advertisement
![Thermocouple Inputs. ETAS ES650 | Manualzz Thermocouple Inputs. ETAS ES650 | Manualzz](http://s1.manualzz.com/store/data/068176823_1-1a67cf55866a93a27275e298eae09d44-360x466.png)
ETAS
5.9.4
Technical Data
Thermocouple Inputs
Resolution
Sampling rate
Measuring range
Cutoff frequency
Input impedance
Maximum inaccuracy T g
Maximum temperature drift
Maximum inaccuracy T i depend ing on the internal resistance of the thermocouple at internal resis tances > 50
T d
21 bits; corresponding to 0.01 K for J-,
K- and N-type thermocouples
0.1 to 10 samples/sec, configurable for each channel
+250 to +1820 °C (type B)
-200 to +1000 °C (type E)
-210 to +1200 °C (type J)
-200 to +1372 °C (type K)
-200 to +1300 °C (type N)
-50 to +1768 °C (type R)
-50 to +1768 °C (type S)
-200 to +400 °C (type T)
10 Hz
> 10 M || 10 nF
0.050 % + 5.00 K for type B
0.050 % + 0.20 K for type E
0.030 % + 0.20 K for type J
0.025 % + 0.25 K for type K
0.035 % + 0.38 K for type N
0.050 % + 2.0 K for type R
0.050 % + 2.0 K for type S
0.050 % + 0.30 K for type T
±0.1 K/K for type B
±0.0040 K/K for type E
±0.004 K/K for type J
±0.005 K/K for type K
±0.008 K/K for type N
±0.0400 K/K for type R
±0.0400 K/K for type S
±0.0050 K/K for type T
±0.0800 K/ for type B
±0.0030 K/ for type E
±0.003 K/ for type J
±0.004 K/ for type K
±0.006 K/ for type N
±0.0300 K/ for type R
±0.0300 K/ for type S
±0.0040 K/ for type T
±1 K Maximum inaccuracy of cold junc tion compensation T cjc
Possible increased measurement error under influence of electromag netic fields in industrial environ ments T emc
Maximum temperature drift of cold junction compensation T d
Maximum allowed input voltage
+16 K /-45 K
±0.008 K/K
32 V DC
Reference temperature T
0
298.15 K).
for T g
, T d
and T d
is 25 °C (equivalent to
ES650.1 - User Guide 33
advertisement
Related manuals
advertisement
Table of contents
- 5 About this Document
- 5 Classification of Safety Messages
- 5 Presentation of Instructions
- 6 Typographical Conventions
- 6 Presentation of Supporting Information
- 6 Scope of supply
- 6 Additional information
- 7 Basic safety notices
- 7 General safety information
- 7 Requirements for users and duties for operators
- 7 Intended use
- 12 Hardware Description
- 12 ES600 Measurement Modules
- 13 ES650.1 A/D and Thermo Module
- 13 Overview
- 14 Features
- 15 Package Contents
- 15 Housing
- 16 Ports and Indicators
- 16 Front Side
- 17 Indicators
- 18 Rear Side
- 19 Maximum Input and Common-Mode Voltages
- 20 Measurement Accuracy
- 22 Getting Started
- 22 Installation and blocking
- 22 General installation recommendations
- 22 Fastening the module onto a carrier system
- 23 Connecting several modules mechanically
- 24 Cabling
- 26 Configuration
- 26 Calibration
- 27 Technical Data
- 27 General Data
- 27 Product labeling
- 28 Fulfilled Standards and Norms
- 28 Environmental Conditions
- 28 Maintenance the Product
- 28 Cleaning the product
- 28 Mechanical Data
- 29 RoHS Conformity
- 29 CE conformity
- 29 UKCA conformity
- 29 KCC conformity
- 29 Taking the Product Back and Recycling
- 30 Use of Open Source software
- 30 System Requirements
- 30 Hardware
- 31 Software
- 32 Electrical Data
- 32 Power Supply
- 32 Host Interface
- 32 Analog Voltage Inputs
- 33 Thermocouple Inputs
- 34 Pin Assignment
- 34 Analog Inputs
- 34 Thermocouple Inputs
- 35 Ethernet Interface, Power Supply
- 35 External Display
- 36 Cables and Accessories
- 36 Measuring Cable
- 36 Measuring Cable with Banana Plugs
- 36 Measuring Cable with BNC Plugs
- 36 Adapter Cable with BNC Plugs
- 37 Ethernet Cable/Power Supply Cable
- 37 Y-Cable (ES650.1 in Stand-alone Mode)
- 38 Connecting Cables to the ES
- 40 Ordering Information
- 40 Accessories
- 41 Calibration
- 43 Configuration
- 44 Contact Information
- 45 Figures
- 46 Index