GammaVision
®
v7 (A66-BW) 783620H / 1013
Iodine Equivalence Sidebar Control Menu
Figure 172 shows the Iodine Equivalence Sidebar’s control menu. To edit an existing file, click
Edit File... and select a file. It will open in a Notepad-like ASCII editor screen (Fig. 173). The formatting does not have to be exactly as shown. Clicking on Save will save any changes to the file (in other words, this is not a Save As...).
Fig. 172.
Fig. 173. Editor Screen for Iodine Equivalence Table.
Destroy clears all values from the current table. Restore abandons any changes and returns the table to its condition before editing (even after using Destroy), as long as the Restore is exe-
cuted before the Iodine Equivalence Sidebar is closed. Close closes the sidebar and table, and makes the currently selected table the internal or working Iodine Equivalence table.
Iodine Equivalence Table Control Menu
The Iodine Equivalence Table’s control menu has the same commands as shown in Fig. 169, including the Print command.
5.5.1.8. DAC (MPC)
The DAC (MPC) table is built using the sidebar shown in Fig. 174; the calculation is described in Section 6.15. To make the table, enter the DAC value (Bq/Kg) and the Isotope name for each isotope to be included in the report. The isotope name must be exactly as given in the analysis
library. Now click Enter to create and display the table (see Fig. 175). Each time another isotope is added, the table updates.
To remove an isotope from the table, click it to select it, then click Delete Isotope on the sidebar. To edit an isotope name or energy, select it with the mouse, make the necessary changes on the sidebar, and click Enter. The table will update.
184