This value is used for the titration direction determination , which is necessary for the start measured value and stop measured value parameters.

Signal drift

Maximum permissible drift for the measured value acceptance, i.e. maximum change of the measured value per minute. For this value, a suitable maximum waiting time is automatically calculated except you already have modified this waiting time.

Measuring mode pH, U and Ipol:

Input range

Selection

Default value

0.1 to 999.0 mV/min off off off

Measured value acceptance will take place after the maximum waiting time has elapsed.

Measuring mode Upol:

Input range

Selection

Default value

0.01 to 99.90 µA/min off off off

Measured value acceptance will take place after the maximum waiting time has elapsed.

Min. waiting time

The measured value is not accepted until the minimum waiting time has elapsed, even if the signal drift has already been reached. The minimum waiting time is only important for drift-controlled measurements.

Input range

Default value

0 to 999999 s

0 s

Max. waiting time

If the signal drift has been switched off or has not yet been reached, then the measured value will be accepted when the maximum waiting time has elapsed. As default value, a suitable waiting time for the signal drift is automatically calculated.

Input range

Default value

0 to 999999 s

1 s

900 Touch Control ■■■■■■■■ 297

29.2 Monotonic equivalence point titrations (MET) ■■■■■■■■■■■■■■■■■■■■■■

29.2.2 Titration parameters

Under [Titration parameters] , the parameters influencing the run of the entire titration are defined.

Titration rate

Three predefined sets of parameters can be selected for the titration rate.

Selection

Default value slow | optimal optimal

| fast | user slow

For titrations in which the finest details are also to be visible. This could however also lead to an increase in noise, which could result in unwanted equivalence points.

optimal

For all standard titrations. The parameters have been optimized for the most frequent applications.

fast

For fast and less critical titrations.

user

The individual titration parameters can be modified.

NOTICE

Select optimal as titration rate when you are developing a new titration method. This parameter is suitable for a lot of titrations and only needs adaptation in special cases.

The settings of the individual titration rates are listed in table 10, page

301 .

Temperature

Temperature entered manually. The temperature is being continuously measured when a temperature sensor is connected and when Temp.

measurement = automatic or continuous is defined (see sensor dialog of the command). This value is used for temperature correction in pH measurements.

Input range

Default value

–20.0 to 150.0 °C

25.0 °C

User-defined parameters

These parameters are only accessible when Titration rate = user .

298 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Volume increment

29 Parameters

The volume that is dosed at each dosing step. High accuracy requires using the correct volume increment.

A good guideline for the volume increment is 0.1 mL. A total of 20 dosing steps is recommended until the equivalence point is reached.

With greater consumption, the titration time can be optimized using the start conditions. Smaller volume increments (smaller than 0.1 mL) are required to determine blank values or to ensure accuracy with highly asymmetric curves. Larger volume increments are advisable for flatter curves.

NOTICE

The accuracy of the evaluation cannot be increased by using smaller increments as the measured value changes between two measuring points are then of the same order of magnitude as the noise.

Input range

Default value

0.00005 to 999.900 mL

0.10000 mL

Dosing rate

Signal drift

Rate at which the volume increments are dosed.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum dosing rate depends on the cylinder volume (see chapter 32, page 492) .

Maximum permissible drift for the measured value acceptance, i.e. maximum change of the measured value per minute. This type of titration is often referred to as equilibrium titration. For this value, a suitable maximum waiting time is automatically calculated except you already have modified this waiting time.

900 Touch Control ■■■■■■■■ 299

29.2 Monotonic equivalence point titrations (MET) ■■■■■■■■■■■■■■■■■■■■■■

NOTICE

A constant measured value is often only reached after a certain time, as mixing and the reaction itself require a certain time. The response time of an electrode can also increase with time, i.e. reaching a constant measured value takes longer and longer. Drift-controlled measured value acceptance is particularly advisable in such cases, as the measured values are only accepted when equilibrium has almost been reached.

Measuring mode pH, U and Ipol:

Input range

Default value

Selection

0.1 to 999.0 mV/min

50.0 mV/min off off

Measured value acceptance will take place after the maximum waiting time has elapsed. This can be useful when the titration reaction proceeds slowly or the electrode is slow to respond.

Measuring mode Upol:

Input range

Default value

Selection

0.01 to 99.90 µA/min

50.00 µA/min off off

Measured value acceptance will take place after the maximum waiting time has elapsed. This can be useful when the titration reaction proceeds slowly or the electrode is slow to respond.

Min. waiting time

The measured value is not accepted until the minimum waiting time has elapsed, even if the signal drift has already been reached. The minimum waiting time is only important for drift-controlled measurements.

Input range

Default value

0 to 999999 s

0 s

Max. waiting time

If the signal drift has been switched off or has not yet been reached, then the measured value will be accepted when the maximum waiting time has elapsed. As default value, a suitable waiting time for the signal drift is automatically calculated.

Input range

Default value

0 to 999999 s

26 s

300 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Table 10 Default values of the predefined titration rates for MET

Titration rate slow optimal

Volume increment

Dosing rate

Signal drift

– pH, U and Ipol

– Upol

Min. waiting time

Max. waiting time

0.05000 mL maximum

20.0 mV/min

20.0 µA/min

0 s

38 s

0.10000 mL maximum

50.0 mV/min

50.0 µA/min

0 s

26 s fast

0.20000 mL maximum

80.0 mV/min

80.0 µA/min

0 s

21 s

29.2.3 Stop conditions

The conditions for canceling the titration are defined under [Stop conditions] .

Stop volume

The titration is stopped when the specified volume has been dosed since the start of the titration. This volume should be adjusted to the size of the titration vessel in order to prevent the contents from running over.

Input range

Default value

Selection

0.00000 to 9999.99 mL

100,000 mL off

Stop meas. value

The titration is canceled when the specified measured value has been reached since the start of the titration.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control instrument used.

Measuring mode pH:

Input range

Selection

Default value

–20.000 to 20.000 off off

Measuring mode U, Ipol:

Input range

Selection

Default value

–2000.0 to 2000.0 mV off off

900 Touch Control ■■■■■■■■ 301

29.2 Monotonic equivalence point titrations (MET) ■■■■■■■■■■■■■■■■■■■■■■

Measuring mode Upol:

Input range –200.0 to 200.0 µA

Selection

Default value off off

Stop EP

The titration is stopped when the specified number of equivalence points has been found.

Input range

Default value

Selection

1 to 9

9 off

Volume after EP

This volume will be added when the number of equivalence points defined under Stop EP has been found. The curve shape after the equivalence point can also be seen this way.

Input range

Selection

Default value

0.01000 to 9999.99 mL off off

Stop time

The titration is stopped when the specified time has elapsed since the start of the titration.

Input range

Selection

Default value

0 to 999999 s off off

Filling rate

Rate at which the dosing cylinder is filled.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum filling rate depends on the cylinder volume (see chapter

32, page 492) .

29.2.4 Potentiometric evaluation

The parameters for the evaluation of the titration curve are defined under

[Potentiometr. evaluation] .

302 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

EP criterion

29 Parameters

The Equivalence point Recognition Criterion (ERC) found is compared with this value. Equivalence points whose ERC is less than the value defined here will be ignored.

Measuring mode pH:

Input range

Default value

0.10 to 9.99

0.50

Measuring mode U, Ipol:

Input range 1 to 999 mV

Default value 30 mV

Measuring mode Upol:

Input range

Default value

0.1 to 99.9 µA

2.0 µA

EP recognition

This parameter allows you to filter out only the equivalence points that are being sought.

Selection

Default value all | greatest | last | ascending | descending | off all all

All equivalence points will be recognized.

greatest

Only the equivalence point with the greatest ERC value, i.e. the steepest jump, will be recognized.

last

Only the last equivalence point will be recognized.

ascending

Only equivalence points with a positive slope of the titration curve will be recognized.

descending

Only equivalence points with a negative slope of the titration curve will be recognized.

off

No evaluation takes place.

Set windows

This parameter allows to recognize equivalence points only in a particular range (so-called window) of the curve. Equivalence points outside this window will not be recognized. A maximum of 9 windows can be

900 Touch Control ■■■■■■■■ 303

29.2 Monotonic equivalence point titrations (MET) ■■■■■■■■■■■■■■■■■■■■■■ defined. Setting windows is advisable when interference and unnecessary equivalence points are to be suppressed.

Selection

Default value

Measured value off

| Volume | off

Measured value

The windows are defined on the measured value axis.

Volume

The windows are defined on the volume axis.

off

The entire titration curve is being evaluated.

Setting windows

The list of defined windows is opened with [Set window] .

When the list is opened for the first time, a window over the entire measured value range or volume range is already defined. The windows must not overlap, they only may lie next to one another. For each window, own criteria can be defined for the equivalence point recognition.

NOTICE

When you define a result variable instead of a numerical value as limit

value (see chapter 32.6, page 496) , then just one window can be

defined.

[New]

[Delete]

[Edit]

Define a new window. This is only possible when not the entire range is covered yet.

Delete the selected window.

Edit the settings of the selected window.

Lower limit

Measured value or volume for the lower limit.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control instrument used.

Measured value (measuring mode pH):

Input range

Default value

–20.000 to 20.000

–20.000

304 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Upper limit

EP criterion

29 Parameters

Measured value (measuring mode U, Ipol):

Input range –2000.0 to 2000.0 mV

Default value –2000.0 mV

Measured value (measuring mode Upol):

Input range

Default value

–200.00 to 200.00 µA

–200.00 µA

Volume:

Input range

Default value

0.00000 to 9999.99 mL

0.00000 mL

Measured value or volume for the upper limit.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control instrument used.

Measured value (measuring mode pH):

Input range –20.000 to 20.000

Default value 20.000

Measured value (measuring mode U, Ipol):

Input range

Default value

–2000.0 to 2000.0 mV

2000.0 mV

Measured value (measuring mode Upol):

Input range –200.00 to 200.00 µA

Default value 200.00 µA

Volume:

Input range

Default value

0.00000 to 9999.99 mL

9999.99 mL

The Equivalence point Recognition Criterion (ERC) found is compared with this value. Equivalence points whose ERC is less than the value defined here will be ignored.

Measuring mode pH:

Input range

Default value

0.10 to 9.99

0.50

900 Touch Control ■■■■■■■■ 305

29.2 Monotonic equivalence point titrations (MET)

Measuring mode U, Ipol:

Input range 1 to 999 mV

Default value 30 mV

Measuring mode Upol:

Input range

Default value

0.1 to 99.9 µA

2.0 µA

EP recognition

■■■■■■■■■■■■■■■■■■■■■■

This parameter allows you to filter out only the equivalence points that are being sought.

Selection

Default value first first

| greatest | last | ascending | descending first

Only the first equivalence point will be recognized.

greatest

Only the equivalence point with the greatest ERC value, i.e. the steepest jump, will be recognized.

last

Only the last equivalence point will be recognized.

ascending

Only equivalence points with a positive slope of the titration curve will be recognized.

descending

Only equivalence points with a negative slope of the titration curve will be recognized.

Only one equivalence point will be recognized in a window. The numbering of the equivalence points (EP) is defined by the numbering of the windows (e.g. EP2 in window 2), so that even if EPs are missing, the calculations will still be carried out with the correctly assigned EP volumes.

U/mV U/mV

EP window 2

EP window 1 EP1

EP2

EP window 1 upper limit 1

+

EP1 lower limit 1

V/mL V/mL

Figure 15 Equivalence point recognition and equivalence point numbering within windows

306 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

In the first example in the above figure, two equivalence points are recognized in two different windows (EP1 in window 1 and EP2 in window 2).

In the second example, two equivalence points are found in one window, but only the first one is recognized. In order to ensure that the user recognizes that more than one equivalence point was found in the set window,

EP1 is marked with a "+" in the result view. In addition, a corresponding message is entered in the message list.

Evaluation and equivalence point criterion with MET

The equivalence points (EPs) are localized by a method based on the Fortuin method which has been adapted by Metrohm for numerical methods.

A search is made for the largest measured value change ( ). The exact EP is determined by using an interpolation factor which depends on the

Δ values before and after .

: EP volume

:

: Dosed total volume before

: Volume increment

Interpolation factor according to Fortuin

For the recognition of the EPs found, the set EP criterion is compared to the ERC (Equivalence point Recognition Criterion) found. The ERC is the sum of the measured value changes before and after the jump:

In certain cases only three or only one summand is taken into account.

EPs whose ERC is smaller than the defined EP criterion will not be recognized. The ERC is displayed in the results dialog for each discovered and recognized EP. If you adjust the EP criterion later in order to recognize more or fewer EPs, then you can reevaluate the determination ( [Recalculate] function in the results dialog).

29.2.5 Control instrument

The control instrument the determination is carried out with is selected under [Control device] . Control instruments are defined under System ▶ Device manager .

NOTICE

This button is only displayed when several control instruments have been configured.

900 Touch Control ■■■■■■■■ 307

29.2 Monotonic equivalence point titrations (MET) ■■■■■■■■■■■■■■■■■■■■■■

Control device

Selection of the control instrument from the list of devices. Only those devices are displayed which are able to carry out the command. The control instruments do not have to be connected. This allows to transfer methods more easily from one system to another.

Selection Selection of configured control instruments

29.2.6 Sensor

The parameters for the sensor are edited under [Sensor] .

Measuring input

Selection of the measuring input the sensor is connected to. The selection is not dependent on whether the control device has one or two measuring interfaces.

Selection

Default value

1 | 2

1

Sensor

Selection of the sensor from the sensor list. The selection depends on the measuring mode. Sensors are defined under System ▶ Sensors . You can also enter a sensor name which is not contained in the sensor list. When a determination is started there is a check whether the sensor is contained in the sensor list.

Selection Selection of configured sensors

I(pol)

The polarization current is the current that is applied to a polarizable electrode during voltametric measurement. This parameter is available only with I(pol) determinations.

The values that are actually adjustable may vary from the values specified here, depending on the hardware. The binding values can be found in the

Technical specifications chapter for the control instrument used.

Input range

Default value

Selection

–125.0 to 125.0 µA (Increment: 2.5)

5.0 µA

–1.0 µA | 1.0 µA

U(pol)

The polarization voltage is the voltage applied to the polarizable electrode during an amperometric measurement. This parameter is available only with U(pol) determinations.

308 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

The values that are actually adjustable may vary from the values specified here, depending on the hardware. The binding values can be found in the

Technical specifications chapter for the control instrument used.

Input range

Default value

–1,250 to 1,250 mV (Increment: 25)

400 mV

Electrode check on | off (Default value: off )

For the following electrodes, an electrode check can be carried out:

■ pH electrodes

Metal electrodes

Ion-selective electrodes

A check is made that the electrode is properly connected and that no short-circuit is present. The electrode check is carried out when this command is started. Note that this electrode check has nothing to do with the electrode test (command ELT ).

Temp. measurement

Type of temperature measurement.

Selection

Default value continuous automatic

| automatic | off continuous

A temperature sensor must be connected. The temperature is measured continuously.

automatic

If a temperature sensor is connected then the temperature will be measured continuously. Otherwise, the temperature entered manually will be used (see dialog of the titration and measuring parameters).

off

The temperature will not be measured. The temperature entered manually is used (see dialog of the titration and measuring parameters).

29.2.7 Dosing device

The parameters for the dosing device are edited under [Dosing device] .

Dosing device

Selection of the MSB connector to which the dosing device is connected.

Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4

900 Touch Control ■■■■■■■■ 309

29.2 Monotonic equivalence point titrations (MET)

Titrant

■■■■■■■■■■■■■■■■■■■■■■

Selection of the titrant from the titrant list. We recommend always selecting the titrant. This ensures that the correct data (titer, concentration, etc.) is always used for the calculation. Titrants are defined under System ▶ Titrants . You can also enter a name which is not contained in the titrant list. When a determination is started there is a check whether the solution is contained in the list.

For buret units with integrated data chip, a check is made in the method run to verify whether the correct titrant has been attached and whether the type of dosing drive matches. For buret units without an integrated data chip, the cylinder volume and the type of dosing drive are checked.

For the selected titrant, the validity of the titer, the working life of the titrant and the time interval for the GLP test of the buret unit are checked at the start of the determination.

Selection

Default value

Selection of configured titrants | not defined not defined not defined

No check takes place.

29.2.8 Stirrer

The parameters for the stirrer are edited under [Stirrer] .

Stirrer

Selection of the MSB connector the stirrer is connected to. Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4 | off off

No stirrer will be used.

Stirring rate

Setting the stirring rate. It can be set in steps of –15 to +15. The default setting 8 corresponds to 1000 rpm. The formula for calculating the rota-

tional speed is specified in chapter 32.4, page 494 . The optimum stirring

rate can be tested in the manual control.

The algebraic sign of the stirring rate changes the direction in which the stirring is done. When the stirrer is viewed from above, this means:

"+": counterclockwise rotation

"–": clockwise rotation

Input range

Default value

–15 to 15

8

310 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Switch off automatically on | off (Default value: on )

If this parameter is activated, the stirrer will be switched off automatically at the end of the titration, measurement, etc.

29.3 Endpoint titrations (SET)

29.3.1 Start conditions

The parameters that are carried out before the start of titration are defined under [Start conditions] .

Pause 1

Waiting time, e.g. for the electrode to settle down before a start volume is added.

Input range

Default value

0 to 999999 s

0 s

Start volume

Volume that is dosed prior to the start of the titration.

Input range

Default value

0.00000 to 9999.99 mL

0.00000 mL

Dosing rate

Rate at which the start volume is dosed.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum dosing rate depends on the cylinder volume (see chapter 32, page 492) .

Pause 2

Waiting time, e.g. for the electrode to settle down after the start or a reaction time after the dosing of a start volume.

Input range

Default value

0 to 999999 s

0 s

Initial measured value

The [Initial meas. value] is measured before carrying out the start conditions . This value is used for the titration direction determination when the titration direction is set to auto and only 1 EP is defined.

900 Touch Control ■■■■■■■■ 311

29.3 Endpoint titrations (SET) ■■■■■■■■■■■■■■■■■■■■■■

Signal drift

Maximum permissible drift for the measured value acceptance, i.e. maximum change of the measured value per minute. For this value, a suitable maximum waiting time is automatically calculated except you already have modified this waiting time.

Measuring mode pH, U and Ipol:

Input range

Selection

Default value

0.1 to 999.0 mV/min off off off

Measured value acceptance will take place after the maximum waiting time has elapsed.

Measuring mode Upol:

Input range

Selection

Default value

0.01 to 99.90 µA/min off off off

Measured value acceptance will take place after the maximum waiting time has elapsed.

Min. waiting time

The measured value is not accepted until the minimum waiting time has elapsed, even if the signal drift has already been reached. The minimum waiting time is only important for drift-controlled measurements.

Input range

Default value

0 to 999999 s

0 s

Max. waiting time

If the signal drift has been switched off or has not yet been reached, then the measured value will be accepted when the maximum waiting time has elapsed. As default value, a suitable waiting time for the signal drift is automatically calculated.

Input range

Default value

0 to 999999 s

1 s

29.3.2 Control parameters

The control parameters for endpoint 1 and endpoint 2 are defined under

[Control parameters] .

Endpoint 1 at

Measured value for the first endpoint.

312 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control instrument used.

Measuring mode pH:

Input range

Selection

Default value

–20.000 to 20.000 off off

Measuring mode U and Ipol:

Input range

Selection

Default value

–2000.0 to 2000.0 mV off off

Measuring mode Upol:

Input range –200.00 to 200.00 µA

Selection

Default value off off

Titration rate

Three predefined sets of parameters can be selected for the titration rate.

Selection

Default value slow | optimal optimal

| fast | user slow

For steep titration curves for which dosing must be carried out in small steps at the endpoint.

optimal

For all standard titrations. The parameters have been optimized for the most frequent applications.

fast

For flat titration curves for which the endpoint is reached only slowly.

user

The individual titration parameters can be modified.

The settings of the individual titration rates are listed in

table 11, page

315

.

User-defined parameters

These parameters are only accessible when Titration rate = user .

Dynamics

This parameter defines the control range before the specified endpoint.

Individual volume steps are dosed in the control range; the dosing is finely controlled. The closer the endpoint, the slower the dosing until the dosing

900 Touch Control ■■■■■■■■ 313

29.3 Endpoint titrations (SET) ■■■■■■■■■■■■■■■■■■■■■■ rate defined under Min. rate has been reached. The larger the control range, the slower the titration. Outside of the control range, dosing is carried out continuously; the dosing rate is defined under Max. rate .

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control instrument used.

Measuring mode pH:

Input range

Default value

Selection

0.001 to 20.000

2.000 off

Measuring mode U and Ipol:

Input range

Default value

Selection

0.1 to 2000.0 mV

100.0 mV off

Measuring mode Upol:

Input range 0.01 to 200.00 µA

Default value

Selection

10.00 µA off

Max. rate

Rate at which dosing is carried out outside of the control range.

Input range

Default value

Selection

0.01 to 166.00 mL/min

10.00 mL/min maximum maximum

The maximum dosing rate depends on the cylinder volume

(see chapter 32, page 492) .

Min. rate

Rate at which dosing is carried out at the very beginning of the titration and in the control range at the end of the titration. This parameter has a decisive influence on the titration rate and thus also on the accuracy. The smaller the selected minimum rate, the slower the titration.

Input range

Default value

0.01 to 9999.00 µL/min

25.00 µL/min

314 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Table 11 Default values of the predefined titration rates for SET

Titration rate slow optimal

Dynamics

– pH

– U und Ipol

– Upol

Max. rate

Min. rate

5.000

300.0 mV

40.00 µA

1.00 mL/min

5.00 µL/min

2.000

100.0 mV

10.00 µA

10.00 mL/min

25.00 µL/min

Stop criterion

Stop criterion fast

0.500

30.0 mV

5.00 µA maximum

50.00 µL/min

The titration is stopped when the endpoint has been reached and this stop criterion has been fulfilled. If no stop criterion has been selected then

the titration will not be stopped. The stop conditions (see chapter 29.3.4, page 317)

always lead to a stop, even if the stop criterion has not been reached.

NOTICE

Stop drift

In older instructions the delay time was usually defined as the stop criterion. The same delay time may however result in different stopping time points – because of different smallest possible volume increments

(depending on the cylinder volume). In contrast, if the stop drift is used, the titration will always be stopped at the same curve slope dV/dt.

Selection

Default value drift | time | off drift drift

The titration is stopped when the stop drift has been reached.

time

The titration is stopped when the endpoint has been exceeded during a certain period of time ( Delay time ).

off

The titration will not be stopped until the stop conditions have been fulfilled.

This parameter can only be edited with Stop criterion = drift .

900 Touch Control ■■■■■■■■ 315

29.3 Endpoint titrations (SET) ■■■■■■■■■■■■■■■■■■■■■■

The titration is stopped when the endpoint and the stop drift have been reached.

Input range

Default value

1 to 999 µL/min

20 µL/min

Delay time

This parameter can only be edited with Stop criterion = time .

When the endpoint has been reached, the specified time is allowed to elapse after the last dosing and the titration is then stopped.

Input range

Default value

0 to 999 s

10 s

Endpoint 2

The control parameters for the second endpoint are defined under [Endpoint 2] . The parameters and input ranges are identical with those for the first endpoint.

29.3.3 Titration parameters

Under [Titration parameters] , the parameters influencing the run of the entire titration are defined.

Titration direction

With the setting "auto", the titration direction is determined automatically from the initial measured value and the set endpoint. It is recommended that you specify whether the titration direction is positive or negative. If two endpoints have been set, then the titration direction will be defined automatically. In this case, the setting will be ignored.

Selection

Default value

+ | | auto auto

+

Positive measured value change by the addition of titrant, i.e. in the direction of a higher pH value, greater voltage or greater current.

-

Negative measured value change by the addition of titrant, i.e. in the direction of a lower pH value, lesser voltage or lesser current.

auto

The titration direction is determined automatically from the initial measured value and the set endpoint.

Extraction time

Minimum duration of the titration. The titration will not be stopped during the extraction time, even if the endpoint has already been reached. The titration is however stopped if a stop condition is fulfilled during this time

316 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

(see chapter 29.3.4, page 317) . The entry of an extraction time may be

advisable, e.g. for the titration of sparingly soluble samples.

Input range

Default value

0 to 999999 s

0 s

Temperature

Temperature entered manually. The temperature is being continuously measured when a temperature sensor is connected and when Temp.

measurement = automatic or continuous is defined (see sensor dialog of the command). This value is used for temperature correction in pH measurements.

Input range

Default value

–20.0 to 150.0 °C

25.0 °C

Time interval MP

Time interval for entering a measuring point in the measuring point list.

The measuring point list is limited to 1000 measuring points.

Input range

Default value

0.1 to 999999.0 s

2.0 s

29.3.4 Stop conditions

The conditions for canceling the titration are defined under [Stop conditions] , if this does not occur automatically. This could be the case when the endpoint set is not reached or if the stop criterion

(see "Stop criterion", page 315) is not fulfilled.

Stop volume

The titration is stopped when the specified volume has been dosed since the start of the titration. This volume should be adjusted to the size of the titration vessel in order to prevent the contents from running over.

Input range

Default value

Selection

0.00000 to 9999.99 mL

100,000 mL off

Stop time

The titration is stopped when the specified time has elapsed following the termination of the start conditions.

Input range

Selection

Default value

1 to 999999 s off off

Filling rate

Rate at which the dosing cylinder is filled.

900 Touch Control ■■■■■■■■ 317

29.3 Endpoint titrations (SET) ■■■■■■■■■■■■■■■■■■■■■■

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum filling rate depends on the cylinder volume (see chapter

32, page 492) .

29.3.5 Conditioning

The conditions required for conditioning are defined under [Conditioning] .

Conditioning on | off (Default value: off )

If this parameter is activated, then the first time the titration is started the working medium will be titrated to the endpoint with the specified control parameters. The status is kept stable. The actual method run does not begin until [ ] has been pressed once more.

Start drift

Conditioning OK will be displayed as soon as this drift has been reached and the titration can be started.

Input range

Default value

1 to 999 µL/min

20 µL/min

Drift correction

The endpoint volume can be corrected by taking the drift value into account. For this, the drift is multiplied with the drift correction time and this value is then subtracted from the endpoint volume. The drift correction time is the time interval between the end of conditioning and the end of the determination.

Selection

Default value auto | manual | off off auto

The value of the current drift is automatically applied at the start of the titration.

manual

If the drift is known throughout a longer period of time, this can be entered manually.

off

No drift correction takes place.

Drift value

This parameter can only be edited with Drift correction = manual .

318 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Drift for manual drift correction.

Input range

Default value

0.0 to 99.9 µL/min

0.0 µL/min

Measured value display on | off (Default value: off )

If this parameter is activated, the currently measured value is displayed during the conditioning.

Conditioning options [Cond. options]

Cond. stop volume

Maximum permissible volume that can be dosed during conditioning.

Conditioning is stopped when the specified volume is dosed. If conditioning is continued by pressing [ ] once again, then the titrant volume that has already been dosed will not be taken into account; i.e. the dosing starts again at zero. The stop volume should be adjusted to the size of the titration cell in order to prevent any overflow.

Input range

Default value

Selection

0.00000 to 9999.99 mL

20.0000 mL off

Cond. stop time

Maximum permissible time over which conditioning may take place. Conditioning is stopped when the specified time has elapsed.

Input range

Selection

Default value

1 to 999999 s off off

Delay reconditioning on | off (Default value: off )

If this parameter is activated, the conditioning will not start before all the method commands have been processed. Otherwise, the conditioning will immediately start after the titration command.

NOTICE

This parameter must be activated if commands follow which must not run simultaneously with the conditioning.

Example: Emptying the titration cell followed by adding new working medium.

900 Touch Control ■■■■■■■■ 319

29.3 Endpoint titrations (SET) ■■■■■■■■■■■■■■■■■■■■■■

29.3.6 Cell

The monitoring of the reagent used is activated or deactivated under

[Cell] .

Reagent monitoring

Selection of the reagent from the list of reagents. The selection depends on the titration mode. Reagents are defined under System ▶ Reagents .

Selection

Default value

Selection of configured reagents off

| off off

The reagent monitoring is not active.

29.3.7 Control instrument

The control instrument the determination is carried out with is selected under [Control device] . Control instruments are defined under System ▶ Device manager .

NOTICE

This button is only displayed when several control instruments have been configured.

Control device

Selection of the control instrument from the list of devices. Only those devices are displayed which are able to carry out the command. The control instruments do not have to be connected. This allows to transfer methods more easily from one system to another.

Selection Selection of configured control instruments

29.3.8 Sensor

The parameters for the sensor are edited under [Sensor] .

Measuring input

Selection of the measuring input the sensor is connected to. The selection is not dependent on whether the control device has one or two measuring interfaces.

Selection

Default value

1 | 2

1

320 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Sensor

29 Parameters

Selection of the sensor from the sensor list. The selection depends on the measuring mode. Sensors are defined under System ▶ Sensors . You can also enter a sensor name which is not contained in the sensor list. When a determination is started there is a check whether the sensor is contained in the sensor list.

Selection Selection of configured sensors

I(pol)

The polarization current is the current that is applied to a polarizable electrode during voltametric measurement. This parameter is available only with I(pol) determinations.

The values that are actually adjustable may vary from the values specified here, depending on the hardware. The binding values can be found in the

Technical specifications chapter for the control instrument used.

Input range

Default value

Selection

–125.0 to 125.0 µA (Increment: 2.5)

5.0 µA

–1.0 µA | 1.0 µA

U(pol)

The polarization voltage is the voltage applied to the polarizable electrode during an amperometric measurement. This parameter is available only with U(pol) determinations.

The values that are actually adjustable may vary from the values specified here, depending on the hardware. The binding values can be found in the

Technical specifications chapter for the control instrument used.

Input range

Default value

–1,250 to 1,250 mV (Increment: 25)

400 mV

Electrode check on | off (Default value: off )

For the following electrodes, an electrode check can be carried out:

■ pH electrodes

Metal electrodes

Ion-selective electrodes

A check is made that the electrode is properly connected and that no short-circuit is present. The electrode check is carried out when this command is started. Note that this electrode check has nothing to do with the electrode test (command ELT ).

900 Touch Control ■■■■■■■■ 321

29.3 Endpoint titrations (SET) ■■■■■■■■■■■■■■■■■■■■■■

Temp. measurement

Type of temperature measurement.

Selection

Default value continuous automatic

| automatic | off continuous

A temperature sensor must be connected. The temperature is measured continuously.

automatic

If a temperature sensor is connected then the temperature will be measured continuously. Otherwise, the temperature entered manually will be used (see dialog of the titration and measuring parameters).

off

The temperature will not be measured. The temperature entered manually is used (see dialog of the titration and measuring parameters).

29.3.9 Dosing device

The parameters for the dosing device are edited under [Dosing device] .

Dosing device

Selection of the MSB connector to which the dosing device is connected.

Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4

Titrant

Selection of the titrant from the titrant list. We recommend always selecting the titrant. This ensures that the correct data (titer, concentration, etc.) is always used for the calculation. Titrants are defined under System ▶ Titrants . You can also enter a name which is not contained in the titrant list. When a determination is started there is a check whether the solution is contained in the list.

For buret units with integrated data chip, a check is made in the method run to verify whether the correct titrant has been attached and whether the type of dosing drive matches. For buret units without an integrated data chip, the cylinder volume and the type of dosing drive are checked.

For the selected titrant, the validity of the titer, the working life of the titrant and the time interval for the GLP test of the buret unit are checked at the start of the determination.

Selection

Default value

Selection of configured titrants | not defined not defined not defined

No check takes place.

322 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

29.3.10 Stirrer

The parameters for the stirrer are edited under [Stirrer] .

Stirrer

Selection of the MSB connector the stirrer is connected to. Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4 | off off

No stirrer will be used.

Stirring rate

Setting the stirring rate. It can be set in steps of –15 to +15. The default setting 8 corresponds to 1000 rpm. The formula for calculating the rotational speed is specified in

chapter 32.4, page 494

. The optimum stirring rate can be tested in the manual control.

The algebraic sign of the stirring rate changes the direction in which the stirring is done. When the stirrer is viewed from above, this means:

"+": counterclockwise rotation

"–": clockwise rotation

Input range

Default value

–15 to 15

8

Switch off automatically on | off (Default value: on )

If this parameter is activated, the stirrer will be switched off automatically at the end of the titration, measurement, etc.

29.4 Volumetric Karl Fischer titrations (KFT)

29.4.1 Start conditions

The parameters that are carried out before the start of titration are defined under [Start conditions] .

Pause 1

Waiting time, e.g. for the electrode to settle down before a start volume is added.

Input range

Default value

0 to 999999 s

0 s

900 Touch Control ■■■■■■■■ 323

29.4 Volumetric Karl Fischer titrations (KFT)

Start volume

■■■■■■■■■■■■■■■■■■■■■■

Volume that is dosed prior to the start of the titration.

Input range

Default value

0.00000 to 9999.99 mL

0.00000 mL

Dosing rate

Rate at which the start volume is dosed.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum dosing rate depends on the cylinder volume

(see chapter 32, page 492) .

Pause 2

Waiting time, e.g. for the electrode to settle down after the start or a reaction time after the dosing of a start volume.

Input range

Default value

0 to 999999 s

0 s

Initial measured value

The [Initial meas. value] is measured before carrying out the start conditions . This value is used for the titration direction determination when the titration direction is set to auto .

Signal drift

Maximum permissible drift for the measured value acceptance, i.e. maximum change of the measured value per minute. For this value, a suitable maximum waiting time is automatically calculated except you already have modified this waiting time.

Measuring mode Ipol:

Input range 0.1 to 999.0 mV/min

Selection

Default value off off off

Measured value acceptance will take place after the maximum waiting time has elapsed.

Measuring mode Upol:

Input range 0.01 to 99.90 µA/min

Selection

Default value off off

324 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters off

Measured value acceptance will take place after the maximum waiting time has elapsed.

Min. waiting time

The measured value is not accepted until the minimum waiting time has elapsed, even if the signal drift has already been reached. The minimum waiting time is only important for drift-controlled measurements.

Input range

Default value

0 to 999999 s

0 s

Max. waiting time

If the signal drift has been switched off or has not yet been reached, then the measured value will be accepted when the maximum waiting time has elapsed. As default value, a suitable waiting time for the signal drift is automatically calculated.

Input range

Default value

0 to 999999 s

1 s

29.4.2 Control parameters

The control parameters for the endpoint are defined under [Control parameters] .

Endpoint at

Measured value for the endpoint.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control device used.

Measuring mode Ipol:

Input range

Default value

Selection

–2,000.0 to 2,000.0 mV

250.0 mV off

Measuring mode Upol:

Input range –200.00 to 200.00 µA

Default value

Selection

25.00 µA off

Titration rate

Three predefined sets of parameters can be selected for the titration rate.

Selection

Default value slow | optimal optimal

| fast | user

900 Touch Control ■■■■■■■■ 325

29.4 Volumetric Karl Fischer titrations (KFT) ■■■■■■■■■■■■■■■■■■■■■■ slow

For samples with a low water content or samples which release their moisture only slowly.

optimal

For all standard titrations. The parameters have been optimized for the most frequent applications.

fast

For uncritical samples with high water content.

user

The individual titration parameters can be modified.

The settings of the individual titration rates are listed in table 12, page

327 .

User-defined parameters

These parameters are only accessible when Titration rate = user .

Dynamics

This parameter defines the control range before the specified endpoint.

Individual volume steps are dosed in the control range, the dosing is finely controlled. The closer the endpoint, the slower the dosing until the volume increment defined under Min. increment has been reached. The larger the control range, the slower the titration. Outside the control range, dosing is carried out continuously, and the dosing rate is defined under Max. rate .

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control device used.

Measuring mode Ipol:

Input range

Default value

Selection

0.1 to 2,000.0 mV

100.0 mV off

Measuring mode Upol:

Input range 0.01 to 200.00 µA

Default value

Selection

10.00 µA off

Max. rate

Rate at which dosing is carried out outside of the control range.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum

326 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters maximum

The maximum dosing rate depends on the cylinder volume (see chapter 32, page 492) .

Min. increment

Minimum volume increment that is dosed at the beginning of the titration and in the control range at the end of the titration. This parameter has a decisive influence on the titration rate and thus also on the accuracy. The smaller the selected minimum increment, the slower the titration.

Input range

Selection

Default value

0.10 to 99.90 µL minimum minimum

Table 12 Default values of the predefined titration rates for KFT

Titration rate slow optimal

Dynamics

– Ipol

– Upol

Max. rate

Min. increment

300.0 mV

40.00 µA

1.00 mL/min minimum (= cylinder volume/10,000)

100.0 mV

10.00 µA maximum minimum (= cylinder volume/10,000)

Stop criterion

Stop criterion fast

30.0 mV

5.00 µA maximum

5.00 µL

The titration is stopped when the endpoint has been reached and this stop criterion has been fulfilled. If no stop criterion has been selected then

the titration will not be stopped. The stop conditions (see chapter 29.4.4, page 329)

always lead to a stop, even if the stop criterion has not been reached.

NOTICE

In older instructions the delay time was usually defined as the stop criterion. The same delay time may however result in different stopping time points – because of different smallest possible volume increments

(depending on the cylinder volume). In contrast, if the stop drift is used, the titration will always be stopped at the same curve slope dV/dt.

Selection

Default value drift | time | rel. drift | off drift

900 Touch Control ■■■■■■■■ 327

29.4 Volumetric Karl Fischer titrations (KFT) ■■■■■■■■■■■■■■■■■■■■■■ drift

The titration is stopped when the stop drift has been reached.

time

The titration is stopped if the endpoint has been exceeded during a certain time period ( Delay time ).

rel. drift

The titration is stopped when the sum of the drift at the start of the titration and the relative stop drift has been reached.

off

The titration will not be stopped until the stop conditions have been fulfilled.

Stop drift

This parameter can only be edited with Stop criterion = drift .

The titration is stopped when the endpoint and the stop drift have been reached.

Input range

Default value

1 to 999 µL/min

20 µL/min

Delay time

This parameter can only be edited with Stop criterion = time .

When the endpoint has been reached, the specified time is allowed to elapse after the last dosing and the titration is then stopped.

Input range

Default value

0 to 999 s

10 s

Relative stop drift

This parameter can only be edited with Stop criterion = rel. drift .

The titration is stopped when the endpoint and the sum of the drift at the start of the titration and the relative stop drift have been reached.

Input range

Default value

1 to 999 µL/min

10 µL/min

29.4.3 Titration parameters

Under [Titration parameters] , the parameters influencing the run of the entire titration are defined.

Titration direction

Selection + | | auto

The default value depends on the measuring mode:

Ipol = , Upol = +

328 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

+

Positive measured value change by the addition of titrant, i.e. in the direction of a greater voltage or greater current.

-

Negative measured value change by the addition of titrant, i.e. in the direction of a lesser voltage or lesser current.

auto

The titration direction is determined automatically from the initial measured value and the set endpoint.

Extraction time

Minimum duration of the titration. The titration will not be stopped during the extraction time, even if the endpoint has already been reached. The titration is however stopped if a stop condition is fulfilled during this time

(see chapter 29.4.4, page 329) . The entering of an extraction time is e.g.

recommended with samples that release water slowly or if a Karl Fischer oven is used.

Input range

Default value

0 to 999999 s

0 s

Temperature

Temperature entered manually. The temperature is being continuously measured when a temperature sensor is connected and when Temp.

measurement = automatic or continuous is defined (see sensor dialog of the command).

Input range

Default value

–20.0 to 150.0 °C

25.0 °C

Time interval MP

Time interval for entering a measuring point in the measuring point list.

The measuring point list is limited to 1000 measuring points.

Input range

Default value

0.1 to 999999.0 s

2.0 s

29.4.4 Stop conditions

The conditions for canceling the titration are defined under [Stop conditions] , if this does not occur automatically. This could be the case when the endpoint set is not reached or if the stop criterion

(see "Stop criterion", page 327) is not fulfilled.

Stop volume

The titration is stopped when the specified volume has been dosed since the start of the titration. This volume should be adjusted to the size of the titration vessel in order to prevent the contents from running over.

900 Touch Control ■■■■■■■■ 329

29.4 Volumetric Karl Fischer titrations (KFT) ■■■■■■■■■■■■■■■■■■■■■■

Input range

Default value

Selection

0.00000 to 9999.99 mL

100,000 mL off

Stop time

The titration is stopped when the specified time has elapsed following the termination of the start conditions.

Input range

Selection

Default value

1 to 999999 s off off

Filling rate

Rate at which the dosing cylinder is filled.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum filling rate depends on the cylinder volume (see chapter

32, page 492) .

29.4.5 Conditioning

The conditions required for conditioning are defined under [Conditioning] .

Conditioning on | off (Default value: on )

If this parameter is activated, then the first time the titration is started the working medium will be titrated to the endpoint with the specified control parameters. The status is kept stable. The actual method run does not begin until [ ] has been pressed once more.

Start drift

Conditioning OK will be displayed as soon as this drift has been reached and the titration can be started.

Input range

Default value

1 to 999 µL/min

20 µL/min

Drift correction

The endpoint volume can be corrected by taking the drift value into account. For this, the drift is multiplied with the drift correction time and this value is then subtracted from the endpoint volume. The drift correction time is the time interval between the end of conditioning and the end of the determination.

330 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Selection

Default value auto | manual | off off auto

The value of the current drift is automatically applied at the start of the titration.

manual

If the drift is known throughout a longer period of time, this can be entered manually.

off

No drift correction takes place.

Drift value

This parameter can only be edited with Drift correction = manual .

Drift for manual drift correction.

Input range

Default value

0.0 to 99.9 µL/min

0.0 µL/min

Measured value display on | off (Default value: off )

If this parameter is activated, the currently measured value is displayed during the conditioning.

Conditioning options [Cond. options]

Cond. stop volume

Maximum permissible volume that can be dosed during conditioning.

Conditioning is stopped when the specified volume is dosed. If conditioning is continued by pressing [ ] once again, then the titrant volume that has already been dosed will not be taken into account; i.e. the dosing starts again at zero. The stop volume should be adjusted to the size of the titration cell in order to prevent any overflow.

Input range

Default value

Selection

0.00000 to 9999.99 mL

20.0000 mL off

Cond. stop time

Maximum permissible time over which conditioning may take place. Conditioning is stopped when the specified time has elapsed.

Input range

Selection

Default value

1 to 999999 s off off

900 Touch Control ■■■■■■■■ 331

29.4 Volumetric Karl Fischer titrations (KFT) ■■■■■■■■■■■■■■■■■■■■■■

Delay reconditioning on | off (Default value: off )

If this parameter is activated, the conditioning will not start before all the method commands have been processed. Otherwise, the conditioning will immediately start after the titration command.

NOTICE

This parameter must be activated if commands follow which must not run simultaneously with the conditioning.

Example: Emptying the titration cell followed by adding new working medium.

29.4.6 Cell

The monitoring of the reagent used is activated or deactivated under

[Cell] .

Reagent monitoring

Selection of the reagent from the list of reagents. The selection depends on the titration mode. Reagents are defined under System ▶ Reagents .

Selection

Default value

Selection of configured reagents off

| off off

The reagent monitoring is not active.

29.4.7 Control instrument

The control instrument the determination is carried out with is selected under [Control device] . Control instruments are defined under System ▶ Device manager .

NOTICE

This button is only displayed when several control instruments have been configured.

Control device

Selection of the control instrument from the list of devices. Only those devices are displayed which are able to carry out the command. The control instruments do not have to be connected. This allows to transfer methods more easily from one system to another.

332 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Selection

29.4.8 Sensor

29 Parameters

Selection of configured control instruments

The parameters for the sensor are edited under [Sensor] .

Measuring input

Selection of the measuring input the sensor is connected to. The selection is not dependent on whether the control device has one or two measuring interfaces.

Selection

Default value

1 | 2

1

Sensor

Selection of the sensor from the sensor list. The selection depends on the measuring mode. Sensors are defined under System ▶ Sensors . You can also enter a sensor name which is not contained in the sensor list. When a determination is started there is a check whether the sensor is contained in the sensor list.

Selection Selection of configured sensors

I(pol)

The polarization current is the current that is applied to a polarizable electrode during voltametric measurement. This parameter is available only with I(pol) determinations.

The values that are actually adjustable may vary from the values specified here, depending on the hardware. The binding values can be found in the

"Technical specifications" chapter for the control instrument used.

Input range

Default value

Selection

–125.0 to 125.0 µA (Increment: 2.5)

50.0 µA

–1.0 µA | 1.0 µA

U(pol)

The polarization voltage is the voltage applied to the polarizable electrode during an amperometric measurement. This parameter is available only with U(pol) determinations.

The values that are actually adjustable may vary from the values specified here, depending on the hardware. The binding values can be found in the

Technical specifications chapter for the control instrument used.

Input range

Default value

–1,250 to 1,250 mV (Increment: 25)

400 mV

900 Touch Control ■■■■■■■■ 333

29.4 Volumetric Karl Fischer titrations (KFT) ■■■■■■■■■■■■■■■■■■■■■■

Electrode check on | off (Default value: off )

For the following electrodes, an electrode check can be carried out:

■ pH electrodes

Metal electrodes

Ion-selective electrodes

A check is made that the electrode is properly connected and that no short-circuit is present. The electrode check is carried out when this command is started. Note that this electrode check has nothing to do with the electrode test (command ELT ).

Temp. measurement

Type of temperature measurement.

Selection

Default value continuous automatic

| automatic | off continuous

A temperature sensor must be connected. The temperature is measured continuously.

automatic

If a temperature sensor is connected then the temperature will be measured continuously. Otherwise, the temperature entered manually will be used (see dialog of the titration and measuring parameters).

off

The temperature will not be measured. The temperature entered manually is used (see dialog of the titration and measuring parameters).

29.4.9 Dosing device

The parameters for the dosing device are edited under [Dosing device] .

Dosing device

Selection of the MSB connector to which the dosing device is connected.

Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4

Titrant

Selection of the titrant from the titrant list. We recommend always selecting the titrant. This ensures that the correct data (titer, concentration, etc.) is always used for the calculation. Titrants are defined under System ▶ Titrants . You can also enter a name which is not contained in the titrant list. When a determination is started there is a check whether the solution is contained in the list.

334 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

For buret units with integrated data chip, a check is made in the method run to verify whether the correct titrant has been attached and whether the type of dosing drive matches. For buret units without an integrated data chip, the cylinder volume and the type of dosing drive are checked.

For the selected titrant, the validity of the titer, the working life of the titrant and the time interval for the GLP test of the buret unit are checked at the start of the determination.

Selection

Default value

Selection of configured titrants | not defined not defined not defined

No check takes place.

29.4.10 Stirrer

The parameters for the stirrer are edited under [Stirrer] .

Stirrer

Selection of the MSB connector the stirrer is connected to. Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4 | off off

No stirrer will be used.

Stirring rate

Setting the stirring rate. It can be set in steps of –15 to +15. The default setting 8 corresponds to 1000 rpm. The formula for calculating the rotational speed is specified in

chapter 32.4, page 494

. The optimum stirring rate can be tested in the manual control.

The algebraic sign of the stirring rate changes the direction in which the stirring is done. When the stirrer is viewed from above, this means:

"+": counterclockwise rotation

"–": clockwise rotation

Input range

Default value

–15 to 15

8

Switch off automatically on | off (Default value: on )

If this parameter is activated, the stirrer will be switched off automatically at the end of the titration, measurement, etc.

900 Touch Control ■■■■■■■■ 335

29.5 Coulometric Karl Fischer titrations (KFC) ■■■■■■■■■■■■■■■■■■■■■■

29.5 Coulometric Karl Fischer titrations (KFC)

29.5.1 Control parameters

The control parameters for the endpoint are defined under [Control parameters] .

Endpoint at

Measured value for the endpoint.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control instrument used.

Input range

Default value

Selection

–1250.0 to 1250.0 mV

50.0 mV off

Titration rate

Three predefined sets of parameters can be selected for the titration rate.

Selection

Default value slow | optimal optimal

| fast | user slow

For samples with a low water content or samples which release their moisture only slowly.

optimal

For all standard titrations. The parameters have been optimized for the most frequent applications.

fast

For uncritical samples with high water content.

user

The individual titration parameters can be modified.

The settings of the individual titration rates are listed in table 13, page

337 .

User-defined parameters

These parameters are only accessible when Titration rate = user .

Dynamics

This parameter defines the control range before the specified endpoint.

Within the control range, the iodine is generated step by step, the generation is finely controlled. The closer the endpoint, the slower the iodine is generated until the rate defined under Min. rate is reached. The larger the control range, the slower the titration. Outside the control range,

336 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters iodine is being continuously generated, the rate is defined under Max.

rate .

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control instrument used.

Input range

Default value

Selection

0.1 to 1250.0 mV

70.0 mV off

Max. rate

Rate at which iodine is generated outside the control range.

Input range

Selection

Default value

1.5 to 2241.0 µg/min maximum maximum

Min. rate

Rate at which iodine is generated at the very beginning of the titration and in the control range at the end of the titration. This parameter has a vital influence on the titration rate and thus also on the accuracy. The smaller the selected minimum rate, the slower the titration.

Input range

Default value

0.3 to 999.0 µg/min

15.0 µg/min

Table 13 Default values of the predefined titration rates for KFC

Titration rate

Dynamics

Max. rate

Min. rate slow

120.0 mV

1000.0 µg/min

0.3 µg/min optimal

70.0 mV maximum

15.0 µg/min

Stop criterion

Stop criterion fast

30.0 mV maximum

30.0 µg/min

The titration is stopped when the endpoint has been reached and this stop criterion has been fulfilled. If no stop criterion has been selected then

the titration will not be stopped. The stop conditions (see chapter 29.5.3, page 339)

always lead to a stop, even if the stop criterion has not been reached.

Selection

Default value drift | time | rel. drift | off rel. drift

900 Touch Control ■■■■■■■■ 337

29.5 Coulometric Karl Fischer titrations (KFC) ■■■■■■■■■■■■■■■■■■■■■■ drift

The titration is stopped when the stop drift has been reached.

time

The titration is stopped if the endpoint has been exceeded during a certain time period ( Delay time ).

rel. drift

The titration is stopped when the sum of the drift at the start of the titration and the relative stop drift has been reached.

off

The titration will not be stopped until the stop conditions have been fulfilled.

Stop drift

This parameter can only be edited with Stop criterion = drift .

The titration is stopped when the endpoint and the stop drift have been reached.

Input range

Default value

1 to 999 µg/min

5 µg/min

Delay time

This parameter can only be edited with Stop criterion = time .

When the endpoint has been reached, the specified time is allowed to elapse after the last dosing and the titration is then stopped.

Input range

Default value

0 to 999 s

10 s

Relative stop drift

This parameter can only be edited with Stop criterion = rel. drift .

The titration is stopped when the endpoint and the sum of the drift at the start of the titration and the relative stop drift have been reached.

Input range

Default value

1 to 999 µg/min

5 µg/min

29.5.2 Titration parameters

Under [Titration parameters] , the parameters influencing the run of the entire titration are defined.

Pause

Waiting time, for example for dissolving the sample. During this time no iodine is generated.

Input range

Default value

0 to 999999 s

0 s

338 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Extraction time

Minimum duration of the titration. The titration will not be stopped during the extraction time, even if the endpoint has already been reached. The titration is however stopped if a stop condition is fulfilled during this time

(see chapter 29.5.3, page 339) . Entering an extraction time may e.g. be

advisable if a Karl Fischer oven is used.

Input range

Default value

0 to 999999 s

0 s

Temperature

Temperature entered manually. The temperature is being continuously measured when a temperature sensor is connected and when Temp.

measurement = automatic or continuous is defined (see sensor dialog of the command).

Input range

Default value

–20.0 to 150.0 °C

25.0 °C

Time interval MP

Time interval for entering a measuring point in the measuring point list.

The measuring point list is limited to 1000 measuring points.

Input range

Default value

0.1 to 999999.0 s

2.0 s

29.5.3 Stop conditions

The conditions for canceling the titration are defined under [Stop conditions] , if this does not occur automatically. This could be the case when the endpoint set is not reached or if the stop criterion

(see "Stop criterion", page 337) is not fulfilled.

Stop time

The titration is stopped when the specified time has elapsed since the start of the titration.

Input range

Selection

Default value

1 to 999999 s off off

29.5.4 Conditioning

The conditions required for conditioning are defined under [Conditioning] .

900 Touch Control ■■■■■■■■ 339

29.5 Coulometric Karl Fischer titrations (KFC) ■■■■■■■■■■■■■■■■■■■■■■

Conditioning on | off (Default value: on )

If this parameter is activated, then the first time the titration is started the working medium will be titrated to the endpoint with the specified control parameters. The status is kept stable. The actual method run does not begin until [ ] has been pressed once more.

Start drift

If the measured drift is smaller than this value for a certain time (so-called stabilizing time), Conditioning OK is displayed and the titration can be started. The stabilizing time is defined in the dialog Conditioning / Conditioning options .

Input range

Default value

1 to 999 µg/min

20 µg/min

Drift correction

The endpoint amount can be corrected taking the drift value into account.

For this, the drift is multiplied with the drift correction time and this value is then subtracted from the endpoint amount. The drift correction time is the time interval between the end of conditioning and the end of the determination.

Selection

Default value auto | manual | off auto auto

The value of the current drift is automatically applied at the start of the titration.

manual

If the drift is known throughout a longer period of time, this can be entered manually.

off

No drift correction takes place.

Drift value

This parameter can only be edited with Drift correction = manual .

Drift for manual drift correction.

Input range

Default value

0.0 to 99.9 µg/min

0.0 µg/min

Measured value display on | off (Default value: off )

If this parameter is activated, the currently measured value is displayed during the conditioning.

340 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Automatic start [Automatic start]

NOTICE

If you use the automatic start function, then we recommend that the start drift be reduced and that a stabilizing time be defined.

Recommendation: start drift ≤ 10 µg/min / stabilizing time ≥ 30 s

Automatic start on | off (Default value: off )

If this parameter is activated, the determination is automatically started when the measured voltage suddenly changes. This setting will be ignored as long as the working medium has not been conditioned.

Threshold value

If the change in voltage is higher than this value, the determination will automatically be started.

Input range

Default value

0 to 999 mV

50 mV

Conditioning options [Cond. options]

Stabilizing time

Waiting time, during which the measured drift has to be smaller than the start drift defined until Conditioning OK is displayed. The start drift is defined in the dialog Edit command / Conditioning .

Input range

Default value

0 to 999999 s

0 s

Cond. stop time

Maximum permissible time over which conditioning may take place. Conditioning is stopped when the specified time has elapsed.

Input range

Selection

Default value

1 to 999999 s off off

Delay reconditioning on | off (Default value: off )

If this parameter is activated, the conditioning will not start before all the method commands have been processed. Otherwise, the conditioning will immediately start after the titration command.

900 Touch Control ■■■■■■■■ 341

29.5 Coulometric Karl Fischer titrations (KFC) ■■■■■■■■■■■■■■■■■■■■■■

NOTICE

This parameter must be activated if commands follow which must not run simultaneously with the conditioning.

Example: Emptying the titration cell followed by adding new working medium.

29.5.5 Cell

The parameters for the titration cell are defined under [Cell] .

Generator electrode

Type of generator electrode.

Selection

Default value without diaphragm without diaphragm

| with diaphragm without diaphragm

Generator electrode without diaphragm.

with diaphragm

Generator electrode with diaphragm.

Generator current

Polarization current at the generator electrode.

Selection 100 mA | 200 mA | 400 mA | auto

400 mA

Default value, when Generator electrode = without diaphragm .

auto

The current is adapted to the conductivity of the reagent and automatically reduced near the endpoint. Default value, when Generator electrode = with diaphragm .

Reagent monitoring

Selection of the reagent from the list of reagents. The selection depends on the titration mode. Reagents are defined under System ▶ Reagents .

Selection

Default value

Selection of configured reagents off

| off off

The reagent monitoring is not active.

342 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

29.5.6 Control instrument

The control instrument the determination is carried out with is selected under [Control device] . Control instruments are defined under System ▶ Device manager .

NOTICE

This button is only displayed when several control instruments have been configured.

Control device

Selection of the control instrument from the list of devices. Only those devices are displayed which are able to carry out the command. The control instruments do not have to be connected. This allows to transfer methods more easily from one system to another.

Selection Selection of configured control instruments

29.5.7 Sensor

The parameters for the sensor are edited under [Sensor] .

Sensor

Selection of the sensor from the sensor list. The selection depends on the measuring mode. Sensors are defined under System ▶ Sensors . You can also enter a sensor name which is not contained in the sensor list. When a determination is started there is a check whether the sensor is contained in the sensor list.

Selection Selection of configured sensors

I(pol)

The polarization current is the current that is applied to a polarizable electrode during the voltametric measurement.

Selection

Default value

5 µA |

10 µA

10 µA | 20 µA | 30 µA

Electrode check on | off (Default value: off )

For polarizable electrodes and the generator electrode, an electrode check can be carried out. The following will be tested:

900 Touch Control ■■■■■■■■ 343

29.5 Coulometric Karl Fischer titrations (KFC) ■■■■■■■■■■■■■■■■■■■■■■

Polarizable electrodes

– Is the electrode connected?

– Is there a short circuit?

The electrode check is carried out when this command is started.

Generator electrode:

– Is the electrode connected?

– Is the generator system defective?

– Is the resistance of the sample solution very high?

The electrode check is carried out continuously during the execution of this command.

Temp. measurement

Type of temperature measurement.

Selection

Default value continuous automatic

| automatic | off continuous

A temperature sensor must be connected. The temperature is measured continuously.

automatic

If a temperature sensor is connected then the temperature will be measured continuously. Otherwise, the temperature entered manually will be used (see dialog of the titration and measuring parameters).

off

The temperature will not be measured. The temperature entered manually is used (see dialog of the titration and measuring parameters).

29.5.8 Stirrer

The parameters for the stirrer are edited under [Stirrer] .

Stirrer

Selection of the MSB connector the stirrer is connected to. Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4 | off off

No stirrer will be used.

Stirring rate

Setting the stirring rate. It can be set in steps of –15 to +15. The default setting 8 corresponds to 1000 rpm. The formula for calculating the rota-

tional speed is specified in chapter 32.4, page 494 . The optimum stirring

rate can be tested in the manual control.

344 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

The algebraic sign of the stirring rate changes the direction in which the stirring is done. When the stirrer is viewed from above, this means:

"+": counterclockwise rotation

"–": clockwise rotation

Input range

Default value

–15 to 15

8

Switch off automatically on | off (Default value: on )

If this parameter is activated, the stirrer will be switched off automatically at the end of the titration, measurement, etc.

29.6 Bromine index determination (BRC)

29.6.1 Control parameters

The control parameters for the endpoint are defined under [Control parameters] .

Endpoint at

Measured value for the endpoint.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control device used.

Input range

Default value

Selection

–1,250.0 to 1,250.0 mV

200.0 mV off

Titration rate

Three predefined sets of parameters can be selected for the titration rate.

Selection

Default value slow | optimal optimal

| fast | user slow

For samples with slow bromine addition.

optimal

For all standard titrations. The parameters have been optimized for the most frequent applications.

fast

For uncritical samples with fast bromine addition.

user

The individual titration parameters can be modified.

900 Touch Control ■■■■■■■■ 345

29.6 Bromine index determination (BRC) ■■■■■■■■■■■■■■■■■■■■■■

The settings of the individual titration rates are listed in table 14, page

346 .

User-defined parameters

These parameters are only accessible when Titration rate = user .

Dynamics

This parameter defines the control range before the specified endpoint.

Within the control range, the bromine is generated step by step, the generation is finely controlled. The closer the endpoint, the slower the bromine is generated until the rate defined under Min. rate is reached. The larger the control range, the slower the titration. Outside the control range, bromine is continuously generated, the rate is defined under Max.

rate .

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control instrument used.

Input range

Default value

Selection

0.1 to 1250.0 mV

400.0 mV off

Max. rate

Rate at which bromine is generated outside the control range.

Input range

Default value

Selection

10.0 to 19876.0 µg/min

1000.0 µg/min maximum

Min. rate

Rate at which bromine is generated at the very beginning of the titration and in the control range at the end of the titration. This parameter has a vital influence on the titration rate and thus also on the accuracy. The smaller the selected minimum rate, the slower the titration.

Input range

Default value

1.0 to 999.0 µg/min

25.0 µg/min

Table 14 Default values for the predefined titration rates for BRC

Titration rate slow optimal

Dynamics

Max. rate

Min. rate

400.0 mV

500.0 µg/min

15.0 µg/min

400.0 mV

1000.0 µg/min

25.0 µg/min fast

400.0 mV

2000.0 µg/min

50.0 µg/min

346 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Stop criterion

29 Parameters

Stop criterion

The titration is stopped when the endpoint has been reached and this stop criterion has been fulfilled. If no stop criterion has been selected then

the titration will not be stopped. The stop conditions (see chapter 29.6.3, page 349)

always lead to a stop, even if the stop criterion has not been reached.

Selection

Default value drift | time | rel. drift | off | drift & time drift & time drift

The titration is stopped when the stop drift has been reached.

time

The titration is stopped if the endpoint has been exceeded during a certain time period ( Delay time ).

rel. drift

The titration is stopped when the sum of the drift at the start of the titration and the relative stop drift has been reached.

off

The titration will not be stopped until the stop conditions have been fulfilled.

drift & time

The titration is stopped when the stop drift has been exceeded during the time set and the endpoint has been reached.

Stop drift

This parameter can only be edited with Stop criterion = drift or drift & time .

The titration is stopped when the endpoint and the stop drift have been reached.

Input range

Default value

1 to 999 µg/min

15 µg/min

Delay time

This parameter can only be edited with Stop criterion = time or drift & time .

When the endpoint has been reached, the specified time is allowed to elapse after the last dosing and the titration is then stopped.

Input range

Default value

0 to 999 s

40 s

900 Touch Control ■■■■■■■■ 347

29.6 Bromine index determination (BRC) ■■■■■■■■■■■■■■■■■■■■■■

Relative stop drift

This parameter can only be edited with Stop criterion = rel. drift .

The titration is stopped when the endpoint and the sum of the drift at the start of the titration and the relative stop drift have been reached.

Input range

Default value

1 to 999 µg/min

5 µg/min

29.6.2 Titration parameters

Under [Titration parameters] , the parameters influencing the run of the entire titration are defined.

Pause

Waiting time, for example for dissolving the sample. During this time no bromine is generated.

Input range

Default value

0 to 999999 s

0 s

Extraction time

Minimum duration of the titration. The titration will not be stopped during the extraction time, even if the endpoint has already been reached. The titration is however stopped if a stop condition is fulfilled during this time

(see chapter 29.6.3, page 349)

.

Input range

Default value

0 to 999999 s

0 s

Temperature

Temperature entered manually. The temperature is being continuously measured when a temperature sensor is connected and when Temp.

measurement = automatic or continuous is defined (see sensor dialog of the command).

Input range

Default value

–20.0 to 150.0 °C

25.0 °C

Time interval MP

Time interval for entering a measuring point in the measuring point list.

The measuring point list is limited to 1000 measuring points.

Input range

Default value

0.1 to 999999.0 s

2.0 s

348 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

29.6.3 Stop conditions

The conditions for canceling the titration are defined under [Stop conditions] , if this does not occur automatically. This could be the case when the endpoint set is not reached or if the stop criterion

(see "Stop criterion", page 347) is not fulfilled.

Stop time

The titration is stopped when the specified time has elapsed since the start of the titration.

Input range

Selection

Default value

1 to 999999 s off off

29.6.4 Conditioning

The conditions required for conditioning are defined under [Conditioning] .

Conditioning on | off (Default value: on )

If this parameter is activated, then the first time the titration is started the working medium will be titrated to the endpoint with the specified control parameters. The status is kept stable. The actual method run does not begin until [ ] has been pressed once more.

Start drift

If the measured drift is smaller than this value for a certain time (so-called stabilizing time), Conditioning OK is displayed and the titration can be started. The stabilizing time is defined in the dialog Conditioning / Conditioning options .

Input range

Default value

1 to 999 µg/min

20 µg/min

Drift correction

The endpoint amount can be corrected taking the drift value into account.

For this, the drift is multiplied with the drift correction time and this value is then subtracted from the endpoint amount. The drift correction time is the time interval between the end of conditioning and the end of the determination.

Selection

Default value auto | manual | off off

900 Touch Control ■■■■■■■■ 349

29.6 Bromine index determination (BRC) ■■■■■■■■■■■■■■■■■■■■■■ auto

The value of the current drift is automatically applied at the start of the titration.

manual

If the drift is known throughout a longer period of time, this can be entered manually.

off

No drift correction takes place.

Drift value

This parameter can only be edited with Drift correction = manual .

Drift for manual drift correction.

Input range

Default value

0.0 to 999.9 µg/min

0.0 µg/min

Measured value display on | off (Default value: off )

If this parameter is activated, the currently measured value is displayed during the conditioning.

Conditioning options [Cond. options]

Stabilizing time

Waiting time, during which the measured drift has to be smaller than the start drift defined until Conditioning OK is displayed. The start drift is defined in the dialog Edit command / Conditioning .

Input range

Default value

0 to 999999 s

0 s

Cond. stop time

Maximum permissible time over which conditioning may take place. Conditioning is stopped when the specified time has elapsed.

Input range

Selection

Default value

1 to 999999 s off off

Delay reconditioning on | off (Default value: off )

If this parameter is activated, the conditioning will not start before all the method commands have been processed. Otherwise, the conditioning will immediately start after the titration command.

350 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

NOTICE

This parameter must be activated if commands follow which must not run simultaneously with the conditioning.

Example: Emptying the titration cell followed by adding new working medium.

29.6.5 Cell

The parameters for the titration cell are defined under [Cell] .

Generator electrode

Type of generator electrode.

Selection

Default value without diaphragm with diaphragm

| with diaphragm without diaphragm

Generator electrode without diaphragm.

with diaphragm

Generator electrode with diaphragm.

Generator current

Polarization current at the generator electrode.

Selection

Default value

100 mA auto

| 200 mA | 400 mA | auto auto

The current is adapted to the conductivity of the reagent and automatically reduced near the endpoint.

Reagent monitoring

Selection of the reagent from the list of reagents. The selection depends on the titration mode. Reagents are defined under System ▶ Reagents .

Selection

Default value

Selection of configured reagents off

| off off

The reagent monitoring is not active.

900 Touch Control ■■■■■■■■ 351

29.6 Bromine index determination (BRC) ■■■■■■■■■■■■■■■■■■■■■■

29.6.6 Control instrument

The control instrument the determination is carried out with is selected under [Control device] . Control instruments are defined under System ▶ Device manager .

NOTICE

This button is only displayed when several control instruments have been configured.

Control device

Selection of the control instrument from the list of devices. Only those devices are displayed which are able to carry out the command. The control instruments do not have to be connected. This allows to transfer methods more easily from one system to another.

Selection Selection of configured control instruments

29.6.7 Sensor

The parameters for the sensor are edited under [Sensor] .

Sensor

Selection of the sensor from the sensor list. The selection depends on the measuring mode. Sensors are defined under System ▶ Sensors . You can also enter a sensor name which is not contained in the sensor list. When a determination is started there is a check whether the sensor is contained in the sensor list.

Selection Selection of configured sensors

I(pol)

The polarization current is the current that is applied to a polarizable electrode during voltametric measurement.

The values that are actually adjustable may vary from the values specified here, depending on the hardware. The binding values can be found in the

"Technical specifications" chapter for the control instrument used.

Input range

Selection

Default value

–125.0 to 125.0 µA (Increment: 2.5)

–1.0 µA | 1.0 µA

1.0 µA

352 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Electrode check on | off (Default value: off )

For polarizable electrodes and the generator electrode, an electrode check can be carried out. The following will be tested:

Polarizable electrodes

– Is the electrode connected?

– Is there a short circuit?

The electrode check is carried out when this command is started.

Generator electrode:

– Is the electrode connected?

– Is the generator system defective?

– Is the resistance of the sample solution very high?

The electrode check is carried out continuously during the execution of this command.

Temp. measurement

Type of temperature measurement.

Selection

Default value continuous automatic

| automatic | off continuous

A temperature sensor must be connected. The temperature is measured continuously.

automatic

If a temperature sensor is connected then the temperature will be measured continuously. Otherwise, the temperature entered manually will be used (see dialog of the titration and measuring parameters).

off

The temperature will not be measured. The temperature entered manually is used (see dialog of the titration and measuring parameters).

29.6.8 Stirrer

The parameters for the stirrer are edited under [Stirrer] .

Stirrer

Selection of the MSB connector the stirrer is connected to. Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4 | off off

No stirrer will be used.

900 Touch Control ■■■■■■■■ 353

29.7 STAT titrations (STAT) ■■■■■■■■■■■■■■■■■■■■■■

Stirring rate

Setting the stirring rate. It can be set in steps of –15 to +15. The default setting 8 corresponds to 1000 rpm. The formula for calculating the rota-

tional speed is specified in chapter 32.4, page 494 . The optimum stirring

rate can be tested in the manual control.

The algebraic sign of the stirring rate changes the direction in which the stirring is done. When the stirrer is viewed from above, this means:

"+": counterclockwise rotation

"–": clockwise rotation

Input range

Default value

–15 to 15

8

Switch off automatically on | off (Default value: on )

If this parameter is activated, the stirrer will be switched off automatically at the end of the titration, measurement, etc.

29.7 STAT titrations (STAT)

29.7.1 Start conditions

The parameters that are carried out before the start of titration are defined under [Start conditions] .

Pause 1

Waiting time, e.g. for the electrode to settle down before a start volume is added.

Input range

Default value

0 to 999999 s

0 s

Start volume

Volume that is dosed prior to the start of the titration.

Input range

Default value

0.00000 to 9999.99 mL

0.00000 mL

Dosing rate

Rate at which the start volume is dosed.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum

354 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters maximum

The maximum dosing rate depends on the cylinder volume (see chapter 32, page 492) .

Pause 2

Waiting time, e.g. for the electrode to settle down after the start or a reaction time after the dosing of a start volume.

Input range

Default value

0 to 999999 s

0 s

Initial measured value

The [Initial meas. value] is measured before carrying out the start conditions . This value is used for the titration direction determination when the titration direction is set to auto .

Signal drift

Maximum permissible drift for the measured value acceptance, i.e. maximum change of the measured value per minute. For this value, a suitable maximum waiting time is automatically calculated except you already have modified this waiting time.

Input range

Selection

Default value

0.1 to 999.0 mV/min off off off

Measured value acceptance will take place after the maximum waiting time has elapsed.

Min. waiting time

The measured value is not accepted until the minimum waiting time has elapsed, even if the signal drift has already been reached. The minimum waiting time is only important for drift-controlled measurements.

Input range

Default value

0 to 999999 s

0 s

Max. waiting time

If the signal drift has been switched off or has not yet been reached, then the measured value will be accepted when the maximum waiting time has elapsed. As default value, a suitable waiting time for the signal drift is automatically calculated.

Input range

Default value

0 to 999999 s

1 s

900 Touch Control ■■■■■■■■ 355

29.7 STAT titrations (STAT) ■■■■■■■■■■■■■■■■■■■■■■

29.7.2 Control parameters

The parameters for the control point are defined under [Control parameters] .

Control point at

Measured value for the control point.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control instrument used.

Measuring mode pH:

Input range

Selection

Default value

–20.000 to 20.000 off off

Measuring mode U:

Input range

Selection

Default value

–2000.0 to 2000.0 mV off off

Titration rate

Three predefined sets of parameters can be selected for the titration rate.

Selection

Default value

50 µL/min |

100 µL/min

100 µL/min | 500 µL/min | user

50 µL/min

For reactions with an expected reaction rate in the range of 50 µL/min.

100 µL/min

For reactions with an expected reaction rate in the range of

100 µL/min.

500 µL/min

For reactions with an expected reaction rate in the range of

500 µL/min.

user

The individual titration parameters can be modified.

The settings of the individual titration rates are listed in table 15, page

358 .

User-defined parameters

These parameters are only accessible when Titration rate = user .

356 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Dynamics

29 Parameters

The control range defines the measured value range before the specified control point. This control point has a vital influence on the titration rate and thus also on the measuring accuracy.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control instrument used.

Measuring mode pH:

Input range

Default value

Selection

0.001 to 20.000

1.000 off

Measuring mode U:

Input range

Default value

Selection

0.1 to 2,000.0 mV

60.0 mV off

Max. rate

The maximum permissible rate at which dosing takes place.

Rule of thumb for the maximum rate in mL/min:

0.005 · expected reaction rate in µL/min

Input range

Default value

Selection

0.01 to 166.00 mL/min

0.75 mL/min maximum maximum

The maximum dosing rate depends on the cylinder volume (see chapter 32, page 492) .

Min. rate

The rate at which dosing is carried out at the very beginning of the titration and in the control range at the end of the titration. This parameter has a decisive influence on the behavior at the control point. The minimum rate also has a decisive influence on how exactly the control point can be kept constant. The smaller the selected minimum rate, the longer it takes until the control point is reached for the first time.

The following rule of thumb should be considered in order to obtain a dosing at the control point as constant (regular) as possible:

Min. rate in µL/min = 0.1 · expected reaction rate in µL/min

Input range

Default value

0.01 to 9999.00 µL/min

10.00 µL/min

900 Touch Control ■■■■■■■■ 357

29.7 STAT titrations (STAT) ■■■■■■■■■■■■■■■■■■■■■■

Table 15 Default values of the predefined titration rates for STAT

Reaction rate

50 µL/min 100 µL/min

Dynamics

– pH

– U

Max. rate

Min. rate

1.000

60.0 mV

0.25 mL/min

5.00 µL/min

1.000

60.0 mV

0.75 mL/min

10.00 µL/min

500 µL/min

Dosing and control range

Outside the control range the dosing rate is mainly characterized by the maximum rate. The parameters Max. rate and Dynamics should be optimized together in such a way that the titration does not overshoot too much when the control point is reached. The Dynamics should be selected in such a way that the measured value remains within the control range while the control point is maintained. A slightly larger control range should be defined for slow reactions (e.g. pH = 3, U = 180 mV). The setpoint value is often set with a SET pretitration before the substrate is added. This means that the STAT titration can be started already with a small deviation from the control point.

29.7.3 Titration parameters

Under [Titration parameters] , the parameters influencing the run of the entire titration are defined.

Titration direction

1.000

60.0 mV

2.00 mL/min

40.00 µL/min

With the setting "auto", the titration direction is determined automatically from the initial measured value and the set control point. It is recommended that you specify whether the change of the measured value is positive or negative.

Selection

Default value

+ | | auto auto

+

Positive measured value change by the addition of titrant, i.e. in the direction of a higher ph value or greater voltage.

-

Negative measured value change by the addition of titrant, i.e. in the direction of a lower pH value or lesser voltage.

auto

The titration direction is determined automatically from the initial measured value and the set control point.

358 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Temperature

29 Parameters

Temperature entered manually. The temperature is being continuously measured when a temperature sensor is connected and when Temp.

measurement = automatic or continuous is defined (see sensor dialog of the command). This value is used for temperature correction in pH measurements.

Input range

Default value

–20.0 to 150.0 °C

25.0 °C

Time interval MP

Time interval for entering a measuring point in the measuring point list.

The measuring point list is limited to 1000 measuring points.

Input range

Default value

0.1 to 999999.0 s

2.0 s

Start time

The measured values are not entered in the measuring point list until this time has elapsed since the start of the titration.

Input range

Default value

0 to 999999 s

0 s

Start meas. value

The measured values are not entered in the measuring point list until this start measured value has been reached.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control instrument used.

Measuring mode pH:

Input range

Selection

Default value

–20.000 to 20.000 off off

Measuring mode U:

Input range

Selection

Default value

–2000.0 to 2000.0 mV off off

Start rate

The measured values are not entered in the measuring point list until the dosing rate has fallen below this value. Checking this parameter is not started until 10 s after the start of the command.

900 Touch Control ■■■■■■■■ 359

29.7 STAT titrations (STAT)

Input range

Selection

Default value

NOTICE

0.01 to 166.00 mL/min off off

■■■■■■■■■■■■■■■■■■■■■■

If several of the parameters Start time , Start meas. value or Start rate have been defined, then all of them must be fulfilled for measured values to be entered in the measuring point list.

29.7.4 Stop conditions

The conditions for canceling the titration are defined under [Stop conditions] .

Stop volume

The titration is stopped when the specified volume has been dosed since the start of the titration. This volume should be adjusted to the size of the titration vessel in order to prevent the contents from running over.

Input range

Default value

Selection

0.00000 to 9999.99 mL

100,000 mL off

Stop time

The titration is stopped when the specified time has elapsed.

Input range

Selection

Default value

0 to 999999 s off off

Since

This parameter can only be edited when a stop time has been defined.

This parameter defines the start of the stop time.

Selection Start | Control point | Last dosing

Start

The stop time starts after the start volume has been dosed and the following pause time has elapsed.

Control point

The stop time starts as soon as the control point has been reached for the first time.

Last dosing

The stop time starts after the last increment has been dosed, i. e. the stop time is reset to zero again after each dosing step.

360 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Stop rate

The titration is stopped when the current titration rate is lower than the value entered. Checking this parameter is not started until 10 s after the start of the command.

Input range

Selection

Default value

0.01 to 166.00 µL/min off off

Filling rate

Rate at which the dosing cylinder is filled.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum filling rate depends on the cylinder volume

(see chapter

32, page 492) .

29.7.5 Monitoring

Under [Monitoring] , the parameters for monitoring the following quantities are defined:

Measured value

Dosing rate (an average rate)

Temperature

Monitoring is not used for parameterization of the control behavior.

Remote signals or RS-232 commands can be assigned to the monitored quantities

(see "Communication", page 365)

. These remote signals or

RS-232 commands are sent if a limit has been infringed. This can be used, for example, to switch a thermostat on/off if necessary.

The parameters are identical for all monitored quantities.

Monitoring on | off (Default value: off )

Enabling and disabling the monitoring function. The monitoring of the dosing rate is not started until 10 s after the start of the titration.

Lower limit

When the monitored quantity falls below this value, the action defined in the following is triggered.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges

900 Touch Control ■■■■■■■■ 361

29.7 STAT titrations (STAT) ■■■■■■■■■■■■■■■■■■■■■■ can be found in the "Technical specifications" chapter for the control instrument used.

Measuring mode pH:

Input range

Default value

–20.000 to 20.000

–20.000

Measuring mode U:

Input range

Default value

–2000.0 to 2000.0 mV

–2000.0 mV

Dosing rate:

Input range

Default value

Temperature:

Input range

Default value

0.00 to 166.00 mL/min

0.00 mL/min

–20.0 to 150.0 °C

–20.0 °C

Lower hysteresis

The lower hysteresis represents the tolerance range for the lower limit (see figure 18, page 365)

. When Action = Wait the dosing does not continue until the monitored quantity has once again exceeded the lower limit incl.

the value entered here.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control instrument used.

Measuring mode pH:

Input range

Default value

0.000 to 20.000

0.020

Measuring mode U:

Input range

Default value

0.0 to 2000.0 mV

2.0 mV

Dosing rate:

Input range

Default value

Temperature:

Input range

Default value

0.00 to 166.00 mL/min

0.20 mL/min

0.0 to 150.0 °C

0.2 °C

362 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Upper limit

29 Parameters

When this value is exceeded, the action defined in the following is triggered.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control instrument used.

Measuring mode pH:

Input range

Default value

–20.000 to 20.000

20.000

Measuring mode U:

Input range

Default value

–2000.0 to 2000.0 mV

2000.0 mV

Dosing rate:

Input range

Default value

Temperature:

Input range

Default value

0.00 to 166.00 mL/min

166.00 mL/min

–20.0 to 150.0 °C

150.0 °C

Upper hysteresis

The upper hysteresis represents the tolerance range for the upper limit

(see figure 18, page 365) . When

Action = Wait , the dosing does not continue until the monitored quantity has once again fallen below the upper limit incl. the value entered here.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control instrument used.

Measuring mode pH:

Input range

Default value

0.000 to 20.000

0.020

Measuring mode U:

Input range

Default value

0.0 to 2000.0 mV

2.0 mV

Dosing rate:

Input range

Default value

0.00 to 166.00 mL/min

0.20 mL/min

900 Touch Control ■■■■■■■■ 363

29.7 STAT titrations (STAT)

Action

■■■■■■■■■■■■■■■■■■■■■■

Temperature:

Input range

Default value

0.0 to 150.0 °C

0.2 °C

Selection of the action which is carried out when the limits of the monitored quantity are infringed.

Selection None | Exit method

Pause | Wait

| Cancel command |

None

No action will be taken.

Exit method

The method is stopped immediately.

Cancel command

The current command is stopped und the next command in the method will be started.

Pause

Adding reagent is stopped until the dosing is continued manually with

[Continue] . If several quantities are monitored, the other quantities will still be monitored in the background.

Wait

Adding reagent is stopped until the value of the monitored quantity is within the limits again (incl. hysteresis). The dosing is continued automatically. If several quantities are monitored, the other quantities will still be monitored in the background.

NOTICE

If the lower limit is infringed while monitoring the dosing rate, the settings Action = Pause and Action = Wait do not make sense because the average dosing rate becomes smaller and smaller during the waiting time. The valid range can never be reached again.

364 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Monitored quantity

Stop

Upper limit

29 Parameters

Lower limit

Time

Figure 16 Action "Exit method" or "Cancel command"

Monitored quantity

Pause

Upper limit

[Continue]

Figure 17 Action "Pause"

Monitored quantity

Wait

Upper hysteresis

Lower limit

Time

Upper limit

900 Touch Control

Lower hysteresis

Lower limit

Time

Figure 18 Action "Wait"

Communication

Under [Communication] , remote signals or RS-232 commands can be assigned to the monitored quantities. These remote signals or RS-232

■■■■■■■■ 365

29.7 STAT titrations (STAT) ■■■■■■■■■■■■■■■■■■■■■■ commands are always sent, independent of the action defined for a limit infringement

(see "Action", page 364)

.

[New]

Define a new assignment.

[Delete]

Delete the selected assignment.

[Edit]

Edit the settings for the selected assignment.

Monitored quantity

Selection of the monitored quantity to which a remote signal or an

RS-232 command is to be assigned.

Selection Measured value any

| Temperature | Dosing rate | any

If several quantities are monitored, the remote signal or RS-232 command set is sent as soon as a limit of one of the monitored quantities is exceeded.

Infringed limit

Indicating the limit whose infringement causes the remote signal or

RS-232 command defined in the following to be sent.

Selection upper | lower | any | back in range upper

The remote signal or the RS-232 command is sent when the upper limit is exceeded.

lower

The remote signal or the RS-232 command is sent when the lower limit is exceeded.

any

The remote signal or the RS-232 command is sent when the upper or the lower limit is infringed.

back in range

The remote signal or the RS-232 command is sent when the value of the monitored quantity is within the limits again (incl. hysteresis).

Interface

Selection of the Remote Box or the serial interface via which the remote signal or the RS-232 command is sent.

Selection Remote Box 1…4 | COM1…8

366 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Output signal

29 Parameters

This parameter can only be edited when a Remote Box is selected under

Interface .

Selection of the signal out of the templates or entering the required bit pattern. Templates are defined under System ▶ Templates ▶ Output lines .

NOTICE

A line set active is not being reset automatically, nor at the end of the determination nor when the monitored quantity is in the range again.

Entering a bit pattern:

0 = line inactive

1 = line active

* = retain line status p = set pulse

The output lines are always numbered from right to left, i.e. with the signal *************1 line 0 is set. With a pulse, the length is set to 200 ms.

If it is required to set pulses with other lengths, you have to define a corresponding template.

Entry

Default value

Selection

Bit patterns containing exactly 14 characters or a max. of 24 characters for the name of the template

**************

Selection of the templates defined

Character string

This parameter can only be edited when a serial interface (COM) is selected under Interface .

Input of the RS-232 command as character string. All characters of the

ASCII code page can be used. Control characters (Esc, FF etc.) have to be entered as a three-place, decimal ASCII code, starting with a slash. Each command is automatically terminated with the ASCII characters CR and

LF .

Entry

Default value

ASCII string with a max. of 24 characters empty

900 Touch Control ■■■■■■■■ 367

29.7 STAT titrations (STAT) ■■■■■■■■■■■■■■■■■■■■■■

29.7.6 Control instrument

The control instrument the determination is carried out with is selected under [Control device] . Control instruments are defined under System ▶ Device manager .

NOTICE

This button is only displayed when several control instruments have been configured.

Control device

Selection of the control instrument from the list of devices. Only those devices are displayed which are able to carry out the command. The control instruments do not have to be connected. This allows to transfer methods more easily from one system to another.

Selection Selection of configured control instruments

29.7.7 Sensor

The parameters for the sensor are edited under [Sensor] .

Measuring input

Selection of the measuring input the sensor is connected to. The selection is not dependent on whether the control device has one or two measuring interfaces.

Selection

Default value

1 | 2

1

Sensor

Selection of the sensor from the sensor list. The selection depends on the measuring mode. Sensors are defined under System ▶ Sensors . You can also enter a sensor name which is not contained in the sensor list. When a determination is started there is a check whether the sensor is contained in the sensor list.

Selection Selection of configured sensors

Electrode check

■ on | off (Default value: off )

For the following electrodes, an electrode check can be carried out: pH electrodes

Metal electrodes

Ion-selective electrodes

368 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

A check is made that the electrode is properly connected and that no short-circuit is present. The electrode check is carried out when this command is started. Note that this electrode check has nothing to do with the electrode test (command ELT ).

Temp. measurement

Type of temperature measurement.

Selection

Default value continuous automatic

| automatic | off continuous

A temperature sensor must be connected. The temperature is measured continuously.

automatic

If a temperature sensor is connected then the temperature will be measured continuously. Otherwise, the temperature entered manually will be used (see dialog of the titration and measuring parameters).

off

The temperature will not be measured. The temperature entered manually is used (see dialog of the titration and measuring parameters).

29.7.8 Dosing device

The parameters for the dosing device are edited under [Dosing device] .

Dosing device

Selection of the MSB connector to which the dosing device is connected.

Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4

Titrant

Selection of the titrant from the titrant list. We recommend always selecting the titrant. This ensures that the correct data (titer, concentration, etc.) is always used for the calculation. Titrants are defined under System ▶ Titrants . You can also enter a name which is not contained in the titrant list. When a determination is started there is a check whether the solution is contained in the list.

For buret units with integrated data chip, a check is made in the method run to verify whether the correct titrant has been attached and whether the type of dosing drive matches. For buret units without an integrated data chip, the cylinder volume and the type of dosing drive are checked.

For the selected titrant, the validity of the titer, the working life of the titrant and the time interval for the GLP test of the buret unit are checked at the start of the determination.

900 Touch Control ■■■■■■■■ 369

29.7 STAT titrations (STAT) ■■■■■■■■■■■■■■■■■■■■■■

Selection

Default value

Selection of configured titrants | not defined not defined not defined

No check takes place.

Tandem dosing

A second dosing device can be defined under [Tandem dosing] in order to enable uninterrupted dosing. Dosing is carried out with a combination of two dosing devices so that the second dosing device is dosing while the first one is being filled and vice versa.

Dosing device

Selection of the MSB connector to which the dosing device is connected.

Every MSB connector will always be displayed.

Selection

Default value

1 | off

2 | 3 | 4 | off

Titrant

Selection of the titrant from the titrant list. We recommend always selecting the titrant. Titrants are defined under System ▶ Titrants . You can also enter a name which is not contained in the titrant list. When a determination is started there is a check whether the solution is contained in list.

For buret units with integrated data chip, a check is made in the method run to verify whether the correct titrant has been attached and whether the type of dosing drive matches. For buret units without integrated data chip, the cylinder volume and the type of dosing drive are checked. For the selected titrant, only the time interval for the GLP test of the buret unit is checked at the start of the determination.

NOTICE

Filling rate

The concentration, validity of the titer and the working life of the titrant are ignored.

Selection

Default value

Selection of configured titrants | not defined not defined not defined

No check takes place.

Rate at which the dosing cylinder is filled.

Input range 0.01 to 166.00 mL/min

370 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Selection

Default value

■ maximum maximum maximum

The maximum filling rate depends on the cylinder volume

(see chapter

32, page 492) .

However, in order to ensure uninterrupted dosing, the following points must be taken into account:

Use the highest possible filling rate in order to keep the filling times as short as possible. The filling rate should be decreased for viscous liquids.

When you use two buret units with different dosing cylinders, the filling rate for the larger sized cylinder must be at the minimum:

= filling rate in mL/min for the larger sized cylinder

= filling rate in mL/min for the smaller cylinder

= cylinder volume in mL of the buret unit of the second dosing device

= cylinder volume in mL of the buret unit of the first dosing device

Example:

Dosing device 1: volume = 20 mL, filling rate = 50 mL/min

Dosing device 2: volume = 50 mL

Filling rate 2 ≥ 50 mL/min · 50 mL / 20 mL ≥ 125 mL/min

The dosing rate must not exceed 75% of the filling rate of the smaller cylinder. These values are listed in the following table, valid at maximum filling rate:

Table 16 Maximum dosing rate for different dosing cylinders

Cylinder volume maximum dosing rate

Exchange unit

2.25 mL/min

Dosing unit

– 1 mL

2 mL

5 mL

10 mL

11.25 mL/min

22.50 mL/min

5.00 mL/min

12.50 mL/min

25.00 mL/min

900 Touch Control ■■■■■■■■ 371

29.7 STAT titrations (STAT) ■■■■■■■■■■■■■■■■■■■■■■

Cylinder volume

20 mL

50 mL maximum dosing rate

Exchange unit Dosing unit

45.00 mL/min

112.50 mL/min

50.00 mL/min

124.50 mL/min

29.7.9 Stirrer

The parameters for the stirrer are edited under [Stirrer] .

Stirrer

Selection of the MSB connector the stirrer is connected to. Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4 | off off

No stirrer will be used.

Stirring rate

Setting the stirring rate. It can be set in steps of –15 to +15. The default setting 8 corresponds to 1000 rpm. The formula for calculating the rota-

tional speed is specified in chapter 32.4, page 494 . The optimum stirring

rate can be tested in the manual control.

The algebraic sign of the stirring rate changes the direction in which the stirring is done. When the stirrer is viewed from above, this means:

"+": counterclockwise rotation

"–": clockwise rotation

Input range

Default value

–15 to 15

8

Switch off automatically on | off (Default value: on )

If this parameter is activated, the stirrer will be switched off automatically at the end of the titration, measurement, etc.

372 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

29.8 Manual titrations (MAT)

A manual titration can be carried out with a Manual Dosing Controller or with the buttons [Dose] and [Cancel command] .

In order for the manual titration to be carried out with the 6.2107.120

Manual Dosing Controller, the latter must be connected to the MSB connector of the Titrando via the 6.2148.010 Remote Box MSB.

Manual titration using the Manual Dosing Controller

1 Press the [GO] key to start the method.

2 Press the [GO] key during the MAT command for dosing.

3 End the MAT command with the [FILL] key.

Manual titration using the buttons

1 Press the key to start the method.

2 Press the [Dose] button during the MAT command for dosing.

3 End the MAT command with the [Cancel command] button.

NOTICE

Start of titration : Because of the technical implementation involved, a few microliters are already dosed at the start of a manual titration. The possibility of deviation from the actual endpoint volume is excluded, however. The already dosed volume is taken into account.

End of titration : A manual titration must always be ended with the

[Cancel command] button (NOT with the [ ] fixed key) or with the

[FILL] key on the Manual Dosing Controller. This is the only way to ensure that subsequent method commands will be carried out.

900 Touch Control ■■■■■■■■ 373

29.8 Manual titrations (MAT) ■■■■■■■■■■■■■■■■■■■■■■

29.8.1 Dosing parameters

Under [Dosing parameters] , the parameters influencing the run of the manual titration are defined.

Dosing ramp

The dosing ramp is a delay in the dosing rate at the time of the start of a dosing step. It is advantageous primarily for manual titrations with indicator, when small volume steps are to be dosed before the equivalence point. The selectable delay time determines after which amount of time the defined dosing rate is to be reached.

Input range

Default value

0 to 10 s

2 s

Dosing rate

The rate at which dosing takes place.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum dosing rate depends on the cylinder volume

(see chapter 32, page 492) .

Time interval MP

Time interval for entering a measuring point in the measuring point list.

The measuring point list is limited to 1000 measuring points.

Input range

Default value

0.1 to 999999.0 s

5.0 s

Filling rate

Rate at which the dosing cylinder is filled.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum filling rate depends on the cylinder volume (see chapter

32, page 492) .

Temperature

Temperature entered manually. The temperature is being continuously measured when a temperature sensor is connected and when Temp.

measurement = automatic or continuous is defined (see sensor dialog

374 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters of the command). This value is used for temperature correction in pH measurements.

Input range

Default value

–20.0 to 150.0 °C

25.0 °C

29.8.2 Control instrument

The control instrument the determination is carried out with is selected under [Control device] . Control instruments are defined under System ▶ Device manager .

NOTICE

This button is only displayed when several control instruments have been configured.

Control device

Selection of the control instrument from the list of devices. Only those devices are displayed which are able to carry out the command. The control instruments do not have to be connected. This allows to transfer methods more easily from one system to another.

Selection Selection of configured control instruments

29.8.3 Sensor

The parameters for the sensor are edited under [Sensor] .

Measuring input

Selection of the measuring input the sensor is connected to. The selection is not dependent on whether the control device has one or two measuring interfaces.

Selection

Default value

1 | 2

1

Sensor

Selection of the sensor from the sensor list. The selection depends on the measuring mode. Sensors are defined under System ▶ Sensors . You can also enter a sensor name which is not contained in the sensor list. When a determination is started there is a check whether the sensor is contained in the sensor list.

Selection Selection of configured sensors

Temp. measurement

Type of temperature measurement.

900 Touch Control ■■■■■■■■ 375

29.8 Manual titrations (MAT) ■■■■■■■■■■■■■■■■■■■■■■

Selection

Default value continuous | automatic | off automatic continuous

A temperature sensor must be connected. The temperature is measured continuously.

automatic

If a temperature sensor is connected then the temperature will be measured continuously. Otherwise, the temperature entered manually will be used (see dialog of the titration and measuring parameters).

off

The temperature will not be measured. The temperature entered manually is used (see dialog of the titration and measuring parameters).

29.8.4 Dosing device

The parameters for the dosing device are edited under [Dosing device] .

Dosing device

Selection of the MSB connector to which the dosing device is connected.

Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4

Titrant

Selection of the titrant from the titrant list. We recommend always selecting the titrant. This ensures that the correct data (titer, concentration, etc.) is always used for the calculation. Titrants are defined under System ▶ Titrants . You can also enter a name which is not contained in the titrant list. When a determination is started there is a check whether the solution is contained in the list.

For buret units with integrated data chip, a check is made in the method run to verify whether the correct titrant has been attached and whether the type of dosing drive matches. For buret units without an integrated data chip, the cylinder volume and the type of dosing drive are checked.

For the selected titrant, the validity of the titer, the working life of the titrant and the time interval for the GLP test of the buret unit are checked at the start of the determination.

Selection

Default value

Selection of configured titrants | not defined not defined not defined

No check takes place.

376 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Tandem dosing

A second dosing device can be defined under [Tandem dosing] in order to enable uninterrupted dosing. Dosing is carried out with a combination of two dosing devices so that the second dosing device is dosing while the first one is being filled and vice versa.

Dosing device

Selection of the MSB connector to which the dosing device is connected.

Every MSB connector will always be displayed.

Selection

Default value

1 | off

2 | 3 | 4 | off

Titrant

Selection of the titrant from the titrant list. We recommend always selecting the titrant. Titrants are defined under System ▶ Titrants . You can also enter a name which is not contained in the titrant list. When a determination is started there is a check whether the solution is contained in list.

For buret units with integrated data chip, a check is made in the method run to verify whether the correct titrant has been attached and whether the type of dosing drive matches. For buret units without integrated data chip, the cylinder volume and the type of dosing drive are checked. For the selected titrant, only the time interval for the GLP test of the buret unit is checked at the start of the determination.

NOTICE

Filling rate

The concentration, validity of the titer and the working life of the titrant are ignored.

Selection

Default value

Selection of configured titrants | not defined not defined not defined

No check takes place.

Rate at which the dosing cylinder is filled.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum

900 Touch Control ■■■■■■■■ 377

29.8 Manual titrations (MAT) ■■■■■■■■■■■■■■■■■■■■■■ maximum

The maximum filling rate depends on the cylinder volume (see chapter

32, page 492) .

However, in order to ensure uninterrupted dosing, the following points must be taken into account:

Use the highest possible filling rate in order to keep the filling times as short as possible. The filling rate should be decreased for viscous liquids.

When you use two buret units with different dosing cylinders, the filling rate for the larger sized cylinder must be at the minimum:

= filling rate in mL/min for the larger sized cylinder

= filling rate in mL/min for the smaller cylinder

= cylinder volume in mL of the buret unit of the second dosing device

= cylinder volume in mL of the buret unit of the first dosing device

Example:

Dosing device 1: volume = 20 mL, filling rate = 50 mL/min

Dosing device 2: volume = 50 mL

Filling rate 2 ≥ 50 mL/min · 50 mL / 20 mL ≥ 125 mL/min

The dosing rate must not exceed 75% of the filling rate of the smaller cylinder. These values are listed in the following table, valid at maximum filling rate:

Table 17 Maximum dosing rate for different dosing cylinders

Cylinder volume maximum dosing rate

1 mL

2 mL

Exchange unit

2.25 mL/min

Dosing unit

5 mL

10 mL

11.25 mL/min

22.50 mL/min

5.00 mL/min

12.50 mL/min

25.00 mL/min

20 mL

50 mL

45.00 mL/min

112.50 mL/min

50.00 mL/min

124.50 mL/min

378 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

29.8.5 Stirrer

The parameters for the stirrer are edited under [Stirrer] .

Stirrer

Selection of the MSB connector the stirrer is connected to. Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4 | off off

No stirrer will be used.

Stirring rate

Setting the stirring rate. It can be set in steps of –15 to +15. The default setting 8 corresponds to 1000 rpm. The formula for calculating the rotational speed is specified in

chapter 32.4, page 494

. The optimum stirring rate can be tested in the manual control.

The algebraic sign of the stirring rate changes the direction in which the stirring is done. When the stirrer is viewed from above, this means:

"+": counterclockwise rotation

"–": clockwise rotation

Input range

Default value

–15 to 15

8

Switch off automatically on | off (Default value: on )

If this parameter is activated, the stirrer will be switched off automatically at the end of the titration, measurement, etc.

29.9 Measurements (MEAS)

29.9.1 Measuring parameters

Under [Measuring parameters] , the parameters influencing the run of the entire measurement are defined.

Signal drift

Maximum permissible drift for the measured value acceptance, i.e. maximum change of the measured value per minute. For this value, a suitable maximum waiting time is automatically calculated except you already have modified this waiting time.

900 Touch Control ■■■■■■■■ 379

29.9 Measurements (MEAS) ■■■■■■■■■■■■■■■■■■■■■■

NOTICE

A constant measured value is often only reached after a certain time, as mixing and the reaction itself require a certain time. The response time of an electrode can also increase with time, i.e., reaching a constant measured value takes longer and longer. Drift-controlled measured value acceptance is particularly advisable in such cases, as the measured values are only accepted when equilibrium has almost been reached.

Measuring mode pH, U, Ipol and Conc:

Input range

Default value

Selection

0.1 to 999.0 mV/min

10.0 mV/min off off

Measured value acceptance will take place after the maximum waiting time has elapsed.

Measuring mode Upol:

Input range

Default value

Selection

0.01 to 99.90 µA/min

10.00 µA/min off off

Measured value acceptance will take place after the maximum waiting time has elapsed.

Measuring mode T:

Input range

Default value

Selection

0.1 to 999.0 °C/min

0.5 °C/min off off

Measured value acceptance will take place after the maximum waiting time has elapsed.

Measuring mode Cond:

Input range

Selection

Default value

0.001 to 999.000 (mS/cm)/min off off off

Measured value acceptance will take place after the maximum waiting time has elapsed.

380 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Min. waiting time

29 Parameters

The measured value is not accepted until the minimum waiting time has elapsed, even if the signal drift has already been reached. The minimum waiting time is only important for drift-controlled measurements.

Input range

Default value

0 to 999999 s

0 s

Max. waiting time

If the signal drift has been switched off or has not yet been reached, then the measured value will be accepted when the maximum waiting time has elapsed. As default value, a suitable waiting time for the signal drift is automatically calculated.

Input range 0 to 999999 s

The default value depends on the measuring mode.

off Selection off

The measurement is continued endlessly.

Stop meas. value

The measurement is canceled when the specified measured value has been reached since the start of the measurement.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control instrument used.

Measuring mode pH:

Input range

Selection

Default value

–20.000 to 20.000 off off

Measuring mode U, Ipol:

Input range

Selection

Default value

–2000.0 to 2000.0 mV off off

Measuring mode Upol:

Input range –200.00 to 200.00 µA

Selection

Default value off off

Measuring mode T:

Input range –20.0 to 150.0 °C

900 Touch Control ■■■■■■■■ 381

29.9 Measurements (MEAS) ■■■■■■■■■■■■■■■■■■■■■■

Selection

Default value off off

Measuring mode Conc:

Input range 0.001 to 999999.999

The unit corresponds to the one of the standards with which the calibration of the sensor has been

Selection

Default value carried out.

off off

Measuring mode Cond:

Input range

Selection

Default value

0.0001 to 999.9999 mS/cm off off

Temperature

This parameter is not available with the MEAS T (temperature measurement) and MEAS Cond (conductivity measurement) commands.

Temperature entered manually. The temperature is being continuously measured when a temperature sensor is connected and when Temp.

measurement = automatic or continuous is defined (see sensor dialog of the command). This value is used for temperature correction in pH and concentration measurements.

Input range

Default value

–20.0 to 150.0 °C

25.0 °C

Time interval MP

Time interval for entering a measuring point in the measuring point list.

The measuring point list is limited to 1000 measuring points.

Input range

Default value

0.1 to 999999.0 s

2.0 s

382 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

NOTICE

The algorithm is only valid for direct measurements (MEAS Conc) if the following conditions are fulfilled:

The sample preparation must be identical for standard solutions and sample. In other words, if aqueous standard solutions were used for calibration, for example, then the sample must also be produced in aqueous medium.

The ratio sample volume / volume auxiliary solution must be identical with the ratio volume standard solution / volume auxiliary solution when calibrating the sensor.

Temperature compensation [Temp. compens.]

The parameters for converting the measured conductivity to a reference temperature are defined under [Temp. compens.] . This button is available only with the MEAS Cond command (conductivity measurement).

Measuring temp.

Temperature entered manually. The temperature is being continuously measured when a temperature sensor is connected and when Temp.

measurement = automatic or continuous is defined (see sensor dialog of the command).

Input range

Default value

–20.0 to 150.0 °C

25.0 °C

Reference temp.

The measured conductivity is converted to this temperature.

Input range

Default value

Selection

–20.0 to 150.0 °C

25.0 °C off off

The temperature compensation is disabled.

Temp. compensation

Type of temperature compensation.

Selection constant | ISO / DIN constant

The temperature compensation is carried out with the value entered for the temperature coefficient.

900 Touch Control ■■■■■■■■ 383

29.9 Measurements (MEAS) ■■■■■■■■■■■■■■■■■■■■■■

ISO / DIN

For ground water, spring water and surface water, the temperature coefficients according to standard ISO 7888:1985 (German version:

DIN EN 27888:1993) are stored in the system.

Temp. coefficient

This parameter can only be edited with Temp. compensation = constant .

Temperature coefficient belonging to the reference temperature entered above. With this temperature coefficient, the conductivity is converted to the reference temperature.

Input range

Default value

0.00 to 9.99 %/°C

0.00 %/°C

29.9.2 Standard addition

Under [Standard addition] , the parameters for the concentration measurement by means of standard addition are defined. This button is available only with the MEAS Conc command (concentration measurement).

Addition

Selection of the addition mode.

Selection

Default value off off

| auto | auto dos | manual off

The concentration measurement is carried out without standard addition, i.e. as a direct measurement.

auto

The standard solution is dosed automatically. The individual volume steps are calculated in the way that a certain voltage difference is reached. This voltage difference ( Delta U ) is defined under Standard addition ▶ Dosing device .

auto dos

The standard solution is dosed automatically. The individual volume steps are defined under Standard addition ▶ Addition increments .

manual

The standard solution is dosed manually. The individual volume steps are defined under Standard addition ▶ Addition increments .

Concentration standard

Concentration of the standard solution used. When Addition = manual is defined, a numerical value has to be entered here, the selection from list creates an error message when starting the determination.

384 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Input range

Selection

Default value

0.00001 to 99999.9 from list from list from list

The standard solution selected under Standard addition ▶ Dosing device is used.

Unit of the standard solution used.

Entry

Selection

Default value max. 24 characters mol/L | mmol/L | ppm | % | g/L | mg/L | µg/L ppm

Number of additions

Number of standard additions. The more addition steps you carry out the more accurate the evaluation becomes.

Input range

Default value

1 to 19

3

Stop volume

Maximum permissible volume for the volume steps. If the sum of the volume steps is higher than this volume, an error message is displayed when starting the determination.

Input range

Default value

0.00 to 9999.90 mL

99.99 mL

Volume aux. solution

Volume of the buffer solution etc. which has to be added to the sample before the measurement can be started.

Input range

Default value

0.00 to 9999.90 mL

0.00 mL

Stirrer control

The parameters for the stirrer control during the standard addition are edited under [Stirrer control] . Ensure that a stirrer has been selected in the Edit command / Stirrer dialog.

Stir solution during measurement on | off (Default value: on )

If this parameter is activated, the stirrer will be switched on and off automatically at the start or end of the measurement.

900 Touch Control ■■■■■■■■ 385

29.9 Measurements (MEAS) ■■■■■■■■■■■■■■■■■■■■■■

Stir before meas.

Pause before meas.

This parameter can only be edited when the parameter Stir solution during measurement is deactivated.

If it is stirred before the measurement, a waiting time can be defined here before starting the measurement. No stirring or measuring occurs during this time.

Input range

Default value

0 to 999999 s

0 s

Dosing device

The parameters for the dosing device are edited under [Dosing device] .

This button is enabled only with Addition = auto or auto dos .

Dosing device

Selection of the MSB connector to which the dosing device is connected.

Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4

Standard

This parameter can only be edited when the parameter Stir solution during measurement is deactivated.

If the stirrer is switched off during the measurement, it will be stirred for the time entered before measuring.

Input range

Default value

0 to 999999 s

0 s

Selection of the standard solution from the titrant list. We recommend always selecting the standard solution. This ensures that the correct data

(titer, concentration, etc.) is always used for the calculation. Standard solutions are defined under System ▶ Titrants . You can also enter a name which is not contained in the titrant list. When a determination is started there is a check whether the solution is contained in list.

For buret units with integrated data chip, a check is made in the method run to verify whether the correct standard solution has been attached and whether the type of dosing drive matches. For buret units without integrated data chip, the cylinder volume and the type of dosing drive are checked. For the selected standard solution, the validity of the titer, the working life of the standard solution and the time interval for the GLP test of the buret unit are checked at the start of the determination.

386 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Selection

Default value

Selection of configured titrants | not defined not defined not defined

No check takes place.

Dosing rate

Three predefined sets of parameters can be selected for the dosing rate.

The maximum dosing rate depends on the cylinder volume

(see chapter

32, page 492)

.

Selection

Default value slow | medium | fast fast slow

The dosing rate is at 10 % of the maximum dosing rate.

medium

The dosing rate is at 50 % of the maximum dosing rate.

fast

The dosing rate is equivalent to the maximum dosing rate.

Delta U

This parameter can only be edited with Addition = auto .

The individual volume steps are calculated in the way that the voltage difference entered is reached.

Input range

Default value

1 to 999 mV

10 mV

Addition increments

The volume of the individual volume steps is defined under [Addition increments] . This button is enabled only with Addition = auto dos or manual .

Volume

Volume of a single volume step.

Input range

Default value

00.000 to 9999.900 mL

00.100 mL

Apply this value for all increments on | off (Default value: off )

If you activate this parameter, the entered volume is applied for all volume steps.

900 Touch Control ■■■■■■■■ 387

29.9 Measurements (MEAS) ■■■■■■■■■■■■■■■■■■■■■■

29.9.3 Control instrument

The control instrument the determination is carried out with is selected under [Control device] . Control instruments are defined under System ▶ Device manager .

NOTICE

This button is only displayed when several control instruments have been configured.

Control device

Selection of the control instrument from the list of devices. Only those devices are displayed which are able to carry out the command. The control instruments do not have to be connected. This allows to transfer methods more easily from one system to another.

Selection Selection of configured control instruments

29.9.4 Sensor

The parameters for the sensor are edited under [Sensor] .

The parameters are valid for all measuring modes except for Temp. measurement . This parameter is not available with the MEAS T command

(temperature measurement).

Measuring input

Selection of the measuring input the sensor is connected to. The selection is not dependent on whether the control device has one or two measuring interfaces.

Selection

Default value

1 | 2

1

Sensor

Selection of the sensor from the sensor list. The selection depends on the measuring mode. Sensors are defined under System ▶ Sensors . You can also enter a sensor name which is not contained in the sensor list. When a determination is started there is a check whether the sensor is contained in the sensor list.

Selection Selection of configured sensors

388 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

I(pol)

The polarization current is the current that is applied to a polarizable electrode during voltametric measurement. This parameter is available only with I(pol) determinations.

The values that are actually adjustable may vary from the values specified here, depending on the hardware. The binding values can be found in the

Technical specifications chapter for the control instrument used.

Input range

Default value

Selection

–125.0 to 125.0 µA (Increment: 2.5)

5.0 µA

–1.0 µA | 1.0 µA

U(pol)

The polarization voltage is the voltage applied to the polarizable electrode during an amperometric measurement. This parameter is available only with U(pol) determinations.

The values that are actually adjustable may vary from the values specified here, depending on the hardware. The binding values can be found in the

Technical specifications chapter for the control instrument used.

Input range

Default value

–1,250 to 1,250 mV (Increment: 25)

400 mV

Electrode check on | off (Default value: off )

For the following electrodes, an electrode check can be carried out:

■ pH electrodes

Metal electrodes

Ion-selective electrodes

A check is made that the electrode is properly connected and that no short-circuit is present. The electrode check is carried out when this command is started. Note that this electrode check has nothing to do with the electrode test (command ELT ).

Temp. measurement

Type of temperature measurement.

Selection

Default value continuous automatic

| automatic | off continuous

A temperature sensor must be connected. The temperature is measured continuously.

900 Touch Control ■■■■■■■■ 389

29.9 Measurements (MEAS) ■■■■■■■■■■■■■■■■■■■■■■ automatic

If a temperature sensor is connected then the temperature will be measured continuously. Otherwise, the temperature entered manually will be used (see dialog of the titration and measuring parameters).

off

The temperature will not be measured. The temperature entered manually is used (see dialog of the titration and measuring parameters).

29.9.5 Stirrer

The parameters for the stirrer are edited under [Stirrer] .

Stirrer

Selection of the MSB connector the stirrer is connected to. Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4 | off off

No stirrer will be used.

Stirring rate

Setting the stirring rate. It can be set in steps of –15 to +15. The default setting 8 corresponds to 1000 rpm. The formula for calculating the rota-

tional speed is specified in chapter 32.4, page 494 . The optimum stirring

rate can be tested in the manual control.

The algebraic sign of the stirring rate changes the direction in which the stirring is done. When the stirrer is viewed from above, this means:

"+": counterclockwise rotation

"–": clockwise rotation

Input range

Default value

–15 to 15

8

Switch off automatically on | off (Default value: on )

If this parameter is activated, the stirrer will be switched off automatically at the end of the titration, measurement, etc.

390 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

29.10 Calibration of sensors (CAL)

29.10.1 Calibration parameters (CAL pH)

Under [Calibration parameters] , the parameters influencing the run of the entire calibration are defined.

Signal drift

Maximum permissible drift for the measured value acceptance, i.e. maximum change of the measured value per minute. For this value, a suitable maximum waiting time is automatically calculated except you already have modified this waiting time.

Input range

Default value

Selection

0.1 to 999.0 mV/min

2.0 mV/min off off

Measured value acceptance will take place after the maximum waiting time has elapsed.

Min. waiting time

The measured value is not accepted until the minimum waiting time has elapsed, even if the signal drift has already been reached. The minimum waiting time is only important for drift-controlled measurements.

Input range

Default value

0 to 999999 s

10 s

Max. waiting time

If the signal drift has been switched off or has not yet been reached, then the measured value will be accepted when the maximum waiting time has elapsed. As default value, a suitable waiting time for the signal drift is automatically calculated.

Input range

Default value

Selection

0 to 999999 s

110 s off

Temperature

Temperature entered manually. The temperature is being continuously measured when a temperature sensor is connected and when Temp.

measurement = automatic or continuous is defined (see sensor dialog of the command).

Input range

Default value

–20.0 to 150.0 °C

25.0 °C

900 Touch Control ■■■■■■■■ 391

29.10 Calibration of sensors (CAL)

Sample Processor

■■■■■■■■■■■■■■■■■■■■■■

If the calibration is carried out with a Sample Processor, the buffers or standards are automatically changed. This is why no request for the calibration temperature appears when starting the calibration. The value entered above is used.

Selection

Default value off | Remote | USB off off

For calibrations without Sample Processor. During the calibration, a request for changing the buffer/standard is displayed.

Remote

For automatic calibrations with Sample Processors, which are connected via the Remote Box.

USB

For automatic calibrations with a Sample Processor that is connected

via USB (see "Definition of the calibration positions", page 498)

.

Buffers

The buffer type and the number of buffers is defined under [Buffers] .

Buffer type

Selection of a predefined buffer series or definition of special buffers. In the case of predefined buffer series, the instrument automatically recognizes which buffer is involved.

Selection

Default value

Metrohm | NIST | DIN | Fisher | Fluka Basel |

Mettler | Merck Tit.

| Beckman | Radiometer |

Custom | Special | Merck Cer.

| Baker | Hamilton | Precisa

Metrohm

Custom

Definition of a custom buffer series. The buffer series is defined under

System ▶ Templates ▶ Custom calib. buffers .

Special

Up to five calibration buffers can be defined in the method. The automatic buffer detection is not activated in this case. The buffers must be measured precisely in the specified sequence.

Merck Cer.

Reference temperature = 25 °C. When using Merck CertiPUR buffers

(20 °C) the buffer type Merck Tit.

must be selected.

Number of buffers

This parameter is not visible when Buffer type = Special is defined.

392 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Number of buffers that are used for calibration. If calibration is accomplished with more than two buffers, then they can be used repeatedly in order to give them more statistical weight. The first two buffers must however always be different from one another.

Selection

Default value

1 | 2 | 3 | 4 | 5

2

Buffer 1 pH

This parameter is visible only when Buffer type = Special .

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control device used.

Input range

Default value

–20.000 to 20.000

7.000

Buffer 2 pH

This parameter is visible only when Buffer type = Special .

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control device used.

Input range

Default value

Selection

–20.000 to 20.000

4.000 off

Buffer 3 pH

This parameter is visible only when Buffer type = Special .

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control device used.

Input range

Selection

Default value

–20.000 to 20.000 off off

Buffer 4 pH

See Buffer 3 pH .

Buffer 5 pH

See Buffer 3 pH .

900 Touch Control ■■■■■■■■ 393

29.10 Calibration of sensors (CAL) ■■■■■■■■■■■■■■■■■■■■■■

Stirrer control

The parameters for the stirrer control during the calibration are edited under [Stirrer control] . Ensure that a stirrer has been selected in the Edit command / Stirrer dialog.

NOTICE

Deactivate the Stir solution during measurement parameter when you carry out the calibration with a Sample Processor. The stirrer control for the tower stirrer is ineffective with these instruments.

Stir solution during measurement on | off (Default value: on )

If this parameter is activated, the stirrer will be switched on and off automatically at the start or end of the measurement.

Stir before meas.

This parameter can only be edited when the parameter Stir solution during measurement is deactivated.

If the stirrer is switched off during the measurement, it will be stirred for the time entered before measuring.

Input range

Default value

0 to 999999 s

0 s

Pause before meas.

This parameter can only be edited when the parameter Stir solution during measurement is deactivated.

If it is stirred before the measurement, a waiting time can be defined here before starting the measurement. No stirring or measuring occurs during this time.

Input range

Default value

0 to 999999 s

0 s

29.10.2 Calibration parameters (CAL Conc)

Under [Calibration parameters] , the parameters influencing the run of the entire calibration are defined.

Signal drift

Maximum permissible drift for the measured value acceptance, i.e. maximum change of the measured value per minute. For this value, a suitable

394 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters maximum waiting time is automatically calculated except you already have modified this waiting time.

Input range

Default value

Selection

0.1 to 999.0 mV/min

2.0 mV/min off off

Measured value acceptance will take place after the maximum waiting time has elapsed.

Min. waiting time

The measured value is not accepted until the minimum waiting time has elapsed, even if the signal drift has already been reached. The minimum waiting time is only important for drift-controlled measurements.

Input range

Default value

0 to 999999 s

10 s

Max. waiting time

If the signal drift has been switched off or has not yet been reached, then the measured value will be accepted when the maximum waiting time has elapsed. As default value, a suitable waiting time for the signal drift is automatically calculated.

Input range

Default value

Selection

0 to 999999 s

110 s off

Temperature

Temperature entered manually. The temperature is being continuously measured when a temperature sensor is connected and when Temp.

measurement = automatic or continuous is defined (see sensor dialog of the command).

Input range

Default value

–20.0 to 150.0 °C

25.0 °C

Sample Processor

If the calibration is carried out with a Sample Processor, the buffers or standards are automatically changed. This is why no request for the calibration temperature appears when starting the calibration. The value entered above is used.

Selection

Default value off | Remote | USB off off

For calibrations without Sample Processor. During the calibration, a request for changing the buffer/standard is displayed.

900 Touch Control ■■■■■■■■ 395

29.10 Calibration of sensors (CAL) ■■■■■■■■■■■■■■■■■■■■■■

Remote

For automatic calibrations with Sample Processors, which are connected via the Remote Box.

USB

For automatic calibrations with a Sample Processor that is connected

via USB (see "Definition of the calibration positions", page 498)

.

Standards

The concentrations of the standards as well as their units are defined under [Standards] .

Unit

Unit of the standard solutions used. For concentration measurements without standard addition, this unit will be used.

Selection

Default value mol/L ppm

| mmol/L | ppm | % | g/L | mg/L | µg/L

Concentration standard 1

Concentration of the standard solution used.

Input range

Default value

00.001 to 999999.999

1,000

Concentration standard 2

Concentration of the standard solution used.

Input range

Selection

Default value

00.001 to 999999.999 off off

Concentration standard 3

See Concentration standard 2 .

Concentration standard 4

See Concentration standard 2 .

Concentration standard 5

See Concentration standard 2 .

Stirrer control

The parameters for the stirrer control during the calibration are edited under [Stirrer control] . Ensure that a stirrer has been selected in the Edit command / Stirrer dialog.

396 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

NOTICE

Deactivate the Stir solution during measurement parameter when you carry out the calibration with an 814 USB Sample Processor or with an 815 Robotic USB Sample Processor XL. The stirrer control for the tower stirrer is ineffective with these instruments.

Stir solution during measurement on | off (Default value: on )

If this parameter is activated, the stirrer will be switched on and off automatically at the start or end of the measurement.

Stir before meas.

This parameter can only be edited when the parameter Stir solution during measurement is deactivated.

If the stirrer is switched off during the measurement, it will be stirred for the time entered before measuring.

Input range

Default value

0 to 999999 s

0 s

Pause before meas.

This parameter can only be edited when the parameter Stir solution during measurement is deactivated.

If it is stirred before the measurement, a waiting time can be defined here before starting the measurement. No stirring or measuring occurs during this time.

Input range

Default value

0 to 999999 s

0 s

29.10.3 Calibration parameters (CAL Cond)

Under [Calibration parameters] , the parameters influencing the run of the entire calibration are defined.

Signal drift

Maximum permissible drift for the measured value acceptance, i.e. maximum change of the measured value per minute.

Input range

Selection

Default value

00.001 to 999,000 (mS/cm)/min off off

900 Touch Control ■■■■■■■■ 397

29.10 Calibration of sensors (CAL) ■■■■■■■■■■■■■■■■■■■■■■ off

Measured value acceptance will take place after the maximum waiting time has elapsed.

Min. waiting time

The measured value is not accepted until the minimum waiting time has elapsed, even if the signal drift has already been reached. The minimum waiting time is only important for drift-controlled measurements.

Input range

Default value

0 to 999999 s

0 s

Max. waiting time

If the signal drift has been switched off or has not yet been reached, then the measured value will be accepted when the maximum waiting time has elapsed.

Input range

Default value

Selection

0 to 999999 s

10 s off off

The measurement is continued endlessly.

Measuring temp.

Temperature entered manually. The temperature is being continuously measured when a temperature sensor is connected and when Temp.

measurement = automatic or continuous is defined (see sensor dialog of the command).

Input range

Default value

–20.0 to 150.0 °C

25.0 °C

Standard

The characteristics of the calibration standard used are defined under

[Standard] . These values will be suggested in the request after the start of the calibration.

Request at calibration start on | off (Default value: off )

When this parameter is activated, then entering the following parameters will be requested after the start.

Conductivity standard

Conductivity of the calibration standard used.

Input range

Default value

0.00000 to 2000.00 mS/cm

12.8800 mS/cm

398 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Reference temp. standard

Reference temperature at which the calibration standard has the conductivity entered above.

Input range

Default value

–20.0 to 150.0 °C

25.0 °C

Temp. coefficient

Temperature coefficient belonging to the reference temperature entered above. With this temperature coefficient, the conductivity is converted to the reference temperature.

Input range

Default value

0.00 to 9.99 %/°C

1.90 %/°C

Stir solution during measurement on | off (Default value: on )

If this parameter is activated, the stirrer will be switched on and off automatically at the start or end of the measurement.

Stir before meas.

Stirrer control

The parameters for the stirrer control during the calibration are edited under [Stirrer control] . Ensure that a stirrer has been selected in the Edit command / Stirrer dialog.

This parameter can only be edited when the parameter Stir solution during measurement is deactivated.

If the stirrer is switched off during the measurement, it will be stirred for the time entered before measuring.

Input range

Default value

0 to 999999 s

0 s

Pause before meas.

This parameter can only be edited when the parameter Stir solution during measurement is deactivated.

If it is stirred before the measurement, a waiting time can be defined here before starting the measurement. No stirring or measuring occurs during this time.

Input range

Default value

0 to 999999 s

0 s

900 Touch Control ■■■■■■■■ 399

29.10 Calibration of sensors (CAL) ■■■■■■■■■■■■■■■■■■■■■■

29.10.4 Control instrument

The control instrument the determination is carried out with is selected under [Control device] . Control instruments are defined under System ▶ Device manager .

NOTICE

This button is only displayed when several control instruments have been configured.

Control device

Selection of the control instrument from the list of devices. Only those devices are displayed which are able to carry out the command. The control instruments do not have to be connected. This allows to transfer methods more easily from one system to another.

Selection Selection of configured control instruments

29.10.5 Sensor

The parameters for the sensor are edited under [Sensor] .

Measuring input

Selection of the measuring input the sensor is connected to. The selection is not dependent on whether the control device has one or two measuring interfaces.

Selection

Default value

1 | 2

1

Sensor

Selection of the sensor from the sensor list. The selection depends on the measuring mode. Sensors are defined under System ▶ Sensors . You can also enter a sensor name which is not contained in the sensor list. When a determination is started there is a check whether the sensor is contained in the sensor list.

Selection Selection of configured sensors

Electrode check on | off (Default value: off )

This parameter is available only for the command CAL pH .

For the following electrodes, an electrode check can be carried out:

■ pH electrodes

400 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Metal electrodes

Ion-selective electrodes

A check is made that the electrode is properly connected and that no short-circuit is present. The electrode check is carried out when this command is started. Note that this electrode check has nothing to do with the electrode test (command ELT ).

Temp. measurement

Type of temperature measurement.

Selection

Default value continuous automatic

| automatic | off continuous

A temperature sensor must be connected. The temperature is measured continuously.

automatic

If a temperature sensor is connected then the temperature will be measured continuously. Otherwise, the calibration temperature entered manually will be used.

off

The temperature will not be measured. The calibration temperature entered manually is used. The temperature is requested after the start, except the determination is carried out with a Sample Processor.

29.10.6 Stirrer

The parameters for the stirrer are edited under [Stirrer] .

Stirrer

Selection of the MSB connector the stirrer is connected to. Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4 | off off

No stirrer will be used.

Stirring rate

Setting the stirring rate. It can be set in steps of –15 to +15. The default setting 8 corresponds to 1000 rpm. The formula for calculating the rotational speed is specified in

chapter 32.4, page 494

. The optimum stirring rate can be tested in the manual control.

The algebraic sign of the stirring rate changes the direction in which the stirring is done. When the stirrer is viewed from above, this means:

"+": counterclockwise rotation

900 Touch Control ■■■■■■■■ 401

29.11 Electrode test for pH electrodes (ELT)

"–": clockwise rotation

Input range

Default value

–15 to 15

8

■■■■■■■■■■■■■■■■■■■■■■

29.11 Electrode test for pH electrodes (ELT)

An automatic electrode test can be carried out for pH electrodes which will permit a qualification of the electrode used. Differentiated specifications regarding characteristic measurement properties of your pH electrode (slope, response time, etc.) will be obtained as the result. The concluding evaluation of the electrode is carried out by testing these results against the specified limit values. Once the electrode test has been successfully completed, the calibration data is stored with the sensor data of the sensor used.

These limit values are saved in the software for the following types of pH electrodes:

Standard electrodes

Gel electrodes

Electrodes with a nonaqueous reference electrolyte

You can, however, also define limits of your own choosing

(see chapter

15.7, page 163) . The precise classification of the measuring chain is deter-

mined by the reference system used. You will find an overview of these

electrode types and the corresponding limit values in table 18, page 406 .

Faulty electrodes often exhibit certain combinations of measurement results of the electrode test which fall outside the intended limits. This makes it possible to have more precise data concerning possible causes of errors and recommended remedial measures.

Buffers

You will require flawless buffers in the pH values 4, 7 and 9 for carrying out the electrode test. We recommend for this purpose the appropriate

Metrohm buffers. The electrode test can, however, in principle also be carried out with any of the buffer sets contained in the memory.

The buffers must be measured in the following sequence:

1.

pH 9

2.

pH 4

3.

pH 7

402 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

NOTICE

Ensure in particular the flawlessness of the pH 9 buffer. It can readily deviate from its specified pH value as the result of absorption of CO

2 from the ambient atmosphere, thus leading to incorrect test results.

Stirrer

A stirrer has to be connected to the titrator for the electrode test.

Temperature

Take care to ensure that the entire electrode test is carried out at a consistently uniform temperature. Work should be accomplished at room temperature wherever possible, in light of the fact that temperature exercises a great influence on response time. The preset limit values are calibrated to 25 °C.

29.11.1 Electrode test parameters

Under [Electrode test param.] , the parameters influencing the sequence of the electrode test are defined.

Buffer type

Selecting a predefined buffer series.

Selection

Default value

Metrohm | NIST | DIN

Merck Tit.

| Radiometer | Merck Cer.

| Baker |

Hamilton | Precisa

Metrohm

| Fluka Basel | Mettler |

Merck Cer.

Reference temperature = 25 °C. When using Merck CertiPUR buffers

(20 °C) the buffer type Merck Tit.

must be selected.

Sample Processor

If the calibration is carried out with a Sample Processor, the buffers or standards are automatically changed. This is why no request for the calibration temperature appears when starting the calibration. The value entered above is used.

Selection

Default value off | Remote | USB off off

For calibrations without Sample Processor. During the calibration, a request for changing the buffer/standard is displayed.

900 Touch Control ■■■■■■■■ 403

29.11 Electrode test for pH electrodes (ELT) ■■■■■■■■■■■■■■■■■■■■■■

Remote

For automatic calibrations with Sample Processors, which are connected via the Remote Box.

USB

For automatic calibrations with a Sample Processor that is connected

via USB (see "Definition of the calibration positions", page 498)

.

Electrode type

Selection of the electrode type.

Selection

Default value

Standard

Standard

| Gel | Non-aqueous | Custom

Standard

Electrodes which contain e.g. KCl solution as reference electrolyte.

Gel

Electrodes which contain Idrolyte as reference electrolyte.

Non-aqueous

Electrodes, which contain a non-aqueous reference electrolyte, e.g.

TEABr in ethylene glycol or LiCl in ethanol.

Custom

Under System ▶ Templates a custom electrode type can be defined.

Temperature

Temperature entered manually. The temperature is being continuously measured when a temperature sensor is connected and when Temp.

measurement = automatic or continuous is defined (see sensor dialog of the command).

Input range

Default value

–20.0 to 150.0 °C

25.0 °C

29.11.2 Control instrument

The control instrument the determination is carried out with is selected under [Control device] . Control instruments are defined under System ▶ Device manager .

NOTICE

This button is only displayed when several control instruments have been configured.

Control device

Selection of the control instrument from the list of devices. Only those devices are displayed which are able to carry out the command. The con-

404 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters trol instruments do not have to be connected. This allows to transfer methods more easily from one system to another.

Selection Selection of configured control instruments

29.11.3 Sensor

The parameters for the sensor are edited under [Sensor] .

Measuring input

Selection of the measuring input the sensor is connected to. The selection is not dependent on whether the control device has one or two measuring interfaces.

Selection

Default value

1 | 2

1

Sensor

Selection of the sensor from the sensor list. The selection depends on the measuring mode. Sensors are defined under System ▶ Sensors . You can also enter a sensor name which is not contained in the sensor list. When a determination is started there is a check whether the sensor is contained in the sensor list.

Selection Selection of configured sensors

Temp. measurement

Type of temperature measurement.

Selection

Default value continuous automatic

| automatic | off continuous

A temperature sensor must be connected. The temperature is measured continuously.

automatic

If a temperature sensor is connected then the temperature will be measured continuously. Otherwise, the calibration temperature entered manually will be used.

off

The temperature will not be measured. The calibration temperature entered manually is used. The temperature is requested after the start, except the determination is carried out with a Sample Processor.

900 Touch Control ■■■■■■■■ 405

29.11 Electrode test for pH electrodes (ELT) ■■■■■■■■■■■■■■■■■■■■■■

29.11.4 Stirrer

The parameters for the stirrer are edited under [Stirrer] .

Stirrer

Selection of the MSB connector the stirrer is connected to. Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4

Stirring rate

Setting the stirring rate. It can be set in steps of –15 to +15. The default setting 8 corresponds to 1000 rpm. The formula for calculating the rota-

tional speed is specified in chapter 32.4, page 494 . The optimum stirring

rate can be tested in the manual control.

The algebraic sign of the stirring rate changes the direction in which the stirring is done. When the stirrer is viewed from above, this means:

"+": counterclockwise rotation

"–": clockwise rotation

Input range

Default value

–15 to 15

8

29.11.5 Additional information

Table 18 Limit values of the three electrode types

Excellent electrode

Rating

Good electrode

Electrode type "Standard"

Streaming potential

Drift

Min. slope

Max. slope

Response time

Lower limit Uoff

Upper limit Uoff

Electrode type "Gel"

Streaming potential

Drift

Min. slope

2.5 mV

2.0 mV/s

96.5%

101.0%

45 s

3.0 mV

2.5 mV/s

96.5%

3.0 mV

2.5 mV/s

96.0%

102.0%

50 s

3.5 mV

3.0 mV/s

96.0%

–15 mV

15 mV

Usable electrode

4.0 mV

3.0 mV/s

95.0%

103.0%

60 s

4.5 mV

4.0 mV/s

95.0%

406 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Max. slope

Response time

Lower limit Uoff

Upper limit Uoff

Electrode type "Non-aqueous"

Excellent electrode

101.0%

60 s

Streaming potential

Drift

Min. slope

Max. slope

Response time

Lower limit Uoff

Upper limit Uoff

3.0 mV

5.0 mV/s

88.0%

120.0%

60 s

Table 19 Possible errors during the electrode test

Test criterion

Buffer pH 9:

The drift in stirred solution is > 1 mV.

Message

General problem

Rating

Good electrode

102.0%

75 s

–20 mV

20 mV

4.5 mV

7.0 mV/s

80.0%

130.0%

75 s

–50 mV

100 mV

Usable electrode

103.0%

90 s

6.0 mV

9.0 mV/s

70.0%

140.0%

90 s

Remedy

Connect an electrode.

Replace the defective electrode cable.

Clean the diaphragm (see leaflet for the electrode).

Replace the electrode.

Replace the electrode.

Buffer pH 9:

–10 mV < U(not stirred) < 10 mV and

The sum of the drift values after 1, 2,

3 and 4 min is < 12 mV/min.

A response time does not fulfill the limit value for the rating Usable electrode .

The pH values are not defined for all the buffers at the temperature measured.

All the slopes fulfill the limit values for the rating Usable electrode .

and

Uoff is outside the preset limits.

Short circuit

Glass membrane / Diaphragm

No buffer data

Unsuitable reference electrode

Clean the diaphragm (see leaflet for the electrode).

Repeat the electrode test at a temperature at which the pH values of all the buffers are defined.

Repeat the electrode test with a suitable reference system.

Replace the contaminated reference electrolyte.

Select Electrode type = Custom and adjust the limits for

Uoff.

900 Touch Control ■■■■■■■■ 407

29.12 Evaluations (EVAL)

Test criterion

The streaming potential is too high.

Message

Contaminated diaphragm

Two slopes do not fulfill the limit values for the rating Usable electrode .

All the slopes do not fulfill the limit values for the rating Usable electrode .

Wrong buffer

Partial short circuit

■■■■■■■■■■■■■■■■■■■■■■

Remedy

Clean the diaphragm (see leaflet for the electrode).

Repeat the electrode test with buffers pH 4, 7 and 9.

Check the temperature sensor.

Enter the correct calibration temperature.

29.12 Evaluations (EVAL)

Various additional evaluations (EVAL commands) can be carried out for titrations and measurements. The evaluation always refers to the last titration or measurement prior to the EVAL command. Only those evaluation commands which are available for the last titration or measurement prior to the EVAL command can be inserted in the list of commands. If a titration or measuring command is deleted before the EVAL command, then it will be shown in red in the list of commands, as the reference is missing.

EVAL commands can also be inserted in the method run at a later time and the determination can be reevaluated ( [Recalculate] function in the results dialog).

Additional evaluations for DET titrations

The following additional evaluations are possible:

EVAL FIX-EP (fixed endpoints)

Measured quantities:

– Measured value

– Time

– Volume

EVAL pK/HNP (pK value / half neutralization potential HNP)

This evaluation is only possible for the measuring modes pH and U .

EVAL MIN/MAX (minimum/maximum)

EVAL BREAK (break points)

Additional evaluations for MET titrations

The following additional evaluations are possible:

EVAL FIX-EP (fixed endpoints)

Measured quantities:

– Measured value

– Time

– Volume

EVAL pK/HNP (pK value / half neutralization potential HNP)

This evaluation is only possible for the measuring modes pH and U .

EVAL MIN/MAX (minimum/maximum)

408 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

EVAL BREAK (break points)

Additional evaluations for SET titrations

The following additional evaluations are possible:

EVAL FIX-EP (fixed endpoints)

Measured quantities:

– Measured value

– Time

– Volume

EVAL MIN/MAX (minimum/maximum)

EVAL RATE (average dosing rate)

Additional evaluations for KFT titrations

The following additional evaluations are possible:

EVAL FIX-EP (fixed endpoints)

Measured quantities:

– Measured value

– Time

– Volume

EVAL MIN/MAX (minimum/maximum)

EVAL RATE (average dosing rate)

Additional evaluations for KFC titrations

The following additional evaluations are possible:

EVAL FIX-EP (fixed endpoints)

Measured quantities:

– Measured value

– Time

– Amount of generated iodine

EVAL MIN/MAX (minimum/maximum)

EVAL RATE (average dosing rate)

Additional evaluations for BRC titrations

The following additional evaluations are possible:

EVAL FIX-EP (fixed endpoints)

Measured quantities:

– Measured value

– Time

– Amount of generated bromine

EVAL MIN/MAX (minimum/maximum)

EVAL RATE (average dosing rate)

Additional evaluations for STAT titrations

The following additional evaluations are possible:

29 Parameters

900 Touch Control ■■■■■■■■ 409

29.12 Evaluations (EVAL)

EVAL FIX-EP (fixed endpoints)

Measured quantities:

– Measured value

– Time

– Volume

EVAL MIN/MAX (minimum/maximum)

EVAL RATE (average dosing rate)

Additional evaluations for measurements

The following additional evaluations are possible:

EVAL FIX-EP (fixed endpoints)

Measured quantities:

– Measured value

– Time

EVAL MIN/MAX (minimum/maximum)

EVAL BREAK (break points)

NOTICE

■■■■■■■■■■■■■■■■■■■■■■

The commands EVAL are not available for concentration measurements

( MEAS Conc ).

29.12.1 Fixed endpoint evaluation (EVAL FIX-EP)

For the fixed endpoint evaluation, the associated values are interpolated from the measuring point list for one quantity (measured value, volume, etc.). Up to nine fixed endpoints can be evaluated with one command.

Fixed quantity

Selection of the quantity to which the associated value is interpolated from the measuring point list.

Selection

Default value

Measured value

Measured value

| Time | Volume | Quantity

Quantity

Amount of iodine/bromine created (only for KFC/BRC).

Fixed EP1 at

The fixed endpoint must lie between the first and the final entry in the measuring point list.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control instrument used.

410 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Measured value (measuring mode pH):

Input range –20.000 to 20.000

Selection

Default value off off

Measured value (measuring mode U, Ipol):

Input range

Selection

Default value

–2,000.0 to 2,000.0 mV off off

Measured value (measuring mode Upol):

Input range –200.00 to 200.00 µA

Selection

Default value off off

Measured value (measuring mode T):

Input range

Selection

Default value

–20.0 to 150.0 °C off off

Measured value (measuring mode Cond):

Input range 0.0001 to 999.9999 mS/cm

Selection

Default value off off

Time:

Input range

Selection

Default value

Volume:

Input range

Selection

Default value

Quantity:

Input range

Selection

Default value

Fixed EP2 at…Fixed EP9 at

See Fixed EP1 at .

0.0 to 999999.9 s off off

0.00000 to 9,999.99 mL off off

0.0 to 99,999.9 µg off off

29 Parameters

900 Touch Control ■■■■■■■■ 411

29.12 Evaluations (EVAL) ■■■■■■■■■■■■■■■■■■■■■■

29.12.2 pK value and half neutralization potential evaluation (EVAL pK/

HNP)

The pK value (pH measuring mode) or the half neutralization potential (U measuring mode) can be determined for DET and MET titrations.

The activities of conjugated acid-base pairs are linked by the so-called

Henderson-Hasselbalch equation: pH = pK a

+ log (a

B

/a

A

)

If the activities of the acid and the conjugated base are equal (a then pH=pK a

A

= a

B

),

. This is the value at the half neutralization point and can be extrapolated from the titration curve. A careful pH calibration is necessary for pK evaluations. Nonetheless, the determined pK value is only an approximation, as the ionic strength is not taken into account. In order to obtain more accurate values, titrations must be carried out with decreasing ionic strength and the results extrapolated to the ionic strength zero.

pK evaluation in aqueous solution is limited to the range 3.5 < pK < 10.5

because of the leveling effect of strong acids and the lack of jumps with very weak acids. pK values of mixtures of acids and polyvalent acids can also be determined.

Measured value

EP2 pK2

EP1 pK1

1/2 1/2 1/2 1/2

Volume

Figure 19 Evaluation of pK value / half neutralization potential

In non-aqueous solutions the half neutralization potential (HNP) is frequently used instead of the pK value. The HNP is evaluated in the same way as the pK value.

No parameters can be edited for the command EVAL pK/HNP .

NOTICE

If a start volume is to be added then it must be smaller than the half endpoint volume.

412 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

29.12.3 Minimum and maximum evaluation (EVAL MIN/MAX)

For the minimum or maximum measured value the associated volume, time and temperature are interpolated from the measuring point list. The evaluation begins as soon as the slope of the curve exceeds a particular threshold value.

Measured value

Maximum

Threshold value

Evaluation

Threshold value

Minimum

Volume / Time

Figure 20 Evaluation of minimum and maximum

Selection whether the minimum or the maximum of the titration curve is being evaluated. With a command either the minimum or the maximum can be evaluated. If both values are needed, a second command has to be defined in the method.

Selection

Default value

Minimum | Maximum

Maximum

The evaluation of the minimum or maximum begins as soon as the slope of the curve exceeds the set threshold value. Use a lower threshold value if the minimum or maximum is not found.

Measuring mode pH:

Input range

Default value

0.1 to 20

1.0

With titrations, the unit is pH/mL, with measurements pH/s.

Measuring mode U, Ipol:

Input range

Default value

1.0 to 2,000.0

25.0

With titrations the unit is mV/mL, with measurements mV/s.

900 Touch Control ■■■■■■■■ 413

29.12 Evaluations (EVAL) ■■■■■■■■■■■■■■■■■■■■■■

Measuring mode Upol:

Input range 0.5 to 10.0

Default value 5.0

With titrations the unit is µA/mL, with measurements µA/s.

Measuring mode T:

Input range

Default value

0.1 to 20.0 °C/s

1.0 °C/s

Measuring mode Cond:

Input range 0.5 to 999.9 (mS/cm)/s

Default value 5.0 (mS/cm)/s

29.12.4 Break point evaluation (EVAL BREAK)

A break point evaluation is used to determine sharp changes of direction in the titration curve. This evaluation is primarily used for photometric and conductivity titrations. The method is based on the search for extremes in the second derivative of the curve.

Measured value

2nd derivative

EP criterion

Break point

Volume / Time

Figure 21 Evaluation of a break point

Measure of the minimum sharpness of the break point. The smaller the EP criterion set, the more break points will be found. As this is a relative value related to the total measured value modification, even small changes in the measured value can be evaluated as a break point for a small measured value range.

Input range

Default value

0.0 to 1.0

0.3

414 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Slope

29 Parameters

Minimum difference between the slope before and after the break point.

The smaller the difference, the more break points will be found.

Input range

Default value

0.0 to 10.0

0.9

Smoothing factor

The higher the smoothing factor, the fewer break points will be found.

Input range

Default value

2 to 20

5

Set windows

This parameter allows to evaluate break points only in a certain range (socalled window) of the curve. Break points outside this window will not be evaluated. Only one window can be defined. Within this window however, all break points are evaluated.

Selection

Default value off | Measured value | Time | Volume off off

The entire titration curve is being evaluated.

Measured value

The window is defined on the measured value axis.

Time

The window is defined on the time axis.

Volume

The window is defined on the volume axis.

Setting windows

The lower and upper limit of the window is defined under [Set window] .

Lower limit

Measured value, time or volume for the lower limit.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control instrument used.

Measured value (measuring mode pH):

Input range

Default value

–20.000 to 20.000

–20.000

900 Touch Control ■■■■■■■■ 415

29.12 Evaluations (EVAL) ■■■■■■■■■■■■■■■■■■■■■■

Measured value (measuring mode U, Ipol):

Input range –2,000.0 to 2,000.0 mV

Default value –2,000.0 mV

Measured value (measuring mode Upol):

Input range

Default value

–200.00 to 200.00 µA

–200.00 µA

Measured value (measuring mode T):

Input range –20.0 to 150.0 °C

Default value –20.0 °C

Measured value (measuring mode Cond):

Input range

Default value

0.0001 to 999.9999 mS/cm

0.0001 mS/cm

Time:

Input range

Default value

0.0 to 999999.9 s

0.0 s

Volume:

Input range

Default value

0.00000 to 9,999.99 mL

0.00000 mL

Upper limit

Measured value, time or volume for the upper limit.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control device used.

Measured value (measuring mode pH):

Input range

Default value

–20.000 to 20.000

20.000

Measured value (measuring mode U, Ipol):

Input range –2,000.0 to 2,000.0 mV

Default value 2,000.0 mV

Measured value (measuring mode Upol):

Input range

Default value

–200.00 to 200.00 µA

200.00 µA

416 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Measured value (measuring mode T):

Input range –20.0 to 150.0 °C

Default value 150.0 °C

Measured value (measuring mode Cond):

Input range

Default value

0.0001 to 999.9999 mS/cm

999.9999 mS/cm

Time:

Input range

Default value

0.0 to 999,999.9 s

999,999.9 s

Volume:

Input range

Default value

0.00000 to 9,999.99 mL

9,999.99 mL

29.12.5 Rate evaluation (EVAL RATE)

The rate evaluation provides the median dosing rate in a defined range

(so-called window). The median dosing rate is determined by means of linear regression through a minimum of three measuring points. If the dosing is carried out with only one dosing device and if more than one cylinder volume is dosed, then the median dosing rate will be reduced as a result of the filling time.

When the command is opened for the first time, a window over the entire time range is already defined. In contrary to the potentiometric evaluation of a DET/MET titration the individual windows may overlap. A maximum of 9 windows can be defined.

[New]

Define a new window.

[Delete]

Delete the selected window.

[Edit]

Edit the settings of the selected window.

Lower limit

Time for the lower limit.

Input range

Default value

0 to 999999 s

0 s

Upper limit

Time for the upper limit.

900 Touch Control ■■■■■■■■ 417

29.13 Calculations ■■■■■■■■■■■■■■■■■■■■■■

Input range

Default value

0 to 999999 s

999999 s

29.13 Calculations

The following calculation commands are available:

CALC Defining calculations of a determination.

CALC LIVE Defining the calculation whose result will be displayed in the live display (so-called live result) during a titration or measurement.

29.13.1 Calculations (CALC)

Calculations are defined with the command CALC . A method may contain a maximum of nine calculation commands. A maximum of nine calculations can be defined per command. A series of variables (raw data from the determination, previously calculated results, etc.) is available for the calculations.

The command contains the following options, among others:

Save result as titer

Save result as common variable

Save result in result table

Define result limits

[New]

[Delete]

418 ■■■■■■■■

The result name is specified in the list for each calculation.

Define a new calculation or load an existing template, see following chapter.

Delete the selected calculation from the list.

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

[Edit]

29 Parameters

Edit the data of the selected calculation

(see chapter 29.13.1.2, page

421) .

29.13.1.1 Loading a result template

The most common calculations are already available, the so-called result templates, in order to facilitate the definition of a new calculation. You also have the option, however, of defining templates of your own (see dialog System / Templates ).

Proceed as follows to load an existing template:

1 Select a result template

In the Sequence / Edit command dialog, tap on the [New] button.

The list of Metrohm result templates is displayed:

900 Touch Control

Select the desired Metrohm result template and tap on [Load template] .

or

Tap on [Custom templates] and select a template you generated yourself.

If a statistic variable SMN# (# = 1…9) is defined in the template which is already used in another calculation, then the following message will be displayed:

■■■■■■■■ 419

29.13 Calculations ■■■■■■■■■■■■■■■■■■■■■■

2 Confirm the message

Tap on [OK] .

The Note dialog is displayed. This note is defined in the dialog Edit result template / Note for wizard .

3 Enter the numerical values of the wildcards

Tap on [Next] .

The list of wildcards is displayed:

420 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Enter the numerical values of the wildcards.

Tap on [Next] .

The editing dialog of the new calculation is displayed:

29 Parameters

The calculation can now be edited further, e.g. define result limits, see following chapter.

29.13.1.2

Result name

Editing a calculation

Calc. formula R1...R9

Shows the calculation formula. A special editor is opened for the definition

(see chapter 29.13.3, page 427)

. The designation R1...R9

corresponds to the result variable. Under [Result variable] this can later be modified.

Entry

Default value max. 100 characters empty

Decimal places

The result name is the text which will be shown in the result display and in the report. The default name corresponds to the result variable.

Entry

Default value max. 24 characters

R1…R9

Number of decimal places used to display the result.

Input range

Default value

0 to 5

2

Result unit

The result unit is displayed and saved together with the result.

Entry max. 10 characters

900 Touch Control ■■■■■■■■ 421

29.13 Calculations

Selection

Default value

■■■■■■■■■■■■■■■■■■■■■■

% | mol/L | mmol/L | g/L | mg/L | mg/mL | mg/100 g | ppm | g | mg | µg | mL | µL | mg/ piece | °C | µg/min | mL/min | µL/min

%

[Note]

Entering a note on the calculation.

[Result variable]

Modifying the result variable.

[Result limits]

Defining the limits for the monitoring of the result.

[Result options]

Defining additional settings for the calculation.

Dialog "Edit calculation / Note"

In this dialog you can enter a short text, e.g. to describe the variables used.

Dialog "Edit calculation / Result variable"

When a new calculation is created, the result variable will be automatically issued. In this dialog, however, it can be modified at any time.

Result variable

To each calculation belongs an unambiguous result variable. With this result variable you can use this result in additional calculations.

Selection R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8 | R9

The selection contains only those result variables which are not issued in this calculation command yet.

Dialog "Edit calculation / Result limits"

For each result, limit values can be defined. These result limits are monitored when the calculation is carried out. If monitoring is activated, the results are shown as follows in the result display:

Green, if the result falls within the limit values

Red, if the result lies outside the limit values

Monitoring result limits on | off (Default value: off )

Enabling and disabling the monitoring function.

422 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Lower limit

When the monitored quantity falls below this value, the action defined in the following is triggered.

Input range

Default value

0 to 9999999999

0

Upper limit

When this value is exceeded, the action defined in the following is triggered.

Input range

Default value

0 to 9999999999

9999999999

Action

Selection of the action, which is carried out when the result limits are exceeded.

Selection

Default value

Display message | Document message | Cancel determination

Display message

Display message

A message is displayed. You can select whether you want to continue with the determination or cancel the run.

Document message

In the determination data it will be documented, that the result limits have been exceeded.

Cancel determination

The determination is stopped.

Dialog "Edit calculation / Result options"

In the dialog Edit calculation / Result options , settings for how to process the calculated result are defined.

Variable for mean value

If the statistic calculations have been activated (see method options), the mean value of the single results will be saved as variable SMN1 to SMN9.

As default value, always the first free variable is being displayed.

Selection SMN1 … SMN9 | off off

For the result not statistic calculations are carried out.

Save as titer

The result can be saved as titer for the selected titrant.

900 Touch Control ■■■■■■■■ 423

29.13 Calculations ■■■■■■■■■■■■■■■■■■■■■■

NOTICE

The titer is assigned to the titrant defined in the last titration command before the calculation. Please note that the CALC command containing the titer assignment is inserted after the titration command with which the titer is determined.

Selection

Default value

Single value | Mean value | off off

Single value

The result of the current determination is saved as titer.

Mean value

If statistic calculations are carried out for the result, the current mean value of the determination series is saved.

off

The result is not saved as titer.

Save as common variable on | off (Default value: off )

The calculated result can be saved as a method-independent variable, a so-called common variable. The result is then also available in other methods for calculations. In all cases, only the single value will be saved, even if the statistics function is enabled.

Variable

Selection of the common variable to which the result is being assigned.

Selection CV01 … CV25

Display result on | off (Default value: on )

If you deactivate this parameter, the result is neither displayed in the result dialog nor printed in the result report. This can be advisable for intermediate results.

Save result in result table on | off (Default value: off )

The calculated result can be saved in the result table. This may be advisable if e.g. the results of all determinations carried out on a particular day are to be displayed clearly. A maximum of nine results from a determination can be saved in the result table.

424 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Precision

Setting, with which accuracy the result is used in additional calculations.

Selection

Default value

Round

Round

| Truncate | Full precision

Round

The result is rounded to the defined number of decimal places (commercial rounding, in accordance with the US Pharmacopeia USP). If the digit at the first dropped decimal place is 1, 2, 3 or 4 , then it will be rounded off; if this digit is 5, 6, 7, 8 or 9 , then it will be rounded up.

Negative digits will be rounded in accordance with their amount, i.e.

away from zero.

Truncate

The result is cut to the number of decimal places defined.

Full precision

The result is used with full accuracy (floating point number either in

"single precision" (32 bit) or in "double precision" (64 bit), according to the standard IEEE 754).

29.13.2 Calculations (CALC LIVE)

The command CALC LIVE can be used to define a calculation, the result of which will be shown in the live display during a titration or measurement. The current volume or measured value, respectively, is used for this purpose. This is helpful, e.g. for Karl Fischer titrations, in order to be able to track the water content directly during the titration.

CAUTION

The command must be inserted directly before the corresponding titration or measuring command.

This command is identical with the CALC command except for the following differences:

The number of variables is limited.

There are no result options.

The result cannot be monitored.

The result variable cannot be modified.

The result will only be displayed in the live display. It appears neither in the results dialog nor in a report.

Result name

With live calculations no custom result name can be defined, the designation LR cannot be modified.

900 Touch Control ■■■■■■■■ 425

29.13 Calculations ■■■■■■■■■■■■■■■■■■■■■■

Calculation formula

Shows the calculation formula. A special editor is opened for the definition

(see chapter 29.13.3, page 427) .

Entry

Default value max. 100 characters empty

Decimal places

Number of decimal places used to display the result.

Input range

Default value

0 to 5

2

Result unit

The result unit is displayed and saved together with the result.

Entry

Selection

Default value max. 10 characters

% mg/100 g | ppm | g | mg | µg | mL | µL | mg/ piece | °C | µg/min | mL/min | µL/min

%

| mol/L | mmol/L | g/L | mg/L | mg/mL |

[Note]

Entering a note on the calculation.

Dialog "Edit calculation / Note"

In this dialog you can enter a short text, e.g. to describe the variables used.

Displaying the live result

Proceed as follows to display the live result in the live display:

1 Starting a determination

Tap on [ ] .

The determination starts and the live display is shown:

2 Define the display options

Tap on the [View] button.

Tap on the [Meas. value options] button.

A maximum of three measured values can be shown in the live display.

Select the option Live result at one of the three parameters.

426 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

3 Display the live display

Tap twice on [ ] .

The current result is displayed,

29 Parameters

29.13.3 Formula editor

The formulas for the calculations are entered with the formula editor. The maximum formula length is 100 characters.

The formula editor contains buttons for entering numbers, mathematical operators and variables. The variables can be selected from lists. This is advisable in order to avoid spelling mistakes. The formula editor is equipped with an automatic syntax check, which is activated when applying the formula. The standard rules of priority apply for the calculation operations.

900 Touch Control

Key

C00

CI

EP

CV

R

TITER

CONC

Description

Sample size

Sample identification CI# (# = 1 - 2)

Volume of endpoint EP# (# = 1 - 9)

Common Variable CV# (# = 1 - 25)

Result R# (# = 1 - 9)

Titer of the titrant selected in the previous command

Concentration of the titrant selected in the previous command

⌫ The character in front of the cursor is deleted.

Delete entry The content of the input field is deleted completely.

■■■■■■■■ 427

29.13 Calculations ■■■■■■■■■■■■■■■■■■■■■■

Key

Description

The cursor within the input field is shifted to the left by one character.

The cursor within the input field is shifted to the right by one character.

If the method contains more than one data-generating command (titration, measurement etc.) before the calculation command, then you will have to indicate the command identification in front of the variable:

Variable

#M.

#C.

#E.

Description

Titration or measuring command

Calculation command

Evaluation command

"#" stands for a sequential number that you must enter manually. The number for each command identification will be counted separately and does not necessarily correspond to the number of the command line.

Example: The variable 1M.TITER

corresponds to the titer of the titrant selected in the first titration command.

CAUTION

The command identifications are not adjusted automatically when additional commands are subsequently inserted into the method. Check the formula for correctness in such cases.

[Variables]

Opening the list of additional variables

(see chapter 29.13.3.1, page 428)

.

[Math. functions]

29.13.3.1

Opening the list with mathematical functions (see chapter 29.13.3.2, page 429)

.

Variables

All of the variables which can be used for calculations are divided up according to subject in the dialog Calculation formula / Variables :

System variables

List of the system-specific variables. These variables describe the current status of the system.

Result and statistics variables

List of the variables which are required for calculating results and statistics for the current determination.

428 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

29.13.3.2

29 Parameters

Common variables

List of common variables. All of the common variables are listed, no matter whether they possess a value or not.

You will find a detailed description of all of the variables in

chapter

29.13.3.3, page 430

.

Mathematical functions

The following mathematical functions can be utilized in a calculation formula:

Function y^z

SQRT(X)

ABS(X)

LN(X)

LOG(X)

FRAC(X)

INT(X)

TST(X,Y)

Description

Power function

Example: 4^2 = 16 square root of X

Example: √ (EP1) absolute value of X

Example: ABS(C00); in order, e.g. with reweighings, to convert the negative sample size to a positive value for later calculations natural logarithm of X decimal logarithm of X

Fraction of X

Example: FRAC(2.5971) = 0.5971

integer part of X

Example: INT(2.5971) = 2

Test function

If invalid variables (e.g. missing endpoints) occur in a calculation, then these can be replaced with a valid value by using this function. In this way invalid results can be avoided.

Syntax:

– X = variable to be tested

– Y = substitute value

900 Touch Control ■■■■■■■■ 429

29.13 Calculations ■■■■■■■■■■■■■■■■■■■■■■

29.13.3.3 Variable list

The following table contains all of the variables which can be used for calculations. For variables having an index (e.g. 'EP1' ), the index must be entered manually. In the following table, this index is characterized with the character "#".

MSS

MSV

MSD

MSM

MIM

MIT

MSA

MSP

EF#

ET#

ED#

ESI#

Variable

C00

CI1, CI2

Description

Sample size

Sample identifications

The sample identifications can be used in calculations only if numerical values are entered.

Duration of the entire determination DD

Titrants

TITER

CONC

Titer of the titrant selected in the titration command

Concentration of the titrant selected in the titration command

Titrations, measurements, calibrations

EP# Volume/quantity of endpoint EP# (# = 1 - 9)

EC#

EM#

Charge at endpoint EP# (# = 1 - 9)

Measured value of endpoint EP# (# = 1 - 9)

ERC of endpoint EP# (# = 1 - 9)

Temperature at endpoint EP# (# = 1 - 9)

Time at endpoint EP# (# = 1 - 9)

Recognition of endpoint EP# (# = 1 - 9)

Endpoint found = 1, no endpoint = 0

MST

MCD

MTM

MDD

Initial measured value, i.e. measured value prior to the processing of the start conditions

Initial temperature, i.e. temperature prior to the processing of the start conditions

Volume for start volume

Volume for start measured value

Volume for start slope

Volume for all the start conditions

Duration of start conditions

Start measured value, i.e. measured value after the processing of the start conditions

Start temperature, i.e. temperature after the processing of the start conditions

Total duration of the titration, measurement or calibration

Type of temperature measurement (Pt1000, NTC or manually)

Format = Text

Duration of effective dosing, i.e. without filling times, pauses

430 ■■■■■■■■ 900 Touch Control

HD#

XIP

XIM

XIT

XID

FD#

HP#

HM#

HT#

MCL

Evaluations

FP#

FM#

FT#

LV

LM

LD

LT

Sensor

MEN

MSL

MVA

MCC

MCM

MCT

MMP

MTS

Variable

MDC

DDC

MCV

MCQ

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Description

Drift for drift correction

Time for drift correction

End volume, i.e. total dosed volume at the end of the titration

End quantity, i.e. total amount of removed water or of generated bromine, respectively, at the end of the titration (in µg)

Final charge, i.e. total charge in mAs at the end of the titration

End measured value

End temperature

Number of measuring points in measuring point list

Stop type

Format = Text

Currently dosed volume of the ongoing determination

Current measured value of the ongoing determination

Current duration of the ongoing titration or measuring command

Current temperature of the ongoing determination

Electrode zero point pH(0) or E(0)

Electrode slope

Electrode variance (mathematically only defined for three standards or more);

CAL Conc: calculated only if calibration is carried out with a minimum of four standards

Cell constant of conductivity measuring cell

Volume of fixed endpoint FP# (# = 1 - 9)

Measured value of fixed endpoint FP# (# = 1 - 9)

Temperature at fixed endpoint FP# (# = 1 - 9)

Time at fixed endpoint FP# (# = 1 - 9)

Volume of pK value / half neutralization potential HP#

Measured value of pK value / half neutralization potential HP#

Temperature at pK value / half neutralization potential HP#

Time at pK value / half neutralization potential HP#

Volume/quantity of minimum measured value

Minimum measured value

Temperature at minimum measured value

Time at minimum measured value

900 Touch Control ■■■■■■■■ 431

29.13 Calculations ■■■■■■■■■■■■■■■■■■■■■■

BT#

BD#

RD#

RDS#

RDC#

RM

RMS

RMC

Variable

XAP

XAM

XAT

XAD

BP#

BM#

Description

Volume/quantity of maximum measured value

Maximum measured value

Temperature at maximum measured value

Time at maximum measured value

Volume of break point BP# (# = 1 - 9)

Measured value of break point BP# (# = 1 - 9)

Temperature at break point BP# (# = 1 - 9)

Time at break point BP# (# = 1 - 9)

Mean rate in time slot # (# = 1 - 9)

Standard deviation for RD# (# = 1 - 9)

Correlation coefficient for RD# (# = 1 - 9)

Mean rate for whole range

Standard deviation for RM

Correlation coefficient for RM

SSR#

SNR#

SSD

SNT

Results, statistics

R# Result (# = 1 - 9)

SMN#

SSA#

Mean value of result R# (# = 1 - 9)

Absolute standard deviation of SMN# (# = 1 - 9)

Relative standard deviation of SMN# in % (# = 1 - 9)

Number of results from which the mean value SMN# was calculated (# = 1 - 9)

Number of determinations that are carried out for the statistics calculations

Statistics status (statistics activated = 1, statistics deactivated = 0)

%SC

%AS

%AC

%AD

Common variables

CV# Common variable (# = 1 - 25)

System variables

%RN Sample number, i.e. number that is increased by one at each start of a determination

Start counter, see dialog More determination data / Properties

Autostart status (autostart activated = 1, autostart deactivated = 0)

Autostart counter, i.e. number of autostarts already carried out

Autostart setpoint counter, i.e. number of autostarts to be carried out

%SS

%SL

%SE

Status of the sample table (sample table activated = 1, sample table deactivated = 0)

Number of the current sample table line

Last sample in the sample table processed (yes = 1, no = 0)

432 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

29.14 Reports (REPORT)

The reports to be printed out are defined with the command REPORT . A maximum of ten reports can be defined for each command. The following reports can be inserted, among others:

Result report

Calculation report

Parameter report

System reports (system settings, sensor list, GLP data etc.)

Method reports (method run, method parameters etc.)

[Report options]

Defining the general settings for the report output.

[Insert]

Add a new report to the list.

[Delete]

Delete the selected report from the list.

[Edit]

Edit the settings of the selected report.

29.14.1 General report options

The settings for the report printout can be adjusted in the dialog Edit command / Report options to indicate, for example, whether a report header, a signature line or a frame is to be printed.

Report header

Defining the output of the report header. The report header contains general information such as device type, serial number, program version and the printing date.

900 Touch Control ■■■■■■■■ 433

29.14 Reports (REPORT) ■■■■■■■■■■■■■■■■■■■■■■

NOTICE

You can also define your own report header, which is printed in addition to this report header ( System ▶ Templates ▶ Report header ).

Selection

Default value off | once | on each page on each page off

The report header will not be printed.

once

The report header will only be printed on the first page.

on each page

The report header will be printed on every page.

Signature line

Output of a special line for date and signature. This line will be printed at the very bottom of every page.

Selection

Default value off | once | on each page off off

No signature line will be printed.

once

The signature line will only be printed on the last page.

on each page

The signature line will be printed on every page.

Frame on | off (Default value: on )

If this parameter is activated, a frame is printed as lateral marking.

29.14.2 Settings of the individual reports

Fixed key [ ] : More reports ▶ Report ▶ Edit

Command REPORT: Report ▶ Edit

Settings can be edited for the following reports (for details, see online help):

Result report

Curve

Definition of the curve size, display of the individual measuring points and grid lines, etc.

434 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Measuring point list

Definition of the method command to which the measuring point list is to be printed.

Calculations

Definition of the accuracy with which the variables used as well as the results calculated are printed.

Statistics short

Definition of the time at which the report is to be printed (with each determination or only at the end of a sample series).

Statistics overview

Definition of the time at which the report is to be printed (with each determination or only at the end of a sample series).

Sample table

Result table

Definition of the time at which the report is to be printed (with each determination, at the end of a sample series or at the end of the sample table).

Used devices

Variables

Monitoring

29.14.3 List of reports

The following reports can be inserted in the command REPORT :

NOTICE

Report

Result report

Curve

Measuring point list

Calculations

Used devices

Variables

Monitoring

Statistics

If you are using a CUSTOM NEOS printer, you can print only the reports that are marked with a # on the display.

Contents

# Report with determination properties, sample data, calculated results, etc. If there are multiple determinations, the statistics will also be printed out.

# Curve report.

# Measuring point list report.

# Details concerning the calculations carried out (parameter settings in accordance with the CALC command and the calculated results).

The devices used for the determination, as displayed in the More determination data / Properties dialog.

All determination variables, as displayed in the results dialog.

Details for the monitored quantities (STAT, DOS only).

900 Touch Control ■■■■■■■■ 435

29.14 Reports (REPORT)

Report

Statistics short

Statistics overview

■■■■■■■■■■■■■■■■■■■■■■

Contents

# Summary of the statistics calculations. The number of determinations, the mean value, and the absolute and relative standard deviations are printed out for each result.

# Detailed statistical overview. The sample data and all individual results are printed out for each determination. The number of determinations, the mean value, and the absolute and relative standard deviations are printed out for each result.

Sample data

Sample table

Result table

Result table

More reports ▶ Method reports

Method sequence

Parameters full

Titration & measuring param.

Modified parameters

List of all determinations in the sample table with the respective sample data, as entered in the sample table.

List of all determinations in the result table with results and with the determination data, as saved in the result table.

Method properties and the list of all method commands.

Method properties and options, all method commands with all parameters.

All parameters which no longer have their default settings will be printed in bold . All parameters which have been modified in comparison to the stored version of the method are indicated by an *.

Method properties; titration, measurement and calibration commands with all parameters.

All parameters which no longer have their default settings will be printed in bold . All parameters which have been modified in comparison to the stored version of the method are indicated by an *.

Method properties, all method parameters which have been modified in comparison to the stored version of the method.

Method properties, all method parameters which no longer have any default settings.

Non default parameters

More reports ▶ System reports

System settings

System settings

Dialog options

Titrants

Titrant list

All titrant data short

All titrant data full

Settings for acoustic signals and accuracy of the measured value display.

Settings for routine dialog and expert dialog.

List of all the titrants configured in the system.

The most important titrant data of all titrants (name, concentration, titer, last titer determination).

All of the titrant data of all of the titrants (working life, titer options, data concerning the exchange unit/dosing unit).

436 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Report

Sensors

Sensor list

All sensor data short

All sensor data full

Device manager

Device list

All device properties

GLP manager

GLP data

Common variables

Common variable list

All common variable properties

Templates

Templates sample data

Result template list

All result templates details

Input/Output lines

Custom calibration buffers

Rack tables

Sample rack list

Miscellaneous

Form feed

29 Parameters

Contents

List of all the sensors configured in the system.

The most important sensor data for all sensors (name, calibration data).

All of the sensor data for all of the sensors (working life, complete calibration data, calibration interval).

List of all devices configured in the system.

Properties of all the devices configured in the system.

All data stored in the GLP manager.

List of all the common variables defined in the system, together with their most important data (name, value, status).

Properties of all common variables (name, value, validity, status).

Sample identification list and sample assignment table.

List with all of the user-generated result templates.

Details of the all of the user-generated result templates (calculation formula, result options, note).

List with all of the defined input and output lines at the remote interface

(name, bit pattern).

Temperature tables for all defined custom calibration buffers.

List with all of the sample racks configured in the system, together with designation, number of positions and rack code.

If this entry is inserted between two reports, then these will each be printed out on a separate page.

900 Touch Control ■■■■■■■■ 437

29.15 Dosing and Liquid Handling ■■■■■■■■■■■■■■■■■■■■■■

29.15 Dosing and Liquid Handling

The following dosing commands are available:

Preparing the exchange unit or dosing unit (PREP)

Rinsing the cylinder and the tubings of the exchange unit / dosing unit.

Emptying the dosing unit (EMPTY)

Dosing a fixed volume (ADD)

Emptying the cylinder and the tubings of the dosing unit.

Dosing a specified volume.

Liquid Handling

(LQH)

Monitored dosing

(DOS)

Carrying out complex dosing tasks with a Dosino.

Dosing with simultaneous monitoring of measured value and temperature.

29.15.1 Preparing an exchange or dosing unit (PREP)

The PREP command is used to rinse and fill air bubble-free the cylinder and tubings of the exchange or dosing unit. You should carry out this function before the first determination or once a day.

Control device

This parameter is only displayed when several control devices are configured. Control devices are defined under System ▶ Device manager .

Selection of the control device from the list of devices. The control devices do not have to be connected. This allows to transfer methods more easily from one system to another.

Selection Selection of configured control devices

Dosing device

Selection of the MSB connector to which the dosing device is connected.

Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4

Titrant

Selection of the titrant from the titrant list. We recommend always selecting the titrant. This ensures that the parameters defined for the titrant for the preparing as well as the tubing parameters (dosing unit only) are used.

Titrants are defined under System ▶ Titrants . You can also enter a name which is not contained in the titrant list. When a determination is started there is a check whether the solution is contained in the list.

438 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

For buret units with integrated data chip, a check is made in the method run to verify whether the correct titrant has been attached and whether the type of dosing drive matches. For buret units without integrated data chip, the cylinder volume and the type of dosing drive are checked. For the selected titrant, the validity of the titer, the working life of the titrant and the time interval for the GLP test of the buret unit are checked at the start of the determination.

Selection

Default value

Selection of configured titrants | not defined not defined not defined

No check takes place. The command is carried out with default parameters

(see Chapter 32.2.2, page 493

and Chapter 32.3.2, page 494)

.

Cylinder volume to

This parameter can only be edited with Titrant = not defined .

Selection of the port via which the titrant is ejected. This setting is only relevant for dosing units, in case of preparing an exchange unit, this setting is ignored.

Selection

Default value

Port 1 | Port 2 | Port 3 | Port 4

Port 1

29.15.2 Emptying a dosing unit (EMPTY)

The EMPTY command is used to empty the cylinder and tubings of the dosing unit.

Control device

This parameter is only displayed when several control devices are configured. Control devices are defined under System ▶ Device manager .

Selection of the control device from the list of devices. The control devices do not have to be connected. This allows to transfer methods more easily from one system to another.

Selection Selection of configured control devices

Dosing device

Selection of the MSB connector to which the dosing device is connected.

Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4

Titrant

Selection of the titrant from the titrant list. We recommend always selecting the titrant. This ensures that the parameters defined for the titrant for

900 Touch Control ■■■■■■■■ 439

29.15 Dosing and Liquid Handling ■■■■■■■■■■■■■■■■■■■■■■ the preparing as well as the tubing parameters (dosing unit only) are used.

Titrants are defined under System ▶ Titrants . You can also enter a name which is not contained in the titrant list. When a determination is started there is a check whether the solution is contained in the list.

For buret units with integrated data chip, a check is made in the method run to verify whether the correct titrant has been attached and whether the type of dosing drive matches. For buret units without integrated data chip, the cylinder volume and the type of dosing drive are checked. For the selected titrant, the validity of the titer, the working life of the titrant and the time interval for the GLP test of the buret unit are checked at the start of the determination.

Selection

Default value

Selection of configured titrants | not defined not defined not defined

No check takes place. The command is carried out with default param-

eters (see Chapter 32.2.2, page 493 and

Chapter 32.3.2, page 494)

.

Air inlet

This parameter can only be edited with Titrant = not defined .

Selection of the port via which air is aspirated. This setting is only relevant for dosing units, in case of preparing an exchange unit, this setting is ignored.

Selection

Default value

Port 1 | Port 2 | Port 3 | Port 4

Port 4

29.15.3 Dosing a specified volume (ADD)

You can dose a specified volume with the command ADD .

29.15.3.1 Dosing parameters

The parameters for the dosing are defined under [Dosing parameters] .

Volume

Volume which is dosed.

Input range

Default value

0.00000 to 99999.9 mL

10.0000 mL

Dosing rate

The rate at which dosing takes place.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum

440 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters maximum

The maximum dosing rate depends on the cylinder volume (see chapter 32, page 492) .

Filling rate

29.15.3.2

Rate at which the dosing cylinder is filled.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum filling rate depends on the cylinder volume

(see chapter

32, page 492) .

Control instrument

The control instrument the determination is carried out with is selected under [Control device] . Control instruments are defined under System ▶ Device manager .

NOTICE

This button is only displayed when several control instruments have been configured.

Control device

29.15.3.3

Selection of the control instrument from the list of devices. Only those devices are displayed which are able to carry out the command. The control instruments do not have to be connected. This allows to transfer methods more easily from one system to another.

Selection Selection of configured control instruments

Dosing device

The parameters for the dosing device are edited under [Dosing device] .

Dosing device

Selection of the MSB connector to which the dosing device is connected.

Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4

Titrant

Selection of the titrant from the titrant list. We recommend always selecting the titrant. This ensures that the correct data (titer, concentration, etc.) is always used for the calculation. Titrants are defined under Sys-

900 Touch Control ■■■■■■■■ 441

29.15 Dosing and Liquid Handling ■■■■■■■■■■■■■■■■■■■■■■ tem ▶ Titrants . You can also enter a name which is not contained in the titrant list. When a determination is started there is a check whether the solution is contained in the list.

For buret units with integrated data chip, a check is made in the method run to verify whether the correct titrant has been attached and whether the type of dosing drive matches. For buret units without an integrated data chip, the cylinder volume and the type of dosing drive are checked.

For the selected titrant, the validity of the titer, the working life of the titrant and the time interval for the GLP test of the buret unit are checked at the start of the determination.

Selection

Default value

Selection of configured titrants | not defined not defined not defined

No check takes place.

Tandem dosing

A second dosing device can be defined under [Tandem dosing] in order to enable uninterrupted dosing. Dosing is carried out with a combination of two dosing devices so that the second dosing device is dosing while the first one is being filled and vice versa.

Dosing device

Selection of the MSB connector to which the dosing device is connected.

Every MSB connector will always be displayed.

Selection

Default value

1 | off

2 | 3 | 4 | off

Titrant

Selection of the titrant from the titrant list. We recommend always selecting the titrant. Titrants are defined under System ▶ Titrants . You can also enter a name which is not contained in the titrant list. When a determination is started there is a check whether the solution is contained in list.

For buret units with integrated data chip, a check is made in the method run to verify whether the correct titrant has been attached and whether the type of dosing drive matches. For buret units without integrated data chip, the cylinder volume and the type of dosing drive are checked. For the selected titrant, only the time interval for the GLP test of the buret unit is checked at the start of the determination.

442 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Filling rate

29 Parameters

NOTICE

The concentration, validity of the titer and the working life of the titrant are ignored.

Selection

Default value

Selection of configured titrants | not defined not defined not defined

No check takes place.

Rate at which the dosing cylinder is filled.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum filling rate depends on the cylinder volume

(see chapter

32, page 492) .

However, in order to ensure uninterrupted dosing, the following points must be taken into account:

Use the highest possible filling rate in order to keep the filling times as short as possible. The filling rate should be decreased for viscous liquids.

When you use two buret units with different dosing cylinders, the filling rate for the larger sized cylinder must be at the minimum:

900 Touch Control

= filling rate in mL/min for the larger sized cylinder

= filling rate in mL/min for the smaller cylinder

= cylinder volume in mL of the buret unit of the second dosing device

= cylinder volume in mL of the buret unit of the first dosing device

Example:

Dosing device 1: volume = 20 mL, filling rate = 50 mL/min

Dosing device 2: volume = 50 mL

Filling rate 2 ≥ 50 mL/min · 50 mL / 20 mL ≥ 125 mL/min

■■■■■■■■ 443

29.15 Dosing and Liquid Handling

29.15.3.4

■■■■■■■■■■■■■■■■■■■■■■

The dosing rate must not exceed 75% of the filling rate of the smaller cylinder. These values are listed in the following table, valid at maximum filling rate:

Table 20 Maximum dosing rate for different dosing cylinders

Cylinder volume maximum dosing rate

Exchange unit Dosing unit

1 mL

2 mL

5 mL

10 mL

20 mL

50 mL

2.25 mL/min

11.25 mL/min

22.50 mL/min

45.00 mL/min

112.50 mL/min

5.00 mL/min

12.50 mL/min

25.00 mL/min

50.00 mL/min

124.50 mL/min

Stirrer

The parameters for the stirrer are edited under [Stirrer] .

Stirrer

Selection of the MSB connector the stirrer is connected to. Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4 | off off

No stirrer will be used.

Stirring rate

Setting the stirring rate. It can be set in steps of –15 to +15. The default setting 8 corresponds to 1000 rpm. The formula for calculating the rota-

tional speed is specified in chapter 32.4, page 494 . The optimum stirring

rate can be tested in the manual control.

The algebraic sign of the stirring rate changes the direction in which the stirring is done. When the stirrer is viewed from above, this means:

"+": counterclockwise rotation

"–": clockwise rotation

Input range

Default value

–15 to 15

8

444 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Switch off automatically on | off (Default value: on )

If this parameter is activated, the stirrer will be switched off automatically at the end of the titration, measurement, etc.

29.15.4 Liquid Handling (LQH)

The command LQH can be used to carry out Liquid Handling tasks with a dosing device of the type Dosino (700 or 800). The manifold options of the Dosino can be exploited in full in combination with a Sample Processor. The four ports of the dosing units can be used at will as output or input ports. This means that not only simple dosing and filling procedures are possible. Complex Liquid Handling tasks such as pipetting or sample transfers can be carried out without difficulty. Several LQH and automation commands are required for this purpose which are best combined into subsequences.

29.15.4.1 Liquid Handling parameters

The parameters for the Liquid Handling function are defined under [Liquid Handling param.] .

Function

Type of Liquid Handling function.

Selection Dose | Fill | Aspirate | Eject | Exchange position | Change port | Compensate | End volume

Dose

The given volume is dosed. The dosing cylinder neither is automatically filled before nor afterwards.

Fill

The dosing cylinder is filled. The valve disc then remains at the selected port.

Aspirate

Liquid is aspirated. The dosing cylinder neither is automatically filled before nor afterwards. The aspirating volume has to be aspirated with a single piston stroke.

Eject

The whole cylinder content is ejected. The piston is, in contrast to the function End volume , lowered to the stop point, i.e. past the maximum volume mark. This function should only be carried out for preparing a dosing unit before the pipetting procedure.

Exchange position

The dosing cylinder is filled. This means that, for example, air can be aspirated via Port 4. The valve disc is then rotated to Port 2 and the dosing drive can be removed from the dosing unit.

900 Touch Control ■■■■■■■■ 445

29.15 Dosing and Liquid Handling ■■■■■■■■■■■■■■■■■■■■■■

Change port

The stopcock is moved to the given port, but no piston movement takes place.

Compensate

Because of the fact that the dosing units are interchangeable, the coupling of the Dosino piston rod (spindle) exhibits a low mechanical tolerance that can be noticed when the piston changes its direction of movement. This tolerance can be compensated with this function. A short piston movement is first made in the same direction as the previous movement, which is then followed by a piston movement in the reverse direction.

End volume

The whole cylinder content is ejected. The piston is lowered to the maximum volume mark i.e. until 10'000 pulses have been dosed. This command should be used for pipetting functions for emptying the cylinder.

In-/Outlet

Selection of the port via which the Liquid Handling command is carried out.

Selection Port 1 | Port 2 | Port 3 | Port 4

Port 1

Default value, when Function = Dose , Aspirate and End volume .

Port 2

Default value, when Function = Fill , Exchange position and

Change port .

Port 4

Default value, when Function = Eject and Compensate .

Volume

This parameter can only be edited with Function = Dose and Aspirate .

Volume which is dosed or aspirated.

Function = Dose:

Input range

Default value

0.00000 to 99999.9 mL

1.00000 mL

Function = Aspirate:

Input range

Default value

0.00000 to 50.0000 mL

1.00000 mL

Flow rate

This parameter can only be edited with Function = Dose , Fill , Aspirate ,

Eject , Exchange position and End volume .

446 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

29.15.4.2

29 Parameters

Rate at which it is dosed or filled.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum dosing and filling rates depend on the cylinder volume

(see chapter 32, page 492) .

Control instrument

The control instrument the determination is carried out with is selected under [Control device] . Control instruments are defined under System ▶ Device manager .

NOTICE

This button is only displayed when several control instruments have been configured.

Control device

29.15.4.3

Selection of the control instrument from the list of devices. Only those devices are displayed which are able to carry out the command. The control instruments do not have to be connected. This allows to transfer methods more easily from one system to another.

Selection Selection of configured control instruments

Dosing device

The parameters for the dosing device are edited under [Dosing device] .

Dosing device

Selection of the MSB connector to which the dosing device is connected.

Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4

Titrant

Selection of the titrant from the titrant list. We recommend always selecting the titrant. This ensures that the correct data (titer, concentration, etc.) is always used for the calculation. Titrants are defined under System ▶ Titrants . You can also enter a name which is not contained in the titrant list. When a determination is started there is a check whether the solution is contained in the list.

900 Touch Control ■■■■■■■■ 447

29.15 Dosing and Liquid Handling ■■■■■■■■■■■■■■■■■■■■■■

For buret units with integrated data chip, a check is made in the method run to verify whether the correct titrant has been attached and whether the type of dosing drive matches. For buret units without an integrated data chip, the cylinder volume and the type of dosing drive are checked.

For the selected titrant, the validity of the titer, the working life of the titrant and the time interval for the GLP test of the buret unit are checked at the start of the determination.

Selection

Default value

Selection of configured titrants | not defined not defined not defined

No check takes place.

29.15.5 Monitored dosing (DOS)

With the DOS command, you can dose and at the same time monitor the measured value and/or the temperature.

29.15.5.1 Dosing parameters

The parameters for the dosing are defined under [Dosing parameters] .

Dosing criteria

Three different dosing types can be selected for the dosing. Two of the following parameters can be specified:

Volume

Dosing rate

Dosing time

The third parameter is calculated according to the following formula:

Volume = Dosing rate · Dosing time

If continuous dosing is necessary, tandem dosing can be used (see chapter

29.7.8, page 369) . Tandem dosing is carried out with a combination of

two dosing devices so that the second dosing device is dosing while the first one is being filled and vice-versa.

Selection Volume/Dosing rate | Volume/Dosing time |

Dosing rate/Dosing time

Volume/Dosing rate

Defining the volume to be dosed and and the dosing rate.

Volume/Dosing time

Defining the volume to be dosed and the dosing time.

Dosing rate/Dosing time

Defining the dosing rate and the dosing time.

448 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Dosing criterion "Volume/Dosing rate"

Single dosing Tandem dosing

V/mL V/mL

29 Parameters

Filling time

Dosing rate Dosing rate t/s

Figure 22 Dosing criterion "Volume/Dosing rate"

Dosing criterion "Volume/Dosing time"

Single dosing Tandem dosing

V/mL V/mL t/s

Filling time t

1 t

2 t/s t

1

+ t

2

Figure 23 Dosing criterion "Volume/Dosing time"

CAUTION t/s

The defined time corresponds to the net dosing time t

1

+t

2

, i.e. the filling time is not included in the total.

If the dosing rate resulting from the settings cannot be reached, then dosing will be carried out at the highest possible rate.

The following formula can be used to estimate whether the maximum dosing rate is being used or not:

:

: Dosing rate in mL/min

Desired flow rate in mL/min

900 Touch Control ■■■■■■■■ 449

29.15 Dosing and Liquid Handling ■■■■■■■■■■■■■■■■■■■■■■

NOTICE

This formula applies only for using a single dosing device. The dosing rate is equal to the flow rate in the case of tandem dosing.

If the required dosing rate exceeds the maximum dosing rate for the

selected exchange unit or dosing unit (see chapter 32, page 492)

, then a larger cylinder volume must be selected.

Dosing criterion "Dosing rate/Dosing time"

Single dosing Tandem dosing

V/mL V/mL

Volume

Dosing rate

Dosing time

Filling time

Dosing rate Dosing rate t

1 t

2 t/s t

1

+ t

2

Figure 24 Dosing criterion "Dosing rate/Dosing time"

Volume which is dosed.

Input range

Default value

0.00000 to 99999.9 mL

10.0000 mL t/s

The rate at which dosing takes place.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum dosing rate depends on the cylinder volume

(see chapter 32, page 492) .

Time while dosing is carried out, i.e. the filling time is not taken into account.

Input range

Default value

0 to 999999 s

100 s

450 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Temperature

29 Parameters

More parameters

Temperature entered manually. The temperature is being continuously measured when a temperature sensor is connected and when Temp.

measurement = automatic or continuous is defined (see sensor dialog of the command). This value is used for temperature correction in pH measurements.

Input range

Default value

–20.0 to 150.0 °C

25.0 °C

Time interval MP

Time interval for entering a measuring point in the measuring point list.

The measuring point list is limited to 1000 measuring points.

Input range

Default value

0.1 to 999999.0 s

5.0 s

Pause

Waiting time, e.g. for the electrode to settle down after the start.

Input range

Default value

0 to 999999 s

0 s

Flow rate of a single dosing device

The cylinder volume and the filling time of the exchange unit or dosing unit must be taken into account for the calculation of the effective flow rate.

Table 21 Maximum flow rates for different cylinder volumes

Cylinder volume Maximum flow rate

1 mL

2 mL

5 mL

10 mL

20 mL

50 mL

Exchange unit approx. 80 mL/h

– approx. 400 mL/h approx. 800 mL/h approx. 1.6 L/h approx. 4 L/h

Dosing unit

– approx. 180 mL/h approx. 450 mL/h approx. 900 mL/h approx. 1.8 L/h approx. 4.5 L/h

Filling times

Filling times are not included in the calculation of the dosing rate by the

Titrando. The filling times (including valve switching) can be estimated according to the following formula:

900 Touch Control ■■■■■■■■ 451

29.15 Dosing and Liquid Handling ■■■■■■■■■■■■■■■■■■■■■■

:

:

:

Filling time in s

: maximum filling rate in mL/min current filling rate in mL/min

Filling time for one cylinder volume:

20 s (exchange unit)

18 s (dosing unit)

The maximum filling rate depends on the cylinder volume of the exchange unit / dosing unit attached

(see chapter 32, page 492) . The approximate

filling times (including valve switching) are given in the table below for various cylinder volumes (exchange unit and dosing unit) at different filling rates:

Cylinder volume

1 mL

2 mL

5 mL

10 mL

20 mL

50 mL max.

24 / –

– / 22

24 / 22

24 / 22

24 / 22

24 / 22

Filling time in s

100 mL/min 50 mL/min

34

28

64

10 mL/min

34

64

124

304

Application example

1 L reagent is to be dosed within 1 hour with a 50 mL exchange unit.

How large is the dosing rate?

Conditions:

: Total duration = 60 min

: Dosing volume = 1,000 mL

: Filling time = 24 s (assumption: max. filling rate)

:

Cylinder volume = 50 mL

Calculation of the number of filling procedures:

452 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

:

: maximum number of filling procedures of the dosing cylinder

Dosing volume = 1,000 mL

Cylinder volume = 50 mL

:

In accordance with this formula, the dosing cylinder must be filled 20 times.

If the division results in a whole number, then the last filling procedure need not be taken into account. In our example, the dosing cylinder must be filled 19 times. There is no dosing during this time, i.e. the actual dosing time is:

Calculation of the actual dosing time: effective dosing time

: Total duration = 60 min

: Number of required filling procedures of the dosing cylinder =

19

: Filling time = 24 s (assumption: max. filling rate)

According to this formula, the actual dosing time is 52.4 min.

This results in the following dosing rate:

Calculation of the required dosing rate:

900 Touch Control

: required dosing rate in mL/min

: Dosing volume = 1,000 mL

: Total duration = 60 min

: Number of required filling procedures of the dosing cylinder =

19

: Filling time = 24 s (assumption: max. filling rate)

According to this formula, dosing must proceed at a dosing rate of 19.1

mL/min.

Parameterization:

■■■■■■■■ 453

29.15 Dosing and Liquid Handling ■■■■■■■■■■■■■■■■■■■■■■

29.15.5.2 Stop conditions

The conditions for canceling the dosing are defined under [Stop conditions] .

Stop volume

The dosing is stopped when the specified volume has been dosed since the start of the dosing. This volume should be adjusted to the size of the titration vessel in order to prevent the contents from running over.

Input range

Selection

Default value

0.00000 to 99999.9 mL off off

Stop time

The dosing is stopped when the specified time has elapsed since the start of the dosing.

Input range

Selection

Default value

0 to 999999 s off off

Filling rate

Rate at which the dosing cylinder is filled.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum filling rate depends on the cylinder volume (see chapter

32, page 492) .

454 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

29.15.5.3 Monitoring

Under [Monitoring] , the parameters for monitoring the following quantities are defined:

Measured value

Temperature

Remote signals or RS-232 commands can be assigned to the monitored quantities

(see "Communication", page 458)

. These remote signals or

RS-232 commands are sent if a limit has been infringed. This can be used, for example, to switch a thermostat on/off if necessary.

The parameters are identical for all monitored quantities.

Measuring mode

Selection of the measuring mode to be monitored.

Selection

Default value pH pH

| U pH

Dosing with monitoring the pH value.

U

Dosing with monitoring the voltage.

Monitoring on | off (Default value: off )

Enabling and disabling the monitoring function.

Lower limit

When the monitored quantity falls below this value, the action defined in the following is triggered.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control device used.

Measuring mode pH:

Input range

Default value

–20.000 to 20.000

–20.000

Measuring mode U:

Input range

Default value

–2,000.0 to 2,000.0 mV

–2,000.0 mV

900 Touch Control ■■■■■■■■ 455

29.15 Dosing and Liquid Handling

Temperature:

Input range

Default value

Lower hysteresis

–20.0 to 150.0 °C

–20.0 °C

■■■■■■■■■■■■■■■■■■■■■■

The lower hysteresis represents the tolerance range for the lower limit (see figure 27, page 458)

. When Action = Wait the dosing does not continue until the monitored quantity has once again exceeded the lower limit incl.

the value entered here.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control device used.

Measuring mode pH:

Input range

Default value

0.000 to 20.000

0.020

Measuring mode U:

Input range

Default value

0.0 to 2,000.0 mV

2.0 mV

Temperature:

Input range

Default value

0.0 to 150.0 °C

0.2 °C

Upper limit

When this value is exceeded, the action defined in the following is triggered.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control device used.

Measuring mode pH:

Input range

Default value

–20.000 to 20.000

20.000

Measuring mode U:

Input range

Default value

–2,000.0 to 2,000.0 mV

2,000.0 mV

Temperature:

Input range

Default value

–20.0 to 150.0 °C

150.0 °C

456 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Upper hysteresis

29 Parameters

The upper hysteresis represents the tolerance range for the upper limit

(see figure 27, page 458) . When

Action = Wait , the dosing does not continue until the monitored quantity has once again fallen below the upper limit incl. the value entered here.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control device used.

Measuring mode pH:

Input range

Default value

0.000 to 20.000

0.020

Measuring mode U:

Input range

Default value

0.0 to 2,000.0 mV

2.0 mV

Temperature:

Input range

Default value

0.0 to 150.0 °C

0.2 °C

Action

Selection of the action which is carried out when the limits of the monitored quantity are infringed.

Selection None | Exit method

Pause | Wait

| Cancel command |

None

No action will be taken.

Exit method

The method is stopped immediately.

Cancel command

The current command is stopped und the next command in the method will be started.

Pause

Adding reagent is stopped until the dosing is continued manually with

[Continue] . If several quantities are monitored, the other quantities will still be monitored in the background.

Wait

Adding reagent is stopped until the value of the monitored quantity is within the limits again (incl. hysteresis). The dosing is continued automatically. If several quantities are monitored, the other quantities will still be monitored in the background.

900 Touch Control ■■■■■■■■ 457

29.15 Dosing and Liquid Handling

Monitored quantity

Stop

■■■■■■■■■■■■■■■■■■■■■■

Upper limit

458 ■■■■■■■■

Lower limit

Time

Figure 25 Action "Exit method" or "Cancel command"

Monitored quantity

Pause

Upper limit

[Continue]

Figure 26 Action "Pause"

Monitored quantity

Wait

Upper hysteresis

Lower limit

Time

Upper limit

Lower hysteresis

Lower limit

Time

Figure 27 Action "Wait"

Communication

Under [Communication] , remote signals or RS-232 commands can be assigned to the monitored quantities. These remote signals or RS-232

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters commands are always sent, independent of the action defined for a limit

infringement (see "Action", page 364) .

[New]

Define a new assignment.

[Delete]

Delete the selected assignment.

[Edit]

Edit the settings for the selected assignment.

Monitored quantity

Selection of the monitored quantity to which a remote signal or an

RS-232 command is to be assigned.

Selection Measured value | Temperature | any any

If several quantities are monitored, the remote signal or RS-232 command set is sent as soon as a limit of one of the monitored quantities is exceeded.

Infringed limit

Indicating the limit whose infringement causes the remote signal or

RS-232 command defined in the following to be sent.

Selection upper | lower | any | back in range upper

The remote signal or the RS-232 command is sent when the upper limit is exceeded.

lower

The remote signal or the RS-232 command is sent when the lower limit is exceeded.

any

The remote signal or the RS-232 command is sent when the upper or the lower limit is infringed.

back in range

The remote signal or the RS-232 command is sent when the value of the monitored quantity is within the limits again (incl. hysteresis).

Interface

Selection of the Remote Box or the serial interface via which the remote signal or the RS-232 command is sent.

Selection Remote Box 1…4 | COM1…8

900 Touch Control ■■■■■■■■ 459

29.15 Dosing and Liquid Handling

Output signal

■■■■■■■■■■■■■■■■■■■■■■

This parameter can only be edited when a Remote Box is selected under

Interface .

Selection of the signal out of the templates or entering the required bit pattern. Templates are defined under System ▶ Templates ▶ Output lines .

NOTICE

A line set active is not being reset automatically, nor at the end of the determination nor when the monitored quantity is in the range again.

Entering a bit pattern:

0 = line inactive

1 = line active

* = retain line status p = set pulse

The output lines are always numbered from right to left, i.e. with the signal *************1 line 0 is set. With a pulse, the length is set to 200 ms.

If it is required to set pulses with other lengths, you have to define a corresponding template.

Entry

Default value

Selection

Bit patterns containing exactly 14 characters or a max. of 24 characters for the name of the template

**************

Selection of the templates defined

Character string

This parameter can only be edited when a serial interface (COM) is selected under Interface .

Input of the RS-232 command as character string. All characters of the

ASCII code page can be used. Control characters (Esc, FF etc.) have to be entered as a three-place, decimal ASCII code, starting with a slash. Each command is automatically terminated with the ASCII characters CR and

LF .

Entry

Default value

ASCII string with a max. of 24 characters empty

460 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

29.15.5.4 Control instrument

The control instrument the determination is carried out with is selected under [Control device] . Control instruments are defined under System ▶ Device manager .

NOTICE

This button is only displayed when several control instruments have been configured.

Control device

Selection of the control instrument from the list of devices. Only those devices are displayed which are able to carry out the command. The control instruments do not have to be connected. This allows to transfer methods more easily from one system to another.

Selection Selection of configured control instruments

29.15.5.5 Sensor

The parameters for the sensor are edited under [Sensor] .

Measuring input

Selection of the measuring input the sensor is connected to. The selection is not dependent on whether the control device has one or two measuring interfaces.

Selection

Default value

1 | 2

1

Sensor

Selection of the sensor from the sensor list. The selection depends on the measuring mode. Sensors are defined under System ▶ Sensors . You can also enter a sensor name which is not contained in the sensor list. When a determination is started there is a check whether the sensor is contained in the sensor list.

Selection Selection of configured sensors

Temp. measurement

Type of temperature measurement.

Selection

Default value continuous automatic

| automatic | off

900 Touch Control ■■■■■■■■ 461

29.15 Dosing and Liquid Handling

29.15.5.6

■■■■■■■■■■■■■■■■■■■■■■ continuous

A temperature sensor must be connected. The temperature is measured continuously.

automatic

If a temperature sensor is connected then the temperature will be measured continuously. Otherwise, the temperature entered manually will be used (see dialog of the titration and measuring parameters).

off

The temperature will not be measured. The temperature entered manually is used (see dialog of the titration and measuring parameters).

Dosing device

The parameters for the dosing device are edited under [Dosing device] .

Dosing device

Selection of the MSB connector to which the dosing device is connected.

Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4

Titrant

Selection of the titrant from the titrant list. We recommend always selecting the titrant. This ensures that the correct data (titer, concentration, etc.) is always used for the calculation. Titrants are defined under System ▶ Titrants . You can also enter a name which is not contained in the titrant list. When a determination is started there is a check whether the solution is contained in the list.

For buret units with integrated data chip, a check is made in the method run to verify whether the correct titrant has been attached and whether the type of dosing drive matches. For buret units without an integrated data chip, the cylinder volume and the type of dosing drive are checked.

For the selected titrant, the validity of the titer, the working life of the titrant and the time interval for the GLP test of the buret unit are checked at the start of the determination.

Selection

Default value

Selection of configured titrants | not defined not defined not defined

No check takes place.

Tandem dosing

A second dosing device can be defined under [Tandem dosing] in order to enable uninterrupted dosing. Dosing is carried out with a combination

462 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters of two dosing devices so that the second dosing device is dosing while the first one is being filled and vice versa.

Dosing device

Selection of the MSB connector to which the dosing device is connected.

Every MSB connector will always be displayed.

Selection

Default value

1 | off

2 | 3 | 4 | off

Titrant

Selection of the titrant from the titrant list. We recommend always selecting the titrant. Titrants are defined under System ▶ Titrants . You can also enter a name which is not contained in the titrant list. When a determination is started there is a check whether the solution is contained in list.

For buret units with integrated data chip, a check is made in the method run to verify whether the correct titrant has been attached and whether the type of dosing drive matches. For buret units without integrated data chip, the cylinder volume and the type of dosing drive are checked. For the selected titrant, only the time interval for the GLP test of the buret unit is checked at the start of the determination.

NOTICE

Filling rate

The concentration, validity of the titer and the working life of the titrant are ignored.

Selection

Default value

Selection of configured titrants | not defined not defined not defined

No check takes place.

Rate at which the dosing cylinder is filled.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum filling rate depends on the cylinder volume

(see chapter

32, page 492) .

However, in order to ensure uninterrupted dosing, the following points must be taken into account:

900 Touch Control ■■■■■■■■ 463

29.15 Dosing and Liquid Handling ■■■■■■■■■■■■■■■■■■■■■■

Use the highest possible filling rate in order to keep the filling times as short as possible. The filling rate should be decreased for viscous liquids.

When you use two buret units with different dosing cylinders, the filling rate for the larger sized cylinder must be at the minimum:

= filling rate in mL/min for the larger sized cylinder

= filling rate in mL/min for the smaller cylinder

= cylinder volume in mL of the buret unit of the second dosing device

= cylinder volume in mL of the buret unit of the first dosing device

Example:

Dosing device 1: volume = 20 mL, filling rate = 50 mL/min

Dosing device 2: volume = 50 mL

Filling rate 2 ≥ 50 mL/min · 50 mL / 20 mL ≥ 125 mL/min

The dosing rate must not exceed 75% of the filling rate of the smaller cylinder. These values are listed in the following table, valid at maximum filling rate:

Table 22 Maximum dosing rate for different dosing cylinders

Cylinder volume maximum dosing rate

Exchange unit Dosing unit

1 mL

2 mL

5 mL

10 mL

20 mL

50 mL

2.25 mL/min

11.25 mL/min

22.50 mL/min

45.00 mL/min

112.50 mL/min

5.00 mL/min

12.50 mL/min

25.00 mL/min

50.00 mL/min

124.50 mL/min

464 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

29.15.5.7 Stirrer

The parameters for the stirrer are edited under [Stirrer] .

Stirrer

Selection of the MSB connector the stirrer is connected to. Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4 | off off

No stirrer will be used.

Stirring rate

Setting the stirring rate. It can be set in steps of –15 to +15. The default setting 8 corresponds to 1000 rpm. The formula for calculating the rotational speed is specified in

chapter 32.4, page 494

. The optimum stirring rate can be tested in the manual control.

The algebraic sign of the stirring rate changes the direction in which the stirring is done. When the stirrer is viewed from above, this means:

"+": counterclockwise rotation

"–": clockwise rotation

Input range

Default value

–15 to 15

8

Switch off automatically on | off (Default value: on )

If this parameter is activated, the stirrer will be switched off automatically at the end of the titration, measurement, etc.

29.16 Communication

The following communications commands are available:

Scanning remote lines (SCAN)

Defining remote signals which are awaited before the next method command is started.

Defining remote signals.

Setting remote lines (CTRL)

Receiving RS-232 commands (SCAN

RS)

Defining RS-232 commands which are awaited before the next method command is started.

900 Touch Control ■■■■■■■■ 465

29.16 Communication ■■■■■■■■■■■■■■■■■■■■■■

Sending RS-232 commands (CTRL

RS)

Defining the RS-232 commands which are sent.

29.16.1 Scanning remote lines (SCAN)

The command SCAN can be used to define input signals at the remote interface which are awaited before the next method command is started.

Control device

This parameter is only displayed when several control devices are configured. Control devices are defined under System ▶ Device manager .

Selection of the control device from the list of devices. The control devices do not have to be connected. This allows to transfer methods more easily from one system to another.

Selection Selection of configured control devices

Remote Box

Selection of the MSB connector the Remote Box is connected to. Every

MSB connector will always be displayed. The remote signal defined is requested on this Remote Box.

Selection

Default value

1 | 2 | 3 | 4

1

Input signal

Selection of the signal out of the templates or entering the required bit pattern. Templates are defined under System ▶ Templates ▶ Input lines .

Entering a bit pattern:

0 = line inactive

1 = line active

* = retain line status

The input lines are always numbered from right to left, i.e. with the signal

*******1 line 0 is expected to be active.

NOTICE

We recommend masking lines that are of no interest or for which no defined condition can be predicted with an asterisk ( * ).

466 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Entry

Default value

Selection

Bit patterns containing exactly 8 characters or a max. of 24 characters for the name of the template

********

Selection of the templates defined

Timeout

When this time interval has expired without recognizing the remote signal, the action defined in the following is triggered.

Input range

Default value

0 to 999 s

0 s

Action

Selection of the action which is carried out when the time interval has expired.

Selection

Default value

Display message | Document message | Cancel determination

Display message

For all three options it is documented in the determination data (see dialog More determination data / Messages ), that the time interval has been expired.

Display message

A message is displayed. You can select whether you want to continue with the determination or cancel the run.

Document message

In the determination data it will be documented, that the time interval has been expired.

Cancel determination

The determination is stopped.

29.16.2 Setting remote lines (CTRL)

The command CTRL can be used to define output signals at the remote interface.

Control device

This parameter is only displayed when several control devices are configured. Control devices are defined under System ▶ Device manager .

Selection of the control device from the list of devices. The control devices do not have to be connected. This allows to transfer methods more easily from one system to another.

Selection Selection of configured control devices

900 Touch Control ■■■■■■■■ 467

29.16 Communication

Remote Box

Output signal

■■■■■■■■■■■■■■■■■■■■■■

Selection of the MSB connector the Remote Box is connected to. Every

MSB connector will always be displayed. The remote signal defined is sent via this Remote Box.

Selection

Default value

1 | 2 | 3 | 4

1

Selection of the signal out of the templates or entering the required bit pattern. Templates are defined under System ▶ Templates ▶ Output lines .

NOTICE

A line set active is not being reset automatically, not even at the end of the determination.

Entering a bit pattern:

0 = line inactive

1 = line active

* = retain line status p = set pulse

The output lines are always numbered from right to left, i.e. with the signal *************1 line 0 is set. With a pulse, the length is set to 200 ms.

If it is required to set pulses with other lengths, you have to define a corresponding template.

NOTICE

We recommend masking lines that are of no interest or for which no defined condition can be predicted with an asterisk ( * ).

Entry

Default value

Selection

Bit patterns containing exactly 14 characters or a max. of 24 characters for the name of the template

**************

Selection of the templates defined

468 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

29.16.3 Scanning the RS-232 interface (SCAN RS)

The command SCAN RS can be used to define RS-232 commands which are awaited before the next method command is started.

Serial port

Selection of the serial interface the peripheral device is connected to. The

RS-232 command defined is requested on this interface.

Selection

Default value

COM1 | COM2 | COM3 | COM4 | COM5 |

COM6 | COM7 | COM8

COM1

Character string

Input of the RS-232 command as character string. All characters of the

ASCII code page can be used. Control characters (Esc, FF etc.) have to be entered as a three-place, decimal ASCII code, starting with a slash. Each command is automatically terminated with the ASCII characters CR and

LF . The * can be used for one or more characters.

Entry

Default value

ASCII string with a max. of 24 characters

!*.R

Timeout

When this time interval has expired without recognizing the RS-232 command, the action defined in the following is triggered.

Input range

Default value

0 to 999 s

0 s

Action

Selection of the action which is carried out when the time interval has expired.

Selection

Default value

Display message | Document message | Cancel determination

Display message

For all three options it is documented in the determination data (see dialog More determination data / Messages ), that the time interval has been expired.

Display message

A message is displayed. You can select whether you want to continue with the determination or cancel the run.

Document message

In the determination data it will be documented, that the time interval has been expired.

900 Touch Control ■■■■■■■■ 469

29.17 Automation ■■■■■■■■■■■■■■■■■■■■■■

Cancel determination

The determination is stopped.

29.16.4 Defining RS-232 commands (CONTROL RS)

The command CONTROL RS can be used to define the RS-232 commands which are sent.

Serial port

Selection of the serial interface the peripheral device is connected to. The

RS-232 command defined is sent via this interface.

Selection

Default value

COM1 | COM2 | COM3 | COM4 | COM5 |

COM6 | COM7 | COM8

COM1

Character string

Input of the RS-232 command as character string. All characters of the

ASCII code page can be used. Control characters (Esc, FF etc.) have to be entered as a three-place, decimal ASCII code, starting with a slash. Each command is automatically terminated with the ASCII characters CR and

LF .

Entry

Default value

ASCII string with a max. of 24 characters

&M;$G

29.17 Automation

29.17.1 Rotating sample rack (MOVE)

The command MOVE is used to move to a rack position or an external position.

Tower

Selection of the tower, with which the command is carried out. Both

Tower 1 and Tower 2 can always be selected, even if your Sample Processor has only one tower.

Selection

Default value

1 | 2

1

1

View from the front, the right tower.

2

View from the front, the left tower.

Destination

Selection of the required target position.

470 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Selection

Default value

Sample | Ext. position | Special beaker | Rack position | Current sample + | Current sample |

Next position | Prev. position | Calibration pos.

| rotate + | rotate | swing + | swing -

Sample

Sample

Rack position, which is defined by the sample variable. The sample variable is automatically increased by 1 after every method run or it can selectively be modified with the command SAMPLE .

Ext. position

One of the four external positions which can be moved to with the

Swing Head. Enter the number of the external position ( 1…4 ) in the field to the right.

Special beaker

Special beakers, which are defined in the rack table, can be directly moved to. Enter the number of the special beaker ( 1…16 ) in field to the right.

Rack position

Any rack position. Enter the number of the rack position ( 1…999 ) in the field to the right.

Current sample +

Starting from the current sample (defined by the sample variable) the rack is moved forward by the number of rack positions ( 1…999 ) entered in the field to the right.

Current sample -

Starting from the current sample (defined by the sample variable) the rack is moved backward by the number of rack positions ( 1…999 ) entered in the field to the right.

Next position

Starting from the current rack position, the rack is moved forward by one position.

Prev. position

Starting from the current rack position, the rack is moved backward by one position.

Calibration pos.

For automatic calibrations with a Sample Processor

(see "Definition of the calibration positions", page 498)

.

rotate +

Moving the rack forward by a certain increment. The rotation increment is defined in the properties of the tower.

rotate -

Moving the rack backward by a certain increment. The rotation increment is defined in the properties of the tower.

900 Touch Control ■■■■■■■■ 471

29.17 Automation ■■■■■■■■■■■■■■■■■■■■■■ swing +

Swinging the robotic arm outward by a certain increment (towards higher angular degrees). The swing increment is defined in the properties of the Swing Head.

swing -

Swinging the robotic arm towards the center of the rack by a certain increment (towards lower angular degrees). The swing increment is defined in the properties of the Swing Head.

Beaker test action

This parameter can only be edited with Destination = Sample , Next position or Prev. position .

Selection of the action which is carried out, when the beaker sensor does not detect any vessel at the sample position moved to. Additionally, enable the beaker sensor in the rack table.

Selection

Default value

Rotate rack | Display message

Display message

Rotate rack

The rack is moved to the next vessel on a sample position. Exception:

If the sample variable is newly defined in the method before this command (command SAMPLE ), then the option Display message is automatically valid.

Display message

A message is displayed. You can select whether you want to continue with the determination or cancel the run.

Options

Shift rate

Rate at which the sample rack is moved.

Input range

Default value

3 to 20 °/s

20 °/s

Shift direction

Direction in which the rack is moved.

Selection

Default value auto auto

| + | auto

A shift direction, with which the shorter way has to be passed, is automatically selected.

+

Counterclockwise rotation.

472 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

-

Clockwise rotation.

Swing rate

Rate at which the robotic arm is swung.

Input range

Default value

10 to 55 °/s

55 °/s

29.17.2 Moving the lift (LIFT)

The lift of a Sample Processor is moved with the command LIFT . The movement can however only be carried out if the Sample Processor is located at a valid rack position. This will not be the case, for example, after a Rack Reset.

Tower

Selection of the tower, with which the command is carried out. Both

Tower 1 and Tower 2 can always be selected, even if your Sample Processor has only one tower.

Selection

Default value

1 | 2

1

1

View from the front, the right tower.

2

View from the front, the left tower.

Lift position

Selecting a predefined lift position or enter any lift position. The predefined lift positions are defined in the device properties of the Sample Processor. They can however also be assigned in the manual control.

Input range

Selection

Default value

0 to 235 mm

Work position | Shift position | Rinse position

Special position | Home position

Work position

|

Home position

The lift is being located at the upper stop position, i.e. at the position 0 mm.

Lift rate

Rate at which the lift is moved.

Input range

Default value

5 to 25 mm/s

20 mm/s

900 Touch Control ■■■■■■■■ 473

29.17 Automation ■■■■■■■■■■■■■■■■■■■■■■

29.17.3 Controlling pumps (PUMP)

The PUMP command is used to control the pumps mounted on or connected to the tower of the sample changer.

Tower

Selection of the tower, with which the command is carried out. Both

Tower 1 and Tower 2 can always be selected, even if your Sample Processor has only one tower.

Selection

Default value

1 | 2

1

1

View from the front, the right tower.

2

View from the front, the left tower.

Pump

Selection of the pump.

Selection

Default value

1 | 2 | 1 + 2

1

1

Pump 1 of the selected tower is switched.

2

Pump 2 of the selected tower is switched.

1 + 2

Both pumps of the selected tower are switched at the same time.

Status/Duration

Switching the pump(s) on and off.

Input range

Selection

Default value

0 to 999 s (Increment: 1)

The pump is running during this time.

on | off on on

The pump is switched on. The pump is running until it is explicitly switched off. If the determination is canceled with the key [ ] , the pump is also switched off.

off

The pump is switched off.

474 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

29.17.4 Resetting the rack (RACK)

The following actions are carried out with the command RACK :

Rack, lift and robotic arm are being reset.

The rack code of the rack attached is being read out and the corresponding rack data is transferred to the Sample Processor.

The sample variable is being reset to the value 1.

Check rack on | off (Default value: off )

Activate this parameter when the rack attached is to be checked. For this you however have to additionally select the rack in the dialog Method options / Start options at the option Check rack .

29.17.5 Defining the sample variable (SAMPLE)

The sample variable describes the current position of the sample on the rack of the Sample Processor. It is automatically increased by 1 after the end of a method run. In the following cases, the sample variable is automatically reset to the value 1:

■ if the autostart counter is reset.

if the sample rack is reset.

The sample variable can be altered specifically with the command SAM-

PLE .

Sample variable

Modifying the sample variable.

Selection

Default value

=

+

| + | -

=

The sample variable corresponds to the number entered in the field

Value .

+

The sample variable corresponds to the current value plus the number entered in the field Value .

-

The sample variable corresponds to the current value minus the number entered in the field Value .

Value

Value by which the current sample variable is to be modified.

900 Touch Control ■■■■■■■■ 475

29.17 Automation ■■■■■■■■■■■■■■■■■■■■■■

NOTICE

The sample variable can only be set on rack positions which are not defined as special beakers.

Input range

Default value

1 to 999

1

29.17.6 Creating a subsequence (SUBSEQ)

It is useful, particularly with more complex tasks, e.g. rinsing sequences or

Liquid Handling, to combine individual command sequences into a single subsequence. Subsequences are managed the same as method commands. A maximum of 99 commands can be combined to form a single subsequence. There is no limit to the number of subsequences within one method.

[Load/ Save]

Loading a saved subsequence or saving the current subsequence.

[Subseq. options]

Defining various settings which correspond to the entire subsequence.

[Insert command]

Insert a new method command. It is inserted before the selected command.

NOTICE

Not all commands are available for subsequences. Commands which cannot be inserted into subsequences are disabled.

[Delete command]

Delete the selected method command.

[Edit command]

29.17.6.1

Edit the selected method command.

Loading / saving a subsequence

[Load/ Save] is used to load stored subsequences or to save new subsequences. This means that they are available for all methods and only have to be created once.

476 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Loading a subsequence

29 Parameters

[Own subseq.]

[Load]

Opening the list of own subsequences.

Loading the selected subsequence.

Saving a subsequence

Custom-made subsequences are saved in the internal memory.

NOTICE

In order to be able to use the subsequences also for other titration systems, you have to create a backup. You can only restore the subsequences from this backup on another system.

File name

File name of the subsequence.

Entry max. 32 characters

[Save]

29.17.6.2

Saving the subsequence to the internal memory.

Subsequence options

In this dialog, the settings can be defined which correspond to the entire subsequence.

900 Touch Control ■■■■■■■■ 477

29.17 Automation

Subsequence type

■■■■■■■■■■■■■■■■■■■■■■

Selection

Default value

Sample sequence | Start sequence | End sequence | Stop sequence | Cond. sequence

Sample sequence

Sample sequence

The subsequence is carried out with every determination.

Start sequence

This subsequence is only carried out at the start of a sample series when the autostart counter = 1.

End sequence

This subsequence is only carried out at the last sample of a series when the autostart counter has reached the setpoint value (number of autostarts).

Stop sequence

The stop sequence is only carried out when canceling a method because of one of the following cases: Manual stop with [ ] , stop because of an error, stop via remote signal.

Cond. sequence

This subsequence is carried out directly before the conditioning. This gives you the possibility to start e.g. a connected Polytron ® high-frequency homogenizer before the conditioning as well as to set its stirring rate. Use the command CONTROL RS for this.

Cycles

[Note]

[Display options]

Number of directly consecutive repetitions of the subsequence.

Input range

Default value

Selection

0 to 999

1

Calibration

Calibration

For automatic calibrations with a Sample Processor

(see chapter 32.7, page 497) .

Entering a note on the subsequence.

Dialog "Subsequence options / Note"

A short text can be entered in this dialog, e.g. for the description of the subsequence commands or for the application purpose.

Definition when the note is being displayed.

478 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Automatically after loading the subsequence on | off (Default value: off )

If this parameter is activated, the note is displayed when loading the subsequence. It can otherwise only be read in this dialog.

29.18 Miscellaneous commands

29.18.1 Controlling a stirrer (STIR)

A stirrer is controlled with the command STIR .

Control device

This parameter is only displayed when several control devices are configured. Control devices are defined under System ▶ Device manager .

Selection of the control device from the list of devices. The control devices do not have to be connected. This allows to transfer methods more easily from one system to another.

Selection Selection of configured control devices

Stirrer

Selection of the MSB connector the stirrer is connected to. Every MSB connector will always be displayed.

Selection

Default value

1

1

| 2 | 3 | 4

Status/Duration

Switching the stirrer on or off.

Input range

Selection

Default value

0 to 999 s (Increment: 1)

During this time the stirrer is running.

on | off on on

The stirrer is switched on. The stirrer runs until it is explicitly switched off. If the determination is canceled with the key [ ] , the stirrer is also switched off.

off

The stirrer is switched off.

Stirring rate

Setting the stirring rate. It can be set in steps of –15 to +15. The default setting 8 corresponds to 1000 rpm. The formula for calculating the rota-

900 Touch Control ■■■■■■■■ 479

29.18 Miscellaneous commands ■■■■■■■■■■■■■■■■■■■■■■

tional speed is specified in chapter 32.4, page 494 . The optimum stirring

rate can be tested in the manual control.

The algebraic sign of the stirring rate changes the direction in which the stirring is done. When the stirrer is viewed from above, this means:

"+": counterclockwise rotation

"–": clockwise rotation

Input range

Default value

–15 to 15

8

29.18.2 Pausing the method run (WAIT)

The method run can be paused with the command WAIT .

Hold sequence on | off (Default value: off )

If this parameter is activated, then the method run is paused until it is continued manually.

Waiting time

This parameter can only be edited when Hold sequence is deactivated.

The method run is automatically continued after this waiting time.

Input range

Default value

0 to 99999 s

30 s

Message on | off (Default value: off )

If this parameter is activated, the text message defined in the following is displayed during the waiting time.

Message text

This parameter can only be edited when Message is activated.

Text which is displayed during the waiting time.

Entry

Default value max. 28 characters empty

29.18.3 Scan data (REQUEST)

The following data can be scanned in the method run with the command

REQUEST :

480 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Sample data

– Sample identification 1

– Sample identification 2

– Sample size (value and unit)

Common variable

You can select thereby whether the sequence should be paused or resumed in the background.

Sample ident.

Selection of the sample identification that is queried in the method run.

Selection

Default value off

ID2 off

| Identification 1 | Identification 2 | ID1 &

Sample size on | off (Default value: on )

If this parameter is activated, then the value for the sample size will be requested.

Sample size unit on | off (Default value: off )

If this parameter is activated, then the unit for the sample size will be requested.

Common variable

Selection of the common variable which is queried in the method run.

Selection

Default value

CV01…CV25 off

| off

Hold sequence on | off (Default value: on )

If this parameter is activated, then the run will be paused during the request. If this parameter is deactivated, then the method continues in the background until the following measurement or titration has been completed.

29.18.4 Defining an acoustic signal (BEEP)

The BEEP command can be used to produce an acoustic signal.

Duration

Approximate playing time in seconds of the permanently stored melody.

900 Touch Control ■■■■■■■■ 481

29.18 Miscellaneous commands ■■■■■■■■■■■■■■■■■■■■■■

Input range

Default value

1 to 9

1

29.18.5 Signing a determination (SIGN)

A determination can be signed with the SIGN command. The method run is automatically paused as soon as this command is reached. At the same time a switch is made to the result view. The run will not be continued until the determination has been signed or the SIGN command has been canceled. You can specify whether a routine user is allowed to cancel the

SIGN command in the setting System ▶ System settings ▶ Dialog options ▶ Routine dialog . No parameters can be edited for this com-

mand. The parameters for signing determinations are described in chapter

17.1, page 184

.

NOTICE

Methods with this command can only be started if you are working with login and password protection.

The command should always be inserted at the end directly before the

REPORT command or at the end of the method sequence if a PC/LIMS report is to be sent automatically.

29.18.6 Canceling the method run (END)

The method run will cancel as soon as the END command is reached. This is a good idea if you wish to test only the first part of a method. No parameters can be edited for this command.

482 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 30 Operation and maintenance

30 Operation and maintenance

30.1 System initialization

In very rare instances, a faulty file system (e.g. because of a program crash) may lead to an impairment of program functioning. The internal file system must be initialized in such cases.

CAUTION

If you carry out a system initialization, then all user data except the methods will be deleted.

We recommend creating a backup of the system at regular intervals in order to avoid data losses.

Proceed as follows for the system initialization:

1 Switching off the instrument

Press the power switch at the rear.

2 Switching on the instrument

Press the power switch at the rear.

An image with a frog will be displayed after a couple of seconds.

While the image is being displayed, press the power switch again and hold it down.

Do not release the power switch until an acoustic signal sounds.

A system message is displayed.

3 Confirm system initialization

Confirm the message with [Yes] .

A second safety prompt is displayed.

Confirm this message with [Yes] as well.

Initialization is started. After the initialization has been completed, the 900 Touch Control is started automatically.

900 Touch Control ■■■■■■■■ 483

31.1 Editing methods ■■■■■■■■■■■■■■■■■■■■■■

31 Troubleshooting

31.1 Editing methods

Problem

The method command cannot be inserted.

The additional evaluations (EVAL commands) cannot be inserted.

An EVAL command is depicted in red in the command list.

Cause

The method command is disabled.

No control instrument which supports the method command is configured in the device manager.

A titration or measuring command to which the evaluation can be applied is missing.

The associated titration or measuring command has been deleted or replaced by a different one, e.g. DET pH by DET U.

Remedy

In the dialog Dialog options / Command list

, activate the desired command (see "Blocking unneeded commands and fixed keys", page 25) .

Add a control instrument which supports the desired method command to the list of devices

(see chapter 11, page 79) .

1. Insert a titration or measuring command.

2. Insert the EVAL command directly after it.

1. Delete the EVAL command.

2. Insert a titration or measuring command.

3. Insert the EVAL command directly after it.

31.2 Sample table

Problem

The method cannot be entered in the sample table.

Cause

The option Use sample assignment table is activated. The method assigned to the sample identification will be loaded automatically at the start of the determination.

Remedy

Deactivate the option

(see chapter 15.1, page

147)

.

31.3 Results/Statistics

Problem

No statistics are carried out for a result.

Cause

Statistics is not activated.

Remedy

In the method options, activate the option

Statistics

(see chapter 16.5, page 172)

.

484 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 31 Troubleshooting

Problem Cause

No statistics variable has been assigned to the calculation.

Statistics is not activated.

Assignment is not carried out automatically with retroactive modifications.

Remedy

In the result options (command CALC ), define a variable for the mean value

(see "Variable for mean value", page 153)

.

In the control dialog, activate the Statistics

option (see chapter 18, page 186)

.

Recalculate the determination manually (see chapter 24, page 214)

.

A result has been removed from the statistics, the mean value has not been reassigned to the common variable or to the TITER variable.

The result is not displayed in the result table.

The column display in the result table is incorrectly configured.

The parameter Save result in result table is not activated.

Modify the settings for the display accordingly

(see chapter 26.1, page 237)

.

In the result options (command CALC ), activate the parameter

(see "Save result in result table", page 154)

.

31.4 Printing

Problem

Tapping on the fixed key [ ] has no effect.

Cause

The fixed key is disabled.

Remedy

In the Dialog options / Fixed keys dialog, activate the Print

option (see chapter 7.2, page 24)

.

A determination is running.

Wait until the determination is finished.

31.5 Manual control

Problem

Tapping on the fixed key [ ] has no effect.

A button in the manual control is disabled.

Cause

The fixed key is disabled.

The required hardware is either not connected or incorrectly connected.

Remedy

In the Dialog options / Fixed keys dialog, activate the Manual control option

(see chapter 7.2, page 24)

.

1. Switch off the 900 Touch Control.

2. Connect the hardware correctly.

3. Switch the 900 Touch Control back on.

900 Touch Control ■■■■■■■■ 485

31.6 Miscellaneous

Problem Cause

The required hardware is being used in the ongoing determination.

■■■■■■■■■■■■■■■■■■■■■■

Remedy

Wait until the determination is finished.

31.6 Miscellaneous

Problem

The 900 Touch Control cannot be switched on.

Cause

The Touch Control is connected to an MSB socket on the control instrument.

Remedy

Plug in the Touch Control at the Controller connector (see manual for control instrument).

31.7 SET titration

Problem

The titration will not be finished.

Cause

The minimum dosing rate is too low.

Remedy

Define Titration rate = user and increase the minimum rate ( Min. rate )

(see chapter 29.3.2, page 312) .

The stop criterion is unsuitable.

The sample is overtitrated.

The control parameters are unsuitable.

Adjust the control parameters

(see chapter

29.3.2, page 312)

:

Increase the stop drift.

Select a short delay time.

Adjust the control parameters

(see chapter

29.3.2, page 312)

:

Select Titration rate = slow .

Define Titration rate = user and increase the control range.

Define Titration rate = user and reduce the maximum rate ( Max. rate ).

Define Titration rate = user and reduce the minimum rate ( Min. rate ).

Stir faster.

Arrange the electrode and buret tip to an optimum .

Replace the electrode.

The titration time is too long.

The electrode responds too slowly.

The control parameters are unsuitable.

Adjust the control parameters

(see chapter

29.3.2, page 312)

:

Select Titration rate = optimal or fast .

486 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Problem

The results are spread widely.

31 Troubleshooting

Cause

Remedy

Define Titration rate = user and reduce the control range.

Define Titration rate = user and increase the maximum rate ( Max. rate ).

Define Titration rate = user and increase the minimum rate ( Min. rate ).

The minimum dosing rate is too high.

Define Titration rate = user and reduce the minimum rate ( Min. rate

) (see chapter 29.3.2, page 312)

.

The control range is too small.

The electrode responds too slowly.

Define Titration rate = user and increase the control range.

Replace the electrode.

31.8 Volumetric Karl Fischer Titration

Problem

The drift is very high during conditioning.

Cause Remedy

The titration cell is leaking.

Check the seals and the septum. Replace if necessary.

Replace the molecular sieve.

The drift becomes greater after each titration.

The titration will not be finished.

The sample releases water very slowly.

A side reaction is taking place.

Adjust the method.

Add solubility promoter.

Increase the temperature (possibly using a

KF oven).

See technical literature.

Use special reagents.

Adjust the method (increase/decrease the temperature, external extraction).

See technical literature.

The pH value is no longer in the optimum range.

Add buffer, see technical literature.

The titration cell is leaking.

Check the seals and the septum. Replace if necessary.

Replace the molecular sieve.

The minimum increment is too low.

Define Titration rate = user and increase the minimum volume increment ( Min. increment )

(see chapter 29.4.2, page 325) .

900 Touch Control ■■■■■■■■ 487

31.8 Volumetric Karl Fischer Titration ■■■■■■■■■■■■■■■■■■■■■■

Problem

The sample is overtitrated.

The solution becomes darker after each titration.

The endpoint is reached too quickly.

The titration times with volumetric titration are constantly longer.

Cause

The stop criterion is unsuitable.

Remedy

Adjust the control parameters

(see chapter

29.4.2, page 325)

:

Increase the stop drift.

Select a short delay time.

See also: The drift becomes greater after each titration.

The increments at the end of the titration are too high.

Define Titration rate = user and reduce the dosing rate ( Max. rate

) (see chapter

29.4.2, page 325)

.

The following experiment provides a clue for the optimum dosing rate: During conditioning, display the drift and add sample without starting the titration. Select a value below the highest drift as dosing rate.

Stir faster.

The amount of methanol in the working medium is too low.

The electrode may be covered.

Replace the working medium.

Reduce the amount of solubility promoter, if working with solvent mixtures, see technical literature.

Wipe off the electrode with ethanol or a suitable solvent.

The electrode may be covered.

The electrode has a short circuit.

The dosing rate outside the control range is too high.

The buffer capacity of the solvent can be exhausted for two-component reagents.

Replace the working medium.

Wipe off the electrode with ethanol or a suitable solvent.

1. Check the Pt wires.

2. Activate the electrode check.

Define Titration rate = user and reduce the dosing rate ( Max. rate

) (see chapter 29.4.2, page 325) .

Replace the working medium.

488 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 31 Troubleshooting

31.9 Coulometric Karl Fischer titration

Problem

The drift is very high during conditioning.

Cause Remedy

The titration cell is leaking.

Check the septum and replace it if necessary.

Replace the molecular sieve.

Check the ground-joint sleeves for frayed edges. If necessary, trim the edges neatly or replace the ground-joint sleeves.

The drift becomes greater after each titration.

The drift is fluctuating.

Deposits containing water can be found in the titration cell.

The reagent is exhausted or contaminated.

The catholyte is old or moist.

Shake the titration cell.

Replace the reagent.

The diaphragm of the generator electrode is contaminated.

A side reaction is taking place.

The molecular sieve on the

KF oven has reached its capacity limit.

The gas flow from the KF oven into the titration cell is too high.

The sample releases water very slowly.

The catholyte must be replaced at least once per week.

Dry the catholyte with KF one-component reagent.

Clean the diaphragm (in accordance with directions in the leaflet for the generator electrode).

Consider using a KF oven.

See technical literature.

Replace the molecular sieve.

Reduce the gas flow (if you are working with the oven method with sample vials, set the gas flow between 40 and 60 mL/min).

Consider using a KF oven.

See technical literature.

The sample solution is poorly stirred.

Set the stirring rate in such a way that the sample solution will become very well mixed.

900 Touch Control ■■■■■■■■ 489

31.9 Coulometric Karl Fischer titration

Problem

The titration time is too long.

The result is too high.

The result is too low.

■■■■■■■■■■■■■■■■■■■■■■

Cause

The Control parameters are set incorrectly.

The drift during conditioning is not stable.

The parameter Start drift is set too high.

Remedy

Reset the Control parameters to default values.

Wait before starting the titration until the drift has become stable.

Set the Start drift lower.

Adjust the Stop drift or use the Relative stop drift .

The parameter is set too low.

is set too low.

Stop drift

The water quantity is too high.

The Control parameters are set incorrectly.

The titration cell is not yet correctly conditioned.

The sample contains oxidizable substances.

The parameter Stop drift

The drift correction is too small (e.g. for the use of a

KF oven or for manual drift correction).

Set the Stop drift higher or use the Relative stop drift .

Observe the sample size and water contents in accordance with .

Set the parameter Dynamics lower and the parameter Max. rate higher.

Shake the titration cell and wait until the drift is stable.

Consider using a KF oven.

See technical literature.

Set the Stop drift higher or use the Relative stop drift .

Start the titration as long as the gas flow still exists between the conditioning vial and the titration cell.

Set the parameter Drift correction to auto (= automatic drift correction).

The sample releases iodine.

Consider using a KF oven.

See technical literature.

The parameter Stop drift is set too high.

The drift correction is too large (e.g. with excessively high start drift, with fluctuating drift or with manual drift correction).

Set the Stop drift lower.

Ensure stable and lower drift.

Set the parameter Drift correction to auto (= automatic drift correction).

490 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Problem

The results are spread widely.

31 Troubleshooting

Cause

The parameter Min. rate is set too low.

Remedy

Set the parameter Min. rate higher.

Reset the Control parameters to default values.

The sample is nonhomogenous.

Homogenize the sample if possible.

Possibly use a larger sample size.

The sample addition is difficult to reproduce.

The drift fluctuates.

Observe the instructions in .

Ensure stable drift.

31.10 STAT titration

Problem

The control point is not kept stable. The measured value is at times too high and at times too low.

The controller "oscillates".

Cause

The control parameters are unsuitable.

The stirring is not optimal.

The arrangement of electrode and buret tip is not optimum.

The cylinder volume of the exchange/dosing unit might be too large.

The control point is not reached for too long.

The control parameters are unsuitable.

Remedy

Adjust the control parameters (see chapter

29.7.2, page 356) :

Reduce the Max. rate and possibly also the Min. rate .

Increase control range .

Check and optimize efficiency of the stirring.

Arrange the electrode and buret tip to an optimum (see manual of the instrument).

Use an exchange/dosing unit with a smaller cylinder volume (yields a smaller volume increment per pulse).

Adjust the control parameters (see chapter

29.7.2, page 356) :

Reduce control range .

increase Min. rate .

increase Max. rate .

900 Touch Control ■■■■■■■■ 491

32.1 Manual Dosing Controller, optional accessory for manual titration ■■■■■■■■■■■■■■■■■■■■■■

32 Appendix

32.1 Manual Dosing Controller, optional accessory for manual titration

Figure 28 Manual Dosing Controller 6.2107.120 – Overview

1 Start/Dosing key [GO]

For starting the method and for dosing during the MAT command.

3 Cancel command button [FILL]

For canceling the running command.

2 Status LED

Indicates whether the instrument is ready.

4 Connector plug

With 25-pin D-sub plug.

492 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 32 Appendix

32.2 Dosing unit

32.2.1 Maximum dosing and filling rate

The maximum dosing rate and maximum filling rate for the dosing unit depend on the cylinder volume:

Cylinder volume

2 mL

5 mL

10 mL

20 mL

50 mL

Maximum rate

6.67 mL/min

16.67 mL/min

33.33 mL/min

66.67 mL/min

166.00 mL/min

Independent of the cylinder volume, values ranging from 0.01 to 166.00

mL/min can always be entered. When the function is carried out the rate will be, if necessary, decreased automatically to the highest possible value.

32.2.2 Default parameters for preparing (PREP) and emptying (EMPTY)

The PREP command is used to rinse and fill the cylinder and tubings of the dosing unit air bubble-free. You should carry out this function before the first determination or once per day.

The command EMPTY is used to empty the cylinders and the tubings of the dosing unit.

If no titrant is selected in the command, preparing and emptying will be carried out with the following default parameters:

The entire cylinder volume is dosed at the maximum dosing rate through Port 1.

The cylinder is filled at the maximum filling rate through Port 2.

The following dimensions apply for the tubings:

– Tubing at Port 1: length = 40.0 cm, diameter = 2 mm

– Tubing at Port 2: length = 25.0 cm, diameter = 2 mm

900 Touch Control ■■■■■■■■ 493

32.3 Exchange unit ■■■■■■■■■■■■■■■■■■■■■■

32.3 Exchange unit

32.3.1 Maximum dosing and filling rate

The maximum dosing rate and maximum filling rate for the exchange unit depend on the cylinder volume:

Cylinder volume

1 mL

5 mL

10 mL

20 mL

50 mL

Maximum rate

3.00 mL/min

15.00 mL/min

30.00 mL/min

60.00 mL/min

150.00 mL/min

Independent of the cylinder volume, values ranging from 0.01 to 166.00

mL/min can always be entered. When the function is carried out the rate will be, if necessary, decreased automatically to the highest possible value.

32.3.2 Default parameters for preparing (PREP)

The PREP command is used to rinse and fill the cylinder and tubings of the exchange unit air bubble-free. You should carry out this function before the first determination or once per day.

If no titrant is selected in the command, preparing will be carried out with the following default parameters:

The entire cylinder volume is dosed twice at the maximum dosing rate.

The cylinder is filled at the maximum filling rate.

32.4 Stirring rate

The stirring rate can be adjusted in steps from –15 to +15.

The approximate rotational speed for the internal magnetic stirrer

(depends on the product version) can be calculated with the following formula:

Rotational speed/min (r/min) = 125 · Stirring rate

Example:

Configured stirring rate: 8

Rotational speed in revolutions per minutes = 125 · 8 = 1,000

494 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 32 Appendix r/min

2000

1500

1000

500

-15 -10 -5 0 5 10 15

Figure 29 Rotational speed depending on the stirring rate

The information on the separately connectable 802 propeller stirrer can be found in the "802 Stirrer" manual.

32.5 Balance

The sample size and the associated unit can be sent from a connected balance. The sample size is transmitted as a number with up to ten characters

(including algebraic sign and decimal point).

Sample size and unit are sent as a single character string. They are separated by a space character. The string is terminated with the ASCII characters CR and LF .

If the balance sends a negative sample size (e.g. when you are reweighing a sample), then the algebraic sign is adopted. The algebraic sign is, however, ignored for the calculations.

NOTICE

With some balances, the sample identification and the method can be sent in addition to the sample size.

Make sure that the balance does not send the sample size until the end.

Mettler AX

For the Mettler AX balance, the fields that contain the sample identification or the method must be designated as follows:

Designation for the field with the method name: METHOD

Designation for the field with sample identification 1: ID1

Designation for the field with sample identification 2: ID2

900 Touch Control ■■■■■■■■ 495

32.6 Result variables as parameter setting ■■■■■■■■■■■■■■■■■■■■■■

32.6 Result variables as parameter setting

For most of the numerical parameters, the result of a calculation can also be utilized in place of a number. This result must be calculated in a preceding calculation command (command CALC ). A typical application is the use of a relative start volume.

The following example is used to explain how this method is used to define a sample size-dependent start volume for a titration. Proceed as follows:

1 Create a new calculation

Insert the calculation command CALC in front of the titration command.

Enter the calculation formula, e.g. R1 = C00*3.

2 Define a result variable as parameter setting

In the titration command, enter the variable R1 as the start volume.

496 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 32 Appendix

The selection of the result variables is displayed in the number editor.

NOTICE

If the method before the command with the result variables applied contains more than one calculation commands (possibly with result variables bearing the same name), then the result variable of the preceding calculation command will always be used.

Please note that the result must be within the input range of the parameter, as otherwise the determination run will be stopped when the parameter is accessed.

32.7 Electrode calibration with Sample Processors

Basic framework for a method for automatic calibrations

If the calibration of electrodes is carried out with a Sample Processor, then the basic framework of the method will appear as follows:

1.

Subsequence with the commands required for the calibration (command SUBSEQ ) a. Command for change of buffer/standard (command MOVE ) b. Command for moving the lift (command LIFT ) c. Calibration command (command CAL pH or CAL Conc ) d. …

2.

In the interest of general comprehensibility, other commands which are required are not listed here. A method template is available for an automatic pH calibration with subsequent measurement.

900 Touch Control ■■■■■■■■ 497

32.7 Electrode calibration with Sample Processors ■■■■■■■■■■■■■■■■■■■■■■

Command parameterization

The following special settings are required with the individual commands:

Command SUBSEQ :

The setting Cycles = Calibration must be adjusted in the Subsequence options dialog.

Command MOVE for changing the buffer/standard:

The only definition made in this command is that a calibration position is to be moved to. The setting Destination = Calibration pos.

must be adjusted for this purpose. The question of which rack positions are used as calibration positions must be defined in the rack table (see below).

Command CAL pH or CAL Conc :

The setting Sample Processor = USB must be adjusted in the Calibration parameters dialog.

Definition of the calibration positions

The positions of the individual buffers/standards on the rack are defined in the rack table of the rack used. The following options are available:

Utilization of sample positions

The desired rack position for the first buffer/standard must be entered in the dialog Sample rack / Edit rack data for the parameter 1. Calibration pos.

. In this case, it is imperative that the remaining buffers/ standards be placed on the rack positions directly following.

Utilization of special beakers

Any rack position can be defined as a special beaker. It is, however, preferable to set them at high rack positions in order to be able to begin sample series at rack position 1. The special beakers are moved to in ascending order.

The list of special beakers is displayed in the dialog Sample rack /

Edit rack data by pressing [Special beakers] . The number of buffers/standards used for calibration will determine the number of special beakers to be defined. We recommend defining special beakers 1-n.

Afterwards, in the dialog Sample rack / Edit rack data , for the parameter 1. Calibration pos.

, the option Special beaker 1 must be selected for the first buffer/standard.

498 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 32 Appendix

32.8 Stored buffer series for pH calibration

The temperature-dependent pH values of several common pH buffers are stored in the Touch Control for automatic buffer recognition during pH calibration. In addition to the Metrohm buffer solutions, other reference buffers and technical buffers are also included in the tables.

The following tables provide an overview of the stored pH(T) series.

pH values printed in bold are the values for the reference temperature of the respective buffer set.

pH values printed in italics are interpolated or extrapolated values, all the others correspond to the manufacturer's specifications.

32.8.1 Metrohm

Table 23 Metrohm buffer solutions

4.09

4.11

4.13

4.15

4.18

4.03

4.04

4.06

4.07

4.00

4.00

4.01

4.02

pH

4.00

3.99

3.99

3.99

3.99

3.99

65

70

75

80

85

45

50

55

60

25

30

35

40

Temp.

(°C)

0

5

10

15

20

6.97

6.97

6.97

6.97

7.00

6.99

6.98

6.98

Metrohm pH

7.00

7.11

7.08

7.06

7.04

7.02

6.98

6.98

6.99

7.00

7.00

8.76

8.74

8.73

8.71

8.70

8.87

8.84

8.81

8.79

9.00

8.96

8.93

8.90

pH

9.00

9.27

9.18

9.13

9.08

9.04

900 Touch Control ■■■■■■■■ 499

32.8 Stored buffer series for pH calibration

Temp.

(°C)

90

95

NOTICE pH

4.00

4.20

4.23

■■■■■■■■■■■■■■■■■■■■■■

Metrohm pH

7.00

7.01

7.02

pH

9.00

8.68

8.67

Update

The values of the individual buffers with the corresponding temperatures are kept up to date as far as possible.

However, they may be changed by the respective manufacturers.

32.8.2 NIST (according to DIN standard 19266, 2015)

Table 24 NIST buffer solutions

NIST (according to DIN standard 19266, 2015-05)

1.694

1.700

1.707

1.715

1.723

1.733

1.743

1.754

1.766

pH

1.679

1.666

1.668

1.670

1.672

1.675

1.679

1.683

1.688

60

65

70

75

80

40

45

50

55

20

25

30

35

Temp.

(°C)

0

5

10

15

6.838

6.836

6.833

6.834

6.836

6.841

6.845

6.852

6.859

pH

6.865

6.984

6.951

6.923

6.900

6.881

6.865

6.853

6.844

4.027

4.038

4.050

4.075

4.091

4.108

4.126

4.145

4.164

pH

4.005

4.000

3.998

3.997

3.998

4.000

4.005

4.011

4.018

9.068

9.040

9.011

8.985

8.962

8.942

8.921

8.903

8.885

pH

9.180

9.464

9.395

9.332

9.276

9.225

9.180

9.139

9.102

11.984

11.841

11.705

11.574

11.449

-

-

-

pH

12.454

-

13.207

13.003

12.810

12.627

12.454

12.289

12.133

500 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 32 Appendix

Temp.

(°C)

85

90

95

NIST (according to DIN standard 19266, 2015-05) pH

1.679

1.779

pH

4.005

4.184

pH

6.865

6.868

pH

9.180

8.868

pH

12.454

-

1.792

1.806

4.205

4.227

6.877

6.886

8.850

8.833

-

-

NOTICE

Update

The values of the individual buffers with the corresponding temperatures are kept as up to date as possible.

However, they may be changed by the respective manufacturers.

The NIST buffers are identical to the buffer solutions that are used in

Chinese pharmacopoeia.

32.8.3 DIN (according to DIN standard 19267, 2012)

Table 25 DIN buffer solutions

DIN (according to DIN standard 19267, 2012-08)

1.09

1.09

1.10

1.10

1.10

1.10

1.11

pH

1.09

1.08

1.08

1.09

1.09

1.11

1.11

1.11

20

25

30

35

40

45

50

Temp.

(°C)

0

5

10

15

55

60

65

4.65

4.65

4.65

4.66

4.66

4.67

4.68

pH

4.65

4.67

4.66

4.66

4.65

4.69

4.70

4.71

3.07

3.06

3.05

3.05

3.04

3.04

3.04

pH

3.06

-

-

3.10

3.08

3.04

3.04

3.04

6.80

6.79

6.78

6.77

6.76

6.76

6.76

pH

6.79

6.89

6.86

6.84

6.82

6.76

6.76

6.76

9.27

9.23

9.18

9.13

9.09

9.04

9.00

pH

9.23

9.48

9.43

9.37

9.32

8.97

8.92

8.90

12.96

12.75

12.61

12.44

12.29

12.13

11.98

pH

12.75

-

-

13.37

13.15

11.84

11.69

11.56

900 Touch Control ■■■■■■■■ 501

32.8 Stored buffer series for pH calibration

Temp.

(°C)

70

75

80

85

90

95

■■■■■■■■■■■■■■■■■■■■■■ pH

1.09

1.11

1.12

1.12

1.12

1.13

-

DIN (according to DIN standard 19267, 2012-08) pH

3.06

3.04

pH

4.65

4.72

pH

6.79

6.76

pH

9.23

8.88

pH

12.75

11.43

3.04

3.05

3.06

3.07

-

4.74

4.75

4.77

4.79

-

6.77

6.78

6.79

6.80

-

8.86

8.85

8.83

8.82

-

11.30

11.19

11.08

10.99

-

NOTICE

Update

The values of the individual buffers with the corresponding temperatures are kept as up to date as possible.

However, they may be changed by the respective manufacturers.

32.8.4 Fisher

Table 26 Fisher buffer solutions

20

25

30

35

Temp.

(°C)

0

5

10

15

40

45

50

55

60

2.00

2.00

2.00

2.02

pH

2.00

-

1.98

1.98

2.02

2.01

2.01

2.01

-

-

4.00

4.00

4.01

4.02

pH

4.00

4.01

3.99

4.00

3.99

4.03

4.04

4.06

4.07

4.09

Fisher

7.02

7.00

6.99

6.98

pH

7.00

7.13

7.10

7.07

7.05

6.97

6.97

6.97

6.97

6.98

pH

10.00

10.34

10.26

10.19

10.12

10.06

10.00

9.94

9.90

9.85

9.81

9.78

9.74

9.70

502 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 32 Appendix

Temp.

(°C)

65

70

75

80

85

90

95 pH

2.00

-

-

-

-

-

-

-

NOTICE pH

4.00

4.11

4.13

4.14

4.16

4.18

4.21

4.23

Fisher pH

7.00

6.99

7.00

7.02

7.03

7.06

7.08

7.11

pH

10.00

9.68

9.65

9.63

9.62

9.61

9.60

9.60

Update

The values of the individual buffers with the corresponding temperatures are kept up to date as far as possible.

However, they may be changed by the respective manufacturers.

32.8.5 Fluka

Table 27 Fluka buffer solutions

Temp.

(°C)

0

5

10

15

20

25

30

35

40

45

50

55 pH

4.000

4.030

4.025

4.020

4.010

4.000

4.000

4.000

4.000

4.000

4.000

4.000

4.000

Fluka pH

7.000

7.130

7.090

7.050

7.020

7.000

6.990

6.980

6.975

6.970

6.965

6.960

6.960

pH

9.000

9.240

9.175

9.110

9.055

9.000

8.965

8.930

8.895

8.860

8.830

8.800

8.775

900 Touch Control ■■■■■■■■ 503

32.8 Stored buffer series for pH calibration

Temp.

(°C)

60

65

70

75

80

85

90

95

NOTICE pH

4.000

4.000

4.000

4.000

4.000

4.000

4.000

4.000

4.000

Fluka pH

7.000

6.960

6.965

6.970

6.975

6.980

6.990

7.000

7.010

■■■■■■■■■■■■■■■■■■■■■■ pH

9.000

8.750

8.730

8.710

8.690

8.670

8.655

8.640

8.620

Update

The values of the individual buffers with the corresponding temperatures are kept as up to date as possible.

However, they may be changed by the respective manufacturers.

32.8.6 Mettler Toledo

Table 28 Mettler Toledo buffer solutions

10

15

20

25

Temp.

(°C)

0

5

30

35

40

45

50

2.01

2.00

2.00

2.00

pH

2.00

2.03

2.02

1.99

1.99

1.98

1.98

1.98

4.00

4.00

4.00

4.01

pH

4.01

4.01

4.01

4.01

4.02

4.03

4.04

4.06

Mettler Toledo pH

7.00

7.12

7.09

7.06

7.04

7.02

7.00

6.99

6.98

6.97

6.97

6.97

9.38

9.32

9.26

9.21

pH

9.21

9.52

9.45

9.16

9.11

9.06

9.03

8.99

pH

11.00

11.90

11.72

11.54

11.36

11.18

11.00

10.82

10.64

10.46

10.28

10.10

504 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 32 Appendix

80

85

90

95

Temp.

(°C)

55

60

65

70

75

2.00

2.00

2.00

pH

2.00

1.98

1.98

1.98

1.99

1.99

NOTICE

4.22

4.26

4.30

4.35

pH

4.01

4.08

4.10

4.13

4.16

4.19

Mettler Toledo pH

7.00

6.98

6.98

6.99

7.00

7.02

7.04

7.06

7.09

7.12

8.83

8.81

8.79

8.77

pH

9.21

8.96

8.93

8.90

8.88

8.85

Update

The values of the individual buffers with the corresponding temperatures are kept up to date as far as possible.

However, they may be changed by the respective manufacturers.

32.8.7 Beckmann

Table 29 Beckmann buffer solutions

10

15

20

25

30

Temp.

(°C)

0

5

35

40

45

4.00

4.00

4.00

4.00

4.01

pH

4.00

4.00

4.00

4.02

4.03

4.05

Beckmann pH

7.00

7.12

7.09

7.06

7.04

7.02

7.00

6.99

6.99

6.98

6.98

pH

10.01

10.32

10.25

10.18

10.12

10.06

10.01

9.97

9.93

9.89

9.86

-

-

-

pH

11.00

-

-

-

-

-

900 Touch Control ■■■■■■■■ 505

32.8 Stored buffer series for pH calibration

75

80

85

90

95

Temp.

(°C)

50

55

60

65

70

NOTICE

4.14

4.16

4.18

4.19

4.21

pH

4.00

4.06

4.08

4.09

4.11

4.12

■■■■■■■■■■■■■■■■■■■■■■

Beckmann pH

7.00

6.97

6.98

6.98

6.99

6.99

7.00

7.00

7.01

7.02

7.03

-

-

-

-

pH

10.01

9.83

-

-

-

-

Update

The values of the individual buffers with the corresponding temperatures are kept up to date as far as possible.

However, they may be changed by the respective manufacturers.

32.8.8 Radiometer Analytical

Table 30 Radiometer Analytical buffer solutions

Radiometer (Analytical)

20

25

30

35

40

Temp.

(°C)

0

5

10

15

1.675

1.679

1.683

1.688

1.694

pH

1.679

1.666

1.668

1.670

1.672

4.001

4.005

4.011

4.018

4.027

pH

4.005

4.000

3.998

3.997

3.998

7.016

7.000

6.987

6.977

6.970

pH

7.000

7.118

7.087

7.059

7.036

9.225

9.180

9.139

9.102

9.068

pH

9.180

9.464

9.395

9.332

9.276

506 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 32 Appendix

70

75

80

85

Temp.

(°C)

45

50

55

60

65

90

95 pH

1.679

1.700

1.707

1.715

1.723

1.732

1.743

1.754

1.765

1.778

1.792

-

NOTICE

4.116

4.137

4.159

4.183

4.210

4.240

Radiometer (Analytical) pH

4.005

4.038

pH

7.000

6.965

4.050

4.064

4.080

4.097

6.964

6.965

6.968

6.974

6.982

6.992

7.004

7.018

7.034

pH

9.180

9.038

9.011

8.985

8.962

8.941

8.921

8.900

8.885

8.867

8.850

-

Update

The values of the individual buffers with the corresponding temperatures are kept as up to date as possible.

However, they may be changed by the respective manufacturers.

32.8.9 Baker

Table 31 Baker buffer solutions

Temp.

(°C)

0

5

10

15

20

25

30

35 pH

4.00

4.00

4.00

4.00

4.00

4.00

4.00

4.01

4.02

pH

7.00

7.13

7.09

7.05

7.03

7.00

6.98

6.98

6.98

Baker pH

9.00

9.23

9.17

9.10

9.05

9.00

8.96

8.91

8.88

pH

10.00

10.30

10.24

10.17

10.11

10.05

10.00

9.96

9.93

900 Touch Control ■■■■■■■■ 507

32.8 Stored buffer series for pH calibration

65

70

75

80

Temp.

(°C)

40

45

50

55

60

85

90

95

4.10

4.12

4.14

4.16

pH

4.00

4.03

4.04

4.05

4.07

4.08

4.19

4.21

-

NOTICE

■■■■■■■■■■■■■■■■■■■■■■

6.97

6.97

6.98

6.98

pH

7.00

6.97

6.97

6.96

6.96

6.96

6.99

7.00

-

Baker

8.71

8.69

8.68

8.66

pH

9.00

8.84

8.81

8.78

8.76

8.73

8.64

8.62

pH

10.00

9.89

9.86

9.82

9.79

9.76

9.74

9.72

9.70

9.68

9.66

9.64

-

Update

The values of the individual buffers with the corresponding temperatures are kept up to date as far as possible.

However, they may be changed by the respective manufacturers.

32.8.10 Hamilton DURACAL

Table 32 Hamilton DURACAL buffer solutions

Temp.

(°C)

0

5

10

15

20

25

30 pH

4.01

-

4.01

4.00

4.00

4.00

4.01

4.01

Hamilton DURACAL pH

7.00

pH

9.21

-

7.09

7.06

7.04

7.02

7.00

6.99

9.26

9.21

9.16

-

9.45

9.38

9.32

pH

10.01

-

10.19

10.15

10.11

10.06

10.01

9.97

508 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 32 Appendix

60

65

70

75

Temp.

(°C)

35

40

45

50

55

80

85

90

95

-

-

-

pH

4.01

4.02

4.03

4.04

4.06

-

-

-

-

-

NOTICE

-

-

-

-

-

-

-

-

6.97

6.97

6.97

-

Hamilton DURACAL pH

7.00

6.98

pH

9.21

9.11

9.06

9.03

8.99

-

-

-

-

-

-

-

-

-

Update

The values of the individual buffers with the corresponding temperatures are kept up to date as far as possible.

However, they may be changed by the respective manufacturers.

32.8.11 Precisa

Table 33 Precisa buffer solutions

Temp.

(°C)

0

5

10

15

20

25 pH

4.00

3.99

3.99

3.99

3.99

3.99

4.00

Precisa pH

7.00

7.11

7.08

7.06

7.04

7.02

7.00

pH

9.00

9.27

9.18

9.13

9.08

9.04

9.00

-

-

-

pH

10.01

9.92

9.86

9.83

9.79

-

-

-

-

-

900 Touch Control ■■■■■■■■ 509

32.8 Stored buffer series for pH calibration

55

60

65

70

Temp.

(°C)

30

35

40

45

50

75

80

85

90

95

NOTICE

4.06

4.07

4.09

4.11

pH

4.00

4.00

4.01

4.02

4.03

4.04

4.13

4.15

4.18

4.20

4.23

6.99

7.00

7.00

7.01

7.02

6.97

6.97

6.98

6.98

Precisa pH

7.00

6.99

6.98

6.98

6.97

6.97

■■■■■■■■■■■■■■■■■■■■■■

8.81

8.79

8.76

8.74

pH

9.00

8.96

8.93

8.90

8.87

8.84

8.73

8.71

8.70

8.68

8.67

Update

The values of the individual buffers with the corresponding temperatures are kept as up to date as possible.

However, they may be changed by the respective manufacturers.

32.8.12 Merck CertiPUR 20 / Titrisol

Table 34 Merck CertiPUR 20 / Titrisol buffer solutions

Merck CertiPUR 20 / Titrisol

Temp.

(°C)

0

5

10

15

20 pH

2.000

2.010

2.010

2.010

2.000

2.000

pH

4.000

4.050

4.040

4.020

4.010

4.000

pH

7.000

7.130

7.070

7.050

7.020

7.000

pH

9.000

9.240

9.160

9.110

9.050

9.000

pH

12.000

12.580

12.410

12.260

12.100

12.000

510 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 32 Appendix

2.010

2.010

2.010

2.010

2.010

pH

2.000

2.000

2.000

2.000

2.000

2.000

2.000

2.000

2.000

2.000

NOTICE

70

75

80

85

90

95

50

55

60

65

Temp.

(°C)

25

30

35

40

45

4.000

4.000

4.000

4.000

4.000

4.000

4.000

4.000

4.000

4.000

Merck CertiPUR 20 / Titrisol pH

4.000

4.010

pH

7.000

6.980

pH

9.000

8.950

4.010

4.010

4.010

4.000

6.980

6.960

6.950

6.950

8.910

8.880

8.850

8.820

6.950

6.950

6.960

6.960

6.960

6.960

6.970

6.980

7.000

7.020

8.790

8.760

8.730

8.715

8.700

8.680

8.660

8.650

8.640

-

Update

The values of the individual buffers with the corresponding temperatures are kept as up to date as possible. However, they may be changed by the respective manufacturers.

The CertiPUR 20 and Titrisol product lines have the identical values.

32.8.13 Merck CertiPUR 25

Table 35 Merck CertiPUR 25 buffer solutions

Merck CertiPUR (25°C)

Temp.

(°C)

0

5

10 pH

4.00

-

4.05

4.04

pH

7.00

-

7.09

7.08

pH

9.00

-

9.22

9.16

pH

10.00

-

10.22

10.16

10.900

10.800

10.700

10.590

10.480

pH

12.000

11.880

11.720

11.670

11.540

11.435

11.330

11.185

11.040

10.970

900 Touch Control ■■■■■■■■ 511

32.8 Stored buffer series for pH calibration

-

-

-

-

-

-

-

-

3.98

3.98

3.98

pH

4.00

4.02

4.01

4.00

3.99

3.98

NOTICE

80

85

90

95

60

65

70

75

40

45

50

55

Temp.

(°C)

15

20

25

30

35

■■■■■■■■■■■■■■■■■■■■■■

-

-

-

-

6.97

6.97

6.97

-

-

-

-

-

Merck CertiPUR (25°C) pH

7.00

7.04

pH

9.00

9.10

7.02

7.00

6.98

6.98

9.06

9.00

8.98

8.93

-

-

-

-

-

-

-

-

8.89

8.86

8.84

-

-

-

-

-

-

-

-

pH

10.00

10.10

10.05

10.00

9.94

9.90

9.86

9.80

9.73

-

Update

The values of the individual buffers with the corresponding temperatures are kept as up to date as possible. However, they may be changed by the respective manufacturers.

512 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 32 Appendix

32.9 Using AuditTrailViewer

When you work in accordance with the FDA Guidance 21 CFR Part 11 , logs with detailed user activities (Audit Trail) are important. The

AuditTrailViewer supports you in your efforts to compile, filter and export logs of this type as TXT files.

First you must install the AuditTrailViewer on your PC. Afterwards, you must create a backup of your Touch Control

(see chapter 12.3, page 127) .

The Audit Trail is generated automatically when you create the backup.

The file name of the Audit Trail is log.madt

. In the following you will find detailed information about how you must proceed.

32.9.1 Installing AuditTrailViewer

You will find the file for installing the AuditTrailViewer on the USB flash drive provided.

900 Touch Control

Proceed as follows:

1 Double-click on the file SetupAuditTrailViewer 1.2.2.exe

.

The dialog Welcome to the AuditTrailViewer Setup Wizard is displayed.

2 Click on [Next >] .

The License Agreement dialog is displayed.

3 Click on [I Agree] to accept the contractual conditions.

The Choose Install Location dialog is displayed. A memory location for the AuditTrailViewer is suggested in this dialog.

4 If you accept the suggested memory location, click on [Next >] .

or

Select a different memory location and click on [Next >] .

The Confirm Installation dialog is displayed.

5 Click on [Install] to install the AuditTrailViewer .

■■■■■■■■ 513

32.9 Using AuditTrailViewer ■■■■■■■■■■■■■■■■■■■■■■

As soon as the installation is completed, the dialog Installation

Complete is displayed.

6 Click on [Next >] .

The dialog Completing the AuditTrailViewer Setup Wizard is displayed.

7 If the AuditTrailViewer should not be started immediately, deactivate the check box Run AuditTrailViewer .

8 Click on [Finish] .

If the preceding instruction step has been skipped, then the

AuditTrailViewer will start.

32.9.2 Opening the Audit Trail

Create a backup of the data and of the settings of your Touch Control

(see chapter 12.3, page 127)

.

1 Open the backup on your PC.

2 Open the file log.madt

with the AuditTrailViewer .

514 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

AuditTrailViewer dialog window

32 Appendix

32.9.3 Contents of the Audit Trail

In the following you will find a description of the dialog window and of the menu bar of the AuditTrailViewer .

Table 36 Dialog window

Column Content

No.

Each entry is numbered in sequence.

900 Touch Control ■■■■■■■■ 515

32.9 Using AuditTrailViewer ■■■■■■■■■■■■■■■■■■■■■■

Column

Symbol

Date

User

Category

Action

Details

Menu bar

Content

Classification of the entry:

Actions that are neither relevant to safety nor alter the determination data.

Actions, e.g. changes to the login options and recalculation of determinations.

Errors that occur, e.g. entry of an incorrect password.

Precise time of the event.

User who triggered the action.

Category to which the entry belongs.

Designation of the action.

Details of the action.

Print

Update

Quick filter

Show all

Print Audit Trail.

An Audit Trail can be printed by means of the symbol.

Update view.

The list is updated automatically only at the time the dialog window is opened.

Filter the Audit Trail.

Display all entries once again.

32.9.4 Filtering the Audit Trail

The entries in the Audit Trail can be filtered. You can set the following filter criteria:

Date

User

Category

Action

Proceed as follows:

1 Click in a cell that contains the desired filter criterion (e.g. Category

= Method ).

516 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 32 Appendix

2

Click on the icon or the menu item Filter ▶ Quick filter .

The only entries which will be displayed are those which correspond to the desired filter criterion.

3

Click on the icon or the menu item Filter ▶ Show all .

All of the entries are displayed once again.

32.9.5 Exporting the Audit Trail

You can export the Audit Trail as .txt file. This way, the Audit Trail can be archived and is available as required for inspections.

Proceed as follows:

1 Use the menu item File ▶ Export to open the dialog window for the export.

2 Enter the path and file name with the extension .txt

.

3 Click on [Open] and then on [Export] .

The Audit Trail is exported.

32.10 Diagnosis

System ▶ Diagnosis

The electronic and mechanical functional groups of Metrohm instruments can and should be checked by specialist personnel from Metrohm as part of a regular maintenance schedule. Please ask your local Metrohm representative regarding the precise terms and conditions involved in concluding a corresponding maintenance agreement.

The following functions are available:

LCD test

Check the display for faulty pixels (see chapter 32.10.1, page 518) .

Temperature control

Measure the operating temperature of the connected control instru-

ment (see chapter 32.10.2, page 519)

.

Format storage medium

Format the external storage medium

(see chapter 32.10.3, page

519) .

900 Touch Control ■■■■■■■■ 517

32.10 Diagnosis ■■■■■■■■■■■■■■■■■■■■■■

Remove storage medium

Remove the external storage medium safely (see chapter 32.10.4, page 519) .

Touch adjustment

Adjust the touch-sensitive screen

(see chapter 32.10.5, page 519)

.

Touch screen test

Test the function of the touch-sensitive screen (see chapter 32.10.6, page 520) .

Software update

Update the software of Touch Control and of the control instruments

(see chapter 32.10.7, page 521)

.

822 Curve Simulator

Simulate titration curves for checking titrators (see chapter 32.10.8, page 525) .

Service

Special functions for the service technician (see chapter 32.10.9, page

527)

.

32.10.1 LCD test

System ▶ Diagnosis ▶ LCD test

You can use the LCD test to check the display for faulty pixels. To do this, various test images are displayed one after the other.

NOTICE

The [ ] fixed key is used to display the preceding test picture; the test can be stopped at any time with the [ ] fixed key.

Proceed as follows:

1

Tap on [LCD test] .

The entire display turns white.

2

Check the display for pixel errors and other irregularities.

Continue the test with [ ] .

Other colors and patterns will be displayed one after the other.

3

Repeat step 2 until the System / Diagnosis dialog is displayed again.

4

Contact your Metrohm representative in the event of faults or irregularities.

518 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 32 Appendix

32.10.2 Temperature monitoring

System ▶ Diagnosis ▶ Temperature control

You can use temperature monitoring to display the operating temperature of a connected control instrument. The temperature is measured in the interior of the housing. The display is updated with the [Update] button.

32.10.3 Formatting an external storage medium

System ▶ Diagnosis ▶ Format storage medium

With this function, you have the option of formatting an external storage medium directly on the 900 Touch Control (quick formatting).

The following file systems can be selected:

FAT

FAT32

ExFAT

CAUTION

If you format the storage medium with the ExFAT file system, it is possible that it will no longer be recognized by a PC. Read the pertinent information from Microsoft Support at http://support.microsoft.com

.

Do not use this file system unless you need to be able to save more than 999 files in a single group.

32.10.4 Removing an external storage medium

System ▶ Diagnosis ▶ Remove storage medium

As long as the stored data is not being accessed (reading/saving data), you can plug in and disconnect the storage medium without difficulty at any time. The [Remove storage medium] function offers additional protection. This ensures that the storage medium cannot be removed while any data is being transferred.

32.10.5 Adjusting the touch screen

System ▶ Diagnosis ▶ Touch adjustment

With time the touch-sensitive screen may not register touches at their precise position. You can readjust the screen in this dialog.

Proceed as follows:

1 In the main dialog, tap on [System] .

900 Touch Control ■■■■■■■■ 519

32.10 Diagnosis ■■■■■■■■■■■■■■■■■■■■■■

2 Tap on [Diagnosis] .

3 Tap on [Touch adjustment] .

A message in English and a cross hair will appear in the center of the screen.

4 Touch a stylus (a special pen-shaped instrument for operating devices with touch screens) to the center of the cross hair and maintain this position until the cross hair moves to a new position on the screen.

5 Repeat this process as long as the cross hair continues to be repositioned on the screen.

A message in English will appear.

6 Tap at any location on the screen to confirm the adjustment.

7 Press [ ] to finish the adjustment.

32.10.6 Testing the touch screen

System ▶ Diagnosis ▶ Touch screen test

You can use this test to check the function of the touch-sensitive monitor screen.

A calibration is not possible with this test. If the Touch screen does not work correctly anymore, it will need to be replaced. For that please contact your Metrohm representative.

NOTICE

You can use the [ ] fixed key to stop the test at any time.

Proceed as follows:

1 Tap on [Touch screen test] .

A red square is displayed at the upper left.

2 Tap on the square as close to the center as possible.

The actual detected point of contact will be displayed by a cross hair, and the deviation from the center will also be displayed on the lower edge of the image.

520 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

The next red square will also be displayed.

32 Appendix

3 Repeat step 2 until all of the areas in the display have been tested.

A list with fixed keys will appear.

4 Tap on all of the fixed keys one after the other in any sequence.

The result is displayed in each case with a short information text, e.g.

[Print] key OK .

5 Finish Touch screen test with [ ] .

32.10.7 Software update (loading program versions and language files)

System ▶ Diagnosis ▶ Software update

New program versions or language files can be loaded from an external storage medium (e.g. USB flash drive). The corresponding files must be stored on the storage medium in the 900/SwUpdates

directory (see

"Directory structure", page 126)

.

NOTICE

It is imperative that you first update the control software of the Touch

Control before you update the device program (firmware) of the control instrument (Titrando, Dosing Interface, etc.).

900 Touch Control ■■■■■■■■ 521

32.10 Diagnosis ■■■■■■■■■■■■■■■■■■■■■■

NOTICE

When updating to software version 5.900.0043, the memory will be reformatted.

Before carrying out the software update, make a backup in order to save your data and the system settings

(see chapter 12.3, page 127) .

Use a separate USB flash drive for the software update. It must be removed from the instrument after the successful software update.

Program files

The files are instrument-specific. The file names are structured as follows:

Control software for 900 Touch Control:

5XXXyyyy.BIN

5XXXyyyy.MBIN

– XXX = instrument type (i.e. "900" for the 900 Touch Control)

– yyyy = program version

Firmware of the control instrument:

5XXXyyyy.BIN

– XXX = instrument type (e.g. 907 for the 907 Titrando)

– yyyy = program version

Language files

Language files may have the following content:

■ one or more additional dialog languages the online help for one or more additional dialog languages supplements for existing dialog languages/online help documents

They can be recognized by means of the two-digit language code in the file name. The file name has the following structure:

5XXXyZZZML.BIN

5XXXyZZZML.MBIN

– XXX = instrument type (i.e. "900" for the 900 Touch Control)

– y = version number of the language package

– ZZZ = program version

522 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

32.10.7.1 Carrying out a software update

CAUTION

32 Appendix

Make sure that the power supply is guaranteed during the entire update process. Otherwise there is the chance that the instrument will no longer be able to be switched on and that it will need to be sent in for repair.

NOTICE

Make sure that no USB/RS-232 adapter is connected!

Proceed as follows:

1 Copying files to an external storage medium

NOTICE

In the case of the program versions and language files for the

Touch Control, it is imperative that both the BIN file and the associated MBIN file be copied.

Copy the files to the "SwUpdates" directory.

900 Touch Control ■■■■■■■■ 523

32.10 Diagnosis ■■■■■■■■■■■■■■■■■■■■■■

Copy also the "MetrohmTools" folder with the "MUpdate-

Tool.exe" file to the external storage medium.

Make sure that the file is stored at the top level of the external storage medium. This also applies to the "900" folder.

524 ■■■■■■■■

For the software update to version 5.900.0043, an additional folder that is on the same level as MetrohmTools with the name SRVPRF900 containing the file

St900_SrPrf.cfg is required.

If this structure is not maintained, then the files will not be found, because the directories are accessed directly during the update process.

Plug in the external storage medium at the Touch Control.

2 Opening the update dialog

Under System ▶ Diagnosis , tap on the button Software update .

3 Selecting the instrument

Tap on the Control device list box and select the instrument to be updated.

4 Selecting the file

Tap on the Binary file list box.

The selection list with the BIN files saved in the "SwUpdates" directory is opened.

Select the required file.

Tap on [Select] .

Update of the Touch Control: Continue with Step 5.

Update of the control instrument: Continue with Step 6.

900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 32 Appendix

5 Starting the update (900 Touch Control)

Tap on the [Start] button (NOT on the [ ] fixed key).

The message 023-102 Program update is displayed.

Confirm the message with [Yes] .

The update process is started, it runs automatically. The device is automatically switched off and back on during this process, possibly several times. No user intervention is required.

After successfully updating the software, remove the external storage medium (USB flash drive) from the instrument.

6 Starting the update (control instruments)

Tap on the [Start] button (NOT on the [ ] fixed key).

If Result = No errors is displayed, then the update has been successful.

Remove the external storage medium (USB flash drive) from the instrument.

Switch the 900 Touch Control off and then back on again.

32.10.8 Simulating titration curves

System ▶ Diagnosis ▶ 822 Curve Simulator

The 822 Titration Curve Simulator (order number 2.822.0010) can be used as a diagnostics tool for making rapid checks on titrators. The device is connected in place of the electrode and simulates various titrations.

Details concerning the connection and operation of the curve simulator can be found in the corresponding manual.

NOTICE

A normal titration method is started. This means that the exchange unit or dosing unit should be emptied or the corresponding buret tip should be placed in a sufficiently large collection container.

900 Touch Control ■■■■■■■■ 525

32.10 Diagnosis ■■■■■■■■■■■■■■■■■■■■■■

1 Connecting the curve simulator

NOTICE

Note that the dosing device is always located as the last device, after the curve simulator, if there is a chain of MSB devices (e.g.

Titrando-MSB – stirrer – curve simulator – dosing device).

If the internal dosing drive on the Titrando is used, then the curve simulator must be connected to MSB connector 2.

Switch off the Touch Control.

Connect the analog output of the curve simulator to the electrode connector of the Titrando.

– Use the 6.2116.020 cable.

– Electrode connector Ind.

: Simulation of an acid-base titration.

– Electrode connector Pol.

: Simulation of a Karl Fischer titration.

Connect the curve simulator to an MSB connector on the

Titrando.

2 Configuring the Touch Control

Switch on the Touch Control.

In the System / Diagnosis dialog, tap on the [822 Curve Simulator] button.

The Diagnosis / 822 Curve Simulator dialog is displayed.

Activate the parameter Send dosing signals to MSB2 and confirm with [Set] .

Tap on [ ] .

The signals from the dosing device will now be copied to the MSB connector 2.

3 Simulating a titration

Create a titration method or load an existing one.

Tap on [ ] .

The titration is started. After the completion of the titration, the results dialog is displayed.

4 Finishing the simulation mode

Switch off the Touch Control.

526 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 32 Appendix

32.10.9 Service

System ▶ Diagnosis ▶ Service

Access to the service functions is password-protected and only accessible to Metrohm service technicians. We recommend that you have maintenance performed on the system regularly by the Metrohm Service Department. You can automatically monitor the time interval after which the

next service visit is due (see chapter 13.5.1, page 139)

.

32.11 Arithmetic algorithms in the Titrando

Some of the algorithms and numerical formats used in the device software are explained in this chapter.

Numerical format

All floating point calculations use the IEEE 754 (1985) numerical format, either in "single precision" (32 Bit) or "double precision" (64 Bit).

Rounding off process

Measured values and results are rounded to the defined number of decimal places (commercial rounding, in accordance with the US Pharmacopeia USP). If the digit at the first dropped decimal place is 1, 2, 3 or 4 , then it will be rounded off; if this digit is 5, 6, 7, 8 or 9 , then it will be rounded up. Negative digits will be rounded in accordance with their amount, i.e. away from zero.

Examples:

2.33

yields 2.3

2.35

yields 2.4

2.47

yields 2.5

–2.38

yields –2.4

–2.45

yields –2.5

Statistics

The arithmetic mean value and the absolute and relative standard deviations of results R, endpoints EP and variable C are calculated:

You can statistically evaluate a maximum of nine results (1 ≤ k ≤ 9) calculated in a determination. A statistical series can contain a maximum of 20 determinations (1 ≤ n ≤ 20).

The following convention applies to the subsequent formulas:

1 ≤ n ≤ 20 and 1 ≤ k ≤ 9.

900 Touch Control ■■■■■■■■ 527

32.11 Arithmetic algorithms in the Titrando

Mean value:

Absolute standard deviation:

■■■■■■■■■■■■■■■■■■■■■■

Relative standard deviation (in %):

Explanations

The individual values are incorporated in the statistics in the selected Precision (Round, Truncate or Full precision).

7 significant places are yielded when the 32 bit numerical format is applied for the floating point number in decimal presentation. For the 64 bit numerical format, there are 15 significant places.

The accuracy can be controlled by the selection of the prefix of the unit

(milli, micro) and the number of decimal places.

Example (64-bit numerical format):

The result displayed, 1234.56789158763 mg/L , has 15 significant places. It should be rounded off to three decimal places according to the above rounding-off process:

1234.568 mg/L .

If the same result is expressed in " g/L " ( 1.23456789158763 g/L ), and is also rounded off to three decimal place, this yields

1.235 g/L .

I.e. you obtain the lowest losses in accuracy with rounding when you select the application and the numerical format in such a way that the numbers displayed have as many places before the decimal point as possible.

A complete recalculation of the statistics using a pocket calculator or PC calculation programs may exhibit deviations. This can be explained by the different binary numerical formats used by these computers.

NOTICE

The above losses of accuracy by rounding off in the range of significant places are only relevant theoretically. Most of the time they are lower by several orders of magnitude than – as an example – the uncertainties resulting from weighing out the sample.

528 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 33 Technical specifications

33 Technical specifications

33.1 Touch screen

Display

Touch panel

Resistance to chemicals

VGA color display (640 pixels x 320 pixels), diagonal approx. 5.7" resistive resistant to the following chemicals (no visible changes after 24 h of duration of action):

■ without protective cover:

– Methanol

– Toluene

– Ethyl acetate

– Acetone

– Dichloromethane

– Hydrochloric acid c(HCl) = 2 mol/L with protective cover (order number 6.2723.300):

– Methanol

– Propanol

– Toluene

– Xylene

– Acetone

– Chloroform

– Formamide

– Sulfuric acid c(H

2

SO

4

) = 2 mol/L

– Hydrochloric acid c(HCl) = 2 mol/L

– Caustic soda c(NaOH) = 2 mol/L

– Composite 5.0

– 1-butanol

– 1-hexanol

– Decanol

900 Touch Control ■■■■■■■■ 529

33.2 Interfaces ■■■■■■■■■■■■■■■■■■■■■■

33.2 Interfaces

Connection cable

USB connector

Standard

Plug

Ethernet connector

Standard

Plug

8-pin, 1025 mm, Mini DIN plug with lock

For connecting USB devices.

USB 1.1 (full speed), high power

Type A

For connection to a data network (LAN).

10BASE-T and 100BASE-TX (IEEE 802.3)

Modular 8P8C (RJ-45) with integrated status LEDs

33.3 Power supply

Voltage

Power consumption

Battery

5 V (nominal)

3.5 W (typical)

1 x lithium manganese dioxide button cell CR2032 (needs to be replaced only in the event of servicing).

The battery must be removed and turned in for proper recycling when the device needs to be disposed of.

33.4 Declarations of Conformity

Up-to-date information on the EU conformity and the requirements for electromagnetic compatibility for your product can be found on the Internet under https://www.metrohm.com/en-us/products-overview/ .

Using the article number, you can download the Declaration of conformity for ...

under the documents of the product.

530 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

33.5 Ambient temperature

Nominal function range

Storage

Transport

+5 to +40 °C

(with 10%–80% rel. humidity, non-condensing)

–20 to +60 °C

–20 to +60 °C

33.6 Dimensions

Poly(butylene terephthalate) (PBT) Material of housing

Touch screen film material

Width

Height

Depth

Weight

Polyester

193 mm

71 mm approx. 180 mm (without cable)

1,493 g

33.7 Storage capacity

128 MB Random access memory

Internal memory

Required memory per method

1 GB of which 900 MB available.

3–6 KB

33 Technical specifications

900 Touch Control ■■■■■■■■ 531

■■■■■■■■■■■■■■■■■■■■■■

34 Accessories

Up-to-date information on the scope of delivery and optional accessories for your product can be found on the Internet. You can download this information using the article number as follows:

Downloading the accessories list

1 Enter https://www.metrohm.com/ into your Internet browser.

2 Enter the article number (e.g. 900 ) into the search field.

The search result is displayed.

3 Click on the product.

Detailed information regarding the product is shown on various tabs.

4 On the Included parts tab, click on Download the PDF .

The PDF file with the accessories data is created.

NOTICE

Once you have received your new product, we recommend downloading the accessories list from the Internet, printing it out and keeping it together with the manual for reference purposes.

532 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ Index

Index

A

Acoustic signals ........................ 40

ADD ....................................... 440

Audit Trail ................................. 39

Archive ............................. 517

Export .............................. 517

Filter ................................. 516

Open ................................ 514

AuditTrailViewer

Installing ........................... 513

Automatic system test ............ 130

Automation ............................ 470

Axial distance ........................... 95

B

Backup ................................... 127

Interval ............................. 139

Balance .......................... 113, 495

Beaker radius

Sample vessel ................... 102

Special beaker .................. 107

Beaker sensor

Sample vessel ................... 103

Special beaker .................. 107

BEEP ....................................... 481

Blank value concentration ......... 71

BRC ...................................... 3, 15

Parameters ....................... 345

Break point

Evaluation ........................ 414

Bromine index determination .... 15

Buffer series ........................... 499

C

CAL ........................................ 391

CALC ...................................... 418

CALC LIVE .............................. 425

Calculation ............................. 418

Add .................................. 418

Delete .............................. 418

Edit .................................. 419

Load template .................. 419

Note ................................. 422

Result options ................... 423

Calibration

History ............................... 72

Calibration buffer ................... 499

Calibration buffers .................. 160

Calibration data ................ 70, 218

Limit values ........................ 75

Calibration interval ................... 77

Calibration parameters

CAL Conc ......................... 394

CAL Cond ......................... 397

CAL pH ............................. 391

Calibration position 1 ............. 103

Cell

BRC .................................. 351

KFC .................................. 342

KFT ................................... 332

SET ................................... 320

Cell constant

Intervention limit ................ 73

Warning limit ...................... 73

Command

ADD ................................. 440

BEEP ................................. 481

BRC .................................. 345

CAL .................................. 391

CALC ................................ 418

CALC LIVE ........................ 425

CONTROL RS .................... 470

CTRL ................................. 467

DET .................................. 278

DOS ................................. 448

ELT ................................... 402

EMPTY .............................. 439

END ................................. 482

EVAL BREAK ..................... 414

EVAL FIX-EP ...................... 410

EVAL MIN/MAX ................ 413

EVAL pK/HNP .................... 412

EVAL RATE ........................ 417

Insert ................................ 171

KFC .................................. 336

KFT ................................... 323

LIFT .................................. 473

LQH ................................. 445

MAT ................................. 373

MEAS ............................... 379

MET ................................. 295

MOVE .............................. 470

PREP ................................. 438

PUMP ............................... 474

RACK ................................ 475

REPORT ............................ 433

REQUEST .......................... 480

SAMPLE ............................ 475

SCAN ............................... 466

SCAN RS ........................... 469

SET ................................... 311

SIGN ................................ 482

STAT ................................. 354

STIR .................................. 479

SUBSEQ ............................ 476

WAIT ................................ 480

Command list ......................... 170

Common variable

Delete .............................. 141

Edit .................................. 141

Properties ......................... 143

Validity ............................. 144

Communication ...................... 465

Conditioning .......................... 266

BRC .................................. 349

KFC .................................. 339

KFT ................................... 330

SET ................................... 318

Control ................................... 186

Control instrument ................... 87

Parameters

291, 307, 320, 332, 343, 352,

368, 375, 388, 400, 404, 441,

447,

Control parameters

461

BRC .................................. 345

KFC .................................. 336

KFT ................................... 325

SET ................................... 312

STAT ................................. 356

Control Remote Box ................. 81

CONTROL RS .......................... 470

Coulometric water content determination according to Karl Fischer

................................................. 15

CTRL ...................................... 467

Curve options ......................... 228

Curve simulator ...................... 525

Curves .................................... 227

Custom logo .......................... 162

D

Delete

Signature .......................... 185

Statistics data ................... 188

DET .............................. 1, 11, 278

Determination

Calibration data ................ 218

900 Touch Control ■■■■■■■■ 533

Index

Cancel .............................. 207

Carry out .......................... 205

Data ................................. 214

Define filters ..................... 225

Delete signatures .............. 185

Endpoint details ................ 217

List of saved determinations

......................................... 223

Load ................................. 222

Measuring point list .......... 217

Message ........................... 218

Properties ......................... 219

Recalculate ....................... 229

Reevaluate ........................ 229

Save ................................. 227

Save automatically ............ 182

Send PC/LIMS report ......... 182

Sign .................................. 184

Titrant data ...................... 218

Determination run .................. 205

Device

Add .................................... 80

Balance ............................ 113

Conductivity Module .......... 87

Delete ................................ 79

Dosing Interface ................. 87

Edit .................................... 79

New ................................... 79

pH Module ......................... 87

Printer .............................. 109

Sample Processor ................ 92

Titrando ............................. 87

USB/RS-232 adapter ......... 115

Device manager ....................... 79

Diagnosis ............................... 517

Dialog language ....................... 22

Dialog options .......................... 24

Dialog settings

System-specific ................... 24

Directory structure .................. 126

DOS ....................................... 448

Dosing ................................... 256

Continuous ....................... 258

Fixed volume .................... 260

Dosing curve .......................... 227

Dosing device

Parameters

293, 309, 322, 334, 369, 376,

441,

447,

462

Dosing key ............................. 492

Dosing parameters

ADD ................................. 440

DOS ................................. 448

MAT ................................. 374

Dosing port .............................. 47

Dosing rate

Maximum ................. 493, 494

Dosing unit

Data ................................... 46

GLP test .............................. 56

Preparing parameters ......... 47

Replace reagent ................ 264

Tubing parameters .............. 49

E

E-mail ....................................... 82

E(0)

Intervention limit ................ 73

Warning limit ...................... 73

Electrode .................................. 66

Parameters

292, 308, 320, 333, 343, 352,

368, 375, 388, 400, 405, 461

Electrode slope

Intervention limit ................ 73

Warning limit ...................... 73

Electrode test parameters ....... 403

Electrode type

Custom ............................ 163

Electrostatic charge .................... 6

ELT ......................................... 402

EMPTY ................................... 439

Default parameters ........... 493

Emptying ................................ 263

Default parameters ........... 493

END ....................................... 482

Endpoint details ...................... 217

ERC ................................ 291, 307

EVAL BREAK ........................... 414

EVAL FIX-EP ............................ 410

EVAL MIN/MAX ...................... 413

EVAL pK/HNP ......................... 412

EVAL RATE .............................. 417

Evaluation

Break point ....................... 414

DET .................................. 290

Fixed endpoint ................. 410

HNP ................................. 412

Maximum ......................... 413

MET ................................. 307

Minimum ......................... 413

pK value ........................... 412

Potentiometric .......... 286, 302

Rate ................................. 417

Set windows .... 287, 303, 415

Evaluations ............................. 408

Exchange unit

Data ................................... 52

■■■■■■■■■■■■■■■■■■■■■■

GLP test .............................. 56

Preparing parameters ......... 54

Tubing parameters .............. 55

Expiry date

Sensor ................................ 69

Titrant ................................ 45

External position

Swing angle ....................... 99

F

Favorite .................................. 190

Add .................................. 191

Delete .............................. 191

Edit .................................. 191

FDA Guidance 21 CFR Part 11

........................................... 28, 39

File

Copy ................................ 123

Properties ......................... 124

Rename ............................ 123

File manager .......................... 121

Filling ..................................... 264

Filling rate

Maximum ................. 493, 494

Filter

Determination .................. 225

Result table ...................... 239

Fixed endpoint

Evaluation ........................ 410

Fixed key .................................... 8

Format storage medium ......... 519

Formula editor ........................ 427

G

GLP manager .......................... 129

GLP test "Measurement" .. 131

GLP test "Titration" ........... 131

System validation .............. 135

Test tool ........................... 130

GLP test

Dosing unit ......................... 56

Exchange unit ..................... 56

Sensor ................................ 73

H

Half neutralization potential

HNP

Evaluation ........................ 412

Evaluation ........................ 412

Hysteresis ....................... 362, 456

I

ID profile

Create ................................ 33

534 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Identification profile

Create ................................ 33

Initialization ............................ 483

Input line

Template .......................... 155

Instrument

Configure ........................... 80

Switch off ........................... 17

Switch on ........................... 16

Touch Control ..................... 81

Intervention limit

Cell constant ...................... 73

E(0) .................................... 73

Electrode slope ................... 73 pH(0) .................................. 73

Titer ................................... 60

K

Key functions ......................... 117

KFC .............................. 2, 15, 336

KFT .............................. 2, 12, 323

L

LCD test ................................. 518

LIFT ........................................ 473

Lift rate .................................. 271

Limit values

Calibration data .................. 75

Sample size ...................... 179

Liquid Handling parameters .... 445

Live display ............................. 210

Live modifications ................... 208

Live parameters ...................... 212

Live result

Note ................................. 426

Load

Determination .................. 222

Sample table .................... 200

Load result template ............... 419

Login

Options .............................. 33

Logout automatically ................ 35

LQH ....................................... 445

M

Manual control ....................... 250

Conditioning .................... 266

Continuous dosing ........... 258

Dosing ............................. 256

Dosing fixed volume ......... 260

Emptying .......................... 263

Filling ............................... 264

Measuring ........................ 252

Preparing .......................... 262

Remote lines ..................... 267

Sample Processor .............. 268

Stirring ............................. 265

Troubleshooting ............... 485

Manual Dosing Controller ....... 492

MAT ............................. 2, 14, 373

Max. swing angle ..................... 97

Maximum evaluation .............. 413

Maximum stroke path .............. 94

MEAS ..................................... 379

Measured value

Display resolution ............... 40

Measurement curve ................ 227

Measuring .............................. 252

Measuring parameters

MEAS ............................... 379

Measuring point list ................ 217

Message

Determination .................. 218

Send e-mail ........................ 82

MET ............................. 2, 11, 295

Method .................................. 167

Command list ................... 170

Create .............................. 167

Delete signatures .............. 185

Edit .................................. 170

Insert command ............... 171

Load ................................. 169

Note ................................. 182

Options ............................ 172

Properties ......................... 180

Sample data ..................... 176

Save ................................. 168

Save as favorite ................ 190

Sign .................................. 184

Start options ..................... 173

Stop options ..................... 175

Troubleshooting ............... 484

Min. beaker radius .................... 94

Minimum evaluation ............... 413

Miscellaneous commands ....... 479

Monitoring

Calibration interval .............. 77

DOS ................................. 455

Limit values calibration data 75

Limit values sample size .... 179

STAT ................................. 361

MOVE .................................... 470

N

Note

Method ............................ 182

Numerical input ........................ 20

Index

O

Operation

General .............................. 17

Output line

Template .......................... 157

P

Paper size ............................... 113

Parameters

ADD ................................. 440

BEEP ................................. 481

BRC .................................. 345

CAL .................................. 391

CALC ................................ 418

CALC LIVE ........................ 425

CONTROL RS .................... 470

CTRL ................................. 467

DET .................................. 278

DOS ................................. 448

ELT ................................... 402

EMPTY .............................. 439

END ................................. 482

EVAL BREAK ..................... 414

EVAL FIX-EP ...................... 410

EVAL MIN/MAX ................ 413

EVAL pK/HNP .................... 412

EVAL RATE ........................ 417

KFC .................................. 336

KFT ................................... 323

LIFT .................................. 473

LQH ................................. 445

MAT ................................. 373

MEAS ............................... 379

MET ................................. 295

Modify live ....................... 212

MOVE .............................. 470

PREP ................................. 438

PUMP ............................... 474

RACK ................................ 475

REPORT ............................ 433

REQUEST .......................... 480

SAMPLE ............................ 475

SCAN ............................... 466

SCAN RS ........................... 469

SET ................................... 311

SIGN ................................ 482

STAT ................................. 354

STIR .................................. 479

SUBSEQ ............................ 476

WAIT ................................ 480

Password

Forgotten ........................... 37

Options .............................. 35

900 Touch Control ■■■■■■■■ 535

Index

PC/LIMS report

Save ................................... 83

Send ................................. 182

PDF settings ........................... 110

Peripheral devices ..................... 79

pH(0)

Intervention limit ................ 73

Warning limit ...................... 73

pK value

Evaluation ........................ 412

Potentiometric evaluation

DET .................................. 286

MET ................................. 302

PREP ....................................... 438

Default parameters ... 493, 494

Parameters for the dosing unit

........................................... 47

Parameters for the exchange

unit .................................... 54

Preparing ................................ 262

Default parameters ... 493, 494

Parameters for the dosing unit

........................................... 47

Parameters for the exchange

unit .................................... 54

Printer .................................... 109

PDF settings ..................... 110

Printing .................................. 242

Troubleshooting ............... 485

Program version

Touch Control ..................... 81

Properties

Determination .................. 219

File ................................... 124

Method ............................ 180

Swing Head ........................ 95

Tower ................................. 93

PUMP ..................................... 474

R

Rack

Copy ................................ 100

Delete .............................. 100

Edit .................................. 100

Load ................................. 100

RACK ..................................... 475

Rack adjustment ..................... 108

Rack name ............................... 93

Rack offset ............................. 103

Rate evaluation ....................... 417

Reagent ................................... 61

Add .................................... 61

Delete ................................ 61

Edit .................................... 61

Replace ............................ 264

Reagent monitoring .................. 62

Reagent replacement ................ 64

Reason ..................................... 38

Modification option ............ 37

Recalculation .......................... 229

Reevaluation ........................... 229

Remote lines

Manual control ................. 267

Report .................................... 242

Print ................................. 242

REPORT .................................. 433

Report header ........................ 162

REQUEST ................................ 480

Resolution

Measured value display ....... 40

Result ..................................... 214

Calculation formula .......... 421

Limits ............................... 422

Save as a common variable

......................................... 145

Template .......................... 151

Result table ............................ 236

Define filters ..................... 239

Delete automatically ......... 240

Save ................................. 241

Result variable ........................ 422

As parameter setting ........ 496

Robotic arm ............................. 97

Max. swing angle ............... 97

Rotation offset ................... 98

Swing direction .................. 98

Swing increment ................ 97

Swing offset ....................... 97

Swing radius ....................... 98

Swing rate ........ 274, 275, 473

Rotation increment ................... 95

Rotation offset ......................... 98

Routine dialog .......................... 26

Standard configuration ....... 26

S

Safety instructions ...................... 5

SAMPLE .................................. 475

Sample assignment table ........ 147

Sample data ........................... 194

Modify live ....................... 208

Settings ............................ 176

Sample identification

Modify designation ........... 176

Modify the designation ..... 204

Sample identification list ......... 147

Sample Processor ................... 268

Sample rack ............................ 100

■■■■■■■■■■■■■■■■■■■■■■

Sample series

Perform ............................ 206

Sample size

Fixed ................................ 176

Limit value ........................ 179

Limits ............................... 177

Send from a balance ......... 495

Sample table .......................... 197

Save as favorite ................ 190

Troubleshooting ............... 484

Save

Determination .................. 227

Result table ...................... 241

Sample table .................... 199

SCAN ..................................... 466

SCAN RS ................................. 469

Sensor ...................................... 66

Add .................................... 67

Calibration data .................. 70

Calibration interval .............. 77

Delete ................................ 67

Edit .................................... 67

Expiry date ......................... 69

GLP test .............................. 73

Parameters

292, 308, 320, 333, 343, 352,

368, 375, 388, 400, 405, 461

Working life ........................ 69

Sensor list ................................. 66

Serial number

Dosing cylinder ............. 47, 53

Dosing unit ......................... 46

Exchange unit ..................... 53

Service ............................... 5, 527

SET ............................... 2, 12, 311

Set the date .............................. 23

Set the time .............................. 23

Sign

Determination .................. 184

Method ............................ 184

SIGN ...................................... 482

Signature

Delete .............................. 185

Software update ..................... 521

Special beakers

Edit .................................. 105

Spooler .................................. 113

Standard addition ................... 384

Start conditions

DET .................................. 278

KFT ................................... 323

MET ................................. 295

SET ................................... 311

STAT ................................. 354

536 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■

Start options ........................... 173

STAT ............................. 2, 13, 354

Statistics ......................... 188, 231

Delete .............................. 188

Troubleshooting ............... 484

Status LED .............................. 492

STIR ........................................ 479

Stirrer

Parameters

294, 310, 323, 335, 344, 353,

372, 379, 390, 401, 406, 444,

465

Stirring ................................... 265

Stirring rate ............................ 494

Stop ....................................... 492

Stop conditions

BRC .................................. 349

DET .................................. 285

DOS ................................. 454

KFC .................................. 339

KFT ................................... 329

MET ................................. 301

SET ................................... 317

STAT ................................. 360

Stop criterion

BRC .................................. 347

KFC .................................. 337

KFT ................................... 327

SET ................................... 315

Stop options ........................... 175

SUBSEQ .................................. 476

Subsequence .......................... 476

Load ................................. 476

Note ................................. 478

Options ............................ 477

Save ................................. 476

Supply voltage ............................ 5

Swing direction ........................ 98

Swing Head

Properties ........................... 95

Swing increment ...................... 97

Swing offset ............................. 97

Swing radius ............................. 98

Swing rate .............. 274, 275, 473

Switch off ................................. 17

Switch on ................................. 16

System initialization ................ 483

System settings ......................... 22

System test ............................. 130

System validation .................... 135

T

Tandem dosing

....................... 370, 377, 442, 462

Temperature monitoring ......... 519

Template

Calibration buffers ............ 160

Electrode type .................. 163

Input line .......................... 155

Output line ....................... 157

Report header .................. 162

Result template ................ 151

Sample assignment table .. 147

Sample identification list ... 147

Templates ............................... 147

Text input ................................. 20

Time zone ................................ 23

Titer

History ............................... 59

Intervention limit ................ 60

Properties ........................... 58

Warning limit ...................... 60

Titrant ...................................... 41

Add .................................... 42

Delete ................................ 42

Edit .................................... 42

Expiry date ......................... 45

Working life ........................ 45

Titrant data ............................ 218

Titrant list ................................. 41

Titration curve ........................ 227

Titration mode

BRC ...................................... 3

DET ...................................... 1

KFC ...................................... 2

KFT ....................................... 2

MAT ..................................... 2

MET ..................................... 2

SET ....................................... 2

STAT ..................................... 2

Titration parameters

BRC .................................. 348

DET .................................. 281

KFC .................................. 338

Index

KFT ................................... 328

MET ................................. 298

SET ................................... 316

STAT ................................. 358

Touch Control ........................... 81

Touch screen ............................ 17

Test .................................. 520

Tower

Properties ........................... 93

Tubing parameters

Dosing unit ......................... 49

Exchange unit ..................... 55

U

Uninterrupted dosing

....................... 370, 377, 442, 462

Update ................................... 521

USB/RS-232 adapter ............... 115

User profile

Create ................................ 33

V

Validation interval ................... 137

Validity

Calibration .......................... 77

Common variable ............. 144

Titer ................................... 58

Valve disk

Shift direction ..................... 52

W

WAIT ...................................... 480

Warning limit

Cell constant ...................... 73

E(0) .................................... 73

Electrode slope ................... 73 pH(0) .................................. 73

Titer ................................... 60

Water determination according to

Karl Fischer ............................... 12

Working life

Sensor ................................ 69

Titrant ................................ 45

Write protection

External memory .............. 125

900 Touch Control ■■■■■■■■ 537

advertisement

Key Features

  • Dynamic equivalence point titration (DET)
  • Monotonic equivalence point titration (MET)
  • Endpoint titration (SET)
  • Water determination according to Karl Fischer (KFT)
  • Titration with constant maintenance of the measured value (STAT)
  • Manual titration (MAT)
  • Coulometric water content determination according to Karl Fischer (KFC)
  • Bromine index determination (BRC)
  • Intuitive touch-sensitive screen for easy operation
  • Comprehensive user administration for secure data management

Related manuals

Frequently Answers and Questions

What is the purpose of the Metrohm 900 Touch Control?
The Metrohm 900 Touch Control is an advanced titration and measuring instrument used for various analytical chemistry applications, including titrations, water content determination, and more.
What are the key advantages of using the Metrohm 900 Touch Control?
The Metrohm 900 Touch Control offers high precision, versatility, and ease of use, making it an ideal choice for both routine and complex analytical tasks.
What types of titration methods does the Metrohm 900 Touch Control support?
The Metrohm 900 Touch Control supports various titration methods, including dynamic equivalence point titration (DET), monotonic equivalence point titration (MET), endpoint titration (SET), and more.
Can I use the Metrohm 900 Touch Control for water content determination?
Yes, the Metrohm 900 Touch Control can perform water content determination according to Karl Fischer (KFT) and coulometric water content determination according to Karl Fischer (KFC).
How do I ensure accurate and reliable results with the Metrohm 900 Touch Control?
The Metrohm 900 Touch Control offers GLP compliance features, ensuring the reliability and auditability of your results.
Download PDF

advertisement

Table of contents