Start conditions. Metrohm 900 Touch Control

Add to My manuals
553 Pages

advertisement

Start conditions. Metrohm 900 Touch Control | Manualzz

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

29.2 Monotonic equivalence point titrations (MET)

29.2.1 Start conditions

The parameters that are carried out before the start of titration are defined under [Start conditions] . The start conditions are processed in the following sequence:

1.

Start volume

2.

Start measured value

3.

Start slope

4.

Pause

Start volume

Volume that is dosed prior to the start of the titration.

Input range

Default value

0.00000 to 9999.99 mL

0.00000 mL

Dosing rate

Rate at which the start volume is dosed.

Input range

Selection

Default value

0.01 to 166.00 mL/min maximum maximum maximum

The maximum dosing rate depends on the cylinder volume (see chapter 32, page 492) .

Pause

Waiting time, e.g. for the electrode to settle down after the start or a reaction time after the dosing of a start volume. The time entered here is only running when all start conditions have been fulfilled.

Input range

Default value

0 to 999999 s

0 s

More start conditions

Start meas. value

Before starting the titration, it will be dosed until this measured value is reached. If the start measured value is reached by the dosing of a start volume, then the titration starts directly.

The values that are actually measurable may vary from the values specified here, depending on the hardware. The binding measurement value ranges can be found in the "Technical specifications" chapter for the control instrument used.

900 Touch Control ■■■■■■■■ 295

29.2 Monotonic equivalence point titrations (MET)

Measuring mode pH:

Input range

Selection

Default value

–20.000 to 20.000 off off

Measuring mode U, Ipol:

Input range

Selection

Default value

–2000.0 to 2000.0 mV off off

Measuring mode Upol:

Input range –200.0 to 200.0 µA

Selection

Default value off off

■■■■■■■■■■■■■■■■■■■■■■

Start slope

Before starting the titration, it will be dosed until this slope (measured value per volume) is reached. If the start slope is reached by the dosing of a start volume, then the titration starts directly.

Measuring mode pH:

Input range

Selection

Default value

00.000 to 9,999 pH/mL off off

Measuring mode U, Ipol:

Input range

Selection

Default value

0 to 999 mV/mL off off

Measuring mode Upol:

Input range 0 to 99 µA/mL

Selection

Default value off off

Dosing rate

Rate to be dosed with until the start measured value or the start slope is reached.

Input range

Default value

Selection

0.01 to 166.00 mL/min

5.00 mL/min maximum maximum

The maximum dosing rate depends on the cylinder volume

(see chapter 32, page 492) .

296 ■■■■■■■■ 900 Touch Control

■■■■■■■■■■■■■■■■■■■■■■ 29 Parameters

Initial measured value

The [Initial meas. value] is measured before carrying out the start conditions . This value is used for the titration direction determination , which is necessary for the start measured value and stop measured value parameters.

Signal drift

Maximum permissible drift for the measured value acceptance, i.e. maximum change of the measured value per minute. For this value, a suitable maximum waiting time is automatically calculated except you already have modified this waiting time.

Measuring mode pH, U and Ipol:

Input range

Selection

Default value

0.1 to 999.0 mV/min off off off

Measured value acceptance will take place after the maximum waiting time has elapsed.

Measuring mode Upol:

Input range

Selection

Default value

0.01 to 99.90 µA/min off off off

Measured value acceptance will take place after the maximum waiting time has elapsed.

Min. waiting time

The measured value is not accepted until the minimum waiting time has elapsed, even if the signal drift has already been reached. The minimum waiting time is only important for drift-controlled measurements.

Input range

Default value

0 to 999999 s

0 s

Max. waiting time

If the signal drift has been switched off or has not yet been reached, then the measured value will be accepted when the maximum waiting time has elapsed. As default value, a suitable waiting time for the signal drift is automatically calculated.

Input range

Default value

0 to 999999 s

1 s

900 Touch Control ■■■■■■■■ 297

advertisement

Key Features

  • Dynamic equivalence point titration (DET)
  • Monotonic equivalence point titration (MET)
  • Endpoint titration (SET)
  • Water determination according to Karl Fischer (KFT)
  • Titration with constant maintenance of the measured value (STAT)
  • Manual titration (MAT)
  • Coulometric water content determination according to Karl Fischer (KFC)
  • Bromine index determination (BRC)
  • Intuitive touch-sensitive screen for easy operation
  • Comprehensive user administration for secure data management

Related manuals

Frequently Answers and Questions

What is the purpose of the Metrohm 900 Touch Control?
The Metrohm 900 Touch Control is an advanced titration and measuring instrument used for various analytical chemistry applications, including titrations, water content determination, and more.
What are the key advantages of using the Metrohm 900 Touch Control?
The Metrohm 900 Touch Control offers high precision, versatility, and ease of use, making it an ideal choice for both routine and complex analytical tasks.
What types of titration methods does the Metrohm 900 Touch Control support?
The Metrohm 900 Touch Control supports various titration methods, including dynamic equivalence point titration (DET), monotonic equivalence point titration (MET), endpoint titration (SET), and more.
Can I use the Metrohm 900 Touch Control for water content determination?
Yes, the Metrohm 900 Touch Control can perform water content determination according to Karl Fischer (KFT) and coulometric water content determination according to Karl Fischer (KFC).
How do I ensure accurate and reliable results with the Metrohm 900 Touch Control?
The Metrohm 900 Touch Control offers GLP compliance features, ensuring the reliability and auditability of your results.
Download PDF

advertisement

Table of contents