Insert CLEAR and START after FILL_BUFFER, as shown here:
783620H / 1013
10.3.2. JOB Commands for List Mode
This section discusses the JOB commands for MCBs that support List Mode (which is discussed in Section 1.6). The SET_LIST and SET_PHA commands switch the selected Detector respectively between the PHA and List modes; the SAVE command supports the
.LIS
file type; and the
SET_RANGE command retrieves a specified time slice of data from an existing
.LIS
file. Note that the SET_RANGE command has two syntaxes: A time slice can be recalled either by specifying an absolute date/time and the desired time-slice duration; or by specifying the starting real time and the desired time-slice duration. Note that while the List Data Range command on the
Calculate menu (Section 5.4.2) can only retrieve time slices to a resolution of 1 second, the
SET_RANGE command syntax supports fractional real time and duration.
The following example uses the commands discussed in the preceding sections plus the list-mode commands to switch “Detector 12" from the standard PHA mode to List Mode, collect data and save it in
.LIS
format as well as the three supported spectrum file formats, recall the a buffer, retrieve a time slice of data from the
RANGE command, save the time slices, and switch the Detector back to PHA mode.
.LIS
file into
.LIS
file using the two syntaxes of the SET_