5.2.11.2. Nuclide Report Calculations. ORTEC A66-BW GammaVision
广告
广告
GammaVision
®
v7 (A66-BW) 783620H / 1013
5.2.11.2. Nuclide Report Calculations
The Nuclide Report displays the activity of up to 9 user-selected peaks. Once the report is set up you can view the Nuclide Report at any time. The peak area calculations are the same as the calculations in GammaVision and other ORTEC software, so the Report contents can be duplicated using the spectra stored in the PC. The calculated value is computed by multiplying the net peak count rate by a user-defined constant. If the constant includes the efficiency and branching ratio, the displayed value will be activity. The nuclide label and the activity units are entered by the user.
The report has this format:
Nuclide
CO-60
CO-60
CO-57
keV
1332.5
1173.2
122.1
uCi/m2 ±%
1.21E+01 10.2
1.09E+01 12.3
1.48E+00 86.2
Calculations
These are the calculations used to generate the Nuclide Report’s Activity, Uncertainty, and
Peak values.
Activity is calculated as follows:
NucCoef is normally the inverse of efficiency times the branching ratio. Note that the efficiency is the ABSOLUTE counting efficiency for the source/detector geometry being used. Thus, in order to get meaningful activity results, as in any counting situation, you need to have efficiency factors which are appropriate to the actual counting geometry. If NucCoef is set to 1, you will get peak count rate on the display.
LiveTime is the current live time.
NetCounts is computed with the following equation: