GammaVision
®
v7 (A66-BW) 783620H / 1013
Initial Directory
Specify the initial directory for the program to use either with directly entered character strings or by selecting from the list of macros shown in Fig. 63. The list is displayed by clicking on the arrow button to the right of the Initial Directory field. The directories on this menu are described in Section 5.1.1.2 under
“Initial Directory” (page 60), except that the reference to file-
Fig. 63. Import Macros.
name applies to the file selected for importation with the File/Import command.
Default
The default extension entered here is used as the extension for the filename in the filename entry dialog. For example, if
TXT
is entered, then the name list in the entry dialog will be
*.TXT
.
Run Options
These three radio buttons (Minimized, Maximized, and Normal Window) are used to select the window for the program. If the program does not have any user dialogs, any option can be selected. If the program needs user inputs, Normal Window should be selected.
Use DataMaster
To use DataMaster as the import program, mark the checkbox then select the file type to be imported from the Default File Type list. Figure 64 shows the list of supported file types. To use the Import Program specified in the upper half of the dialog, just unmark the Use DataMaster box.
5.1.1.4. Directories
This tab (Fig. 65) gives you the option of designating where GammaVision’s various file types should be stored. Otherwise, all spectra, libraries,
.JOB
.ROI
files,
files, reports, correction table files,
Fig. 64. Choose the DataMaster
Import File Type.
sample type files, and calibrations will be stored in
C:\User
. When you generate any of Gamma-
Vision’s many File Types, the program reads the file extension and stores it according to the location(s) specified on this tab. For instance, if you specify a location for the Spectra file type, all
.SPC
,
.CHN
, and
.SPE
files will be saved there by default. Similarly, if you specify a location
62