Homing the pipette head
For information about...
See...
Using the Jog/Teach tab controls to move the pipette head
•
“Homing the pipette head” on page 84
•
“Jogging the pipette head” on page 86
•
“Using the Move and Approach commands” on page 89
Moving the pipette head manually
The four axes of the Bravo instrument
“Moving the head manually” on page 72
Homing the pipette head
About this topic
This topic describes when and how to home the pipette head.
About homing
The pipette head has a defined home position for each axis of motion. Homing sends the pipette head to the home position for the axes. Home the pipette head to reset the axes. For example, if you notice the Bravo Platform is not moving to locations or teachpoints accurately, home the pipette head.
Homing the pipette head is typically done automatically on first initialization after startup.
There are four homing axes.
Moves the … Homing in this axis …
X
Y
Z
W
Pipette head to its home position in the horizontal (left- right) axis. This position is near the middle of the x- axis range.
Pipette head to its home position for the horizontal (forwardbackward) axis. This position is near the middle of the y- axis range.
Pipette head to its home position in the vertical (up- down) axis. This position is near the top of the z- axis range.
Internal pipette head to its home position in the vertical aspirate- dispense axis. This position is near the low- to- mid
w- axis range.
84 Bravo Automated Liquid-Handling Platform User Guide
5 Using Bravo Diagnostics
Homing the pipette head
Homing the pipette head
WARNING
Keep away from the Bravo Platform when the pipette head is moving or about to move, especially in the z-axis direction. The pipette head might not stop immediately in a collision and a pipette tip could pierce your hand.
To home the pipette head:
1
In
Bravo Diagnostics, click the Jog/Teach tab.
2
To home the pipette head for all axes, click
Home XYZ.
The pipette head moves to the home position.
The axis order of homing is z—> x—> y.
Note: To home the pipette head in a single axis, click
Home [ axis]. Make sure you home the z- axis first so that the pipette head avoids the labware.
Related information
For information about...
Opening Bravo Diagnostics
See...
“Opening Diagnostics” on page 26
Bravo Automated Liquid-Handling Platform User Guide 85
advertisement
Related manuals
advertisement
Table of contents
- 7 Preface
- 8 About this guide
- 9 Accessing Automation Solutions user information
- 15 Introduction to the Bravo Platform
- 16 Safety notes
- 17 About the Bravo Platform
- 19 Hardware overview
- 23 Connection panel description
- 26 Pipette heads and pin tools
- 29 Software overview
- 31 Setting up the Bravo Platform
- 32 Workflow for setting up the Bravo Platform
- 33 Starting up and shutting down
- 35 Creating or adding a Bravo device
- 40 Opening Diagnostics
- 42 Creating and managing profiles
- 46 Initializing the Bravo Platform
- 50 Setting teachpoints
- 57 Verifying the gripper setup
- 59 Configuring external robot access
- 69 Preparing for a protocol run
- 70 Workflow for preparing a protocol run
- 70 Emergency stops
- 71 Planning for the protocol run
- 72 About performing dry runs
- 73 Maintaining the Bravo Platform
- 74 Routine maintenance
- 75 Cleaning the Bravo Platform
- 76 Cleaning up after a run
- 77 Changing the Bravo head
- 83 Replacing the fuse
- 84 Retracting tip box stripper pins
- 86 Moving the head manually
- 87 Recovering from a head collision
- 89 Troubleshooting hardware problems
- 91 Troubleshooting hardware-related error messages
- 92 Reporting problems
- 95 Using Bravo Diagnostics
- 96 About the Jog/Teach tab
- 98 Homing the pipette head
- 100 Jogging the pipette head
- 102 Changing the pipette head speed
- 103 Using the Move and Approach commands
- 105 Fine-tuning the gripper movements
- 111 Clearing motor faults and checking the head type
- 112 Performing a task using Bravo Diagnostics
- 115 Accessories and platepads
- 116 Accessories overview for the Bravo Platform
- 118 Installing the Accessories Hub
- 121 Setting up a Microplate Vacuum Alignment Station
- 126 Setting up the Orbital Shaking Station
- 132 Using the teach plate to set the teachpoint for an accessory
- 134 Setting up the Vacuum Filtration Station
- 147 Configuring a platepad for delidding
- 148 Installing an Alignment Station
- 150 Setting up a Nested Rack Insert
- 153 Using the Manual Fill Reservoir
- 154 Setting up the Thermal Station
- 157 Setting up thermal and shaking stations (Inheco controller)
- 165 (Bravo SRT only) Setting up an LT tip box location
- 169 Autofilling accessories
- 170 Autofilling station overview and setup workflow
- 171 Configuring an autofilling station for the Bravo Platform
- 176 Setting up a Weigh Station for the Bravo Platform
- 179 Setting up the Barcode Reader
- 180 About setting up the Barcode Reader
- 182 Installing or removing the Barcode Reader
- 183 Creating a profile for the Barcode Reader
- 185 Specifying the Barcode Reader location
- 187 Testing and optimizing barcode scanning
- 193 Setting up the AssayMAP Bravo Platform
- 194 AssayMAP Bravo Platform overview
- 197 Bravo 96AM Head overview
- 202 AssayMAP cartridge overview
- 204 About installing the AssayMAP Bravo Platform
- 205 Installing and using the Bravo 96AM Head
- 213 Installing the 96AM Wash Station
- 217 Setting up the AssayMAP Bravo Platform
- 220 Setting teachpoints for the AssayMAP Bravo Platform
- 223 Verifying and adjusting teachpoints for the accessory locations
- 231 Opening the Protein Sample Prep Workbench
- 232 Requirements for partial-head liquid handling
- 237 Diagnostics quick reference
- 238 Bravo Diagnostics dialog box
- 239 Configuration tab quick reference
- 242 Accessory Diagnostics dialog box
- 243 External Robots tab quick reference
- 244 IO tab quick reference
- 246 Jog/Teach tab quick reference
- 250 Gripper tab quick reference
- 252 Profiles tab quick reference
- 256 Processes tab quick reference
- 259 Aspirate task parameters
- 263 Dispense task parameters
- 265 Dispense to Waste task parameters
- 269 Mix task parameters
- 272 Pump Reagent parameters
- 273 Shake task parameters
- 275 Vacuum Filtration Station task parameters
- 278 Wash Tips task parameters
- 281 Pin Tool task parameters
- 285 AM Aspirate task parameters
- 290 AM Dispense task parameters
- 295 AM Mix task parameters
- 299 Index