GammaVision
®
v7 (A66-BW) 783620H / 1013
Fig. 324. The Peak Info Window for an
ROI.
11.1.4.10. Show Hover Window
When you select this command, a checkmark is displayed by this menu item to indicate that it is in hover-window mode. In this mode, the Peak Info window opens when the mouse pointer is paused over a peak for approximately 1 second, and closes when the pointer is moved away from the peak. To turn off hover mode, select Show Hover Window again to remove the checkmark.
11.1.4.11. Sum Spectrum
This sums the gross counts in the area selected with the rubber rectangle or, if you have not selected an area, the gross counts in the entire spectrum. The results are displayed on the status line, and indicate the span of channels summed.
11.1.4.12. Print Graph
This command prints performs a “quick print” of the contents of the Expanded Spectrum
Window, using the currently selected printer and print settings (these will be either the default printer and print settings for this PC or the printer and settings used most recently during this
GVPlot session). To change printers and/or print properties, use the Print... command on the File menu (page 427).
11.1.4.13. Properties
This opens the Graph Properties dialog, which is discussed in Section 11.1.3.3.
434