Chlorine HR (KI) with Tablet. Lovibond MD 100


Add to My manuals
292 Pages

advertisement

Chlorine HR (KI) with Tablet. Lovibond MD 100 | Manualzz

64

GB Methods

CLHr

Chlorine HR with Tablet

5 – 200 mg/l Cl

2

Insert the adapter for 16 mm vials.

0.0.0

Zero

Test

CLHr

RESULT

Fill a clean vial (16 mm Ø) with 8 ml of the water sample and perform zero calibration (see “Operation”).

Add one CHLORINE HR (KI) tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod.

Add one ACIDIFYING GP tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod.

Close the vial tightly with the cap and swirl gently several times until the tablets are dissolved.

Place the vial in the sample chamber making sure that the marks are aligned.

Press the [ZERO/TEST] key.

The method symbol flashes for approx. 3 seconds.

The result is shown in the display in mg/l Chlorine.

MD100_7 12/2016

GB Methods

Notes:

1. Oxidising agents interfere as they react in the same way as Chlorine.

Reagent

Set ACIDIFYING GP/

CHLORINE HR (KI)

CHLORINE HR (KI)

ACIDIFYING GP

Form of reagent/Quantity

Tablet / per 100 inclusive stirring rod

Tablet / 100

Tablet / 100

Order-No.

517721BT

513000BT

515480BT

MD100_7 12/2016 65