- No category
advertisement
1 Introduction to the Bravo Platform
Connection panel description
Connection panel description
Bravo rear connection panel
The following figure and table describe the connection panel at the rear of the
Bravo Platform.
Figure Connection panel on back of Bravo Platform
1 2 3 4 5
00220
Bravo
Back
Bravo Automated Liquid-Handling Platform User Guide 9
1 Introduction to the Bravo Platform
Connection panel description
Item Feature
1
Description
Ethernet port Connects the Ethernet card in the controlling computer to the Bravo Platform.
2
3
4
5
Serial port
Pendant port Connects the pendant to the Bravo safety interlock circuit.
If the Bravo Platform is integrated with the
Agilent Automation Control Unit, this 6- pin pendant port connects to an E- STOP DEVICE port on the Automation Control Unit.
The safety interlock circuit must be closed for the
Bravo Platform to operate. The pendant’s disable button interrupts this circuit. The safety interlock circuit can also be fitted with a light curtain to shut off power to the Bravo Platform if the light boundary is breached.
Fuse holder
An RS- 232 serial port that provides serial connection to the Bravo Platform.
If you connect the Bravo Platform using the serial port, you do not need to connect using the
Ethernet port.
AC power entry
Houses the AC inlet fuse and a spare fuse for the
Bravo Platform. For details on the fuse type, see the
Bravo Automated Liquid Handling Platform
Safety and Installation Guide
.
Connects the Bravo power cord to an AC outlet with a grounded circuit.
Pump I/O port
The following figure shows the Pump I/O port on the back of the Bravo
Platform. The table below describes the Pump I/O port in detail.
10 Bravo Automated Liquid-Handling Platform User Guide
Figure Pump I/O port (Bravo Platform back view)
00212c
Bravo
Acc_IO
1 Introduction to the Bravo Platform
Connection panel description
1
Item Feature
1
Description
Pump I/O port An RJ- 45 I/O port that connects the Pump
Module to the Bravo Platform. The connection is made with a straight- through shielded Cat- 5 or
Cat- 6 (Ethernet) cable.
IMPORTANT
This is not an Ethernet port and should only be used to connect Automation
Solutions accessories to the Bravo Platform.
Related information
For more information about...
Connecting the Bravo Platform
The pendant disable button
How to replace the fuse
Installing a Light Curtain
See...
Bravo Automated Liquid Handling
Platform Safety and Installation
Guide
“Replacing the fuse” on page 69
Bravo Automated Liquid Handling
Platform Safety and Installation
Guide
Bravo Automated Liquid-Handling Platform User Guide 11
advertisement
Related manuals
advertisement
Table of contents
- 7 Preface
- 8 About this guide
- 9 Accessing Automation Solutions user information
- 15 Introduction to the Bravo Platform
- 16 Safety notes
- 17 About the Bravo Platform
- 19 Hardware overview
- 23 Connection panel description
- 26 Pipette heads and pin tools
- 29 Software overview
- 31 Setting up the Bravo Platform
- 32 Workflow for setting up the Bravo Platform
- 33 Starting up and shutting down
- 35 Creating or adding a Bravo device
- 40 Opening Diagnostics
- 42 Creating and managing profiles
- 46 Initializing the Bravo Platform
- 50 Setting teachpoints
- 57 Verifying the gripper setup
- 59 Configuring external robot access
- 69 Preparing for a protocol run
- 70 Workflow for preparing a protocol run
- 70 Emergency stops
- 71 Planning for the protocol run
- 72 About performing dry runs
- 73 Maintaining the Bravo Platform
- 74 Routine maintenance
- 75 Cleaning the Bravo Platform
- 76 Cleaning up after a run
- 77 Changing the Bravo head
- 83 Replacing the fuse
- 84 Retracting tip box stripper pins
- 86 Moving the head manually
- 87 Recovering from a head collision
- 89 Troubleshooting hardware problems
- 91 Troubleshooting hardware-related error messages
- 92 Reporting problems
- 95 Using Bravo Diagnostics
- 96 About the Jog/Teach tab
- 98 Homing the pipette head
- 100 Jogging the pipette head
- 102 Changing the pipette head speed
- 103 Using the Move and Approach commands
- 105 Fine-tuning the gripper movements
- 111 Clearing motor faults and checking the head type
- 112 Performing a task using Bravo Diagnostics
- 115 Accessories and platepads
- 116 Accessories overview for the Bravo Platform
- 118 Installing the Accessories Hub
- 121 Setting up a Microplate Vacuum Alignment Station
- 126 Setting up the Orbital Shaking Station
- 132 Using the teach plate to set the teachpoint for an accessory
- 134 Setting up the Vacuum Filtration Station
- 147 Configuring a platepad for delidding
- 148 Installing an Alignment Station
- 150 Setting up a Nested Rack Insert
- 153 Using the Manual Fill Reservoir
- 154 Setting up the Thermal Station
- 157 Setting up thermal and shaking stations (Inheco controller)
- 165 (Bravo SRT only) Setting up an LT tip box location
- 169 Autofilling accessories
- 170 Autofilling station overview and setup workflow
- 171 Configuring an autofilling station for the Bravo Platform
- 176 Setting up a Weigh Station for the Bravo Platform
- 179 Setting up the Barcode Reader
- 180 About setting up the Barcode Reader
- 182 Installing or removing the Barcode Reader
- 183 Creating a profile for the Barcode Reader
- 185 Specifying the Barcode Reader location
- 187 Testing and optimizing barcode scanning
- 193 Setting up the AssayMAP Bravo Platform
- 194 AssayMAP Bravo Platform overview
- 197 Bravo 96AM Head overview
- 202 AssayMAP cartridge overview
- 204 About installing the AssayMAP Bravo Platform
- 205 Installing and using the Bravo 96AM Head
- 213 Installing the 96AM Wash Station
- 217 Setting up the AssayMAP Bravo Platform
- 220 Setting teachpoints for the AssayMAP Bravo Platform
- 223 Verifying and adjusting teachpoints for the accessory locations
- 231 Opening the Protein Sample Prep Workbench
- 232 Requirements for partial-head liquid handling
- 237 Diagnostics quick reference
- 238 Bravo Diagnostics dialog box
- 239 Configuration tab quick reference
- 242 Accessory Diagnostics dialog box
- 243 External Robots tab quick reference
- 244 IO tab quick reference
- 246 Jog/Teach tab quick reference
- 250 Gripper tab quick reference
- 252 Profiles tab quick reference
- 256 Processes tab quick reference
- 259 Aspirate task parameters
- 263 Dispense task parameters
- 265 Dispense to Waste task parameters
- 269 Mix task parameters
- 272 Pump Reagent parameters
- 273 Shake task parameters
- 275 Vacuum Filtration Station task parameters
- 278 Wash Tips task parameters
- 281 Pin Tool task parameters
- 285 AM Aspirate task parameters
- 290 AM Dispense task parameters
- 295 AM Mix task parameters
- 299 Index