783620H / 1013 5. MENU COMMANDS
for the Table Files file type, GammaVision will save your energy (
.ENT
) and efficiency (
.EFT
) calibration tables and table for DAC (
.DAC
), geometry (
.GEO
) and other corrections in that folder.
To change the path (Location) of a particular File Type, click the desired file type, then
Modify.... This will open a standard file-recall dialog. Choose the desired location, creating a new folder if necessary, and click Open. When all path changes have been completed, click OK to put them into effect or else Cancel to retain the previous settings.
Fig. 65. Directories Tab.
5.1.2. Recall...
This command reads a spectrum or List Mode file into a new buffer window. It opens a standard
Windows file-open dialog (Fig. 66), allowing you to select the file to be recalled. The spectrum files are created by GammaVision’s Save and Save As... functions and by any other programs that can produce the
.CHN
,
.SPC
, or
.SPE
format (e.g., MAESTRO, ISOTOPIC). The buffer is resized to the memory size of the recalled spectrum.
Note the Show Description checkbox on the lower left of the dialog. Mark this to display the sample description, format, and spectrum size of each file without having to open it.
If the maximum eight buffer windows are currently open, GammaVision will ask if you wish to close the oldest buffer. Answering No will cancel the recall operation and the oldest buffer will remain onscreen. Answering Yes will close the oldest buffer and open a new buffer containing
63