TECHNICAL DATA. EKOM DK50 2V, DK50 Plus Mobile, DK50 2V/M MOBILE, DK50 2V S, DK50 PLUS

Add to My manuals
104 Pages

advertisement

TECHNICAL DATA. EKOM DK50 2V, DK50 Plus Mobile, DK50 2V/M MOBILE, DK50 2V S, DK50 PLUS | Manualzz

TECHNICAL DATA

TECHNICAL DATA

Compressors are designed to operate in dry, ventilated and indoor dust-free rooms with the following climatic conditions:

Temperature

Relative humidity from +5 °C to +40°C max. 70%

Working pressure 6 – 8 bar

Nominal voltage

Frequency a)

Performance at 6 bar (FAD)

Working pressure

Max. current

Motor power b)

V, Hz l/min bar

A kW

DK50 PLUS

230, 50

70

6.0 – 8.0

4.1

0.55

Air tank volume

Air quality - filtration

Maximum operating pressure of safety valve

Noise level at 5 bar (L pA

)

 l m bar

25

-

11.0

Operating mode

PDP drying performance at 7 bar

Time to fill air tank from 0 to

7 bar

Dimensions (net) w x d x h

Net weight dB

%

°C s mm kg

≤57

S1-100

-

149

550x718x995

100 c)

Classification under EN 60601-1 Class I. a) State the compressor variant in the order b) For other range of pressure consult with the supplier c) Weight value is only informative data and applies only to a product without any accessories

Dependence of compressor capacity on working pressure

NP-DK50 PLUS, 2V-MOBILE-A-4_12-2019-MD 14 12/2019

TECHNICAL DATA

Working pressure 6 – 8 bar

Nominal voltage

Frequency a)

Performance at 6 bar

(FAD)

Working pressure b)

DK50 2V DK50 2V/M

V, Hz 230, 50 115, 60 230, 50 115, 60 l/min bar

135

6.0 – 8.0

104

6.0 – 8.0

Max. current

Motor power

A kW

8.8

1.2

20.1

1.1

8.9

1.2

20.2

1.1

Air tank volume

Air quality - filtration

Maximum operating pressure of safety valve

Noise level at 5 bar (L pA

) l

 m bar dB

25

-

11.0

25

0,3

11.0

Operating mode %

°C

≤62

S1-100

S3-60 d)

≤64

-

≤62 ≤64

S1-100

S3-60 d)

 +3°C

PDP drying performance at 7 bar

Time to fill air tank from 0 to 7 bar

Dimensions (net) w x d x h

Net weight s mm kg

61

550x718x995

104 c)

78

550x718x995

110 c)

Classification under EN60601-1 Class I. a) State the compressor variant in the order b) For other range of pressure consult with the supplier c) Weight value is only informative data and applies only to a product without any accessories d) Information about the weight is for informative purposes only and applies to the product without any additional accessories

Dependence of compressor capacity on working pressure

12/2019 15 NP-DK50 PLUS, 2V-MOBILE-A-4_12-2019-MD

TECHNICAL DATA

Working pressure 8 – 10 bar

Nominal voltage

Frequency a)

Performance at 8 bar

(FAD)

Working pressure b)

V, Hz l/min bar

Max. current

Motor power

A kW

Air tank volume

Air quality - filtration

Maximum operating pressure of safety valve

Noise level at 5 bar (L pA

)

Operating mode

PDP – drying performance at 7 bar

Tie to fill air tank from 0 to 7 bar

Dimensions (net) w x d x h

Net weight l

 m bar dB

%

°C s mm kg

DK50 2V

230, 50

110

8.0 – 10.0

86

1.2

25

-

11.0

≤62

S1-100

-

96

550x718x995

107 c)

Classification under EN 60601-1 Class I. a) State the compressor variant in the order b) For other range of pressure consult with the supplier c) Weight value is only informative data and applies only to a product without any accessories

Dependence of compressor capacity on working pressure

NP-DK50 PLUS, 2V-MOBILE-A-4_12-2019-MD 16 12/2019

TECHNICAL DATA

FAD correction of capacity for altitude

Capacity given in the form of FAD („Free Air Delivery“) applies to the following conditions:

Altitude 0 m.n.m.

Atmospheric pressure 101325 Pa

Temperature

Relative humidity

20°C

0%

To calculate FAD compressor capacity in dependence on altitude, it is necessary to apply correction factor according to the following table:

Altitude [m.n.m.]

FAD correction factor

0 -1500

1

1501 - 2500

0,8

2501 - 3500

0,71

3501 - 4500

0,60

12/2019 17 NP-DK50 PLUS, 2V-MOBILE-A-4_12-2019-MD