The DAC units displayed at the top of the sidebar are the currently defined analysis units on the Systems tab under
Analyze/Settings/Sample Type.... If a
.DAC
file using different units is recalled, the dialog in Fig. 176 opens, warning of this difference. Click Yes to open the dialog for the conversion factor (Fig. 177). Enter the factor and click OK to scale the table values and put the table in the analysis units. If No is selected, the analysis and table will be in different units and the DAC/MPC calculation will produce unknown results.
5. MENU COMMANDS
Fig. 174. DAC (MPC) Sidebar.
Fig. 175. DAC (MPC) Table.
Fig. 176. Warning Message.
To save the current table, click Save.... This opens a filesave dialog. Enter a filename and click Save; GammaVision will append the default
.DAC
file extension.
To display an existing table, click Recall..., select the desired file, and click on Open.
Fig. 177. Conversion Factor.
DAC (MPC) Sidebar Control Menu
The DAC (MPC) Sidebar contains the same commands as the menu in Figure 172. To edit an existing file, click Edit File... and select a file. It will open in a Notepad-like ASCII editor
(Fig. 178). If you make any changes, click Save before closing the editor window.
Destroy clears all values from the current table. Restore abandons any changes and returns the table to its condition before editing (even after using Destroy), as long as the Restore is exe-
cuted before the DAC (MPC) Sidebar is closed. Close closes the sidebar and table, and makes the currently selected table the internal or working DAC (MPC) table.