Placement and Calibration of Sensors. TSI Quest Quest Environmental Monitor EVM Series, EVM-7
Add to My manuals178 Pages
advertisement
43 Calibrate Sensors
Placement and Calibration of Sensors
Calibrate Sensors
CHAPTER
4
Placement and Calibration of Sensors
Before you begin your measurement sampling, it is recommended to calibrate the sensors you will be using. This chapter focuses on inserting and removing sensors, calibrating each sensor, viewing past calibrations, and how to verify your calibration set points. (Reference the following sections which apply to your installed sensor components.)
Calibrating sensors include:
Particulate sensor
Pump flow rate calibration
CO
2
and O
2
sensor
Toxic sensors
Photo-Ionization Detector (PID) sensor (for VOCs)
Temperature (temp) and Relative humidity (RH) sensors
Inserting and Removing Sensors
1.
The sensors you purchased with your EVM will be inserted, factory calibrated, and ready for usage.
To understand how to insert or replace a sensor this is discussed in the procedures below.
NOTE
: To order new sensors when expired, see Appendix A , Replacement and optional parts.)
Inserting sensors
2.
Ensure the instrument is turned off and the external power cord is disconnected. Next, remove the sensor bar by unscrewing two screws from the top compartment with a screw driver.
3.
Lift the cover off and set aside. (The location of the sensors is indicated in Figure 4-2.)
4.
Remove the black manifold casing.
Manifold
Sensor housing
Figure 4-1: Sensor housing and manifold
EVM Series User Manual
44 Placement and Calibration of Sensors
Calibrating Particulates
5.
To insert a sensor, gently push the sensor’s pins into the socket pin of the circuit board so it aligns accordingly. It will fit securely.
6.
To remove a sensor, pull the senor up horizontally and remove it from the sensor chamber.
7.
When completed, replace the black manifold over the sensors and then place the cover over the sensor housing. Tighten the screws back into place.
Air inlet cover
Screw 1
Dial-in Impactors
Toxic sensor
Sensor bar- Remove screw 1 & screw 2 and lift off cover.
Temperature/
Relative Humidity sensor
(Dry bulb thermometer measures ambient air temperature)
Screw 2
PID sensor
CO
2
Sensor
Figure 4-2: Sensors’ locations
NOTE: The Temperature/ Relative Humidity sensor is not user removable.
NOTE: The user will receive a message on the display when a new sensor is detected or an existing sensor is no longer detected.
Calibrating Particulates
The EVM is factory calibrated to the “Arizona Road Dust” (also called “ISO Road dust” ISO 12103-1 A2
Fine) standard.
Before you measure particulates, it is recommended to calibrate to the environment you are in. If you are in an ARD environment, you would adjust the impactor to correlate with a specific particulate setting, run a calibration based on the particulates in that environment, and compare the results against the
ARD calibration. If you are in any other environment, you would calibrate based on the environment you are in by determining a correction factor (also called “K-factor”) to the correct variation from the ARD calibration standard.
EVM Series User Manual
45 Placement and Calibration of Sensors
Impactor and use
The EVM supports eight different user profiles. Each profile stores a customizable profile name field, and a correction factor. (Setting-up profiles/correction factors is discussed in the Set-up chapter,
“Setting profiles for particulates”).
The following sections will walk you through how the impactor operates, how to adjust the impactor, how to run a zero calibration, and lastly running a gravimetric calibration.
Impactor and use
The impactor is a device used to remove larger sized particulates from the air-stream you are measuring.
How does it operate?
Air travels through the impactor and the heavy particulates will stick to the greased plates within the impactor (1). The Pump will circulate the air as it passes through the tubing in the back of the instrument (2). The smaller, lighter particulates will pass through the optical engine (3) and collects in the gravimetric filter/cassettes (4).
2. Pump – circulates the air through the tubing
1. Filters through impactor
3. Passes through optical engine
4. Collects in Grav. filter/cassettes
Figure 4-3: How the impactor functions
NOTE: Depending on the environment (i.e., a dense, dusty manufacturing plant), you will want to ensure the impactor is well greased and cleaned before using. Additionally, it is recommended to monitor this process while measuring. You may need to stop your measurements, clean and grease the impactors, and then resume measuring. To clean and
grease the impactors, see Chapter 7 “Cleaning and maintenance”.
EVM Series User Manual
46 Placement and Calibration of Sensors
Adjusting the Impactor/Turret Setting
Adjusting the Impactor/Turret Setting
When measuring, the industrial processes (such as sanding, cutting, crushing, grinding, and transportation) cause air pollutants. The size of the particles will enter the human body in different areas, affecting the nasal, throat, and/or lung area.
The impactor setting is used to trap and filter out the larger size particles based on your cut-point (or impactor setting). The range is programmed between 1 μm to over 100 μm. For example, if you are measuring coal dust (aimed at the lung area), you would adjust your impactor setting to PM10. The following are the user-selectable cut-points:
PM2.5, PM4, PM10, or PM (all dust ranging from 0 μm-100 μm)
NOTE: The particulate size is written on top of the turret. For PM there is no marking but is noted by the (raised) air chamber.
Adjust impactor/turret for particulate logging
1.
Using the keypad navigate to the particulate or composite measurement screen.
How to?
From the Start screen, press the Right arrow key.
PM 2.5 & profile displayed
Impactor setting updates as the turret is rotated
(when in stop mode)
Figure 4-4: PM2.5 impactor setting example
EVM Series User Manual
47 Placement and Calibration of Sensors
Particulates Calibration (Zero Cal)
2.
Twist the turret (red casing around the impactor) to the left or right until you see the appropriate setting on the display. It will align to the impactor text and arrow on top of the housing.
NOTE: The actual diameter of the air inlet hole is directly to the left of the impactor text positioning.
Air inlet hole for PM2.5 Air inlet hole for PM4
PM2.5 position
Air inlet hole for PM10
PM4 position
Air inlet hole for PM
PM10 position PM position (no text marking)
Position used for zero calibration
Figure 4-5: Adjusting impactor/turret setting
3.
If you see “ < > PM ERR ”, this indicates the impactor is not in the correct position and that it is between two positions. Adjust the turret, until the correct reading is displayed on the screen.
Particulates Calibration (Zero Cal)
A zero calibration is recommended the first time you measure particulates in the run mode. For particulate calibration, place a zero filter (also called a
HEPA filter) with attached coupler into the vent of the turret. The impactor setting must be set at PM in order to conduct a zero cal.
NOTE: For the span on the particulate calibration, refer to the
“Gravimetric Filtering” instructions to determine the gain factor from the correction factor.
Particulates zero cal
1.
Insert the attached coupler and zero filter into the turret. Ensure the impactor is well-greased
before calibrating which requires a thin film of grease. (See “ Impactor care ” for details).
EVM Series User Manual
48 Placement and Calibration of Sensors
Particulates Calibration (Zero Cal)
Zero/HEPA filter
Used for particulate zero cal.
Zero filter with coupler and vent opening
Insert the Zero Filter via the coupler into the “2.5” vent opening.
Turret
Turn the turret so “PM” appears on the display of the EVM. The raised section will align with the
“impactor” text.
Figure 4-6: Particulate zero cal with HEPA filter
2.
Adjust the turret/impactor to PM (refer to “Turret” in Figure 4-6).
3.
From the Start screen, press ▲ / ▼ Arrows until Calibration is selected and press Enter .
Calibration – select to open to the Calibration set screen
Figure 4-7: Selecting calibration screen
4.
In the Calibration screen, press ▲ / ▼ Arrows and select PM (for particulate matter).
NOTE : If the impactor setting is incorrect, the instrument will flash a warning sign indicating that you need to turn the impactor to “PM” before proceeding.
EVM Series User Manual
49 Placement and Calibration of Sensors
Particulates Calibration (Zero Cal)
5.
Press the Cal soft key. (The pump will turn on.)
NOTE: The HISTR and VERIFY soft keys will not be present if a user calibration and a calibration verify have not been performed.
PM
History soft key
Calibration soft key
Verify soft key
Figure 4-8: Particulate cal screen
6.
The ready screen with the PM, Profile will appear. Press the Start soft key.
Start soft key
Figure 4-9: Particulate calibration and ready screen
EVM Series User Manual
50 Placement and Calibration of Sensors
Particulates Calibration (Zero Cal)
7.
Allow the level to stabilize (approximately 1–3 minutes). It will state Waiting (A) and Set soft key will appear when it has stabilized (B).
A B
Zero will appear when it has stabilized
Current Level reading
Set soft key will appear when stabilized
Figure 4-10: Particulate calibration and waiting screen
8.
Press the Set soft key to set the Zero cal point.
9.
The instrument will show a Pause state with the intended and actual values.
Save soft key
Figure 4-11: Particulate calibration pause screen
EVM Series User Manual
51 Placement and Calibration of Sensors
Particulate Advanced Calibration (Gravimetric)
10.
Press the Save soft key to save your calibration (or press Cancel and Retry to repeat this process.)
A screen pop-up will state “Zero cal set point” when completed (or “User canceled calibration”). A pass (cancel) screen will be displayed.
Retry
Actual calibration value
Exit
Figure 4-12: Zero Cal set for Particulates
Press Retry soft key to repeat the zero cal procedures.
11.
Press Exit soft key to close this screen. To return to the start screen, press On/Off/Esc or press the Cal soft key to continue calibrating the sensors.
NOTE: A zero cal is performed for the instrument and will be used for all profiles.
Particulate Advanced Calibration (Gravimetric)
With the EVM, you can store up to eight different profiles (also called correction factors) to reflect the environment or building zone you plan to measure.
The instrument is factory calibrated to Arizona Road dust. In some situations, you may want to perform a gravimetric calibration in order to determine the correction factor for an alternative environment. This is an optional procedure.
NOTE: If you do not plan on performing an advanced particulate calibration with gravimetric sampling.
With a gravimetric calibration, you determine the mass of the particulate you want to measure. The particulate mass is then used to calculate a correction factor to the environment you will be measuring.
The steps below outline a gravimetric calibration.
(For additional specifications on a gravimetric calibration, please refer to the NIOSH 0600 standard.)
EVM Series User Manual
advertisement
Related manuals
advertisement
Table of contents
- 9 List of Figures
- 10 List of Tables
- 12 Particulate Sampling (Aerosols/Dust Vapors)
- 12 Particulate Path
- 13 Particulate Sizes
- 14 Particulates and Cut-Point
- 14 Collecting/Sensing the Particulates
- 14 Gravimetric Sampling
- 14 Pump
- 14 Gas Sensors Sampling
- 15 Gas Sensors and Path
- 16 Volatile Organic Compounds (VOCs) Sampling
- 16 Temperature, Relative Humidity, Dew Point
- 16 What the Standards Say
- 17 Resources
- 18 EVM Series
- 18 Applications
- 19 Models Explained
- 21 Checking the Equipment
- 22 Display and Keypad
- 22 Diagram
- 22 Keypad description
- 23 Turning on
- 23 Start Screen Explanation
- 24 Navigating
- 24 Indicators/Icons on the display
- 25 Turning Off
- 25 Instrument Information
- 25 Unit Information Screen
- 27 TSI Detection Management Software DMS
- 27 Communicating
- 27 Powering and Outputs on the EVM
- 27 Battery
- 28 Charging
- 28 Outputs
- 29 Setting up Parameters via the Keypad
- 29 Time and Date Setting
- 30 Checking Battery Power
- 32 Display: Language, Contrast, and Key Click
- 33 Setting Logging and Viewing Run Time
- 35 PID Sensor Setup
- 37 Changing Units (Temperature, Particulates, PID, and Air Velocity)
- 37 Setting up Particulate Settings
- 38 Setting Profiles (Particulate Correction Factors)
- 38 How Particulate Factors are Adjusted and Calculated
- 38 Dust Correction Factors (Profiles)
- 40 Particulate Pump: Enable or Disable
- 40 Resetting the Gravimetric Filter
- 41 Auxiliary Input/Output
- 41 Digital Output
- 43 Air Velocity
- 43 Analog-out
- 44 Restoring Default Factory Settings
- 46 Backlight Setting
- 46 Resetting the EVM
- 46 Lock/Secure Run and/or Setup
- 48 Unlock run and/or setup
- 48 Auto-run
- 48 Auto-run and timed-run mode
- 49 Auto-Run with Date Mode
- 50 Auto-Run with Day of Week (DOW) Mode
- 53 Placement and Calibration of Sensors
- 53 Calibrating sensors include:
- 53 Inserting and Removing Sensors
- 54 Calibrating Particulates
- 55 Impactor and use
- 56 Adjusting the Impactor/Turret Setting
- 57 Particulates Calibration (Zero Cal)
- 61 Particulate Advanced Calibration (Gravimetric)
- 62 Performing a Gravimetric Calibration
- 63 Pump Flow Rate Calibration
- 64 Toxic Gas Sensors
- 65 Cross Sensitivity
- 65 How Sensors Operate
- 65 Quality Calibrations
- 65 About Sensors and Cal Storage
- 65 Calibration History and Sensor Serial Numbers
- 66 Applying a Gas Sensor
- 66 Calibrating Toxic Gas Sensors
- 67 Toxic Gas Zero Calibration
- 67 Ozone Calibration for Zero and Span
- 69 Toxic Gas Span Calibration
- 70 Calibrating CO2 AND O2
- 70 O2 Accuracy
- 73 Calibrating Photo-Ionization Detector (PID) Sensor
- 73 PID parts per million (ppm) zero calibration
- 73 PID parts per billion (ppb) zero calibration
- 74 PID Span Calibration
- 76 Calibrating Relative Humidity and Temperature
- 77 Calibration History
- 78 Verify Calibration and Drifting Analysis
- 62 Gravimetric Calibration Overview
- 81 Overview of Running a Session
- 81 Quick Help
- 81 Measuring
- 82 Understanding What Data is Stored
- 82 Summary Data
- 82 Logged Data/Time History
- 83 Running and Stopping
- 84 Measurement Screens and Navigation
- 84 Unit defined
- 85 About the Measurement Soft Keys
- 86 Particulate Measurement Screen
- 86 Impactors and Particulate Measuring
- 86 Extending Run-Time for PM Measurements
- 87 Changing Particulate Measurements
- 88 Relative Humidity and Temperature Measurements
- 88 Relative Humidity and Temperature Screen
- 89 Gas Measurements
- 89 Gas Screen
- 90 Composite Measurement
- 90 Composite Screens
- 91 Air Velocity Meaursurement Screen
- 92 Summary Data
- 92 Viewing Details in the Measurement Screen
- 93 Trend Data
- 93 Viewing trend data in the measurement screen
- 94 Past Session
- 94 Summary Data Screen
- 96 File System Menu
- 96 Deleting Sessions Individually
- 97 Delete All Sessions
- 98 Memory Remaining and File Properties
- 99 Estimated Run Time
- 101 Environmental Monitor Communication
- 103 Downloading Data with the EVM
- 103 EVM: Viewing Data
- 103 Selecting a session/study
- 104 EVM Panel Layout View (PLV) Page
- 106 EVM Reports and Printing
- 107 Customizing Reports
- 108 EVM Setup
- 108 Saving and Sending Configurations
- 109 General EVM Setup
- 109 EVM General Parameters Table
- 110 EVM and Logging Settings
- 111 EVM Logging Parameters Table
- 112 Timed-Run and Auto-Run Setting
- 112 Date and Auto-Run Setting
- 113 Day of Week and Auto-Run Setting
- 114 EVM and Security Settings
- 114 EVM and Triggering Settings
- 116 EVM and Particulate Settings
- 116 EVM and Particulates Setup
- 117 EVM Firmware Update
- 119 Overview
- 119 PID Sensor
- 120 Removing and Installing PID Sensor
- 121 Maintenance and Cleaning Overview of PID
- 121 Cleaning PID Sensor
- 123 Optical Engine: Maintenance and Cleaning
- 126 Impactor Care
- 127 Impactors and When To Clean
- 127 Gravimetric Cassette and Filter Paper
- 129 Particulate Sensor
- 129 Volatile Organic Compounds (VOCs) Gas Detector (PID Sensor)
- 130 Carbon Dioxide Sensor
- 130 Relative Humidity Sensor
- 130 Temperature Sensor
- 131 Toxic Gas Sensors
- 131 Air Velocity Accessory (Optional – Air Probe 10)
- 131 Displayed Data
- 132 Electrical Characteristics
- 132 Environmental effects
- 132 Physical characteristics
- 132 Measurements units
- 133 Operating Modes
- 133 Time Constants
- 133 Special Functions
- 134 Estimated Logged Run Times
- 134 Standard EVM-7 (Memory and logged time)
- 135 Standard EVM-4 (Memory and logged time)
- 137 Standard EVM-3 (Memory and logged time)
- 138 Accessories: Replacement and Optional Parts
- 141 Customer Service
- 141 Technical Support Contacts
- 141 Service Contact Information
- 142 Returning for Service
- 142 Calibration
- 143 Warranty
- 151 PID sensor and correction factors tables
- 161 Frequently Asked Questions (FAQs)
- 161 Particulate Measurements
- 163 Gravimetric Sample (optional for particulate calibration)
- 164 Toxic Sensor Measurements
- 165 Pump/Airflow Measurements
- 166 Digital Output (using stereo plug & diagram)
- 167 PID Questions (for Volatile organic compound measurements)
- 168 Power
- 168 Alarm Outputs
- 169 General: Mechanical
- 169 CO2 Sensor and how to reorder
- 170 CO2 Sensor and how to reorder (continued)
- 170 Calibrating Sensors
- 173 System Errors Displayed on the EVM Series