advertisement
ETAS
3.5.2
Hardware Description
Indicators
The three indicators (LEDs) are located in the top left corner on the front panel.
They indicate the following operational states of the ES650.1:
Indicator
ON
Indication green
Operational State
The unit is powered on. green, blinking The unit is in standby mode.
TX
ER t green, pulse, intermittent brightness
The unit is synchronizing with the
ES600.
t yellow Host connection exists. yellow, blinking Data are being transferred. red
(duration > 5 s) red, bright blink ing
-
An error has occurred. Measurement accuracy may be out of specification.
Please send the unit to ETAS for service.
A serious error has occurred. Measure ments are no longer possible. Please send the unit to ETAS for repair. t red, pulse, inter mittent bright ness
Device identification activated from
INCA. t
When enabling the device identification in INCA, the ER LED lights up for 5 s with intermittent brightness. This allows you to identify the device selected in
INCA within a device block.
Serial number
The serial number is located on the front side of the module. This number is important for the unique addressing of the module in the application software.
ES650.1 - User Guide 17
advertisement
Related manuals
advertisement
Table of contents
- 5 About this Document
- 5 Classification of Safety Messages
- 5 Presentation of Instructions
- 6 Typographical Conventions
- 6 Presentation of Supporting Information
- 6 Scope of supply
- 6 Additional information
- 7 Basic safety notices
- 7 General safety information
- 7 Requirements for users and duties for operators
- 7 Intended use
- 12 Hardware Description
- 12 ES600 Measurement Modules
- 13 ES650.1 A/D and Thermo Module
- 13 Overview
- 14 Features
- 15 Package Contents
- 15 Housing
- 16 Ports and Indicators
- 16 Front Side
- 17 Indicators
- 18 Rear Side
- 19 Maximum Input and Common-Mode Voltages
- 20 Measurement Accuracy
- 22 Getting Started
- 22 Installation and blocking
- 22 General installation recommendations
- 22 Fastening the module onto a carrier system
- 23 Connecting several modules mechanically
- 24 Cabling
- 26 Configuration
- 26 Calibration
- 27 Technical Data
- 27 General Data
- 27 Product labeling
- 28 Fulfilled Standards and Norms
- 28 Environmental Conditions
- 28 Maintenance the Product
- 28 Cleaning the product
- 28 Mechanical Data
- 29 RoHS Conformity
- 29 CE conformity
- 29 UKCA conformity
- 29 KCC conformity
- 29 Taking the Product Back and Recycling
- 30 Use of Open Source software
- 30 System Requirements
- 30 Hardware
- 31 Software
- 32 Electrical Data
- 32 Power Supply
- 32 Host Interface
- 32 Analog Voltage Inputs
- 33 Thermocouple Inputs
- 34 Pin Assignment
- 34 Analog Inputs
- 34 Thermocouple Inputs
- 35 Ethernet Interface, Power Supply
- 35 External Display
- 36 Cables and Accessories
- 36 Measuring Cable
- 36 Measuring Cable with Banana Plugs
- 36 Measuring Cable with BNC Plugs
- 36 Adapter Cable with BNC Plugs
- 37 Ethernet Cable/Power Supply Cable
- 37 Y-Cable (ES650.1 in Stand-alone Mode)
- 38 Connecting Cables to the ES
- 40 Ordering Information
- 40 Accessories
- 41 Calibration
- 43 Configuration
- 44 Contact Information
- 45 Figures
- 46 Index