GammaVision
®
v7 (A66-BW) 783620H / 1013
8.4. Creating a QA Database
If you select the Settings... command on the QA submenu and the “Unable to Access QA Database” message is displayed (Fig. 298), GammaVision cannot find the
QA database file
GvQa32.Mdb
in the folder in which it was originally created (i.e., it has been moved, renamed, or deleted). By default, GammaVision creates the QA database file in
C:\User
.
If you have moved the QA database file to another folder, you can either (1) reestablish connection with it
Fig. 298. Cannot Find QA
Database; Do You Wish to Create a New One?
and its contents, or (2) create a new, empty QA database.
If you choose to create a new QA database, we recommend that you modify the name of the old database file to prevent data loss due to overwriting.
To create a new, empty copy of
GvQa32.Mdb
, answer Yes. This will open the Select Database
File dialog (Fig. 299). Click OK to accept the default filename and location; or select a new location and/or filename, then click OK. A blank Settings dialog will open, as described in
Section 8.2.1, so you can establish new settings for the currently selected detector.
To reconnect to your existing QA database file, answer Yes to open the Select Database
File dialog, and click Browse. This will open a standard Windows file-open dialog. Locate the desired QA database file, click Save, and confirm that you wish to overwrite the file.
Note that this does not overwrite (delete) the
contents of your database file. The Settings dialog will open, as described in Section 8.2.1, and will display the settings for the currently selected detector.
Fig. 299. Accept Default Settings or Choose
Location and Name of New QA Database File.
374