783620H / 1013 5. MENU COMMANDS
!
The best match, if any, from the library
And, if a match is found in the library:
!
The activity, calculated using the net area, the live time, and the efficiency
ROI# RANGE (keV) GROSS NET +/- CENTROID FWHM FW(1/10) LIBRARY (keV) Bq +/-
1 63.26 67.86 4231 490 116 65.47 0.96 2.11 No close library match.
2 135.95 140.91 5898 482 145 138.80 0.66 1.40 Hf181 136.28 0.76 0.23
3 507.40 513.06 8632 5445 140 510.22 2.73 4.56 Sr85 514.00 0.74 0.02
4 1455.81 1463.59 2088 1588 69 1460.68 2.07 3.97 Br82 1474.90 2.72 0.12
5 2610.29 2620.19 818 609 48 2614.98 2.83 4.55 No close library match.
Fig. 182. Example of Column-Format Report.
5.5.4. Entire Spectrum in Memory...
This selection initiates an analysis of the entire spectrum in the active buffer or Detector window
(if the Detector is not acquiring data), generating
.RPT
,
.UFO
, and
.AN1
files. The analysis is performed in the background and the display is available for continued interactive use. The spectrum is moved to disk for the analysis so the spectrum in the Detector or buffer can be changed if needed. When the analysis is completed, you are notified and the report is generated. If any errors occur, the error number is displayed along with the spectrum name. The errors corresponding to the error numbers are explained in Appendix C.
If the Program radio button on the Report tab of Analyze/Settings/Sample Type... has been selected, with
Notepad.exe
designated as the output program, GammaVision will open Windows
Notepad and display the analysis report. The software will not accept inputs while the report display is shown. When you exit Notepad, control returns to GammaVision. At this point, if you select Analyze/Display Analysis Results, the
.UFO
file for the spectrum just analyzed will automatically be displayed.
5.5.5. Spectrum on Disk...
This command analyzes a spectrum stored on disk, using the analysis parameters stored in the spectrum file. A standard file-open dialog is displayed. A background analysis is performed as soon as you open the spectrum filename, and a
.UFO
and
.RPT
file are generated. When the analysis is complete a message is displayed on the information line. If any errors occur, the error number is displayed along with the spectrum name. The numbers are explained in
Appendix C.
193